The BS-800M/ABS800/BA-800M Chemistry Analyzer is designed for clinical chemistry laboratory use, making direct quantitative measurements of Na+(sodium), K+(potassium), Cl- (chloride) in serum, plasma and urine samples, and Urea Nitrogen in serum samples. Additionally, other various chemistry tests may be adaptable to the analyzer depending on the reagent used to induce a photometric reaction. The BS-800M/ABS800/BA-800M ISE Kit is for the in vitro quantitative determination of Sodium (Na+), Potassium (K+), and Chloride (Cl-) concentrations in serum, plasma and urine samples on the The BS-800M/ABS800/BA-800M Chemistry Analyzer. Sodium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases involving electrolyte imbalance. Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of disease conditions characterized by, low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders. Urea Nitrogen (BUN) measurements are used to aid in the determination of liver and kidney function and other diseases associated with protein catabolism.
Device Story
Automated clinical chemistry analyzer; performs in vitro photometric assays and ion-selective electrode (ISE) measurements. Inputs: serum, plasma, or urine samples. Photometric unit uses tungsten halogen lamp, diffraction grating spectrophotometer, and photodiode detector to measure absorbance; ISE module uses ion-selective electrodes for electrolyte concentration. System is computer-controlled; features rack feeder, reagent/sample disks, and positive displacement pipettor. Used in clinical laboratories by trained personnel. Output: quantitative concentration results for Na+, K+, Cl-, and BUN. Results integrated into LIS; used by clinicians to monitor electrolyte balance, diagnose metabolic disorders, and assess organ function. Benefits: high-throughput automated testing, improved diagnostic efficiency.
Clinical Evidence
Bench testing only. Performance validated via precision (repeatability/total precision), linearity, limit of detection (LoB/LoD/LoQ), and analytical specificity (interference studies). Method comparison against predicate (N=120-124 per analyte) showed high correlation (r > 0.99). Matrix comparison (serum vs. lithium heparin plasma) confirmed equivalency.
Technological Characteristics
Automated chemistry analyzer; photometric and ISE sensing principles. Photometric unit: tungsten halogen lamp, diffraction grating, photodiode detector. ISE module: Na+, K+, Cl-, and reference electrodes. Throughput: 800 photometric tests/hr, 1200 tests/hr with ISE. Reaction unit: 165 glass non-disposable cuvettes, 37°C. Connectivity: barcode identification. Power: 110/115V. Operating environment: 15-30°C, 35-85% humidity.
Indications for Use
Indicated for clinical laboratory use for quantitative measurement of Na+, K+, Cl- in serum, plasma, and urine, and Urea Nitrogen in serum. Used to monitor electrolyte balance, diagnose/treat electrolyte/metabolic disorders, and assess liver/kidney function.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Predicate Devices
BS-200 Chemistry Analyzer with ISE Module (k072018)
Pointe Scientific BUN (Liquid) Reagent Set (k972671)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. ZHENG ZHE ENGINEER OF TECHNICAL REGULATION DEPARTMENT MINDRAY BUILDING, KEJI 12TH RD SOUTH, HI-TECH INDUSTRIAL PARK, NANSHAN, SHENZHEN, 518057 P.R. CHINA
June 9, 2016
Re: K160370
Trade/Device Name: BS-800M/ ABS800/BA-800M ISE KIT BS-800M Chemistry Analyzer BA-800M Chemistry Analyzer ABS-800 Chemistry Analyzer
Regulation Number: 21 CFR 862.1665 Regulation Name: Sodium Test System Regulatory Class: II Product Code: JGS, CGZ, CGZ, CEM, CDQ, JJE Dated: April 15, 2016 Received: May 2, 2016
Dear Zeng Zhe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{1}------------------------------------------------
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809). please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Katherine Serrano -S
For:
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K160370
#### Device Name
BS-800M Chemistry Analyzer, BA-800M Chemistry Analyzer, ABS800 Chemistry Analyzer, BS-800M/ABS800/BA-800M ISE Kit
#### Indications for Use (Describe)
The BS-800M/ABS800/BA-800M Chemistry Analyzer is designed for clinical chemistry laboratory use, making direct quantitative measurements of Na+(sodium), K+(potassium), Cl- (chloride) in serum, plasma and urine samples, and Urea Nitrogen in serum samples. Additionally, other various chemistry tests may be adaptable to the analyzer depending on the reagent used to induce a photometric reaction.
