ORAGENE-DX OGD-500; ORAGENE-DX OGD-575; ORAGENE-DX OXD-525; ORAGENE-DX OYD-500

K110701 · DNA Genotek, Inc. · OYJ · Dec 2, 2011 · Clinical Chemistry

Device Facts

Record IDK110701
Device NameORAGENE-DX OGD-500; ORAGENE-DX OGD-575; ORAGENE-DX OXD-525; ORAGENE-DX OYD-500
ApplicantDNA Genotek, Inc.
Product CodeOYJ · Clinical Chemistry
Decision DateDec 2, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Intended Use

Oragene•Dx is intended for use in the non-invasive collection of saliva samples. DNA from the saliva sample is isolated, stabilized, and suitable for use in FDA cleared molecular diagnostic applications. Saliva may be collected by spitting directly into the Oragene•Dx container or may be transferred into the Oragene•Dx container using a sponge. Saliva samples collected using Oragene•Dx are stabilized and can be transported and/or stored long term at ambient conditions.

Device Story

Oragene·Dx is a non-invasive saliva collection system comprising a tube, stabilizing liquid, and optional collection sponges. User spits or uses sponge to deposit saliva into tube; stabilizing liquid lyses cellular/nuclear membranes to release and preserve DNA. Used in clinical or home settings; operated by patients or healthcare providers. Stabilized samples are transported/stored at ambient conditions for up to 12 months. Healthcare providers extract DNA via alcohol precipitation for downstream molecular diagnostic testing. Device ensures sample integrity, enabling reliable genetic analysis and facilitating clinical decision-making based on molecular diagnostic results.

Clinical Evidence

No clinical trials; bench testing only. Reproducibility studies (n=15 donors, 42 tests) and sample volume tolerance studies (n=60 donors, 240 tests) evaluated DNA yield, concentration, and purity (A260/A280). Genotyping accuracy was validated using the eSensor Warfarin Sensitivity Saliva Test, achieving 99.8% agreement with bidirectional sequencing after re-testing. Stability studies confirmed DNA integrity for 12 months at ambient temperature for OGD-500/OGD-575/OYD-500 and 3 months for OXD-525. Interfering substances (endogenous/exogenous) showed no impact on genotyping results.

Technological Characteristics

Plastic collection tube containing nucleic acid stabilization solution. Non-invasive collection via spitting or sponge. Manual operation. No energy source. Compatible with standard alcohol precipitation DNA extraction methods. Shelf-life up to 24 months at room temperature.

