LIQUICHEK URINALYSIS CONTROL (BI-LEVEL), CONTROL LEVEL1, LEVEL 2,MINIPAK, 435,436,437,435X
Applicant
Bio-Rad, Diagnostics Grp.
Product Code
JJW · Clinical Chemistry
Decision Date
Apr 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
Liquichek Urinalysis Control is intended for use as assayed quality control urine to monitor the precision of urinalysis test procedures for the analytes listed in the package insert.
Device Story
Liquichek Urinalysis Control is a liquid quality control material derived from human urine; supplemented with human electrolytes, simulated leukocytes, animal-derived constituents, chemicals, preservatives, and stabilizers. Used in clinical laboratories to monitor the precision of urinalysis test procedures (dipstick and microscopic). Healthcare providers use the control to verify the performance of urinalysis analyzers; results are compared against assigned target values and ranges provided by the manufacturer. The control helps ensure the accuracy and reliability of patient urinalysis results.
Clinical Evidence
No clinical data. Stability studies were performed to validate shelf life (30 months at 2-8°C) and open-vial stability (30 days at 18-25°C).
Technological Characteristics
Liquid matrix derived from human urine; contains human erythrocytes, simulated leukocytes, animal-derived constituents, pure chemicals, and 5-chloro-2-methyl-2H-isothiazol-3-one preservative. Designed for use with standard urinalysis dipstick and microscopic testing procedures.
Indications for Use
Indicated for use as an assayed quality control urine to monitor the precision of urinalysis test procedures in clinical laboratory settings. No specific patient population or contraindications apply as this is a quality control material.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
K965171 — LIQUICHEK URIANLYSIS CONTROL (CAT. NO. 420,421,422,425,426) · Bio-Rad · Jan 13, 1997
K072835 — LIQUICHEK MICROALBUMIN CONTROL LEVEL 1, LEVEL 2, CONTROL MINIPAK · Bio-Rad, Diagnostics Grp. · Dec 11, 2007
K971954 — LIQUICHEK URINE CHEMISTRY CONTROL (CAT.NO. 397,398) · Bio-Rad · Jun 11, 1997
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k070848
B. Purpose for Submission:
Premarket Notification 510(k) of intention to manufacture and market the Bio-Rad Liquichek Urinalysis Control
C. Measurand:
Not Applicable – Urinalysis control material
D. Type of Test:
Quality control material
E. Applicant:
Bio-Rad Laboratories
F. Proprietary and Established Names:
Bio-Rad Laboratories, Liquichek™ Urinalysis Control
G. Regulatory Information:
1. Regulation section:
21 CFR §862.1660 Quality control material (assayed and unassayed)
2. Classification:
Class I, (reserved)
3. Product code:
JJW
4. Panel:
{1}
75, Chemistry
## H. Intended Use:
1. Intended use(s):
See Indications for use below
2. Indication(s) for use:
Liquichek Urinalysis Control is intended for use as assayed quality control urine to monitor the precision of urinalysis test procedures for the analytes listed in the package insert.
3. Special conditions for use statement(s):
For prescription Use
4. Special instrument requirements:
The specific analyzers are listed in the package insert.
## I. Device Description:
The Bio-Rad Liquichek Urinalysis Control is prepared from human urine with added human electrolytes, simulated leukocytes, constituents of animal origin, pure chemicals, preservatives and stabilizers. The control is provided in liquid form for convenience and packaged as followed:
Bilevel 12x12 mL (6 per level)
Level 1 12x12 mL
Level 2 12x12 mL
Bilevel MiniPak 2x12 (1 per level)
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Bio-Rad Liquichek Urinalysis Control
2. Predicate 510(k) number(s):
k031231
{2}
3
3. Comparison with predicate:
The Bio-Rad Liquichek Urinalysis Control k070848 were compared to the previously cleared Bio-Rad Liquichek Urinalysis Control 510(k) k031231. The table below lists the similarities and differences between the Predicate and Proposed device.
| Characteristics | Bio-Rad Liquichek™ Urinalysis Control (New Device) k070848 | Bio-Rad Liquichek™ Urinalysis Control (Predicate Device) k031231 |
| --- | --- | --- |
| Similarities | | |
| Intended Use | Liquichek Urinalysis Control is intended for use as an assayed quality control urine to monitor the precision of urinalysis test procedures for the analytes listed in this package insert | Liquichek Urinalysis Control is intended for use as an assayed quality control urine to monitor the precision of laboratory dipstick and microscopic testing procedures for analytes listed in this package insert |
| Form | Liquid | Liquid |
| Matrix | Urine | Urine |
| Storage | 2° C to 8° C Until expiration date | 2° C to 8° C Until expiration date |
| Open Vial | 30 days at room temperature (18-25° C) | 30 days at room temperature (18-25° C) |
| Preservatives | 5-chloro-2-methyl-2H-isothiazol-3-one | 5-chloro-2-methyl-2H-isothiazol-3-one |
| Squeezeer Caps | Approved for use | Approved for use |
| Differences | | |
| Analytes | Same analytes as the predicate device with the additional claims for color and clarity | Does not have claims for color and clarity |
K. Standard/Guidance Document Referenced (if applicable):
None were referenced.
L. Test Principle:
Not applicable. This submission is for clearance of control material.
{3}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The Bio-Rad Liquichek Urinalysis Control is prepared from human urine with added human electrolytes, simulated leukocytes, constituents of animal origin, pure chemicals, preservatives and stabilizers. The manufacturer develops the target values and independent laboratories using the same protocol assay the material for each analyte/methodology for 20 times over a period of at least 10 days. The data are analyzed for each lot and ranges for each analyte/methodology are assigned by the manufacturer.
Stability studies have been performed to determine the open vial stability and shelf life for the Liquichek Urinalysis Control. Product claims are as follows:
Open vial: Once control is opened and stored tightly capped, all analytes will be stable for 30 days at room temperature (18-25° C).
Shelf life: 30 months when stored at (2-8° C).
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
{4}
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Specific ranges for each analyte/methodology are listed in the package insert. The manufacturer recommends in the labeling that each laboratory establish its own parameters of precision.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.