The BS-800M/ABS800/BA-800M ISE Kit is for the in vitro quantitative determination of Sodium (Na+), Potassium (K+), and Chloride (Cl-) concentrations in serum, plasma and urine samples on the The BS-800M/ABS800/BA-800M Chemistry Analyzer.
Sodium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases involving electrolyte imbalance.
Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of disease conditions characterized by, low or high blood potassium levels.
Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.
Urea Nitrogen (BUN) measurements are used to aid in the determination of liver and kidney function and other diseases associated with protein catabolism.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
The assigned 510(k) number is: K160370
#### Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
#### ● Contact Person:
#### Zeng Zhe
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
#### ● Date Prepared:
June 2, 2016
#### Name of the device:
- Trade/Proprietary Name:
BS-800M Chemistry Analyzer, BA-800M Chemistry Analyzer, ABS800 Chemistry Analyzer, BS-800M/ABS800/BA-800M ISE Kit (BS-800M, BA-800M and ABS800 are the same analyzers except the appearance, logo and name of the models. For convenience of explanation, the BS-800M Chemistry Analyzer is
represented of the three in this summary.)
- Common Name: Clinical Chemistry Analyzer (with optional ISE Module)
- Classification Number/Class: ●
75JJE, Class I 75CDQ, Class II
{4}------------------------------------------------
75CEM, Class II 75CGZ, Class II 75JGS, Class II
# Legally Marketed Predicate Device:
K072018
BS-200 Chemistry Analyzer, SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD
K972671 BUN (LIQUID) REAGENT SET, POINTE SCIENTIFIC, INC.
### Description:
The BS-800M/BA-800M/ABS800 Chemistry Analyzer is an automated clinical chemistry analyzer capable of performing various in vitro photometric assays. The BUN (LIQUID) REAGENT SET was cleared under K972671 and is the chosen assay to demonstrate performance for the photometric unit. The BS-800M Chemistry Analyzer has an optional Ion-Selective Electrode (ISE) module which measures the concentration of the electrolytes, sodium, potassium, and chloride, in samples using ion selective electrode technology.
### Intended Use/ Indication for Use:
The BS-800M/ABS800/BA-800M Chemistry Analyzer is designed for clinical chemistry laboratory use, making direct quantitative measurements of Na+(sodium), K+ (potassium), Cl- (chloride) in serum, plasma and urine samples, and Urea Nitrogen in serum samples. Additionally, other various chemistry tests may be adaptable to the analyzer depending on the reagent used to induce a photometric reaction.
The BS-800M/ABS800/BA-800M ISE Kit is for the in vitro quantitative determination of Sodium (Na+), Potassium (K+), and Chloride (Cl-) concentrations in serum, plasma and urine samples on the The BS-800M/ABS800/BA-800M Chemistry Analyzer.
Sodium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases involving electrolyte imbalance.
Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of disease conditions characterized by, low or high blood potassium levels.
Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.
Urea Nitrogen (BUN) measurements are used to aid in the determination of liver and kidney function and other diseases associated with protein catabolism.
{5}------------------------------------------------
# Comparison of Technological Characteristics:
Substantial equivalence has been demonstrated between the BS-800M Chemistry Analyzer and BS-200 Chemistry Analyzer. Both of them utilize absorbance photometry to perform and output quantitative results for kinetic and endpoint clinical chemistries. For analytes, BS-800M Chemistry Analyzer and BS-200 Chemistry Analyzer determine the concentration of unknown samples from a standard curve generated with known analyte concentrations. The BS-800M Chemistry Analyzer and BS-200 Chemistry Analyzer both utilize Ion-Selective Electrodes technology to measures the concentration of the electrolytes, sodium, potassium, and chloride in samples.
