Anura Mobile Core SDK (AMC-SDK) is designed to use Transdermal Optical Imaging (TOI) for non-invasive spot (discrete) measurement of pulse rate and breathing rate (chest wall movements) when the subject is relaxed, still, and seated upright. AMC-SDK can be used for general healthcare in adults 18 years of age or older, but is not designed to treat or diagnose patients. The measurement results should complement, but not replace, the user's professional medical care and/or medication. If abnormalities are detected during the measurement with the SDK, users are advised to consult a medical professional. AMC-SDK is not intended to be the sole method of checking the physical health of a subject. AMC-SDK is not designed to function as an apnea monitor or to detect cessation of breathing.
Device Story
Software development kit (SDK) for integration into mobile applications; uses Transdermal Optical Imaging (TOI) via smartphone cameras to capture facial and upper body video; extracts blood flow and chest wall movement signals; processes data to estimate pulse rate and breathing rate; output displayed to user; intended for use by patients or healthcare professionals in clinical or home settings; provides supplemental vital sign data for general healthcare; aids in identifying need for professional consultation; benefits include non-contact, convenient spot-checking of vitals.
Clinical Evidence
Clinical validation study with 57 subjects (18-74 years old) comparing AMC-SDK to GE Healthcare (Datex-Ohmeda) S5 Multi-parameter Monitor. Pulse rate RMSD was 2.5 bpm (objective < 3 bpm). Breathing rate RMSD was 2.3 brpm (objective < 2 brpm). Study population included diverse skin types, BMI, and comorbidities (COPD, CHF, hypertension, asthma, diabetes).
Technological Characteristics
Software as a Medical Device (SaMD); utilizes Transdermal Optical Imaging (TOI) via standard mobile device cameras; operates on iOS and Android platforms; requires ambient indoor lighting (>100 Lux); measures pulse rate (45-105 bpm) and breathing rate (8-39 brpm); signal processing and computational algorithms estimate vitals from video stream; 30-second measurement window.
Indications for Use
Indicated for non-invasive spot measurement of pulse rate and breathing rate in adults 18+ years old. Intended for use when the subject is relaxed, still, and seated upright. Not for use in patients requiring critical care or continuous monitoring. Not for apnea detection or as a sole method of health assessment.
Regulatory Classification
Identification
The device uses software algorithms to analyze video signal and estimate pulse rate, heart rate, breathing rate, and/or respiratory rate. This device is not intended to independently direct therapy.
Special Controls
In combination with the general controls of the FD&C Act, the software for optical camerabased measurement of pulse rate, heart rate, breathing rate, and/or respiratory rate is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) A software description and the results of verification and validation testing based on a comprehensive hazard analysis and risk assessment must include:
(i) A full characterization of the software technical parameters, including algorithms;
(ii) If required image acquisition hardware is not included with the device, full specifications of the hardware requirements and testing to demonstrate the specified hardware ensures adequate data for validated and accurate measurements;
(iii) A description of the expected impact of all applicable sensor acquisition hardware characteristics and associated hardware specifications;
(iv) A description of all mitigations for user error or failure of any subsystem components (including signal detection, signal analysis, data display, and storage) on output accuracy; and
(v) Software documentation must include a cybersecurity vulnerability and management process to assure software functionality.
(2) Clinical data must be provided. This assessment must fulfill the following:
(i) The clinical data must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified.
(ii) The assessment must demonstrate output consistency using the expected range of data sources and data quality encountered in the intended use population and environment.
(iii) The assessment must compare device output with a clinically accurate patient-contacting relevant comparator device in an accurate and reproducible manner.
(3) A human factors and usability engineering assessment must be provided that evaluates the risk of improper measurement.
(4) Labeling must include:
(i) A description of what the device measures and outputs to the user;
(ii) Warnings identifying sensor acquisition factors or subject conditions or characteristics (garment types/textures, motion, etc.) that may impact measurement results;
(iii) Guidance for interpretation of the measurements, including a statement that the output is adjunctive to other physical vital sign parameters and patient information;
(iv) The expected performance of the device for all intended use populations and environments; and
(v) Robust instructions to ensure correct system setup.
