UA-Cellular® Complete is an assayed chemistry and cellular urine control for evaluating the accuracy and precision of automated procedures that measure urinary sediment and chemistry parameters on the Sysmex® UF-1000i™ Automated Urine Particle Analyzer and the Siemens® Clinitek Atlas Automated Urine Chemistry Analyzer utilizing the Clinitek Atlas 10 Reagent Pak. The list of assayed parameters includes: Sysmex UF-1000i:RBC(/μL), WBC//μL), Epithelial (/uL), Cast, Bacteria (/μL), Crystals, Conductivity (mS/cm) Siemens Clinitek Atlas with Atlas 10 Reagent Pak: Glucose (mg/dL), Bilirubin (As Measured), Ketones (mg/dL), Specific Gravity (As Measured), Blood (As Measured), pH (As Measured), Protein (mg/dL), Urobilinogen (EU/dL), Nitrite (As Measured), Leukocytes (As Measured), Color (As Measured), Clarity (As Measured)
Device Story
UA-Cellular Complete is a tri-level, in vitro diagnostic control material; contains stabilized human-source red/white blood cells, stabilized bacteria, and simulated urine sediments in a preservative medium. Used by clinical laboratory personnel to verify performance (calibration, precision, accuracy) of Sysmex UF-1000i and Siemens Clinitek Atlas analyzers prior to patient testing. Provides values within expected ranges for chemistry and cellular parameters; ensures instrument alignment. Packaged in 4 oz. amber plastic bottles. Human source materials tested non-reactive for HBsAg, HIV, HCV, and syphilis.
Clinical Evidence
No clinical data. Analytical performance verified via value assignment studies across three external sites and internal testing. Precision and stability (shelf-life and open-vial) verified using real-time studies on the Sysmex UF-1000i and Siemens Clinitek Atlas platforms. Assay data met ±3 Standard Deviation requirements.
Technological Characteristics
Tri-level liquid control; contains stabilized human-source RBCs, WBCs, bacteria, and simulated sediments. Preservative medium. Packaged in 4 oz. amber plastic bottles. Storage 2-10°C. Open-vial stability 30 days; shelf-life 60 days. Standard referenced: CLSI EP05-A2.
Indications for Use
Indicated for use as an assayed chemistry and cellular urine control to evaluate accuracy and precision of automated urinalysis procedures on Sysmex UF-1000i and Siemens Clinitek Atlas analyzers. Intended for clinical laboratory use.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
MAS UA Control (k023928)
UFII Control (k080887)
Related Devices
K170091 — UA-Cellular Complete · Streck · Feb 9, 2017
K090201 — UA-CELLULAR FOR IQ · Streck · Mar 27, 2009
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k131444
B. Purpose for Submission:
New device
C. Measurand:
Quality control material for the following parameters in urine samples: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, specific gravity, turbidity, color and microscopic determination of RBC, WBC, epithelial cells, casts, bacteria, crystals, and conductivity.
D. Type of Test:
Not applicable
E. Applicant:
Streck
F. Proprietary and Established Names:
UA-Cellular Complete
G. Regulatory Information:
1. Regulation section:
862.1660, Quality control material, (assayed, and unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJW
4. Panel:
Chemistry (75)
{1}
H. Intended Use:
1. Intended use(s):
See indications for use statements below.
2. Indication(s) for use:
UA-Cellular Complete is an assayed chemistry and cellular urine control for evaluating the accuracy and precision of automated procedures that measure urinary sediment and chemistry parameters on the Sysmex® UF-1000i™ Automated Urine Particle Analyzer and the Siemens® Clinitek Atlas Automated Urine Chemistry Analyzer utilizing the Clinitek Atlas 10 Reagent Pak.
The list of assayed parameters includes:
Sysmex UF-1000i: RBC(/μL), WBC(/μL), Epithelial (/μL), Cast, Bacteria (/μL), Crystals, Conductivity (mS/cm)
Siemens Clinitek Atlas with Atlas 10 Reagent Pak: Glucose (mg/dL), Bilirubin (As Measured), Ketones (mg/dL), Specific Gravity (As Measured), Blood (As Measured), pH (As Measured), Protein (mg/dL), Urobilinogen (EU/dL), Nitrite (As Measured), Leukocytes (As Measured), Color (As Measured), Clarity (As Measured)
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Sysmex® UF-1000i™ Automated Urine Particle Analyzer
Siemens® Clinitek Atlas Automated Urine Chemistry Analyzer
I. Device Description:
UA-Cellular Complete is an in vitro diagnostic product designed specifically for use with the Siemens Clinitek Atlas Automated Urine Chemistry Analyzer using the Clinitek Atlas 10 Reagent Pak and the Sysmex UF-1000i Automated Urine Particle Analyzer. UA-Cellular Complete is a tri-level control material which contains the following: stabilized mammalian red blood cells (human source) and white blood cells, stabilized bacteria, and simulated urine sediments in a preservative medium. Analyte levels are adjusted with appropriate chemicals. The product is packaged in 4 oz. amber plastic bottles with foil-lined flip-top-caps. All human source material used to manufacture this product was non-reactive for antigens to Hepatitis B (HBsAg), negative by tests for antibodies to HIV (HIV-1/HIV-s) and Hepatitis C (HCV), non-reactive for HIV-1 RNA and HCV RNA by licensed NAT, and non-reactive to Serological Test for Syphilis (STS) using techniques specified by the U.S. Food and Drug Administration.