| Comparison<br>Section | BS-800M | BS-200<br>(Predicate Device) | |
|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| 510(K) Number | K160370 | K072018 | |
| Intended use | The BS-800M/ABS800/BA-800M<br>Chemistry Analyzer is designed for<br>clinical chemistry laboratory use,<br>making direct quantitative<br>measurements of Na+(sodium), K+<br>(potassium), Cl- (chloride) in serum,<br>plasma and urine samples, and Urea<br>Nitrogen in serum samples.<br>Additionally, other various<br>chemistry tests may be adaptable to<br>the analyzer depending on the<br>reagent used to induce a photometric<br>reaction. | The BS-200 Chemistry Analyzer is<br>designed for clinical laboratory use,<br>making direct quantitative<br>measurements of Na+ (sodium), K+<br>(potassium), Cl-(chloride) in serum,<br>plasma and urine samples and<br>Glucose in serum samples.<br>Additionally, other various<br>chemistry assays may be adaptable<br>to the analyzer depending on the<br>reagent used to induce a photometric<br>reaction. | |
| Parameter(photometric) | BUN | BUN | |
| Parameter(ion selective<br>electrode) | Na+, K+, Cl⁻ | Na+, K+, Cl⁻ | |
| Comparison on photometric assay Chart 1: BS-800M and BS-200 analyzer | | | |
| Feature | BS-800M | BS-200<br>(Predicate Device) | Same<br>(S)/Different<br>(D) |
| 510(K) | K160370 | K072018 | / |
# Comparison Table of BS-800M and Predicate Device:
{6}------------------------------------------------
| 1 | System Function | | | | |
|---|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|---|
| | System Control | Automatic, computer controlled | Automatic, microprocessor controlled | S | |
| | LIS external connectivity capability | Yes | Yes | S | |
| | Calibration/QC | Automatic and Manual calibration/QC | Automatic and Manual calibration/QC | S | |
| | Barcode | Yes | Yes | S | |
| 2 | Throughput (Max) | | | | |
| | 800 photometric tests per hour<br>1200 tests per hour with ISE | | 200 photometric tests per hour<br>330 tests per hour with ISE | | D |
| 3 | Configuration | | | | |
| | Analyzing unit, the Rack Feeder System, Operation unit, Output unit | | Analytical unit, Operation unit, Output unit | | D |
| 4 | Principle of Analysis | | | | |
| | Mode of detection | Photometric | Photometric | S | |
| | Analytical methods | Endpoint<br>Fixed-time<br>Kinetic | Endpoint<br>Fixed-time<br>Kinetic | S | |
| | Calibration methods | Linear calibration and nonlinear calibration | Linear calibration and nonlinear calibration | S | |
| 5 | Optical Measurement Unit | | | | |
| | Measurement Modes | Absorbance | Absorbance | S | |
| | Optical Modes | Monochromatic, Bichromatic | Monochromatic, Bichromatic | S | |
| | Photometer | Multi-wavelength, diffraction grating spectrophotometer | Multi-wavelength, Light transmission mode of the filter | D | |
| | Wavelength | 340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm and 800nm | 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm | D | |
| | Linear absorbance range | 0-3.4 absorbance | 0-4.0 absorbance | D | |
| | Light Source | Tungsten halogen lamp | Tungsten halogen lamp | S | |
| | Detector | Photodiode | Photodiode | S | |
| 6 | Reaction Unit | | | | |
| | Reaction cuvettes | Glass, 165 non-disposable | Plastic, 80 disposable | D | |
| | Reaction volume | $100{\text{\textasciitilde}}360\mu L$ | $180{\text{\textasciitilde}}500\mu L$ | D | |