Indications for Use

Indicated for non-invasive collection of saliva samples for DNA isolation and stabilization in adults 18 years or older. For professional use only. Cleared for use with genotyping tests that have obtained FDA clearance for use with saliva samples obtained with these collection devices.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC - 2 2011 K110701 # 510(k) Summary This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1900 and CFR 807.92. | Date: | 30 November 2011 | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | DNA Genotek Inc.<br>2 Beaverbrook Road, Ottawa,<br>Ontario K2K 1L1 Canada | | Contact: | Chantal Hemens-Davis<br>Vice President Operations, Quality & Regulatory Affairs<br>(Official Correspondent)<br>Tel: (613) 723-5757 Ext. 238, Fax: (613) 368-4628<br>Email: chantal.hemens-davis@dnagenotek.com | | Device Proprietary<br>Names | Oragene®·Dx<br>Formats: OGD-500, OGD-575, OXD-525, OYD-500 | | Common names | Kit for collection of Human DNA, Saliva kit, Sample collection kit | | Regulatory Classification | Regulation: 21CFR 862.1675<br>Panel: Chemistry (75)<br>Classification: Class II<br>Product Code: OYJ DNA Specimen Collection, Saliva | | Predicate Device | Vacutainer® Brand PPT™ Plasma Preparation Tube (k972075) | ### DEVICE DESCRIPTION The Oragene·Dx family of products offers reliable collection, transportation and long-term room temperature storage of human DNA from saliva. Oragene Dx is a non-invasive alternative for collecting high quality and quantity DNA for use in molecular diagnostic applications. All formats consist of a collection tube, stabilizing liquid and optional sponges for assisted collection. After saliva is collected, the stabilizing liquid is mixed with the sample. Saliva can be delivered directly by spitting or using provided sponges to transfer saliva into the device. {1}------------------------------------------------ ### DNA Stabilization, Transportation and Storage Upon contacting saliva cells, the stabilizing liquid lyses cellular and nuclear membranes to release and stabilize nucleic acids (DNA). Samples can be immediately processed, transported or stored for future use. Oragene Dx samples are stable at room temperature for up to 12 months. Device and sample integrity are preserved during typical ambient transport and storage conditions. ### Sample Processing and Testing DNA extraction from Oragene-Dx can be performed using alcohol precipitation for the purpose of molecular diagnostic applications. ### Intended Use Oragene·Dx is intended for use in the non-invasive collection of saliva samples. DNA from the saliva sample is isolated, stabilized, and suitable for use in FDA cleared molecular diagnostic applications. Saliva may be collected by spitting directly into the Oragene-Dx container or may be transferred into the Oragene Dx container using a sponge. Saliva samples collected using Oragene-Dx are stabilized and can be transported and/or stored long term at ambient conditions. ### Indications for Use See "Intended Use", above. {2}------------------------------------------------ # COMPARISON TO MATERIAL AND TECHNOLOGICAL FEATURES OF THE PREDICATE DEVICE The following table outlines the similarities and differences between Vacutainer® Brand PPT™ Plasma Preparation Tube (predicate) and Oragene-Dx Collection devices. | Principle,<br>Materials and<br>Technology | Vacutainer® Brand PPT™ Plasma<br>Preparation Tube | Oragene-Dx | Similar | Different | |------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----------| | Description of<br>Intended Use | The Vacutainer® Brand PPT™<br>Plasma Preparation Tube with<br>EDTA anticoagulant and a gel<br>barrier material are evacuated<br>blood collection tubes which<br>provide a means of collecting,<br>processing and transporting blood<br>in a closed plastic tube. When the<br>Tube is used together with<br>Vacutainer® Brand Needles and<br>Holders, it is a closed system for<br>the collection of venous blood with<br>the same indications identified<br>here.