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s):
MAS UA Control
Sysmex UF-1000i with Urinalysis WAM
2. Predicate 510(k) number(s):
k023928
k080887
3. Comparison with predicate:
| Similarities and Differences | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate 1 (k023928 – MAS UA Control) | Predicate 2 (k080887 – UFII Control) |
| Intended Use | Same | Same | Assayed control material for evaluating the accuracy and precision of automated procedures that measure urinary analytes |
| Measurands | Sysmex UF-1000i: RBC, WBC, Epithelial, Cast, Bacteria, Crystals, Conductivity
Siemens Clinitek Atlas with Atlas 10
Reagent Pak: Glucose Bilirubin, Ketones, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes, Color, Clarity | Glucose; Bilirubin,Ketones, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocyte, Creatinine, Color, Appearance, Crystals | RBC, WBC, Epithelial Cells, Cast, Bacteria, Conductivity |
| Open-Vial Stability | 30 days | 6 weeks at 18-25°C
3 months at 2-8°C | 30 days |
| Closed-Vial Stability | 60 days | 2 years | 6 months |
| Reagents | Stabilized mammalian red blood cells and white blood cells, stabilized | UA Control is a liquid stable control material prepared from human urine. Analyte levels | Latex Control Particles |
{3}
| Similarities and Differences | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate 1 (k023928 – MAS UA Control) | Predicate 2 (k080887 – UFII Control) |
| | bacteria, and simulate urine sediments in a preservative medium. Analyte levels are adjusted with appropriate chemicals. | are adjusted with various pure chemicals and human source materials. UA Control also contains preservatives and stabilizers. | |
| Storage Conditions | 2-10°C | 2-8°C | 2-10°C |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition
L. Test Principle:
UA-Cellular Complete is intended to be run by the clinical laboratory to collectively verify the performance attributes (i.e. calibration, precision and accuracy) of the intended instrument platforms, prior to running clinical patient specimens. The comprehensive tri-level UA-Cellular Complete control serves to align six stabilized urine particle types (i.e. red blood cells, white blood cells, cast, crystals, bacteria and epithelial) with the ten urine chemistry parameters.
Level 1-PC Control – Serves as the positive chemistry control for the urine chemistry parameters that do not align with a corresponding urine particle type (i.e. glucose, bilirubin, ketones and urobilinogen).
Level 1-PCM Control – Serves as the positive chemistry control for the urine chemistry parameters that serve to align with a corresponding urine particle type.
Level NC Control – Serves as the true “negative” control for all qualitative and semi-qualitative urine chemistry measurands and the quantitative urine particle types.
UA-Cellular Complete provides values within the expected ranges indicated on the assay sheet.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{4}
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Value assignment:
Value Assignment for the UA-Cellular® Complete parameters was based on data collected across three external sites and data collected internally at Streck. Data was collected across three separately manufactured lots. Each site involved in testing provided a 10-run reproducibility study for the tri-level control on each lot (n=40 per level). Four instruments of the Siemens® Clinitek Atlas and Sysmex® UF-1000i™ were utilized to collect data. Four operators were used in the value assignment study. Assay data collected met the sponsor’s ±3 Standard Deviation requirements.
Stability:
Shelf-Life and Open Vial Stability testing protocols and acceptance criteria were described and found to be adequate.
Shelf-Life – The 60-day claim was verified in real-time stability studies on the Sysmex® UF-1000i™ and Siemens® Clinitek Atlas with all three lots of device.
Open-vial – The open-vial stability claim was verified on both the Sysmex® UF-1000i™ Automated Urine Particle Analyzer and the Siemens® Clinitek Atlas Automated Urine Chemistry Analyzer using the Clinitek Atlas 10 Reagent Pak.
Open-vial stability testing was performed real-time and is 30 days.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
{5}
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.