| | Optical path | 5mm | 5mm | S | |
| | Reaction temperature | 37°C | 37°C | S | |
| 7 | Sample and Reagent System | | | | |
| | Sample disk | 140 positions. 45 positions<br>respectively for outer two<br>circles of outer rings and 25<br>positions respectively for<br>inter two circles of inter<br>rings | 40 sample tube positions on<br>the outer circle | D | |
| | Reagent disk | 120 positions. 50 positions<br>for inner circle and 70<br>positions for outer circle. | 40 reagent bottle positions<br>on the inner circle | D | |
| | Pipttor System | Positive displacement<br>stepper motordrive | Positive displacement<br>stepper motor drive | S | |
| | Refrigerator temperature | 2-8°C | 4-15°C | D | |
| | Sample Dispense | $1.5\mu L -50\mu L$ | $3\mu L -45\mu L$ | D | |
| | Reagent Dispense | $15\mu L-300\mu L$ | $30\mu L-450\mu L$ | D | |
| 8 | POWER | | | | |
| | Input | $110/115V{\text{\textasciitilde}}\pm 10\%, 60Hz\pm 1$<br>Hz | $100-130V{\text{\textasciitilde}},50/60\pm 1 Hz$ | D | |
| | Consumption | 3800VA | 1000 VA | D | |
| 9 | Operating environmental conditions | | | | |
| | Temperature | 15°C to 30°C | 15°C to 30°C | S | |
| | Humidity | 35% to 85%,<br>non-condensing | 35% to 80%,<br>non-condensing | S | |
| | Comparison on Electrolytes assay Chart 2: BS-800M ISE module and BS-200 ISE module | | | | |
| | Feature | BS-800M ISE Kit | BS-200 ISE Kit (Predicate<br>Device) | Same<br>(S)/Different<br>(D) | |
| | 510(K) | K160370 | K072018 | / | |
| 1 | Indication for use/Intended Use | | | | |
| | ISE Kit | The BS-800M/ ABS800/<br>BA-800M ISE Module is<br>for the in vitro quantitative<br>determination of Sodium | The BS-200 ISE Module is<br>for the in vitro quantitative<br>determination of Sodium<br>(Na⁺), Potassium (K⁺), and | S | |
| | | (Na+), Potassium (K+), and<br>Chloride (Cl-)<br>concentrations in serum,<br>plasma and urine samples<br>on the the BS-800M/<br>ABS800/<br>BA-800M<br>Chemistry Analyzer. | Chloride (Cl-)<br>concentrations in serum,<br>plasma and urine samples<br>on the the BS-200<br>Chemistry Analyzer. | | |
| | Sodium | Sodium measurements<br>monitor electrolyte balance<br>and in the diagnosis and<br>treatment of diseases<br>involving electrolyte<br>imbalance. | Sodium measurements<br>monitor electrolyte balance<br>and in the diagnosis and<br>treatment of diseases<br>involving electrolyte<br>imbalance. | S | |
| | Potassium | Potassium measurements<br>monitor electrolyte balance<br>and in the diagnosis and<br>treatment of disease<br>conditions characterized by,<br>low or high blood potassium<br>levels. | Potassium measurements<br>monitor electrolyte balance<br>and in the diagnosis and<br>treatment of disease<br>conditions characterized by,<br>low or high blood potassium<br>levels. | S | |
| | Chloride | Chloride measurements are<br>used in the diagnosis and<br>treatment of electrolyte and<br>metabolic disorders. | Chloride measurements are<br>used in the diagnosis and<br>treatment of electrolyte and<br>metabolic disorders. | S | |
| 2 | System Function | | | | |
| | Method Principle | Ion Selective Electrode | Ion Selective Electrode | S | |
| | Electrode | Na+ electrode, K+<br>electrode, Cl- electrode,<br>reference electrode | Na+ electrode, K+<br>electrode, Cl- electrode,<br>reference electrode, Space<br>electrode | D | |
