AS/3 LIGHT MONITOR

K961247 · Datex Division Instrumentarium Corp. · DPS · Oct 23, 1996 · Cardiovascular

Device Facts

Record IDK961247
Device NameAS/3 LIGHT MONITOR
ApplicantDatex Division Instrumentarium Corp.
Product CodeDPS · Cardiovascular
Decision DateOct 23, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2

Indications for Use

AS/3 Light Monitor is intended for stationary and mobile indoor monitoring of a patient’s vital signs.

Device Story

Portable patient monitor; measures ECG, invasive blood pressure, SpO2, temperature, and NIBP. Configurations: LM (ECG, SpO2, 1 temp, NIBP) or LMP (ECG, SpO2, 2 invasive pressures, 2 temps, NIBP). Operates on battery or mains power. Used in clinical settings for continuous vital sign monitoring. Provides real-time physiological data to clinicians to support patient assessment and clinical decision-making.

Clinical Evidence

Bench testing only. Device tested against established specifications of the predicate AS/3 Compact Monitor to confirm equivalent performance.

Technological Characteristics

Portable patient monitor; measures ECG, invasive/non-invasive blood pressure, SpO2, and temperature. Powered by battery or mains. Utilizes same measurement techniques and algorithms as predicate AS/3 Compact Monitor.

Indications for Use

Indicated for patients requiring stationary or mobile indoor monitoring of vital signs, including ECG, invasive blood pressure, SpO2, temperature, and non-invasive blood pressure (NIBP).

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Submission Summary (Full Text)

{1} Datex K961247 510(k) SUMMARY OCT 23 1996 27 March, 1996 Datex-Engström Division Instrumentarium Corp. P.O. Box 446 FIN-00101 Helsinki Finland Tel: +358 0 39411 Fax: +358 0 1463310 Contact person: Rauno Ruoho ## PRODUCT NAMES Device name: AS/3™ Light Monitor Common name: Hemodynamic measurement device ## DESCRIPTION OF THE DEVICE AS/3 Light Monitor is a portable patient monitor which measures ECG, 2 invasive blood pressures, oxygen saturation SpO₂, 2 temperatures and non-invasive blood pressure NIBP. The measurement technics are similar to the AS/3 Compact Monitor with ESTP module and NIBP module. See section 3 for details. AS/3 Light Monitor will be available in two following configurations: - LM which measures ECG, SpO₂, 1 temperature and NIBP - LMP which measures ECG, SpO₂, 2 invasive blood pressures, 2 temperatures and NIBP. The monitor can be driven with battery or with mains power. ## INTENDED USE AS/3 Light Monitor is intended for stationary and mobile indoor monitoring of a patient’s vital signs. ## SUBSTANTIAL EQUIVALENCE In terms of overall functions and parameters measured, the AS/3™ Light Monitor is substantially equivalent in safety and effectiveness to Datex AS/3™ Compact Monitor with ESTP and NIBP modules. {2} Measurement parameters are the same as in Datex AS/3™ Compact Monitor with ESTP and NIBP modules. Measurement units are redesigned, but measurement technics and algorithms are the same. The specifications of the monitors were compared and AS/3™ Light Monitor was tested against the specifications. The results showed that AS/3™ Light Monitor is substantially equivalent with AS/3™ Compact Monitor.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...