GE Healthcare (Datex-Ohmeda) S5 Multi-parameter Monitor
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 11, 2026
Nuralogix Corporation
Jessica Mavadia-Shukla
Director of Regulatory Strategy & Health Science
250 University Ave.
Suite 209
Toronto, ON M5H3E5
Canada
Re: K253650
Trade/Device Name: Anura Mobile Core SDK (M1)
Regulation Number: 21 CFR 870.2785
Regulation Name: Software For Optical Camera-Based Measurement Of Pulse Rate, Heart Rate,
Breathing Rate, And/Or Respiratory Rate
Regulatory Class: Class II
Product Code: QME, BZQ
Dated: June 10, 2026
Received: June 10, 2026
Dear Jessica Mavadia-Shukla:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253650 - Jessica Mavadia-Shukla
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253650 - Jessica Mavadia-Shukla
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**JENNIFER W. SHIH -S**
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K253650 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Anura Mobile Core SDK (M1) | | | |
| Please provide your Indications for Use below. | | | ? |
| Anura Mobile Core SDK (AMC-SDK) is designed to use Transdermal Optical Imaging (TOI) for non-invasive spot (discrete) measurement of pulse rate and breathing rate (chest wall movements) when the subject is relaxed, still, and seated upright. | | | |
| AMC-SDK can be used for general healthcare in adults 18 years of age or older, but is not designed to treat or diagnose patients. The measurement results should complement, but not replace, the user's professional medical care and/or medication. If abnormalities are detected during the measurement with the SDK, users are advised to consult a medical professional. | | | |
| AMC-SDK is not intended to be the sole method of checking the physical health of a subject. | | | |
| AMC-SDK is not designed to function as an apnea monitor or to detect cessation of breathing. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
K253650 510(k) Summary
# 510(k) Summary
[As per 21 CFR 807.92]
## Submitter's Information
| Applicant | NuraLogix Corporation |
| --- | --- |
| Address | 250 University Avenue, Suite 209 Toronto, Ontario M5H 3E5 Canada |
| Correspondent | Jessica Mavadia-Shukla, PhD Director of Regulatory Strategy & Health Sciences Phone: +1 (732)-688-7843 Email: compliance@nuralogix.ai |
| Date Prepared | June 9^{th}, 2026 |
## Device Names and Classification
### Proposed Device
| Device Propriety Name | Anura Mobile Core SDK |
| --- | --- |
| Common or Usual Name | Anura Mobile Core SDK or AMC-SDK |
| Classification Name | Pulse Rate and Breathing Rate Measuring Software |
| Primary Product Code | |
| Product Code | QME |
| Regulation Number | 21 CFR 870.2785 |
| Regulation Description | Software For Optical Camera-based Measurement of Pulse Rate, Heart Rate, Breathing Rate, And/Or Respiratory Rate |
| Regulation Class | Class II |
| Review Panel | Cardiovascular |
| Secondary Product Code | |
| Product Code | BZQ |
| Regulation Number | 21 CFR 868.2375 |
| Regulation Description | Breathing Frequency Monitor |
| Regulation Class | Class II |
| Review Panel | Anesthesiology |
### Predicate Devices
| Trade Name | FH vitals SDK | FH vitals SDK-RR |
| --- | --- | --- |
| Manufacturer | FaceHeart Corporation | FaceHeart Corporation |
| 510(k) Number | K223622 | K243966 |
| Date Cleared | September 1, 2023 | April 9, 2025 |
### Device Description
Anura Mobile Core SDK (AMC-SDK) is a software development kit designed to facilitate video-based, non-invasive, contactless measure of pulse rate and breathing rate. The software uses cameras to detect the user's face and upper body, extracts blood flow and movement information from image stream, and then with signal
NuraLogix
Page 1 of 8
{5}
## **K253650 510(K) Summary**
processing and computational algorithms to estimate the pulse rate and breathing rate, and returns the results of such measurements. The software is intended to be integrated into applications that are installed on commercial mobile devices (including smartphones) equipped with cameras and internet connectivity. It can be deployed on Android and iOS platforms.
It is the responsibility of the final device manufacturer to integrate AMC-SDK correctly and to obtain the necessary regulatory approval/clearance for the final device. It is also the responsibility of the final device manufacturer to ensure that their device or software application has the correct regulatory approval/clearance and meets the required standards for their intended use. Any regulatory clearances and related tests that AMC-SDK has undergone do not apply to devices or software integrating AMC-SDK. These processes and tests should be repeated at the system level.