| | Reference reagent | MR Serum Standard, MR<br>Urine Standard, MR Buffer<br>Solution, MR Detergent<br>Solution, MR Na/K Check<br>Solution | Reagent Pack, Urine<br>Diluent, Cleaning Solution, | D | |
| | ISE Internal<br>Standard | None | None | S | |
| | Sample Volume | 22 µL total for all three tests | 70 µL Serum, plasma mode;<br>140 µL Urine mode | D | |
| | ISE Throughput<br>Rate | 600 tests/hour | 300 tests/hour Serum,<br>plasma mode; | D | |
| | Test Environment | 198 tests/hour Urine mode<br>Clinical Lab | | S | |
| 3 | Calibration | | | | |
| | Calibrator | MR Serum Standard for<br>Serum, Plasma mode;<br>MR Urine standard for<br>Urine mode; | ISE reagent pack for all<br>three sample types | D | |
| | Standardization | Na: NIST standard material<br>SRM919<br>K: NIST standard material<br>SRM 918<br>Cl: NIST standard material<br>SRM 919 | Na: NIST standard material<br>SRM956<br>K: NIST standard material<br>SRM 956<br>Cl: NIST standard material<br>SRM 956 | D | |
| | Calibrator Stability | 5°C~35°C, 12 months of<br>shelf-life, 8 weeks of in-use<br>stability | 4°C~25°C, 24 months of<br>shelf-life | D | |
| | Calibrator Matrix | Buffered Aqueous matrix | Buffered Aqueous matrix | S | |
| | Calibrator Form | liquid | liquid | S | |
| | Number of<br>Calibrator levels | Two for serum/Plasma and<br>Two for Urine | Two for serum/plasma/urine | S | |
| | Calibration<br>Frequency | Daily | 8 hours | D | |
| 4 | Performance Characters | | | | |
| | Analytical<br>Measuring<br>(mmol/L) | Serum/Plasma<br>Na: 100-200<br>K: 1-8<br>Cl: 50-150<br><br>Urine<br>Na: 10-400<br>K: 5-200<br>Cl: 15-400 | Serum/Plasma<br>Na: 113-194<br>K: 1.1-8.6<br>Cl: 53-154<br><br>Urine<br>Na: 27-372<br>K: 13-184<br>Cl: 42-422 | D | |
| | Reference Interval | Serum(Adults):<br>Sodium: 136-145 mmol/L<br>Potassium: 3.5-5.1 mmol/L<br>Chloride: 98-107 mmol/L<br>Plasma(Adults):<br>Sodium: 136-145 mmol/L<br>Potassium: 3.4-4.5 mmol/L<br>Chloride: 98-107 mmol/L<br>Urine. 24 hour (Adults): | Serum(Adults):<br>Sodium: 136-145 mmol/L<br>Potassium: 3.5-5.1 mmol/L<br>Chloride: 98-107 mmol/L<br>Plasma(Adults):<br>Sodium: 136-145 mmol/L<br>Potassium: 3.4-4.5 mmol/L<br>Chloride: 98-107 mmol/L<br>Urine. 24 hour (Adults): | S | |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| | Sodium: 40-220 mmol/24h | Sodium: 40-220 mmol/24h | |
|------------------------------------|--------------------------|--------------------------|---|
| | Potassium:25-125 mol/24h | Potassium:25-125 mol/24h | |
| | Chloride:110-250 mol/24h | Chloride:110-250 mol/24h | |
| NSI Interferences<br>concentration | Bilirubin | Bilirubin | D |
| | Na/K/Cl for three sample | Na/K/Cl for three sample | |
| | types: 40 mg/dL | types: 20 mg/dL | |
| | | | |
| | Hemoglobin | Hemoglobin | |
| | Na/Cl for three sample | Na/K/Cl for three sample | |
| | types: 500 mg/dL | types: 500 mg/dL | |
| | Urine K: 500 mg/dL | | |
| | | Lipemia | |
| | Lipemia | Na/K/Cl for three sample | |
| | Na/K/Cl for three sample | types: 1000 mg/dL | |
| | types: 1000 mg/dL | | |
| | | | |
| | Ascorbic acid | | |
| | Na/K/Cl for three sample | | |
| | types: 30 mg/dL | | |
### Performance Characteristics:
Performance testing of the BS-800M Chemistry Analyzer consisted of running the FDA previously cleared assay and the ISE module on the BS-800M to evaluate precision, linearity, and method comparison, Limits of Detection and Limits of Quantitation, interference, ISE plasma sample type studies.
