K191692 · Imedrix Inc. (Formerly Piitech Inc.) · DPS · Jan 10, 2020 · Cardiovascular
Device Facts
Record ID
K191692
Device Name
KardioScreen
Applicant
Imedrix Inc. (Formerly Piitech Inc.)
Product Code
DPS · Cardiovascular
Decision Date
Jan 10, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
KardioScreen is a portable and mobile device intended to acquire a resting, diagnostic ECG using surface electrodes from adult patients and pediatric patients age 14-21 years. The device is intended to be used by doctors and/or trained healthcare professionals in a hospital, medical professional's facility, clinic, physician's office or any medical outreach center. The device can acquire, display, analyze and print either 6-lead or 12-lead electrocardiograms which can be exported for external use. The recorded ECG signals can help users to analyze and diagnose heart disease, but the device is not meant as a sole means of diagnosis.
Device Story
KardioScreen is a portable, battery-operated electrocardiograph for clinical use by physicians or trained healthcare professionals. It acquires resting 6-lead or 12-lead ECG signals via surface electrodes connected to a KardioScreen Converter. The converter processes signals and transmits data via Bluetooth Low Energy (BLE) to an Android tablet running the KardioScreen application. The application displays, analyzes, and prints ECG waveforms for clinical review. The device aids in heart disease diagnosis but is not a sole diagnostic tool. It is intended for use in hospitals, clinics, and medical outreach centers.
Clinical Evidence
No clinical data were submitted. Substantial equivalence is supported by non-clinical testing, including electrocardiograph safety and performance, electrical safety, EMC, wireless co-existence, software verification and validation, usability validation, and biocompatibility.
Technological Characteristics
Portable, battery-operated electrocardiograph. Uses surface electrodes (disposable or optional reusable clamp/bulb). Connectivity via Bluetooth Low Energy (BLE) to Android tablet. Defibrillation-proof. Complies with electrical safety and EMC standards.
Indications for Use
Indicated for adult and pediatric patients (ages 14-21) for the acquisition of resting, diagnostic 6-lead or 12-lead ECGs to assist in the analysis and diagnosis of heart disease.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
January 10, 2020
iMedrix Inc. (formerly Piitech Inc.) Srikanth Jadcherla Chief Executive Officer 1525 McCarthy Blvd., Suite 1059 Milpitas, California 95035
Re: K191692
Trade/Device Name: KardioScreen Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: December 11, 2019 Received: December 12, 2019
Dear Srikanth Jadcherla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191692
Device Name KardioScreen
#### Indications for Use (Describe)
KardioScreen is a portable and mobile device intended to acquire a resting, diagnostic ECG using surface electrodes from adult patients and pediatric patients age 14-21 years. The device is intended to be used by doctors and/or trained healthcare professionals in a hospital, medical professional's facility, clinic, physician's office or any medical outreach center. The device can acquire, display, analyze and print either 6-lead or 12-lead electrocardiograms which can be exported for external use. The recorded ECG signals can help users to analyze and diagnose heart disease, but the device is not meant as a sole means of diagnosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Imedrix. The logo consists of a square made up of smaller squares in different shades of blue and pink. To the right of the square is the word "imedrix" in a blue sans-serif font. The "i" is lowercase, and the rest of the letters are lowercase.
## I. SUBMITTER
iMedrix Inc. (formerly Piitech Inc.) 1525 McCarthy Blvd., Suite 1059 Milpitas, CA 95035
Phone: 866-463-3749 Fax: 866-463-3749
Contact Person: Srikanth Jadcherla, Chief Executive Officer
Date Summary Prepared: December 11, 2019
## II. DEVICE
| Proprietary Name: | KardioScreen |
|----------------------|--------------------|
| Classification Name: | Electrocardiograph |
| Product Code: | DPS |
| Regulation: | 21 CFR 870.2340 |
| Regulatory Class: | II |
## III. PREDICATE DEVICE
The primary predicate device is Edan Instruments, Inc.'s PADECG, which was cleared under K161302.
This predicate device has not been to subject of a recall.
# IV. DEVICE DESCRIPTION
KardioScreen is a portable battery-operated electrocardiograph designed to acquire either a 6lead or 12-lead resting, diagnostic electrocardiogram (ECG) using surface electrodes from adult patients and pediatric patients aged 14-21 years. The device consists of the following three (3) parts:
- 1. Electrodes and patient cable with KardioScreen Converter
- 2. KardioScreen device
- 3. KardioScreen application running on Android tablet
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Imedrix. The logo consists of a square made up of smaller squares in different shades of blue and pink, followed by the word "imedrix" in blue lowercase letters. The logo is simple and modern, and the colors are calming and professional.
The device is supplied with patient cables with lead wires that connect to any commercially available disposable ECG electrode. Optional reusable clamp electrodes can be used for the limb leads and bulb electrodes for the chest leads.
Once the electrodes have been connected to the patient, the ECG cable with KardioScreen Converter is then connected to the "ECG" port on the KardioScreen device. This module receives the electrical ECG signals, performs processing on the signals and then wirelessly transmits the ECG data through a Bluetooth Low Energy (BLE) interface to a mobile application running on an Android tablet. The ECG waveforms are displayed on the KardioScreen application for viewing and analysis. From the Android tablet, the ECG report can be printed or exported for external use.
## V. INTENDED USE / INDICATIONS FOR USE
KardioScreen is a portable and mobile device intended to acquire a resting, diagnostic ECG using surface electrodes from adult patients and pediatric patients age 14-21 years. The device is intended to be used by doctors and/or trained healthcare professionals in a hospital, medical professional's facility, clinic, physician's office or any medical outreach center. The device can acquire, display, analyze and print either 6-lead or 12-lead electrocardiograms which can be exported for external use. The recorded ECG signals can help users to analyze and diagnose heart disease, but the device is not meant as a sole means of diagnosis.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
iMedrix's KardioScreen and Edan Instruments, Inc.'s PADECG have the same intended use, same clinical setting, same target user, same ECG measurements, and same pace detection. Both devices comply with the electrical safety, EMC and electrocardiograph safety and performance standards and are defibrillation-proof.
Both devices transmit wirelessly to a mobile device, though KardioScreen is designed to work with Android tablets while PADECG works with iOS devices. KardioScreen can also capture a 6lead ECG, in addition to the standard 12-lead ECG that PADECG captures. Both devices are designed for adult and pediatric populations, though KardioScreen restricts pediatric use to ages 14 and older.
These differences in technological characteristics do not raise new issues of safety or effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Imedrix. The logo consists of a square made up of smaller squares in different shades of blue and pink. To the right of the square is the word "imedrix" in blue lowercase letters. The "i" is the same shade of blue as the darkest square in the logo, while the "x" is a lighter shade of blue.
### VII. PERFORMANCE DATA - NON-CLINICAL TESTING
No animal testing was submitted to support the substantial equivalence of KardioScreen to PADECG. The following non-clinical testing was performed to support the substantial equivalence of KardioScreen to the predicate device:
- Electrocardiograph safety and performance
- Electrical safety
- . Electromagnetic compatibility (EMC)
- . Wireless co-existence testing
- Software verification and validation ●
- . Usability validation
- Biocompatibility .
#### VIII. PERFORMANCE DATA - CLINICAL TESTING
No clinical data were submitted to support the substantial equivalence of KardioScreen to PADECG.
#### IX. CONCLUSIONS
Based on the indications for use, technological characteristics, results of non-clinical testing, and comparison to the predicate device, KardioScreen has been shown to be substantially equivalent to a legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.