### **Indications for Use**
Anura Mobile Core SDK (AMC-SDK) is designed to use Transdermal Optical Imaging (TOI) for non-invasive spot measurement of pulse rate and breathing rate (chest wall movement) when the subject is seated and still.
AMC-SDK can be used for general healthcare in adults 18 years of age or older, but is not designed to treat or diagnose patients. The measurement results provided by the SDK should complement, but not replace, the user's professional medical care and/or medication. If abnormalities are detected during the measurement with the SDK, users are advised to consult a medical professional.
AMC-SDK is not intended to be the sole method of checking the physical health of a subject.
AMC-SDK is not designed to function as an apnea monitor or to detect cessation of breathing.
### **Comparison with Predicate Devices (Substantial Equivalence)**
The subject device AMC-SDK is substantially equivalent to the predicate devices with respect to indications for use, measurement method, and performance claims. The subject device and the predicate devices all employ contactless remote measuring of patient vital information. The primary difference between the subject and predicate devices is the hardware each system is compatible with. Detailed comparison information is provided below.
---NuraLogix
Page 2 of 8
{6}
K253650 510(K) Summary
| Table 3. Comparison of Anura Mobile Core SDK with Predicate Devices | | | | |
| --- | --- | --- | --- | --- |
| | Subject Device Anura Mobile Core SDK | Primary Predicate FH vitals SDK | Secondary Predicate FH vitals SDK | Comparison |
| 510(k) # | K253650 | K223622 | K243966 | N/A |
| Indications for Use | Anura Mobile Core SDK (AMC-SDK) uses Transdermal Optical Imaging (TOI) for non-invasive spot (discrete) measurement of pulse rate and breathing rate (chest wall movements) when the subject is relaxed, still, and seated upright. AMC-SDK can be used for general healthcare in adults 18 years of age or older, but is not designed to treat or diagnose patients. The measurement results should complement, but not replace, the user's professional medical care and/or medication. If abnormalities are detected during the measurement, users are advised to consult a medical professional. AMC-SDK is not intended to be the primary or sole method of checking the physical health of a subject. AMC-SDK is not designed to function as an apnea monitor or to detect cessation of breathing. | The FH Vitals SDK is designed to measure the pulse rate based on the given facial video stream. It is intended for non-invasive spot measurements of pulse rate when the subject is still. This SDK is not intended for use in patients with known or suspected heart arrhythmias. While the SDK can be used for general healthcare, it is not designed to treat patients. The pulse rate measurement results provided by the FH Vitals SDK should complement, but not replace, the user's usual professional medical care and/or medication. If abnormalities are detected during the measurement with the FH Vitals SDK, users are advised to consult a medical professional. The FH Vitals SDK is indicated for use in humans 18 years of age or older who do not require critical care or continuous vital signs monitoring. This software should not be the primary or sole method for assessing an individual's health. | FH Vitals SDK-RR is a software-only respiratory rate measurement tool intended to be integrated into third-party software applications on compatible mobile devices, laptops, or computers. It is intended for spot checking of Respiration Rate (RR) in an automatic contactless manner by analyzing chest wall movement from video input when the subject is still and properly positioned in front of the camera in a well-lit environment. FH Vitals SDK-RR is intended for use under the supervision of a healthcare professional, either in clinical or home environments, and is not intended for continuous monitoring, apnea detection, or as the sole method for evaluating physical health. It is intended to serve as a supplemental tool to assist in the overall assessment of the patient. The software is indicated for use on individuals aged 18 years and older who do not require critical care or continuous vital sign monitoring. | Similar – AMC-SDK, and FH vitals SDK are non-contacting, non-invasive spot checking devices for pulse rate (FH vitals) and breathing rate (chest wall movements). AMC-SDK includes both pulse rate and respiration rate features which are separately cleared via 2 510(K)s for FH Vitals in K223622 and K243966. |