A correlation analysis between the BS-800M Chemistry Analyzer and BS-200 Chemistry Analyzer yielded the following results:
| Analyte | Unit | Sample<br>Range | N | Slope | Intercept | Correlation<br>Coefficient |
|-----------|--------|-----------------|-----|-------|-----------|----------------------------|
| BUN | mg/dL | 5.4-149.5 | 122 | 0.98 | 0.81 | 0.9997 |
| Serum Na+ | mmol/L | 100.2-196.4 | 124 | 1.03 | -5.95 | 0.9966 |
| Serum K+ | mmol/L | 1.40-7.70 | 121 | 1.04 | -0.09 | 0.9994 |
| Serum Cl- | mmol/L | 50.4-149.4 | 124 | 1.00 | 0.15 | 0.9993 |
| Urine Na+ | mmol/L | 12.1-395.4 | 120 | 0.99 | -2.52 | 0.9997 |
| Urine K+ | mmol/L | 5.1-194.7 | 120 | 0.98 | -1.00 | 0.9997 |
| Urine Cl- | mmol/L | 15.2-381.3 | 120 | 0.97 | 2.06 | 0.9988 |
And the bias at the medical decision points of method comparison yielded the following
{11}------------------------------------------------
| Analyte | Unit | Medical decision points | Bias at the medical decision points<br>(Difference/Difference%) | | |
|-----------|--------|-------------------------|-----------------------------------------------------------------|--------------|--------------|
| | | | Point 1 | Point 2 | Point 3 |
| BUN | mg/dL | 6,26,50 | 0.696/11.6% | 0.317/1.2% | -0.138/-0.3% |
| Serum Na+ | mmol/L | 115,135,150 | -2.713/-2.4% | -2.149/-1.6% | -1.727/-1.2% |
| Serum K+ | mmol/L | 3.0,5.8,7.5 | 0.032/1.1% | 0.144/2.5% | 0.212/2.8% |
| Serum Cl- | mmol/L | 90,112 | 0.245/0.3% | 0.267/0.2% | / |
| Urine Na+ | mmol/L | 40,112 | -2.759/-6.9% | -3.837/-1.7% | / |
| Urine K+ | mmol/L | 25,125 | -1.418/-5.7% | -3.085/-2.5% | / |
| Urine Cl- | mmol/L | 110,250 | -1.112/-1.0% | -5.147/-2.1% | / |
results:
The total precision test of BS-800M yielded the following results:
| Analyte | Unit | Sample | n | Mean | Repeatability | | Within-Device<br>Precision | |
|-----------|--------|-------------------|----|-------|---------------|------|----------------------------|------|
| | | | | | SD | CV% | SD | CV% |
| BUN | mg/dL | Control<br>pool 1 | 80 | 11.0 | 0.28 | 2.5% | 0.33 | 3.0% |
| | mg/dL | Control<br>pool 2 | 80 | 46.4 | 0.57 | 1.2% | 0.94 | 2.0% |
| Serum Na+ | mmol/L | Control<br>pool 1 | 80 | 136.1 | 0.35 | 0.3% | 0.83 | 0.6% |
| | mmol/L | Control<br>pool 2 | 80 | 166.1 | 0.46 | 0.3% | 1.12 | 0.7% |
| Serum K+ | mmol/L | Control<br>pool 1 | 80 | 3.72 | 0.006 | 0.2% | 0.021 | 0.6% |
| | mmol/L | Control<br>pool 2 | 80 | 6.10 | 0.013 | 0.2% | 0.033 | 0.5% |
| Serum CL- | mmol/L | Control<br>pool 1 | 80 | 89.4 | 0.21 | 0.2% | 0.38 | 0.4% |
| | mmol/L | Control<br>pool 2 | 80 | 107.5 | 0.23 | 0.2% | 0.44 | 0.4% |
Page 9 of 14
{12}------------------------------------------------
| | | pool 2 | | | | | | |
|-----------|--------|-------------------|----|-------|------|------|------|------|
| Urine Na+ | mmol/L | Control<br>pool 1 | 80 | 87.1 | 0.22 | 0.3% | 0.44 | 0.5% |
| Urine Na+ | mmol/L | Control<br>pool 2 | 80 | 148.6 | 0.36 | 0.2% | 0.74 | 0.5% |
| Urine K+ | mmol/L | Control<br>pool 1 | 80 | 35.1 | 0.08 | 0.2% | 0.21 | 0.6% |
| Urine K+ | mmol/L | Control<br>pool 2 | 80 | 68.8 | 0.16 | 0.2% | 0.41 | 0.6% |
| Urine Cl- | mmol/L | Control<br>pool 1 | 80 | 90.4 | 0.20 | 0.2% | 0.40 | 0.4% |
| Urine Cl- | mmol/L | Control<br>pool 2 | 80 | 133.9 | 0.17 | 0.1% | 0.48 | 0.4% |
