ISE Reagents, Glucose, CRP Latex, DxC 500 AU Clinical Chemistry Analyzer
Applicant
Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd.
Product Code
JGS · Clinical Chemistry
Decision Date
Jul 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1665
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. The Glucose test system is for the quantitative measurement of glucose in human serum, plasma, urine and cerebrospinal fluid on Beckman Coulter AU/DxC AU analyzers. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and of pancreatic islet cell carcinoma. System reagent for the quantitative determination of C-Reactive Protein in human serum and plasma on Beckman Coulter AU/DxC AU Analyzers. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. Measurements may also be useful as an aid in the identification of individuals at risk for future cardiovascular disease. High sensitivity CRP (hsCRP) measurements, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes. Reagents for the quantitative determination of Sodium, Potassium and Chloride concentrations in human serum, plasma and urine on the Beckman Coulter ISE modules. The sodium test system is intended for the quantitative measurement of sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. The potassium test system is intended for the quantitative measurement of potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. The chloride test system is intended for the quantitative measurement of the level of chloride in plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Device Story
DxC 500 AU Clinical Chemistry Analyzer is a fully automated, random-access, low-to-mid volume clinical chemistry analyzer. It processes serum, plasma, urine, and CSF samples. Input samples are identified via barcode or manual entry. The system utilizes three primary analytical methods: ion selective electrodes (ISE) for electrolytes (Na+, K+, Cl-); colorimetric enzymatic assay for glucose; and latex agglutination for CRP. The device transforms sample inputs into quantitative analyte concentrations using specific reagents, calibrators, and QC materials. The system features a refrigerated reagent compartment and photometric detection (340-800 nm). Results are displayed to healthcare providers to aid in clinical decision-making for conditions like diabetes, electrolyte imbalances, and inflammatory states. The device benefits patients by providing rapid, accurate diagnostic data for disease management.
Clinical Evidence
No clinical studies were performed. Evidence consists of analytical performance testing, including precision (CLSI EP05-A3), linearity (CLSI EP06), interference (CLSI EP07/EP37), and method comparison (CLSI EP09c) against the predicate device. Precision studies (N=80) showed low CVs across all analytes. Method comparison (N>100 per analyte) demonstrated high correlation (R > 0.99) with the predicate. Matrix comparison studies confirmed suitability of various plasma anticoagulants.
Technological Characteristics
Automated photometric and ISE clinical chemistry analyzer. Materials: crown ether membrane electrodes (Na+, K+), PVC membrane (Cl-), latex particles coated with rabbit anti-CRP antibodies. Energy: Halogen lamp (340-800 nm). Connectivity: Barcode-based sample identification. Sterilization: N/A. Software: Embedded control system for automated analysis, calibration, and QC management.
Indications for Use
Indicated for quantitative measurement of sodium, potassium, chloride, glucose, and C-reactive protein (CRP) in human serum, plasma, urine, and cerebrospinal fluid (glucose only). Used for diagnosis/treatment of electrolyte, metabolic, and inflammatory disorders, including diabetes, cardiovascular risk assessment, and infection evaluation. Prescription use only.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Predicate Devices
ISE Reagents, Glucose, CRP Latex, DxC 700 AU Clinical Chemistry Analyzer (K161837)
Related Devices
K161837 — ISE Reagent, Glucose, CRP Latex, DxC 700 AU Clinical Chemistry Analyzer · Beckman Coulter, Inc. · Dec 16, 2016
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
July 20, 2023
Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. Tracy Jin Sr. Regulatory Affairs No. 181 West Su Hong Road. Suzhou Industrial Park Suzhou, Jiangsu, CN 215021 China
Re: K220977
Trade/Device Name: ISE Reagents, Glucose, CRP Latex, DxC 500 AU Clinical Chemistry Analyzer Regulation Number: 21 CFR 862.1665 Regulation Name: Sodium Test System Regulatory Class: Class II Product Code: JGS, CEM, CGZ, CFR, NQD, DCN, JJE Dated: January 12, 2023 Received: January 12, 2023
Dear Tracy Jin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Paula V. Caposino -2
Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220977
### Device Name
DxC 500 AU Clinical Chemistry Analyzer; Glucose; CRP Latex; ISE Reagents
### Indications for Use (Describe)
The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
The Glucose test system is for the quantitative measurement of glucose in human serum, plasma, urine and cerebrospinal fluid on Beckman Coulter AU/DxC AU analyzers. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and of pancreatic islet cell carcinoma.
