CLINITEK STATUS+ANALYZER AND CLINITEK STATUS CONNECT SYSTEM, MODELS 1780, 1797
Applicant
Siemens Healthcare Diagnostics
Product Code
JIL · Clinical Chemistry
Decision Date
Oct 9, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1340
Device Class
Class 2
Intended Use
The Clinitek Status®+ Urine Chemistry Analyzer is a portable analyzer. It is designed to read only Siemens Reagent Strips for Urinalysis and Clinitest® hCG tests. This analyzer is intended for the measurement of the following in urine: Albumin, Bilirubin, Blood (Occult), Creatinine, Glucose, Ketone, Leukocytes, Nitrite, pH, Protein, Protein-to-Creatinine Ratio, Albumin-to-Creatinine Ratio, Specific Gravity, Urobilinogen, and human Chorionic Gonadotropin (hCG). These measurements are used to assist diagnosis in the following areas: - Kidney Function - Urinary tract infections - Metabolic disorders (e.g. diabetes mellitus) - Liver Function - Pregnancy Tests performed using the Clinitek Status®+ Analyzer are intended for in vitro diagnostic use only. The Clinitek Status®+ Analyzer is intended for near patient (point-of-care) facilities and centralized laboratory locations.
Device Story
Portable urine chemistry analyzer; reads Siemens Reagent Strips and Clinitest hCG tests. Optical system uses six LEDs, light guide, mirror, lens, and detector to measure reflected light from test strips/cassettes. Microprocessor converts light intensity into clinical results. Features include automatic strip identification, humidity-exposure quality check, user-selectable reporting, and access restrictions. Connect System variant adds barcode entry, QC management, and network connectivity. Used in point-of-care and lab settings by healthcare professionals. Provides objective, automated analysis to assist clinical diagnosis of metabolic, renal, and hepatic conditions, and pregnancy.
Clinical Evidence
Internal performance study evaluated 150 urine specimens comparing the modified device (Clinitek Status+) against the predicate (Clinitek Status). Testing covered urinalysis strips (Multistix 10SG) and hCG cassettes. Results demonstrated percent positive agreement (relative sensitivity) and percent negative agreement (relative specificity) met prescribed requirements. Software verification and validation plans, including risk management, were provided.
Technological Characteristics
Reflectance photometer; six LEDs, light guide, mirror, lens, detector. Portable form factor. Connectivity: optional LIS/HIS via network module. Software: embedded firmware for instrument control and data processing. Calibration: automated via white reflectance bar on test strip table. Sterilization: N/A (in vitro diagnostic instrument).
Indications for Use
Indicated for use in near-patient (point-of-care) and centralized laboratory settings to measure urine analytes (Albumin, Bilirubin, Blood, Creatinine, Glucose, Ketone, Leukocytes, Nitrite, pH, Protein, Protein/Creatinine ratio, Albumin/Creatinine ratio, Specific Gravity, Urobilinogen, and hCG) to assist in diagnosing kidney function, urinary tract infections, metabolic disorders (e.g., diabetes), liver function, and pregnancy.
Regulatory Classification
Identification
A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.
Predicate Devices
Clinitek Status Urine Chemistry Analyzer (K031947)
Clinitek Status Urine Chemistry Analyzer (K032563)
Related Devices
K142543 — Mission U120 Ultra Urine Analyzer, Mission Liquid Urine Controls, Mission Liquid Diptube Urine Control · ACON Laboratories, Inc. · May 11, 2015
K031947 — BAYER CLINITEK STATUS ANALYZER · Bayer Healthcare, LLC · Dec 19, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary - updated
10 91216
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
#### Submitter information
| Contact person: | Noor Malki<br>Head, Point of Care Regulatory Affairs | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------|--------------|
| Address: | Siemens Healthcare Diagnostics<br>Point of Care Products<br>2 Edgewater Drive<br>Norwood, MA 02062 | OCT - 9 2009 |
| Phone: | 781-269-3401 | |
| Date summary prepared: | September 25, 2009 | |
| Device Information<br>Proprietary Name: | Clinitek Status®+ Analyzer<br>Clinitek Status® Connect System | |
| Common name: | Urine Chemistry Analyzer | |
| Main classification name: | Automated Urinalysis System | |
| Main classification number: | 21 CFR 862.2900, Class I | |
Main classification panel: Clinical Chemistry and Clinical Toxicology
### Predicate Devices
| Element | Predicate Device |
|---------------|----------------------------------------|
| Device Name | Clinitek Status |
| Common name | Urine Chemistry Analyzer |
| 510(k) Number | K031947 and K032563 |
| Manufacturer | Siemens Healthcare Diagnostics Limited |
### Device Description
The Clinitek Status®+ Analyzer, a device modification to the current Clinitek Status®, is a portable easy to use urine chemistry analyzer. It is designed to read only Siemens Reagent Strips for Urinalysis and Clinitest® hCG tests.
Identical to the Clinitek Status Analyzer, the Clinitek Status+ analyzer is intended for the measurement of the following in urine: Albumin, Blood (Occult), Creatinine, Glucose, Ketone, Leukocytes, Nitrite, pH, Protein-to-Creatinine Ratio, Albumin
{1}------------------------------------------------
to-Creatinine Ratio, Specific Gravity, Urobilinogen, and human Chorionic Gonadotropin (hCG).
The enhanced Analyzer identifies automatically all instrument-read Siemens strip types and adds a quality check to assess if the strip has been compromised by humidity exposure.