The linearity test of BS-800M yielded the following results:
| Analyte | Unit | Slope | Intercept | Correlation<br>Coefficient | Linear Range<br>Tested | Claimed<br>Linear<br>Range |
|--------------|--------|--------|-----------|----------------------------|------------------------|----------------------------|
| BUN | mg/dL | 0.9999 | 0.0053 | 0.9996 | 4.5-162.6 | 5-150 |
| Serum<br>Na+ | mmol/L | 0.9996 | 0.0783 | 1.0000 | 41.9-224.4 | 100-200 |
| Serum<br>K+ | mmol/L | 0.9997 | -0.0003 | 0.9999 | 0.61-9.34 | 1-8 |
| Serum<br>Cl- | mmol/L | 0.9998 | 0.0389 | 0.9999 | 11.0-162.4 | 50-150 |
| Urine<br>Na+ | mmol/L | 1.0000 | 0.0139 | 0.9999 | 7.5-469.5 | 10-400 |
| Urine K+ | mmol/L | 1.0001 | 0.0175 | 0.9999 | 3.1-254.4 | 5-200 |
| Urine<br>Cl- | mmol/L | 1.0000 | 0.0370 | 0.9999 | 10.9-439.4 | 15-400 |
The detection limit studies test of BS-800M yielded the following results:
| Analyte | Unit | LoB | LoD | LoQ |
|-----------|--------|------|------|------|
| BUN | mg/dL | 0.4 | 0.9 | 4.2 |
| Serum Na+ | mmol/L | 1.9 | 3.5 | 39.5 |
| Serum K+ | mmol/L | 0.04 | 0.05 | 0.54 |
Page 10 of 14
{13}------------------------------------------------
| Serum Cl- | mmol/L | 0.5 | 0.6 | 9.5 |
|-----------|--------|-----|-----|-----|
| Urine Na+ | mmol/L | 0.9 | 1.4 | 6.9 |
| Urine K+ | mmol/L | 0.1 | 0.2 | 2.4 |
| Urine Cl- | mmol/L | 0.2 | 0.3 | 9.8 |
The Interference test of BS-800M yielded the following results:
Effects of bilirubin, hemoglobin, lipemia, ascorbic acid are tested, yielded the following results:
| Analyte | Analyte<br>concentration<br>(mmol/L) | Bilirubin<br>level(mg/dL) | Bias(mmol/L) | Comments |
|-----------|--------------------------------------|---------------------------|----------------|----------|
| BUN | 8.5<br>39.5 | 40 | +0.0<br>-0.1 | NSI |
| Serum Na+ | 134.0<br>154.8 | 40 | -0.4<br>+0.4 | NSI |
| Serum K+ | 3.23<br>5.94 | 40 | -0.01<br>+0.00 | NSI |
| Serum Cl- | 89.7<br>114.7 | 40 | -0.1<br>-0.2 | NSI |
| Urine Na+ | 45.9<br>224.1 | 40 | +0.6<br>-1.3 | NSI |
| Urine K+ | 26.9<br>127.3 | 40 | +0.0<br>-0.4 | NSI |
| Urine Cl- | 113.9<br>253.5 | 40 | +0.1<br>-1.5 | NSI |
| | Results of the bilirubin interference testing | |
|--|-----------------------------------------------|--|
| | | |
Results of the hemoglobin interference testing
| Analyte | Analyte concentration (mmol/L) | Hemoglobin level(mg/dL) | Bias(mmol/L) | Comments |
|-----------|--------------------------------|-------------------------|--------------|----------|
| BUN | 8.6<br>40.0 | 500 | +0.4<br>-0.1 | NSI |
| Serum Na+ | 129.4<br>153.1 | 500 | +0.9<br>+0.9 | NSI |
| Serum K+ | 3.15 | 125 | +0.30 | NSI |
| | | ≥250 | +0.61 | SI |
| Serum K+ | 5.81 | 125 | +0.33 | NSI |
| | | ≥250 | +0.68 | SI |
Page 11 of 14
{14}------------------------------------------------
| Serum Cl- | 88.8<br>113.5 | 500 | +2.0<br>+2.6 | NSI |
|-----------|----------------|-----|--------------|-----|
| Urine Na+ | 45.1<br>223.4 | 500 | +1.5<br>+2.2 | NSI |
| Urine K+ | 29.3<br>135.0 | 500 | +1.5<br>+0.1 | NSI |
| Urine Cl- | 113.1<br>261.3 | 500 | +2.5<br>+1.8 | NSI |
# Results of the lipemia interference testing
| Analyte | Analyte<br>concentration<br>(mmol/L) | Lipemia<br>level(mg/dL) | Bias(mmol/L) | Comments |