System reagent for the quantitative determination of C-Reactive Protein in human serum and plasma on Beckman Coulter AU/DxC AU Analyzers. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. Measurements may also be useful as an aid in the identification of individuals at risk for future cardiovascular disease. High sensitivity CRP (hsCRP) measurements, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes. Reagents for the quantitative determination of Sodium, Potassium and Chloride concentrations in human serum, plasma and urine on the Beckman Coulter ISE modules.
The sodium test system is intended for the quantitative measurement of sodium in serum, plasma, and urine.
Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
The potassium test system is intended for the quantitative measurement of potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
The chloride test system is intended for the quantitative measurement of the level of chloride in plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the words "BECKMAN" and "COULTER" stacked on top of each other in black, bold font. The logo is simple and modern, with a focus on the company name.
### 1.0 Submitted By
Tracy Jin Sr. Regulatory Affairs Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. No. 181 West Su Hong Road, Suzhou Industrial Park, Suzhou, Jiangsu, CN 215021 Telephone: 86-512- 6895 5129 Fax: 86-512-6742 1069
#### 2.0 Date of Preparation
July 19, 2023
### 3.0 Device Name(s)
### DxC 500 AU Clinical Chemistry Analyzer 3.1
| Proprietary Name: | DxC 500 AU Clinical Chemistry Analyzer |
|----------------------|----------------------------------------------------------|
| Common Name: | DxC 500 AU |
| 510 (K) Number: | K220977 |
| Class: | Class I |
| Classification Name: | Discrete photometric chemistry analyzer for clinical use |
| Regulation Number: | 21 CFR § 862.2160 |
| Product Code: | JJE |
### 3.2 ISE Reagents
| Proprietary Name: | ISE Reagents |
|-------------------|--------------|
| Common Name: | ISE Reagents |
Class: Class II Classification Name/Regulation Number/Product Code:
Sodium test system 21 CFR § 862.1665 [JGS] Potassium test system 21 CFR § 862.1600 [CEM] Chloride test system 21 CFR § 862.1170 [CGZ]
#### 3.3 Glucose
| Proprietary Name: | Glucose |
|----------------------|---------------------|
| Common Name: | Glucose |
| Class: | Class II |
| Classification Name: | Glucose test system |
| Regulation Number: | 21 CFR § 862.1345 |
| Product Code: | CFR |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or ribbon. To the right of the circle are the words "BECKMAN" and "COULTER" stacked vertically, both in a bold, sans-serif font. The text is black, providing a strong contrast against the white background.
### 3.4 CRP Latex
Proprietary Name: CRP Latex Common Name: CRP Latex Class: Class II Classification Name: C-reactive protein immunological test system Regulation Number: 21 CFR § 866.5270 Product Code: DCN, NQD
### 4.0 Predicate Devices
| Candidate(s) | Predicate | Manufacturer |
|----------------------------------------------------|--------------------------------------------------------------|--------------------------|
| DxC 500 AU Clinical<br>Chemistry Analyzer | DxC 700 AU Clinical Chemistry<br>Analyzer<br>(K161837) | Beckman Coulter,<br>Inc. |
| ISE Reagents:<br>Sodium, Potassium and<br>Chloride | ISE Reagents:<br>Sodium, Potassium and Chloride<br>(K161837) | |
| Glucose | Glucose<br>(K161837) | |
| CRP Latex | CRP Latex<br>(K161837) | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the words "BECKMAN" and "COULTER" stacked on top of each other in black font. The logo is simple and modern, with a focus on the company name.
### 5.0 Device Description
#### 5.1 DxC 500 AU Clinical Chemistry Analyzer
The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer carries out automated analysis of serum, plasma, urine samples and other body fluids and automatically generates results. The device is an automated photometric clinical analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. Electrolyte measurement is performed using a single cell lon Selective Electrode (ISE) which is also common among the other members of the AU family.