| Device Type | Software as a Medical Device (SaMD) | SaMD | SaMD | Identical |
| Type of Use | Prescription Use | Prescription Use | Prescription Use | Identical |
| Use Population | Adults not requiring critical care | Adults not requiring critical care | Adults ≥18, not requiring critical care | Identical |
| Use Environment | Hospitals, clinics, long-term care facilities, and telemedicine-based home use | Unknown, presumably telemedicine-based home use | Clinical or home environments | Similar – comparable settings across devices |
| Monitoring | Spot-checking | Spot-checking | Spot-checking | Identical |
| Measurement Site | Pulse rate: surface of facial skin Breathing rate: upper body | Pulse rate: surface of facial skin Breathing rate: N/A | Chest wall movement | Identical (Pulse Rate) – AMC-SDK and FH vitals SDK (K223622) take measurements on the face Identical (Breathing Rate) – AMC-SDK and FH Vitals-RR (K243966) both take measurements of the upper body (including the chest). |
| Measuring Distance | Measuring between 30 to 91 cm (or 12 to 36 inches) camera distance from face | Measuring distance of 50 to 150 cm (or 0.5 to 1.5m) depends on camera | Mobile/PC-based use; 0.5–1.5m distance, indoor use with general restrictions | Similar – All devices measure vitals remotely; AMC-SDK operates at a closer distance than FH vitals |
| Environmental Conditions | Lighting: ambient indoor lighting, generally above 100 Lux | Illumination above 300 Lux depends on camera | 300 lux+ | Similar lighting to FH vitals |
| Operating Restriction | • Required to stay steady during measurement • Support minor movement such as breathing, as long as the signal passes quality checks • Required to be seated | • Required to stay steady during measurement • Support minor movement (speak or nod slightly) | • Designed for still subjects | Identical |
NuraLogix
Page 3 of 8
{7}
K253650 510(K) Summary
Table 3. Comparison of Anura Mobile Core SDK with Predicate Devices
| | Subject Device Anura Mobile Core SDK | Primary Predicate FH vitals SDK | Secondary Predicate FH vitals SDK | Comparison |
| --- | --- | --- | --- | --- |
| Physiological parameters measured | Pulse rate and breathing rate (chest wall movements) | Pulse rate | Breathing rate (chest wall movements) | Identical |
| Pulse Rate Measurement Range | 45-105 beats per minute (bpm) | 50-180 bpm | N/A | Similar – AMC-SDK measures a slightly narrower range than FH Vitals based upon the clinical data that supports AMC-SDK. |
| Pulse Rate Performance (Error Level) | ± 3 bpm | ± 3 bpm | N/A | Identical |
| Breathing Rate Measurement Range | 8-39 breaths per minute (brpm) | N/A – Does not measure breathing rate | 5-36 bpm | Similar. |
| Breathing Rate Performance (Error Level) | ± 2.3 brpm | N/A – Does not measure breathing rate | ± 2 brpm | Similar – both device measures accurately ≤ 2.5 brpm |
| Measurement Window | 30 seconds | 60 seconds (1 minute) | 60 seconds (1 minute) | Similar – AMC-SDK takes a shorter time to measure than FH vitals |
| Compatibility with Hardware – Computing | AMC-SDK is a software product that exclusively supports the following systems: • Apple mobile devices with ARMv8-A CPU architecture, Apple A10 or more performant CPU and Apple iOS 18 or newer compatible OS • Android mobile devices with ARMv8-A CPU architecture, Qualcomm Snapdragon 855 or more performant CPU and Android 13 or newer compatible OS | The pulse rate measurement feature of FH Vitals SDK is a software product that exclusively supports the following three operating systems: • Windows 10 • Android 12 • iOS 12 | This SDK is strictly a software-only product and is compatible with common operating systems, including Windows 10, Android 14, and iOS 17. | Similar |
| Compatibility with Hardware – Camera | Item iPhone 13 iPhone XR Samsung S23 | Item C920+ Laptop iPhone 13 Pro iPhone 13 Pro Max Samsung S22 Ultra Samsung S22+ | It has been tested and is compatible with the following cameras: • Logitech C930 webcam • Samsung S24+, S24+ smartphone front-facing cameras • iPhone 15 Pro, and iPhone 15 Pro Max front-facing cameras | Similar to FH vitals, where AMC-SDK uses camera on mobile phones |