In addition, the enhanced Analyzer provides reporting flexibility by allowing users to select specific test parameters to report from on a given strip and provides enhanced security to allow for user/operator access restrictions.
When the connector is attached to the enhanced analyzer (Clinitek Status+), the model is referred to as Clinitek Status Connect System. Additional functionality is enabled including bar code data entry, quality control management, and connectivity to hospital network.
## Statement of Intended Use
The Clinitek Status®+ Analyzer Urine Chemistry Analyzer is a portable easy to use analyzer. It is designed to read only Siemens Reagent Strips for Urinalysis and Clinitest® hCG tests.
This analyzer is intended for the measurement of the following in urine: Albumin, Bilirubin, Blood (Occult), Creatinine, Glucose, Ketone, Leukocvtes, Nitrite, pH, Protein, Protein-to-Creatinine Ratio, Albumin-to-Creatinine Ratio, Specific Gravity, Urobilinogen, and human Chorionic Gonadotropin (hCG).
These measurements are used to assist diagnosis in the following areas:
- . Kidney Function
- . Urinary tract infections
- . Metabolic disorders (e.g. diabetes mellitus)
- . Liver Function
- . Pregnancv
Tests performed using the Clinitek Status®+ Analyzer are intended for in vitro diagnostic use only.
The Clinitek Status®+ Analyzer is intended for near patient (point-of-care) facilities and centralized laboratory locations.
## Summary of Technological Characteristics
The Clinitek Status+ and Clinitek Status Connect System operating principle, technical platform, instrument and reagent analytical method, and intended use remain the same as the current Clinitek Status analyzer. In addition, it uses the same test strips and cassettes currently used on the Clinitek Status analyzer.
{2}------------------------------------------------
### 510(k) Summary - updated
The optical system consists of six light emitting diodes, a light quide, a mirror, a lens and a detector. Light from the LEDs travels along the light quide and is reflected off the calibration bar, strip or cassette onto the mirror. It is then directed through an aperture on the lens, from where it is focused onto the detector. The light intensity detected is converted into electrical impulses, which are processed by the instrument's microprocessor and converted into clinically meaningful results.
### Assessment of Performance
An internal study was conducted to demonstrate the performance of the Clinitek Status+ Analyzer and Clinitek Status Connect System (modified device) and assess its substantial equivalence against the Clinitek Status (predicate device). The scope of the study covered the impact of the proposed software enhancements on performance with both urinalysis strips (Multistix® 10SG) and hCG cassettes (Clinitest® hCG).
Over the study duration, 150 urine specimens were evaluated. The testing was performed to ensure that the software enhancement has not impacted current performance. Based on data analysis, the Urinalysis method comparison results for all 5 of the test conditions meet prescribed performance requirements for percent positive agreement (relative sensitivity) and percent negative agreement (relative specificity) relative to the predicate device (Clinitek Status). Also, performance for both hCG levels tested meet prescribed limits.
In addition information on Software Development Life Cycle including software requirements specifications, risk management report, and overall verification and validation plans were included to provide additional assurance of device performance.
#### Conclusion
In conclusion, the modified device (Clinitek Status+ and Clinitek Status Connect System) has the same Technological Characteristics and Intended Use as the predicate. The data presented is a summary of risk management activities, internal laboratory performance evaluation and software development information which all provide assurance that the modified device is substantially equivalent to the currently marketed Clinitek Status.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
Siemens Healthcare Diagnostics c/o Noor Malki Senior Manager, Regulatory Affairs (POC & Walpole sites) 2 Edgewater Drive, Norwood, MA 02062
OCT - 9 2009
Re: k091216
Trade/Device Name: Clinitek Status+Analyzer And Clinitek Status Connect System, Models 1780, 1797
Regulation Number: 21 CFR 862.1340
Regulation Name: Urinary glucose (nonquantitative) test system.
Regulatory Class: II
Product Code: JIL, JIP, JHI, JFY, JIR, JJB, JIN, LJX, JMT, CEN, JRE, CDM, KOO Dated: September 25, 2009
Received: October 1, 2009
Dear: Ms. Malki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
AC
Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510(k) Number (if known): K091216
Device Name: Clinitek Status®+ Analyzer
Clinitek Status® Connect System (consist of Clinitek Status®+ Analyzer with Connector attached).
Indication For Use:
The Clinitek Status®+ Urine Chemistry Analyzer is a portable easy to use analyzer. It is designed to read only Siemens Reagent Strips for Urinalysis and Clinitest® hCG tests.
This analyzer is intended for the measurement of the following in urine: Albumin, Bilirubin, Blood (Occult), Creatinine, Glucose, Ketone, Leukocytes, Nitrite, pH, Protein, Protein-to-Creatinine Ratio, Albumin-to-Creatinine Ratio, Specific Gravity, Urobilinogen, and human Chorionic Gonadotropin (hCG).
These measurements are used to assist diagnosis in the following areas:
- Kidney Function
Urinary tract infections
Metabolic disorders (e.g. diabetes mellitus) .
Liver Function
Pregnancy
Tests performed using the Clinitek Status®+ Analyzer are intended for in vitro diagnostic use only.
The Clinitek Status®+ Analyzer is intended for near patient (point-of-care) facilities and centralized laboratory locations.
Prescription Use X (21 CFR Part 801 Subpart D)
Over the
(21 CFR
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
And/Or
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Carol Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) R 091216
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.