<br><br>Blood collected in a tube<br>containing EDTA anticoagulant and<br>gel barrier material is used<br>primarily to provide undiluted<br>plasma for use in molecular<br>diagnostic test methods including<br>but not limited to PCR (Polymerase<br>Chain Reaction) and bDNA<br>(branched DNA). The specimen<br>may also be used for other testing<br>that requires and undiluted plasma<br>sample as determined by the<br>laboratory. | Oragene-Dx is intended for use in<br>the non-invasive collection of saliva<br>samples. DNA from the saliva<br>sample is isolated, stabilized, and<br>suitable for use in FDA cleared<br>molecular diagnostic applications.<br>Saliva may be collected by spitting<br>directly into the Oragene-Dx<br>container or may be transferred into<br>the Oragene-Dx container using a<br>sponge. Saliva samples collected<br>using Oragene-Dx are stabilized and<br>can be transported and/or stored<br>long term at ambient conditions. | X | | | Analyte | Various (including DNA) | DNA | X | | | Sample<br>collection | Biological samples collected<br>through venous puncture into a<br>sterile plastic evacuated<br>collection tube | Non-invasive collection of biological<br>samples delivered into a non-sterile<br>plastic collection tube | | X | | Formats | Multiple | Multiple | X | | | Tube material | Plastic | Plastic | X | | | Sample source | Venous blood | Human saliva | | X | | Additive | EDTA and gel barrier material | Nucleic acid stabilization solution | | X | | Principle,<br>Materials and<br>Technology | Vacutainer® Brand PPT™ Plasma<br>Preparation Tube | Oragene-Dx | Similar | Different | | Transport and<br>Stability | Store unfilled tubes at 4-25°C.<br>Limited excursion temperatures up<br>to 40°C, for a cumulative time not<br>to exceed 10 days, are acceptable. | Pre-collection Oragene-Dx kits can be<br>transported at temperatures ranging<br>from -20°C to 50°C | | X | | | Whole blood may be stored in the<br>BD PPT™ Tube up to six (6) hours<br>prior to centrifugation. | Post-collection Oragene-Dx samples<br>can be transported at temperatures<br>ranging from -20°C to 50°C<br>Pre-collection Oragene-Dx kits can be<br>stored at room temperature for up<br>to 24 months<br>Post-collection Oragene-Dx samples<br>can be stored at room temperature<br>for up to 12 months (OGD-500, OGD-<br>575, OYD-500) and 3 months for<br>OXD-525 | | | | Suitability<br>for use with<br>molecular<br>diagnostic<br>applications | Used for testing plasma in<br>molecular diagnostics. | Stabilized DNA can be used in<br>molecular diagnostic testing. | X | | # Table 25 – Similarities between Predicate and Oragene·Dx {3}------------------------------------------------ # Substantial Equivalence Discussion The similarities in intended use, materials, technological characteristics show that Oragene Dx is substantially equivalent to Vacutainer® Brand PPT™ Plasma Preparation Tube. The differences tabulated above do not affect the safety and performance of Oragene Dx. Oragene Dx performance has been validated using GenMark Diagnostics' FDA cleared eSensor® Warfarin Sensitivity Saliva Test. ### PERFORMANCE CHARACTERISTICS # Performance Characteristics of Oragene · Dx by Format The DNA concentration and total sample DNA yield was determined by fluorescence and the A260/A280 ratio was determined by UV absorbance. These parameters were evaluated across all formats (OGD-500, OGD-575, OXD-525 and OYD-500). Data from a total of 450 samples from 245 unique donors is used in support of the performance characteristics for OGD-500. A subset of 45 donors is used in support of the performance characteristics for OXD-525 and OYD-500. A subset of 43 donors is used in support of the performance characteristics for OGD-575. All samples were extracted using an alcohol precipitation method. Downstream performance was demonstrated on the eSensor® Warfarin Sensitivity Saliva Test. {4}------------------------------------------------ | | | OGD-500 | OGD-575 | OXD-525 | OYD-500 | |-----------------------|----------------|---------------|---------------|---------------|---------------| | Samples