### 5.2 ISE Reagents
The ISE module for Na+, K+, and Cl- employs crown ether membrane electrodes for sodium and potassium and a molecular oriented PVC membrane for chloride that are specific for each ion of interest in the sample. An electrical potential is developed according to the Nernst Equation for a specific ion. When compared to the Internal Reference Solution, this electrical potential is translated into voltage and then into the ion concentration of the sample.
### 5.3 Glucose
In this Beckman Coulter procedure, glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6phosphate dehydrogenase (G6P-DH) specifically oxidizes G-6-P to 6phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD+) to nicotinamide adenine dinucleotide, reduced (NADH). The change in absorbance at 340/660 nm is proportional to the amount of glucose present in the sample.
### 5.4 CRP Latex
The CRP Latex reagent is an in vitro diagnostic device that consists of ready to use buffer and latex particles coated with rabbit anti-CRP antibodies. In this procedure, the measurement of the rate of decrease in light intensity transmitted through particles suspended in solution is the result of complexes formed during the immunological reaction between the CRP of the patient serum and rabbit anti-CRPantibodies coated on latex particles. Two measuring range settings are available: Normal application (CRP Concentrations ranging between 5.0-480 mg/L) and Highly Sensitive (Cardiac) Application- (CRP concentrations ranging between 0.2-80mg/L).
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or ribbon. To the right of the circle are the words "BECKMAN" and "COULTER" stacked vertically, both in a bold, sans-serif font. The text is black, providing a strong contrast against the white background.
### 6.0 Indications for Use
### DxC 500 AU Clinical Chemistry Analyzer 6.1
The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
### 6.2 ISE Reagents (Sodium, Potassium and Chloride)
Reagents for the quantitative determination of Sodium, Potassium and Chloride concentrations in human serum, plasma, and urine on the Beckman Coulter ISE modules.
The sodium test system is intended for the quantitative measurement of sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
The potassium test system is intended for the quantitative measurement of potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
The chloride test system is intended for the quantitative measurement of the level of chloride in plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
### 6.3 Glucose
The Glucose test system is for the quantitative measurement of glucose in human serum, plasma, urine and cerebrospinal fluid on Beckman Coulter AU/DxC analyzers. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red oval shape with two white curved lines inside, resembling a stylized wave or ribbon. To the right of the oval, the words "BECKMAN" and "COULTER" are written in bold, black, sans-serif font, with "BECKMAN" on the top line and "COULTER" on the bottom line.
### 6.4 CRP Latex
System reagent for the quantitative determination of C-Reactive Protein in human serum and plasma on Beckman Coulter AU/DxC AU Analyzers. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. Measurements may also be useful as an aid in the identification of individuals at risk for future cardiovascular disease. High sensitivity CRP (hsCRP) measurements, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes.
### 7.0 Comparison to the Predicate
The DxC 500 AU Clinical Chemistry Analyzer is a member of Beckman Coulter's AU/DxC AU series of automated photometric analyzers that includes the DxC 700 AU Clinical Chemistry Analyzer (K161837) to which the substantial equivalence is claimed. The devices share common design characteristics and modes of operation. The key features are summarized in Tables 7.1 through 7.4.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the company name in black text. The word "BECKMAN" is on top of the word "COULTER".
