DIASCREEN REAGENT STRIPS FOR URINALYSIS, MODELS D122OO (2GK), D12100 (1K), D11100 (1G)

K033851 · Hypoguard USA, Inc. · JIL · Mar 2, 2004 · Clinical Chemistry

Device Facts

Record IDK033851
Device NameDIASCREEN REAGENT STRIPS FOR URINALYSIS, MODELS D122OO (2GK), D12100 (1K), D11100 (1G)
ApplicantHypoguard USA, Inc.
Product CodeJIL · Clinical Chemistry
Decision DateMar 2, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1340
Device ClassClass 2

Intended Use

DiaScreen® Reagent Strips for Urinalysis are intended for the qualitative and semi-quantitative measurement of glucose and ketones in urine. They are intended for both professional and over-the-counter use.

Device Story

Plastic strips coated with dry reagents; used for qualitative/semi-quantitative detection of glucose and ketones in urine. Principle: chemical reaction between urine analyte and dry reagent causing visual color change. Operation: dip-and-read or midstream collection; visual comparison against color chart. Used in professional settings or home by patients. Output: color-coded concentration level. Clinical utility: aids diagnosis/management of diabetes, metabolic disorders, and ketosis; enables patient self-monitoring.

Clinical Evidence

Clinical data provided to support over-the-counter use. Study compared performance between home users and professional users, demonstrating that home users obtain accuracy substantially equivalent to professional users.

Technological Characteristics

Plastic strips with dry reagent pads. Sensing principle: colorimetric chemical reaction between reagent and specific analyte (glucose or ketone). Visual readout via color chart. No electronic components or software.

Indications for Use

Indicated for qualitative and semi-quantitative measurement of glucosuria and ketonuria in patients. Used for diagnosis and treatment of carbohydrate metabolism disorders (diabetes mellitus, hypoglycemia, hyperglycemia) and acidosis/ketosis, and for monitoring patients on ketogenic diets or with diabetes.

Regulatory Classification

Identification

A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR - 2 2004 ## 510(k) Summary | | 1. SUBMITTED BY: | Bruce A. MacFarlane, Ph.D.<br>5182 West 76th Street<br>Edina, MN 55439<br>USA<br>Summary prepared: 26 November 2003 | |--|----------------------------|---------------------------------------------------------------------------------------------------------------------| | | 2. NAME OF DEVICES: | | | | Trade Name: | DiaScreen® Reagent Strips for Urinalysis | | | Common Names/Descriptions: | Urine Test Strips | | | Classification Names: | Ketones (nonquantitative) test system<br>Urinary glucose (nonquantitative) test system | | | Regulatory Status: | Class II | | | 3. PREDICATE DEVICE: | DiaScreen® Reagent Strips for Urinalysis | - 3. PREDICATE DEVICE: - 4. DEVICE DESCRIPTION: DiaScreen® Reagent Strips for Urinalysis are plastic strips coated with one or more dry reagents. Each reagent is specific for a particular analyte. Reagent areas are spaced along the test strip, with a physical gap between reagent areas. Three configurations are designed for both professional and home use: detecting one or both of the following analytes: - · glucose - ketone . The dry reagents change color in reaction to the presence, and concentration, of their associated analyte. The strips are read visually against a color chart. ## 5. INTENDED USE: DiaScreen® Reagent Strips for Urinalysis are dip-and-read test strips intended for use as an in vitro diagnostic aid using urine specimens. DiaScreen Reagent Strips provide gualitative and semiquantitative tests for ketone, and glucose by visual comparison to a color chart. The tests provided are considered routine urinalysis. {1}------------------------------------------------ # 6. Summary of TECHNOLOGICAL CHARACTERISTICS DiaScreen® Reagent Strips for Urinalysis are plastic strips coated with one or more dry reagents. Each reagent is specific for a particular analyte. Reagent areas are spaced along the test strip, with a physical gap between reagent areas. Several product along the tool comp, man commercially available, detecting one or more of the following analytes: specific gravity, pH, blood, leukocytes, nitrite, protein, bilirubin, ketone (acetoacetic acid), glucose, and urobilinogen. The dry reagents change color in (acction to the presence, and concentration, of their associated analyte. The present premarket notification encompasses the ketone and glucose analytes only. ### 7. NON-CLINICAL TESTING Not applicable. #### 8. CLINICAL TESTING Clinical data indicate that home users can obtain accuracy which is substantially equivalent to that obtained by professional users. #### 9. CONCLUSIONS FROM TESTING Testing demonstrated that performance of the test strips with new labeling is appropriate for Over-the-Counter home usage. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 MAR - 2 2004 Bruce A. MacFarlane, Ph.D. VP, Regulatory Affairs And Quality Systems Hypoguard USA, Inc. 5182 West 76th Street Edina, MN 55439 Re: k033851 Trade/Device Name: Dia Screen® Reagent Strips for Urinalysis Regulation Number: 21 CFR 862.1340 Regulation Name: Urinary glucose (non-quantitative) test system Regulatory Class: Class II Product Code: JIL; JIN Dated: November 26, 2003 Received: December 17, 2003 Dear Dr. MacFarlane: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to conimer be prox & may 2011 and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may) arores or your softhe Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Codc of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I ederal the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ #### Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Jean M. Cooper, MS, DVM. Yean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K033851 Device Name: DiaScreen® Reagent Strips for Urinalysis Indications For Use: DiaScreen® Reagent Strips for Urinalysis are intended for the qualitative and semi-Diasoicon "Reagone Burgo 101 Original (glucose in urine). These measurements are qualitiative measurement of gracessaria (grace metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia. DiaScreen® Reagent Strips for Urinalysis are intended for the qualitative and semi-Diantitative measurement of ketones in urine. Identification of ketones is used in qualifiative incasurement of acidosis or ketosis and for monitoring patients on ketogenic diets and patients with diabetes. Testing can be done by dipping the Reagent Strips into a specimen container or by holding them directly in the urine stream. The glucose and ketone strips are intended for both professional and over-thecounter use. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Carol C Benson **Division Sign-Off** of In Vitro Dlag Page 1 of 1 510(k)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...