| | Camera Aperture Size f/2.2 f/2.2 f/2.2 | Camera FOV 78° 77° 80° 80° | | |
| | Streaming 30 fps | Camera Sensor Size 1/3" 1/3.6" ½.82" 1/3.24" | | |
| | | Camera Aperture Size F2.0 F2.2 F2.2 F2.2 | | |
| | | Streaming VGA @ 30 fps | | |
| Compatibility with Hardware – Accessories | Item iPhone 13 iPhone XR Samsung S23 | Item C920+ Laptop iPhone 13 Pro iPhone 13 Pro Max Samsung S22 Ultra Samsung S22+ | | Similar to FH vitals, where AMC-SDK can be installed on off the shelf mobile phones |
| | CPU Apple A15 Bionic (5 nm) Apple A12 Bionic (7 nm) Qualcomm SM8550 AC Snapdragon 8 Gen 2 | CPU Intel i7-1160G7 A15 1 x Cortex X2 3 x Cortex A710 4 x Cortex A510 Adreno 730 | | |
| | GPU Apple GPU (4-core graphics) Adreno 740 | GPU Intel Iris Xe Graphics | | |
| | OS iOS 17 iOS 15+ Android 14 | | | |
| | Memory 4GB 3GB 8GB | | | |
NuraLogix
Page 4 of 8
{8}
K253650 510(K) Summary
Table 3. Comparison of Anura Mobile Core SDK with Predicate Devices
| | Subject Device Anura Mobile Core SDK | Primary Predicate FH vitals SDK | | | | Secondary Predicate FH vitals SDK | Comparison |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | OS | Windows 10 | iOS 12 | Android 12 | | |
| | | Memory Required | 500 MB and above | | | | |
| **Software Language** | Python, Go, SQL, TypeScript, Node.js, ReactJS, Java, Kotlin, Swift, Objective-C, C++, C; use of 3^{rd} party libraries | C++/Java/Objective-C++ | | | | | Different |
NuraLogix
Page 5 of 8
{9}
K253650 510(K) Summary
# **Non-Clinical/Bench Testing**
Non-clinical performance testing was conducted for Anura Mobile Core SDK (AMC-SDK) to ensure device performance across the intended range of pulse rate and breathing rate. The testing consists of software verification and validation, Human Factors (Usability) testing, and Simulation Testing of Pulse Rate.
In addition, AMC-SDK was tested in accordance with the special controls for optical camera-based measurement of pulse rate, heart rate, breathing rate, and/or respiratory rate (21 CFR 870.2785). The testing demonstrated that the technological and performance criteria were comparable to the predicate devices.
The software's compatibility was verified on specific devices, ensuring that the final application functions correctly. The tested devices are as follows:
- Samsung S23
# Human Factors (Usability) Testing
A human factors summative study was conducted on 30 participates, including 15 lay users (patients), and 15 healthcare professionals (8 doctors and 7 nurses). The results demonstrate that AMC-SDK controls use-related risks for the intended users, uses, and use environments.
# **Clinical Testing**
Clinical performance of Anura Mobile Core SDK (AMC-SDK) was tested to support the safety and performance of the device, and to ensure device performance and device accuracy across the intended range of pulse rate and breathing rate. A clinical validation study was conducted with 57 subjects ranged from 18 to 74 years of age.
| Clinical Validation Study Subject Demographics | | |
| --- | --- | --- |
| | Characteristic | Count |
| **Sample Size** | | 57 |
| **Sex at Birth** | Male | 30 |
| | Female | 27 |
| **Age Group** | ≤ 50 | 36 |
| | > 50 | 21 |
| **Body Mass Index (BMI) (kg/m²)** | Normal (< 25.0) | 16 |
| | Overweight (25.0-29.9) | 24 |
| | Obese (> 30.0) | 17 |
| **Skin Type (Fitzpatrick Scale)** | Light (Type I & II) | 13 |
| | Medium (Type III & IV) | 25 |
| | Dark (Type IV & V) | 19 |
| **Race & Ethnicity** | White | 26 |
| | Black/African American | 6 |
| | Asian | 5 |
| | Hispanic (Ethnicity) | 7 |
| | Other | 4 |
| | Not Reported | 16 |
| **Smoker** | Yes | 9 |
| | No | 48 |
NuraLogix
Page 6 of 8
{10}
K253650 510(K) Summary
| Clinical Validation Study Subject Demographics | | |
| --- | --- | --- |
| Characteristic | | Count |
| **Disease History / Comorbidity** | Chronic obstructive pulmonary disease (COPD) | 4 |
| | Congestive Heart Failure (CHF) | 3 |
| | Hypertension | 10 |
| | Asthma | 11 |
| | Diabetes | 4 |
The study evaluated the AMC-SDK device performance measuring pulse and breathing rate under expected conditions. Pulse and breathing rates were compared to the standard contact device, GE Healthcare (Datex-Ohmeda) S5 Multi-parameter Monitor, as the reference device.