| | Predicate Device: | Predicate Device: |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | DxC 700 AU Clinical<br>Chemistry Analyzer<br>(K161837) | DxC 500 AU Clinical<br>Chemistry Analyzer<br>(K220977) |
| Intended Use: | Automated chemistry<br>analyzer that measures<br>analytes in samples, in<br>combination with appropriate<br>reagents, calibrators, quality<br>control (QC) material and<br>other accessories. For in vitro<br>diagnostic use only.<br>Applications include<br>colorimetric, latex<br>agglutination and ion<br>selective electrode. | Same (removal of test<br>applications only):<br>Automated chemistry<br>analyzer that measures<br>analytes in samples, in<br>combination with appropriate<br>reagents, calibrators, quality<br>control (QC) material, and<br>other accessories. This<br>system is for in vitro<br>diagnostic use only. |
| Classification: | Analyzer, chemistry<br>(photometric, discrete), for<br>clinical use has been<br>classified as Class I, JJE by<br>the Clinical Chemistry and<br>Clinical Toxicology Devices<br>Panel, (21 CFR 862.2160). | Same |
| Sample Handling | | |
| Sample Containers | 10 sample tubes on a Rack | 7 sample tubes on a Rack |
| Sample Volume | 1.0 to 25.0 µL | Same |
| Sample Types | Serum, plasma, urine, CSF,<br>and whole blood (HbA1c<br>only) | Same |
| Sample Input | Sample Rack, STAT Table,<br>Direct-line Sample Aspiration<br>(For Automation<br>Connections) | Sample Rack, STAT Table |
| | Predicate Device: | Candidate Device: |
| Feature | DxC 700 AU Clinical<br>Chemistry Analyzer<br>(K161837) | DxC 500 AU Clinical<br>Chemistry Analyzer<br>(K220977) |
| Single Sample<br>Replicate Analysis | Ability to request test<br>replicates for one sample, up<br>to 20 replicates per sample<br>per test. | Same |
| Sample Analysis | | |
| Wavelength (nm) | Halogen Lamp<br>340 to 800 nm<br>13 wavelengths:<br>340, 380, 410, 450, 480, 520,<br>540, 570, 600, 660, 700, 750<br>and 800 nm (maximum of 2<br>wavelengths) | Same |
| Type of Measurement | • End-point assay<br>• Rate assay<br>• Fixed point assay<br>• Electrode method (ISE) | Same |
| Throughput | Maximum 800 photometric<br>tests/hour or 1200 tests/hour<br>(photometric + ISE) | Maximum 400 photometric<br>tests/hour or 800 tests/hour<br>(photometric + ISE) |
| Reagent Handling | | |
| Reagent Identification | Barcode and fixed position in<br>the reagent carousel | Same |
| Refrigerated Reagent<br>On-board Capacity | System has two Reagent<br>Refrigerators:<br>Reagent 1: 60 bottle capacity<br>Reagent 2: 48 bottle capacity | System has one Reagent<br>Refrigerator with 76 bottle<br>capacity (Reagent 1+Reagent<br>2) |
| | Predicate Device: | Candidate Device: |
| Feature | DxC 700 AU Clinical Chemistry Analyzer<br>(K161837) | DxC 500 AU Clinical Chemistry Analyzer<br>(K220977) |
| ISE Reagents | Three ISEs | Same |
| Reagent Storage Temperature | Refrigeration temperature: 4 to 12 °C (39.2 to 53.6 °F) | Same |
| Reagent Loading during Analysis | Reagent bottles can be changed while the instrument continues to measure patient samples. A reagent bottle can be added or changed within 5 minutes of request. | Reagent bottles can be changed while the instrument continues to measure patient samples. A reagent bottle can be added or changed within 5 minutes 30 seconds of request. |
| Computers/OS | | |
| CPU | Intel Processor<br>running<br>Windows 7 64bit | Intel Processor<br>running<br>Windows 10 64bit |
| Display/Monitor | 24" display<br>with touch screen | 21.5" display<br>with touch screen |
| Handheld Barcode Reader | Provided for reading 1D and 2D barcodes | Same |
| Software | | |
| GUI Application | User interface which will be common to new Beckman Coulter IVD systems | Re-engineered GUI; Latest common GUI design to Beckman Coulter IVD systems |
