The COR12 ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on a screen or printed on paper. The COR12 ECG can be used for applications in patients age 4 years and older and a weight of 20 kg or higher. The COR12 ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital).
Device Story
COR12 ECG is a 12-lead surface electrocardiograph; collects ECG signals via standard electrodes using impedance measurement; transmits data via Bluetooth 2.1+EDR (SPP protocol) to external customer-defined host software for display or printing. Device lacks onboard display; features single power-on button; auto-shuts off after 10 minutes inactivity. Operated by trained healthcare professionals (nurses, doctors) in clinical settings (doctor's office, hospital). Not for home, EMS, or intracranial use. Provides ECG data to clinicians to support diagnostic decision-making; benefits patients through routine cardiac monitoring.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance verified through compliance with ISO 62304 (software life cycle), IEC 60601-1 (general safety), IEC 60601-2-25 (ECG safety), and IEC 60601-1-2 (EMC).
Technological Characteristics
12-lead surface ECG; impedance measurement principle; 1.94μV/bit resolution at 500 Hz sampling; 0.05–150 Hz bandpass filter; 4000 Hz pacemaker detection. Connectivity: Bluetooth 2.1+EDR. Power: 1x AA alkaline or NiMH battery. Dimensions: 8.0 x 9.3 x 2.1 cm; Weight: 200g. Electrodes: 4mm gilded snap. Type CF protection against electric shock.
Indications for Use
Indicated for measuring surface ECG in patients age 4 years and older weighing 20 kg or higher. Used for routine ECG collection under resting and stress conditions by trained healthcare professionals in clinical facilities.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{1}------------------------------------------------
December 28, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Corscience GmbH & Co. % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K183369
Trade/Device Name: COR12 ECG Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: December 2, 2018 Received: December 4, 2018
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Arielle Drummond -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K183369
Device Name
## COR12 ECG
Indications for Use (Describe)
The COR12 ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on a screen or printed on paper. The COR12 ECG can be used for applications in patients age 4 years and older and a weight of 20 kg or higher. The COR12 ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
K183369
Corscience
| 510(k) Summary | | |
|----------------------------|---------------------------------------------------|--|
| Page 1 of 5 | | |
| 26-Nov-18 | | |
| Submitter: | Karlheinz Trost | |
| | Head of Quality Management and Regulatory Affairs | |
| | Corscience GmbH & Co. KG | |
| | Hartmannstraße 65 | |
| | D-91052 Erlangen, Germany | |
| | E-mail - Karlheinz.Trost@corscience.de | |
| | Tel +49-9131-977986-518 | |
| Submission Correspondent: | Paul Dryden | |
| | ProMedic, LLC | |
| | 131 Bay Point Dr NE | |
| | St. Petersburg, FL 33704 | |
| | E-mail - paul.dryden@promedic.cc | |
| | Tel - 239-307-6061 | |
| Proprietary or Trade Name: | COR12 ECG | |
| Common/Usual Name: | Electrocardiograph | |
| Classification Name: | 21 CFR 870.2340 | |
| | DPS – Electrocardiograph | |
| | Class II | |
| Predicate Devices: | K150810 – CareFusion - Vyntus ECG | |
#### Device Description
The COR12 ECG is intended for the recording of 12-lead ECG data of a patient by trained staff, e.g., nurse, doctor's assistant or the doctor. This can be performed as a resting or a stress ECG.
It transmits the ECG data in a specified format by Bluetooth® transmission to a customer-defined host software running with its own display, which is not part of the COR12. There is no display of ECG waveforms on the COR12.
Limitation:
- It is not suitable to be used with High Frequency surgery devices. •
- . Not intended for intracranial use.
- . Not intended for use in an EMS environment (Emergency Medical Services Environment).
- . Not intended for use in home healthcare environments.
It has a single power-ON button to start data collecting and automatically turns off after 10 minutes of inactivity.
#### Indications for Use
The COR12 ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on a screen or printed on paper. The COR12 ECG can be used for applications in patients age 4 years and older and a weight of 20 kg or higher. The COR12 ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital).
{5}------------------------------------------------
#### 510(k) Summary Page 2 of 5 26-Nov-18
### Comparison to Predicate
We present in Tables 5.1 – 5.2 a comparison of the subject device compared to the predicate.