The primary objective of the study was to assess the accuracy of pulse rate and estimated breathing rate (upper body movements) measurements made using AMC-SDK, calculated as the root mean square difference (RMSD) across all measurements when compared to the reference device (summarized in Table 2). The pulse rate RMSD was found to be 2.5 beats per minute (bpm), below the study objective of 3 bpm. The breathing rate RMSD was found to be 2.3 breaths per minutes (brpm), slightly above the study objective of 2 brpm.
#### Consideration of Special Controls Guidance
In combination with the general controls of the FD&C Act, AMC-SDK was also tested and demonstrated compliance in accordance with the following special controls per 21 CFR 870.2785 for the software for optical camera-based measurement of pulse rate, heart rate, breathing rate, and/or respiratory rate.
| Special Controls | Evidence of Compliance |
| --- | --- |
| (1) A software description and the results of verification and validation testing based on a comprehensive hazard analysis and risk assessment must include: - (i) A full characterization of the software technical parameters, including algorithms; - (ii) If required image acquisition hardware is not included with the device, full specifications of the hardware requirements and testing to demonstrate the specified hardware ensures adequate data for validated and accurate measurements. - (iii) A description of the expected impact of all applicable sensor acquisition hardware characteristics and associated hardware specifications; - (iv) A description of all mitigations for user error or failure of any subsystem components (including signal detection, signal analysis, data display, and storage) on output accuracy; and - (v) Software documentation must include a cybersecurity vulnerability and management process to assure software functionality. | Software documentation included a description of the software technical parameters, including algorithm, specifications of compatible hardware and the impact of these specifications on the device, as well as testing of all mitigations for user error. All system components were assessed for potential cybersecurity vulnerabilities. Additional third-party penetration testing was conducted to identify any cybersecurity vulnerabilities. Any detected vulnerabilities were addressed by design and/or testing. Information regarding cybersecurity vulnerability and management process are provided in the Cybersecurity section. |
NuraLogix
Page 7 of 8
{11}
K253650 510(K) Summary
| Special Controls | Evidence of Compliance |
| --- | --- |
| (2) Clinical data must be provided. This assessment must fulfill the following: (i) The clinical data must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified. (ii) The assessment must demonstrate output consistency using the expected range of data sources and data quality encountered in the intended use population and environment. (iii) The assessment must compare device output with a clinically accurate patient-contacting relevant comparator device in an accurate and reproducible manner. | A clinical validation study was conducted on 57 subjects (41 CRO study participants, 16 internal study participants) representative of the intended population and demonstrated the expected range and accuracy of AMC-SDK as compared to the clinically accurate patient-contacting comparator devices. Details of the clinical validation study is provided in the Clinical Testing section. |
| (3) A human factors and usability engineering assessment must be provided that evaluates the risk of improper measurement. | A human factors (usability) summative study was conducted with 30 subjects: 15 healthcare professionals and 15 patients that showed minimal user errors, close calls, or use difficulties. |
| (4) Labeling must include: (i) A description of what the device measures and outputs to the user; (ii) Warnings identifying sensor acquisition factors or subject conditions or characteristics (garment types/textures, motion, etc.) that may impact measurement results; (iii) Guidance for interpretation of the measurements, including a statement that the output is adjunctive to other physical vital sign parameters and patient information; (iv) The expected performance of the device for all intended use populations and environments; and (v) Robust instructions to ensure correct system setup. | Labelling requirements are incorporated in the labelling of AMC-SDK. |
### Conclusion
Anura Mobile Core SDK is shown through comparison and performance testing to be substantially equivalent to the identified predicate devices. Specifically, the subject device has the same intended use and technological characteristics as the predicate devices. Hence, Anura Mobile Core SDK is substantially equivalent to its predicate devices.
NuraLogix
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.