| User Help | HTML version of IFU is accessible with built-in interactive video. | HTML version of IFU is accessible in User Help |
| | Predicate Device: | Candidate Device: |
| Feature | DxC 700 AU Clinical<br>Chemistry Analyzer | DxC 500 AU Clinical<br>Chemistry Analyzer |
| | (K161837) | (K220977) |
| Laboratory<br>Information System<br>(LIS) Interfaces | Similar with LIS ASTM 1394<br>and AU unique LIS protocol | Similar with LIS ASTM 1394<br>protocol |
| Laboratory<br>Automation System<br>(LAS) Interfaces | Beckman LAS interface for<br>Direct Track Sampling system | Not supported |
| Real-Time Solutions<br>(RTS) support | Interface to RTS via<br>PROService protocol which is<br>made by Beckman Coulter<br>Remote software installation<br>capabilities. | Interface to RTS via remote<br>service protocol which is<br>made by Beckman Coulter |
| External printers | USB and LAN interfaced<br>printers | USB interfaced printers |
| | Predicate Device: | Candidate Device: |
| Feature | ISE – Sodium, Potassium<br>and Chloride<br>(K161837) | ISE –Sodium, Potassium<br>and Chloride<br>(K220977) |
| Intended Use | Reagent for the quantitative<br>determination of Sodium,<br>Potassium and Chloride<br>concentrations in human<br>serum and urine on the<br>Beckman Coulter ISE<br>modules. | Same |
| Measurement | Quantitative | Same |
| Instrument Required | DxC 700 AU Beckman<br>Coulter Analyzer | DxC 500 AU Beckman<br>Coulter Analyzer |
| Methodology | Indirect ISE | Same |
| Reagent form and<br>storage | Liquid, on-board storage | Same |
| Specimen Type | Serum, plasma, and urine | Same |
| Calibrator | Low Serum Standard (Cat #<br>AUH1014) and High Serum<br>Standard (Cat # AUH1015)<br>Low/High Urine Standard<br>(Cat # AUH1016) | Same |
| Calibration Stability | Calibrate daily. | Same |
| Reagent stability | ISE reagents are stable up to<br>90 days when stored in<br>accordance with the IFU<br>instructions. | Same |
| | Predicate Device: | Candidate Device: |
| Feature | ISE – Sodium, Potassium<br>and Chloride | ISE–Sodium, Potassium<br>and Chloride |
| | (K161837) | (K220977) |
| Analytical Range | Serum<br>Na+ 50 – 200 mEq/L<br>K+ 1.0 – 10.0 mEq/L<br>Cl- 50 – 200 mEq/L | Same |
| | Urine<br>Na+ 10 – 400 mEq/L<br>K+ 2.0 – 200.0 mEq/L<br>Cl- 15 – 400 mEq/L | |
| Interferences | Bilirubin (Na/K/Cl):<br>NSI* up to 40 mg/dL. | Same |
| | Lipemia (Na/K/Cl):<br>NSI up to 500 mg/dL. | |
| | Hemoglobin (Na):<br>NSI up to 250 mg/dL. | |
| | Hemoglobin (K):<br>NSI up to 70 mg/dL. | |
| | Hemoglobin (Cl):<br>NSI up to 500 mg/dL. | |
| Precision<br>(serum, urine) | Within-run CV < 3%<br>Total Precision CV < 5%. | Same |
## Table 7.1 Predicate Device Comparison
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or ribbon. To the right of the circle are the words "BECKMAN" in a bold, sans-serif font, stacked above the word "COULTER," which is also in a bold, sans-serif font.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Beckman Coulter logo. The logo consists of a red circle with two white curved lines inside, followed by the words "BECKMAN" in bold, and "COULTER" in bold below it. The text is in black.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or ribbon. To the right of the circle are the words "BECKMAN" in a bold, sans-serif font, stacked above the word "COULTER," which is also in a bold, sans-serif font.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the text "BECKMAN" in bold black letters above the text "COULTER" in bold black letters. The logo is simple and modern, and the colors are eye-catching.
## Table 7.2 ISE (Sodium, Potassium & Chloride) Predicate Device Comparison
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or heartbeat. To the right of the circle are the words "BECKMAN" stacked on top of "COULTER" in a bold, sans-serif font. The overall design is clean and corporate, conveying a sense of innovation and reliability.