#### Table 5.1 - Indications and Classification Comparison of Subject and Predicates
| | Subject Device | K150810 | Comment |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Product Name | COR12 ECG | Vyntus/Sentry Suite Product Line<br>Vyntus ECG | |
| Product Code | DPS | DPS and BZC | DPS is a secondary<br>product code and |
| CFR | 870.2340 | 868.1880 | CFR in the<br>predicate |
| Classification Name | Electrocardiograph | Pulmonary Function Data<br>Calculator | |
| Indications for Use | The COR12 ECG is intended for<br>measuring the surface<br>electrocardiogram (ECG) of a<br>patient. The acquired ECG can be<br>recorded and displayed on a screen<br>or printed on paper. | The Vyntus ECG is intended for<br>measuring the surface<br>electrocardiogram (ECG) of a<br>patient. The acquired ECG can be<br>recorded and displayed on the<br>screen or printed on paper. | Identical |
| Data collection only | Yes | Yes | |
| Test conditions | The COR12 ECG is intended to be<br>used for routine ECG collection,<br>recording both under resting and<br>stress conditions. | The Vyntus ECG is intended to be<br>used for routine ECG collection,<br>recording both under resting and<br>stress conditions. | Identical |
| Population | The patient population for this<br>device is age 4 years and older and a<br>weight of 20 kg or higher. | The patient population for this<br>device is age 4 years and older and<br>a weight of 20 kg or higher. | Identical |
| Environments of Use | The measurement is performed by<br>trained healthcare professionals<br>under the direction of a physician in<br>healthcare facilities (e.g. the<br>doctor's office or hospital). | The measurement is performed by<br>trained healthcare professionals<br>under the direction of a physician<br>in healthcare facilities (e.g. the<br>doctor's office or hospital). | Identical |
{6}------------------------------------------------
K183369
Corscience
# 510(k) Summary
#### Table 5.2 – Comparison of Technical Specifications
| | Subject Device – COR12 ECG | Primary Predicate - Vyntus ECG – K150810 |
|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Feature | Value | |
| Records ECG only | Yes | Yes |
| Data output | Recorded and displayed on screen or printed | Recorded and displayed on screen or printed |
| Records | Under resting and stress conditions | Under resting and stress conditions |
| Dimensions W x H x D | 8.0 x 9.3 x 2.1 cm (3.3 x 3.7 x 0.8 in) | 8.0 x 9.3 x 2.1 cm (3.3 x 3.7 x 0.8 in) |
| Weight (without battery) | 200 g (0.4 lbs) | 200 g (0.4 lbs) |
| 12-channel Surface ECG Recording | Yes | Yes |
| Electrodes | Standard ECG electrodes | Standard ECG electrodes |
| Connection of the electrodes | 4 mm snap, gilded | 4 mm snap, gilded |
| Technology of ECG signal | Impedance measurement | Impedance measurement |
| Accuracy of the ECG signal | 1,94μV/bit resolution (at sampling rate of 500 Hz) | 1,94μV/bit resolution (at sampling rate of 500 Hz) |
| Filter | Bandpass 0.05 Hz – 150 Hz No<br>line filter (50 Hz / 60 Hz) | Bandpass 0.05 Hz – 150 Hz No<br>line filter (50 Hz / 60 Hz) |
| Pacemaker detection | 4000 Hz | 4000 Hz |
| Temperature range operation | T = 10 – 37 °C | T = 10 – 37 °C |
| Ambient pressure range operation | 700 to 1060 hPa (525 to 795 mmHg) | 700 to 1060 hPa (525 to 795 mmHg) |
| Humidity (operation, storage, transport) | 5 – 95% RH (not condensing) | 5 – 95% RH (not condensing) |
| Power supply | 1x AA Alkaline (professional or rechargeable)<br>battery or rechargeable NiMH battery | 1x AA Alkaline (professional or rechargeable)<br>battery or rechargeable NiMH battery |
| Runtime | > 5h with AA battery<br>> 8h with rechargeable NiMH battery (2850 mAh) | > 5h with AA battery<br>> 8h with rechargeable NiMH battery (2850 mAh) |
| Current consumption at 1.5 V<br>- Operation / Idle | 230 mA / 185 mA | 230 mA / 185 mA |
| Classification according to 60601-1<br>- Protection type against electric shock<br>- Protection level against electric shock | Device with internal power supply<br>Type CF | Device with internal power supply Type CF |
{7}------------------------------------------------
# K183369
Corscience
| 510(k) Summary<br>Page 4 of 5<br>26-Nov-18 | | |
|----------------------------------------------------------------------------------------------------------|--------------------------------------------------|--------------------------------------------------|
| Recovery Time after defibrillation | < 8 seconds | < 8 seconds |
| Electromagnetic compatibility (EMC)<br>according to 60601-1-2 and 60601-2-25<br>- Emission<br>- Immunity | CISPR 11 Class B<br>IEC 61000-4 parts 2, 3, 6, 8 | CISPR 11 Class B<br>IEC 61000-4 parts 2, 3, 6, 8 |
| Data transmission | Bluetooth 2.1+EDR Protocol: SPP. | Bluetooth 2.1+EDR Protocol: SPP. |
#### Discussion of the Comparison and Differences
As presented in Tables 5.1 to 5.2, we have compared the subject device, Vyntus ECG cleared under K150810 for equivalence of:
#### Indications, Patient Population and Environment of Use -
Both devices are the same indications for use which the surface electrocardiogram (ECG) of a patient. The aquired ECG can be recorded and displayed on a screen or printed on paper recording ECG signals.
The patient population and environment of use are identical.
- The patient population from 4 years on and a minimum weight of 20 kg. ●
- · The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital)..
Discussion - The subject and predicate device are identical.
Prescription-Use Only – Both devices are prescription-use only.
Discussion – There is no difference in the type of device as a prescription device; thus, they are substantially equivalent.
Design, Technology and Principle of Operation – The fundongy, and principle of operation is measuring ECG signals via impedance from a surface electrode is identical.
Discussion - The subject and predicate devices have the identical technological characteristics for measuring ECG signals.
#### Performance - Non-Clinical Testing Summary -
Biocompatibility - The only component which in in patient is the off-the-shelf electrodes. These are 510(k) cleared and we have provided an example of one in the submission.
{8}------------------------------------------------
#### 510(k) Summary Page 5 of 5 26-Nov-18
Bench Testing - The subject and predicate devices are identical as the supplier of the predicate and thus the technical specifications are identical.
- Software life cycle ISO 62304 ●
- IEC 60601-1
- IEC 60601-2-25
- EMC Compatibility IEC 60601-2
#### Substantial Equivalence Conclusion
The COR12 ECG is substantially equivalent to the predicate in all aspects of substantial equivalence comparison, as it is identical to the predicate, K150810. There are no differences.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.