|-----------|--------------------------------------|-------------------------|----------------|----------|
| BUN | 8.3<br>38.2 | 1000 | -1.2<br>-1.1 | NSI |
| Serum Na+ | 125.3<br>147.6 | 1000 | -0.5<br>+0.4 | NSI |
| Serum K+ | 3.08<br>5.64 | 1000 | +0.05<br>+0.07 | NSI |
| Serum Cl- | 85.5<br>111.5 | 1000 | +0.0<br>-0.1 | NSI |
| UrineNa+ | 40.9<br>215.8 | 1000 | -0.1<br>-0.8 | NSI |
| Urine K+ | 28.1<br>127.8 | 1000 | +0.1<br>-0.6 | NSI |
| UrineCl- | 110.8<br>232.9 | 1000 | -0.2<br>-0.8 | NSI |
# Results of the ascorbic acid interference testing
| Analyte | Analyte<br>concentration<br>(mmol/L) | Ascorbic Acid<br>level(mg/dL) | Bias(mmol/L) | Comments |
|-----------|--------------------------------------|-------------------------------|----------------|----------|
| BUN | 9.0<br>41.1 | 30 | -0.3<br>-0.4 | NSI |
| Serum Na+ | 132.9<br>153.4 | 30 | -0.3<br>+1.0 | NSI |
| Serum K+ | 3.19<br>5.89 | 30 | +0.00<br>+0.01 | NSI |
| Serum Cl- | 89.1<br>114.0 | 30 | -0.2<br>-0.4 | NSI |
Page 12 of 14
{15}------------------------------------------------
| Urine Na+ | 43.3<br>223.3 | 30 | +0.4<br>+0.8 | NSI |
|-----------|----------------|----|--------------|-----|
| Urine K+ | 29.7<br>125.0 | 30 | +0.1<br>-0.1 | NSI |
| Urine Cl- | 116.2<br>251.4 | 30 | +0.0<br>+0.3 | NSI |
There is no significant interference (NSI) observed when the concentrations of interference materials (including drugs) is below the ones in the following table
| Interferents | Level tested(mg/dL) |
|-----------------------|---------------------|
| Lipemia | 1000 |
| Bilirubin | 40 |
| Ascorbic Acid | 30 |
| Imipramine | 0.15 |
| Procainamide | 15 |
| Nortriptyline | 0.26 |
| Hydroxytyramine | 50.7 |
| Ibuprofen | 512 |
| Valproic acid | 75 |
| Chlorpromazine | 6 |
| Salicylic acid | 72 |
| Acetylsalicylic acid | 1205 |
| Erythromycin | 7.6 |
| Ethosuximide | 30.6 |
| Acetaminophen | 242 |
| Benzalkonium Chloride | 10.4 |
| Ampicillin | 6 |
There was significant interference for hemoglobin and Potassium Thiocynate.
- · Avoid Hemolyzed samples for potassium. Hemolyzed samples may give incorrect elevated potassium. Intracellular potassium concentration is 30-50 fold greater than that of extracellular serum or plasma.
- · Potassium thiocynate increases potassium by 0.55 mmol/L at the concentration of 3.30 mmol/L and by 0.58 mmol/L at the concentration of 5.39 mmol/L.
- Potassium thiocynate increases chloride by12.3 mmol/L at the concentration of 90.9 mmol/L and by12.6 mmol/L at the concentration of 114.6 mmol/L.
The BS-800M's sample type studies between serum and plasma of Na , K , Cl test yielded the following results, which proved the sample type plasma can also apply to ISE test:
{16}------------------------------------------------
| Analyte | Unit | N | Sample<br>Range | Slope | Intercept | Correlation<br>Coefficient |
|---------|--------|----|-----------------|-------|-----------|----------------------------|
| Na+ | mmol/L | 51 | 105.2-195.9 | 1.000 | -0.33 | 0.9989 |
| K+ | mmol/L | 51 | 1.45-7.87 | 0.966 | -0.13 | 0.9930 |
| Cl- | mmol/L | 51 | 54.2-148.1 | 0.996 | 0.62 | 0.9997 |
# Conclusion:
The data demonstrates that the BS-800M Chemistry Analyzer is substantially equivalent to BS-200 Chemistry Analyzer.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.