Tested | | 450 | 43 | 45 | 45 | | Yield (µg) | Mean ± SD | 58.52 ± 47.02 | 13.50 ± 8.84 | 50.10 ± 42.38 | 56.05 ± 46.84 | | | Median | 48.44 | 10.96 | 33.35 | 37.28 | | | 95% of samples | ≥ 13.1 | ≥ 3.8 | ≥ 14.8 | ≥ 13.0 | | Concentration (ng/µL) | Mean ± SD | 68.11 ± 55.27 | 41.12 ± 24.59 | 88.89 ± 74.51 | 65.38 ± 55.94 | | | Median | 55.27 | 33.20 | 60.22 | 42.11 | | | 95% of samples | ≥ 16.0 | ≥ 11.2 | ≥ 27.6 | ≥ 14.8 | | A260/A280 ratio | Mean ± SD | 1.7 ± 0.1 | 1.7 ± 0.1 | 1.7 ± 0.1 | 1.7 ± 0.1 | | | Median | 1.7 | 1.7 | 1.7 | 1.7 | | | 98% of samples | 1.5 - 2.0 | 1.5 - 1.9 | 1.5 – 1.9 | 1.5 – 1.9 | Table 26 – Summary of DNA Yield Concentration and A260/A280 Ratio Results for Oragene-Dx Performance Characteristics | Table 27 - Summary of eSensor® Warfarin Sensitivity Saliva Test Results for Oragene-Dx Format | | |-----------------------------------------------------------------------------------------------|--| | Comparison | | | Format | SNP | Samples<br>tested | Correct<br>calls | Incorrect<br>calls | No-calls | % Correct<br>calls | |---------------|---------|-------------------|------------------|--------------------|----------|--------------------| | OGD-500 | 2C9*2 | 45 | 45 | 0 | 0 | 100% | | | 2C9*3 | 45 | 45 | 0 | 0 | 100% | | | VKORC1 | 45 | 45 | 0 | 0 | 100% | | OGD-575 | 2C9*2 | 43 | 40 | 0 | 3† | 93.0% | | | 2C9*3 | 43 | 40 | 0 | 3† | 93.0% | | | VKORC1 | 43 | 40 | 0 | 3† | 93.0% | | OXD-525 | 2C9*2 | 45 | 45 | 0 | 0 | 100% | | | 2C9*3 | 45 | 45 | 0 | 0 | 100% | | | VKORC1 | 45 | 45 | 0 | 0 | 100% | | OYD-500 | 2C9*2 | 45 | 45 | 0 | 0 | 100% | | | 2C9*3 | 45 | 45 | 0 | 0 | 100% | | | VKORC 1 | 45 | 45 | 0 | 0 | 100% | | After re-test | | | | | | | | OGD-575 | 2C9*2 | 43 | 43 | 0 | 0 | 100% | | | 2C9*3 | 43 | 43 | 0 | 0 | 100% | | | VKORC1 | 43 | 43 | 0 | 0 | 100% | † First-pass no-call results were due to eSensor® control failures. : . . {5}------------------------------------------------ #### SAMPLE VOLUME TOLERANCE A total of 240 samples were collected using OGD-500 with modified "fill to" lines (the volume of stabilizing liquid was kept constant at 2 mL in each of the devices) in order to simulate both under and over spitting. Collected sample volumes ranged from as low as 0.58 mL of saliva to as much as 3.64 mL of saliva with a median collection volume of 2 mL. As expected the DNA yield was dependent on collected volume and neither the A260/A280 ratio nor performance was affected by under or over spitting. All samples contained sufficient quality to be used on the eSensor® Warfarin Sensitivity Saliva Test. All 240 samples gave a correct call after re-testing. Image /page/5/Figure/3 description: The image is a histogram showing the number of samples versus volume in mL. The x-axis represents volume in mL, ranging from 0.0 to 4.0, while the y-axis represents the number of samples, ranging from 0 to 30. The histogram shows the distribution of samples across different volume ranges, with the highest number of samples occurring between 1.0 and 2.5 mL. Figure 1 -- Collected Saliva Volume in Sample Volume Tolerance Study # Table 28 - Summary of eSensor® Warfarin Sensitivity Saliva Test Results After Re-testing for Sample Volume Tolerance Study | Range of collected saliva<br>volume (mL) | Samples<br>tested | Correct<br>calls | Incorrect<br>calls | No-calls | % Correct<br>calls | |------------------------------------------|-------------------|------------------|--------------------|----------|--------------------| | 0.58 – 3.64 | 240 | 240 | 0 | 0 | 100% | " One first-pass no-call which was resolved upon re-testing. {6}------------------------------------------------ ### REPRODUCIBILITY The OGD-500 device reproducibility study was conducted at three sites (two external and one internal). Three samples (collected using three lots of OGD-500) from each of ten donors, covering all possible genotypes for three alleles for the eSensor® Warfarin Sensitivity Saliva Test, were tested in triplicate by four different operators at the