*NSI – No significant interference (Chloride ±2.5%, Potassium ±0.25 mEq/L, Sodium ±2mEq/L)
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, and the words "BECKMAN COULTER" in black, with "BECKMAN" above "COULTER". The font is bold and sans-serif.
| | Predicate Device: | Candidate Device: |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Glucose<br>(K161837) | Glucose<br>(K220977) |
| Intended Use | The Glucose test system is<br>for the quantitative<br>measurement of glucose in<br>human serum, plasma, urine,<br>and cerebrospinal fluid on<br>Beckman Coulter AU<br>analyzers. | Same; updated instrument<br>family branding only.<br><br>The Glucose test system is<br>for the quantitative<br>measurement of glucose in<br>human serum, plasma, urine<br>and cerebrospinal fluid on<br>Beckman Coulter AU/DxC AU<br>analyzers. |
| Measurement | Quantitative | Same |
| Instrument Required | DxC 700 AU Beckman<br>Coulter Analyzer. | DxC 500 AU<br>Beckman Coulter Analyzer |
| Methodology | Photometric | Same |
| Reagent form and<br>storage | Liquid, on-board storage | Same |
| Specimen Type | Serum, plasma, urine, and<br>cerebrospinal fluid | Same |
| Calibrator | Chemistry Calibrator (Cat #<br>DR0070)<br><br>Urine Calibrator (Cat #<br>DR0090) | Same |
| Calibration Stability | 30 days | Same |
| Onboard Stability | 30 days refrigerated | Same |
| Feature | Predicate Device:<br>Glucose<br>(K161837) | Candidate Device:<br>Glucose<br>(K220977) |
| Analytical Range | Serum, plasma, and CSF:<br>10 - 800 mg/dL<br>Urine:<br>10 - 700 mg/dL | Same |
| Interferences | Bilirubin:<br>NSI* up to 40 mg/dL.<br>Lipemia:<br>NSI up to 700 mg/dL.<br>Hemoglobin:<br>NSI up to 500 mg/dL | Same |
| Precision<br>(serum, urine) | Within-run CV < 3%<br>Total Precision CV < 3%. | Same |
# Table 7.3 Glucose Predicate Device Comparison
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or heartbeat. To the right of the circle are the words "BECKMAN" stacked on top of "COULTER" in a bold, sans-serif font. The overall design is clean and corporate, conveying a sense of innovation and reliability.
*NSI – No significant interference designates recovery within 10% of the initial value.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the company name "BECKMAN COULTER" in bold, black letters. The word "BECKMAN" is stacked on top of the word "COULTER".
| Feature | Predicate Device:<br>CRP Latex<br>(K161837) | Candidate Device:<br>CRP Latex<br>(K220977) |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | System reagent for the<br>quantitative determination of<br>C-Reactive Protein in human<br>serum and plasma on<br>Beckman Coulter AU<br>Analyzers. Measurement of<br>CRP is useful for the<br>detection and evaluation of<br>infection, tissue injury,<br>inflammatory disorders and<br>associated diseases.<br>Measurements may also be<br>useful as an aid in the<br>identification of individuals at<br>risk for future cardiovascular<br>disease. High sensitivity CRP<br>(hsCRP) measurements,<br>when used in conjunction with<br>traditional clinical laboratory<br>evaluation of acute coronary<br>syndromes, may be useful as<br>an independent marker of<br>prognosis for recurrent<br>events, in patients with stable<br>coronary disease or acute<br>coronary syndromes. | Same; updated instrument<br>family branding only.<br>System reagent for the<br>quantitative determination<br>of C-Reactive Protein in<br>human serum and plasma<br>on Beckman Coulter<br>AU/DxC AU analyzers.<br>Measurement of CRP is<br>useful for the detection and<br>evaluation of infection,<br>tissue injury, inflammatory<br>disorders and associated<br>diseases. Measurements<br>may also be useful as an<br>aid in the identification of<br>individuals at risk for future<br>cardiovascular disease.<br>High sensitivity CRP<br>(hsCRP) measurements,<br>when used in conjunction<br>with traditional clinical<br>laboratory evaluation of<br>acute coronary syndromes,<br>may be useful as an<br>independent marker of<br>prognosis for recurrent<br>events, in patients with<br>stable coronary disease or<br>acute coronary syndromes. |
| Feature | Predicate Device:<br>CRP Latex<br>(K161837) | Candidate<br>Device:<br>CRP Latex<br>(K220977) |
| Measurement | Quantitative | Same |
| Instrument Required | DxC 700 AU<br>Beckman…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.