QME · Software For Optical Camera-Based Measurement Of Pulse Rate, Heart Rate, Breathing Rate, And/Or Respiratory Rate

Cardiovascular · 21 CFR 870.2785 · Class 2

Overview

Product CodeQME
Device NameSoftware For Optical Camera-Based Measurement Of Pulse Rate, Heart Rate, Breathing Rate, And/Or Respiratory Rate
Regulation21 CFR 870.2785
Device ClassClass 2
Review PanelCardiovascular

Identification

The device uses software algorithms to analyze video signal and estimate pulse rate, heart rate, breathing rate, and/or respiratory rate. This device is not intended to independently direct therapy.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the software for optical camerabased measurement of pulse rate, heart rate, breathing rate, and/or respiratory rate is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) A software description and the results of verification and validation testing based on a comprehensive hazard analysis and risk assessment must include: (i) A full characterization of the software technical parameters, including algorithms; (ii) If required image acquisition hardware is not included with the device, full specifications of the hardware requirements and testing to demonstrate the specified hardware ensures adequate data for validated and accurate measurements; (iii) A description of the expected impact of all applicable sensor acquisition hardware characteristics and associated hardware specifications; (iv) A description of all mitigations for user error or failure of any subsystem components (including signal detection, signal analysis, data display, and storage) on output accuracy; and (v) Software documentation must include a cybersecurity vulnerability and management process to assure software functionality. (2) Clinical data must be provided. This assessment must fulfill the following: (i) The clinical data must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified. (ii) The assessment must demonstrate output consistency using the expected range of data sources and data quality encountered in the intended use population and environment. (iii) The assessment must compare device output with a clinically accurate patient-contacting relevant comparator device in an accurate and reproducible manner. (3) A human factors and usability engineering assessment must be provided that evaluates the risk of improper measurement. (4) Labeling must include: (i) A description of what the device measures and outputs to the user; (ii) Warnings identifying sensor acquisition factors or subject conditions or characteristics (garment types/textures, motion, etc.) that may impact measurement results; (iii) Guidance for interpretation of the measurements, including a statement that the output is adjunctive to other physical vital sign parameters and patient information; (iv) The expected performance of the device for all intended use populations and environments; and (v) Robust instructions to ensure correct system setup.

Recent Cleared Devices (12 of 12)

RecordDevice NameApplicantDecision DateDecision
K253650Anura Mobile Core SDK (M1)Nuralogix CorporationJun 11, 2026SESE
K260066PanopticAI Vital Signs (1.6.1-22)PanopticAI Technologies LimitedMay 21, 2026SESE
K251200Vital SignsOxehealth LimitedFeb 2, 2026SESE
K243687Vital SignsOxehealth LimitedAug 27, 2025SESE
K250078Informed Vital Core Application (IVC App) (v2.0.0.2.0.0)Mindset Medical, Inc.May 30, 2025SESE
K240890PanopticAI Vital SignsPanopticAI technologies LimitedDec 23, 2024SESE
K241633Informed Vital Core Application (IVC App)Mindset Medical, Inc.Nov 18, 2024SESE
K232804FibriCheckQompium NVJun 7, 2024SESE
K223622FaceHeart Vitals Software Development Kit (FH vitals SDK)Faceheart Corp.Sep 1, 2023SESE
K220899Oxehealth Vital SignsOxehealth LimitedApr 29, 2022SESE
K211906Vital SignsOxehealth LimitedJul 20, 2021SESE
DEN200019Oxehealth Vital SignsOxehealth LimitedMar 26, 2021DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...