three sites. Each operator extracted DNA from each sample using the same alcohol precipitation method, followed by determination of DNA concentration and A260/A280 ratio for all samples by an independent operator at one of the sites. Four operators at three sites tested the extracted DNA samples on the eSensor® Warfarin Sensitivity Saliva Test. The following table summarizes the yield, DNA concentration and A260/A280 ratio results obtained by four operators at the three sites. | | | Operator 1 | Operator 2 | Operator 3 | Operator 4 | Combined | |--------------------------|----------------|------------------|------------------|------------------|------------------|------------------| | Samples tested | | 87* | 87* | 90 | 90 | 354 | | Yield (µg) | Mean ± SD | 74.89 ±<br>68.00 | 76.68 ±<br>61.76 | 69.59 ±<br>57.24 | 77.40 ±<br>68.36 | 74.62 ±<br>63.79 | | | Median | 57.31 | 66.32 | 60.95 | 57.62 | 60.39 | | | 95% of samples | ≥ 23.23 | ≥ 26.56 | ≥ 18.13 | ≥ 26.02 | ≥ 23.47 | | Concentration<br>(ng/µL) | Mean ± SD | 86.76 ±<br>84.43 | 87.84 ±<br>70.76 | 80.21 ±<br>67.78 | 90.20 ±<br>86.69 | 86.24 ±<br>77.61 | | | Median | 63.83 | 74.43 | 68.82 | 66.20 | 68.58 | | | 95% of samples | ≥ 25.87 | ≥ 29.58 | ≥ 20.42 | ≥ 28.98 | ≥ 26.74 | | A260/A280 | Mean ± SD | 1.9 ± 0.1 | 1.8 ± 0.1 | 1.9 ± 0.1 | 1.8 ± 0.1 | 1.9 ± 0.1 | | | Median | 1.9 | 1.8 | 1.9 | 1.8 | 1.9 | | | 95% of samples | 1.6 - 2.3 | 1.6 - 2.1 | 1.7 - 2.0 | 1.5 - 2.0 | 1.6 - 2.2 | ### Table 29 – Summary of Device Reproducibility DNA Concentration, Yield and A260/A280 Results * One sample [3 aliquots] from a donor was excluded due to failure to meet incoming study screening criteria. {7}------------------------------------------------ # Table 30 – Summary of eSensor® Warfarin Sensitivity Saliva Test Results for Device Reproducibility Study Stratified by Site and Operator ・ : | Site | Operator | SNP | Samples<br>tested | Correct<br>calls | Incorrect<br>calls | No-calls* | % Correct<br>calls | |--------|------------|-------|-------------------|------------------|--------------------|-----------|--------------------| | Site 1 | Operator 1 | 2C9*2 | 87 | 86 | 0 | 1 | 98.9% | | | | 2C9*3 | 87 | 86 | 0 | 1 | 98.9% | | | | VKOR | 87 | 86 | 0 | 1 | 98.9% | | Site 1 | Operator 2 | 2C9*2 | 87 | 86 | 0 | 1 | 98.9% | | | | 2C9*3 | 87 | 86 | 0 | 1 | 98.9% | | | | VKOR | 87 | 86 | 0 | 1 | 98.9% | | Site 2 | Operator 3 | 2C9*2 | 90 | 87 | 1+ | 2 | 96.7% | | | | 2C9*3 | 90 | 88 | 0 | 2 | 97.8% | | | | VKOR | 90 | 87 | 1+ | 2 | 96.7% | | Site 3 | Operator 4 | 2C9*2 | 90 | 43 | 0 | 47 | 47.8% | | | | 2C9*3 | 90 | 43 | 0 | 47 | 47.8% | | | | VKOR | 90 | 43 | 0 | 47 | 47.8% | : First-pass #### After re-testing and investigation | | Operator | SNP | Samples<br>tested | Correct<br>calls | Incorrect<br>'calls | No-calls | % Correct<br>calls | |--------|------------|-------|-------------------|------------------|---------------------|----------|--------------------| | Site 1 | Operator 1 | 2C9*2 | 87 | 87 | 0 | 0 | 100% | | | | 2C9*3 | 87 | 87 | 0 | 0 | 100% | | | | VKOR | 87 | 87 | 0 | 0 | 100% | | | Operator 2 | 2C9*2 | 87 | 87 | 0 | 0 | 100% | | | | 2C9*3 | 87 | 87 | 0 | 0 | 100% | | | | VKOR | 87 | 87 | 0 | 0 | 100% | | Site 2 | Operator 3 | 2C9*2 | 90 | 90 | 0 | 0 | 100% | | | | 2C9*3 | 90 | 90 | 0 | 0 | 100% | | | | VKOR | 90 | 90 | 0 | 0 | 100% | | Site 3 | Operator 4 | 2C9*2 | 90 | 90 | 0 | 0 | 100% | | | | 2C9*3 | 90 | 90 | 0 | 0 | 100% | | | | VKOR | 90 | 90 | 0 | 0 | 100% | ² Incorrect call due to operator error resolved upon investigation. * 46 first-run no-calls were due to two runs (23 samples per run) invalidated due to DNA Contamination Monitor (DCM) failures. The other first-pass no calls were low signal for the 2C9*2 allele (three), positive control failure (one) and contractictory score at the 2C9*3 allele (one). {8}------------------------------------------------ # STABILITY Shelf-life conditions were evaluated by storing unused devices (OGD-500, OXD-525 and OYD-500) at room temperature, 6°C ± 4°C or -20°C ± 5°C for up to 24 months. OGD-575 was not tested as the chemistry is identical to the OGD-500 format. Devices were exposed to multiple freeze/thaw cycles of -20°C ± 5°C/50°C ± 5°C. In a separate study, devices were sent by standard mail to donors for collection and the donors mailed their samples directly to the processing laboratories. At all study time points a subset of devices were evaluated for physical and chemical properties to ensure the product specifications remained within acceptable tolerances. Another subset of devices was used to collect saliva from which DNA was extracted and analyzed for vield and A260/A280 ratio. The data supports shelf-life claims of room temperature storage for up to 24 months. Additionally, the devices may be transported by standard mail at temperatures ranging from -20°C ± 5°C. Post-collection sample stability was evaluated by having 30 donors each self-collect four saliva samples for each Oragene Dx format (OGD-500, OXD-525 and OYD-500), for a study total of 360 samples. Samples were stored at room temperature, 6℃ ± 4℃ or -20℃ ± 5℃ for 12 months or at 50℃ ± 5℃ for 3 months. At all study time-points, DNA was extracted and analyzed for yield and A260/A280 ratio. Samples stored at room temperature were analyzed for microbial content using a real-time PCR-based assay. A sub-population of samples (564 DNA samples across all time-points) were tested on the eSensor® Warfarin Sensitivity Saliva Test. The data supports post-collection storage of OGD-500, OGD-575 and OYD-500 at -20°C to room temperature for up to 12 months and at 50°C ± 5°C for 3 months. The data supports post-collection storage of OXD-525 at room temperature for up to 3 months. All formats stored at room temperature for 12 months exhibited no significant change in microbial content. ### INTERFERING SUBSTANCES Both endogenous (amylase, hemoglobin, lgA, total protein) and exogenous (eating, drinking, chewing gum, mouthwash, smoking) potentially interfering substances were added separately to samples from donors with known genotypes. Addition of tested substances had no effect as demonstrated through testing on the eSensor® Warfarin Sensitivity Saliva Test. All samples gave a correct call on first-pass. # CONCLUSION The submitted information in this premarket notification is complete and supports the safety and effectiveness of the Oragene Dx family of collection devices. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure. 10903 New Hampshire Avenue Silver Spring, MD 20993 DNA Genotek, Inc. c/o Chantal Hemens-Davis 2 Beaverbrook Road Ottawa. Ontario K2K 1L1 Canada DEC - 2 2011 Re: k110701 Trade Name: Oragene•Dx OGD-500, Oragene•Dx OGD-575, Oragene•Dx OXD-525, Oragene.Dx OYD-500 Regulation Number: 21 CFR §862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Codes: OYJ Dated: October 24, 2011 Received: October 25, 2011 Dear Ms. Hemens-Davis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {10}------------------------------------------------ Page 2 - If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Virro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Sincerely yours, signature Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {11}------------------------------------------------ ### Indications for Use Form 510(k) Number (if known): k110701 Device Name: Oragene®•Dx Collection Devices: OGD-500; OGD-575; OXD-525; OYD-500 Indications for Use: Oragene-Dx is intended for use in the non-invasive collection of saliva samples. DNA from the saliva sample is isolated, stabilized, and suitable for use in FDA cleared molecular diagnostic applications. Saliva may be collected by spitting directly into the Oragene-Dx container or may be transferred into the Oragene·Dx container using a sponge. Saliva samples collected using Oragene·Dx are stabilized and can be transported and/or stored long term at ambient conditions. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Signature Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K. 11 070
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...