EDAN i15 Blood Gas and Chemistry Analysis System, EDAN i15 Calibrant Fluid Pack, EDAN i15 Blood Gas and Electrolyte Control, EDAN i15 Hematocrit Control
EDAN i15 Blood Gas and Chemistry Analysis System, EDAN i15 Calibrant Fluid Pack, EDAN i15 Blood Gas and Electrolyte Control, EDAN i15 Hematocrit Control
Applicant
Edan Instruments, Inc.
Product Code
CHL · Clinical Chemistry
Decision Date
Jul 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1120
Device Class
Class 2
Intended Use
The i15 Blood Gas and Chemistry Analysis System (including Blood Gas and Chemistry Analyzer, Calibrant Fluid Pack, Test Cartridge) is a portable, automated system that measures pH and blood gases (pCO2, pO2), electrolytes (Na+, K+, Ca++, Cl-) and hematocrit in arterial and venous whole blood samples with lithium heparin or calcium balanced heparin. The system is intended for in-vitro diagnostic use only by trained health care professionals in a laboratory environment, near patient or point-of-care settings.
Device Story
Portable, automated, microprocessor-based system for in-vitro diagnostic analysis of whole blood. Inputs: lithium- or calcium-balanced heparinized whole blood (140 µL) via syringe or capillary. Operation: disposable Test Cartridge integrates sensors, flow path, and waste; Calibrant Fluid Pack provides calibration solutions. Principle: potentiometric (pH, pCO2, Na+, K+, Ca++, Cl-), amperometric (pO2), and conductivity (Hct). Output: quantitative results for measured parameters and various calculated clinical indices (e.g., HCO3, BE, AnGap) displayed on a color touch screen. Used in clinical labs or point-of-care settings by trained professionals. Results assist in diagnosing/treating acid-base, oxygenation, and electrolyte disorders. System includes external multi-analyte quality controls for verification.
Clinical Evidence
Bench testing only. Precision studies (n=80 per level) and method comparison studies (n=488 samples) against the predicate device were conducted. Linearity, interference, and detection limit studies (LoB/LoD/LoQ for Ca++) were performed per CLSI guidelines. No clinical prospective/retrospective studies were required.
Technological Characteristics
Potentiometric (pH, PCO2, Na+, K+, Cl-, Ca++), amperometric (PO2), and conductivity (Hct) sensors. Single-use disposable cartridges with barcode-based calibration. Analyzer includes CPU, electronic interfaces, Wi-Fi/wired connectivity, and barcode scanner. Calibrant fluid pack contains NIST-traceable standards. Aqueous controls provided for quality verification.
Indications for Use
Indicated for trained healthcare professionals in lab or point-of-care settings to measure pH, blood gases (pCO2, pO2), electrolytes (Na+, K+, Ca++, Cl-), and hematocrit in heparinized arterial/venous whole blood. Used for diagnosis/treatment of acid-base/oxygenation disturbances, electrolyte imbalances, and hematological states (e.g., anemia, erythrocytosis).
Regulatory Classification
Identification
A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Predicate Devices
Siemens (Bayer) Rapidpoint 400 System (k002738)
RNA Medical QC823 Blood Gas, Electrolyte, Metabolite Control (k943754)
RNA Medical QC900 Hematocrit Control (k955630)
Reference Devices
Radiometer ABL 800
Related Devices
K061974 — GEM PREMIER 4000 WITH IQM, MODEL 4000; GEM CVP 1 AND 2 WITH CO-OX; GEM CVP 3 AND 4 HEMATOCRIT · Instrumentation Laboratory CO · Sep 15, 2006
K240998 — ABL90 FLEX PLUS System · Radiometer Medicals Aps · Dec 13, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2016
EDAN INSTRUMENTS, INC. C/O RANDY BYRD RANDY BYRD, LLC 1 BLUEBERRY HILL WAY PEPPERELL MA 01463-4200
Re: K151978 Trade/Device Name: EDAN i15 Blood Gas and Chemistry Analysis System, EDAN i15 Calibrant Fluid Pack, EDAN i15 Blood Gas and Electrolyte Control, EDAN i15 Hematocrit Control Regulation Number: 21 CFR 862.1120 Regulation Name: Blood Gases (pCO2, pO2) and Blood pH System Regulatory Class: II Product Code: CHL, JGS, CEM, JFP, CGZ, GKF, JIX, JJS, GLK Dated: July 8, 2016 Received: July 11, 2016
Dear Randy Byrd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements
{1}------------------------------------------------
as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Katherine Serrano -S
For:
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151978
#### Device Name
EDAN i15 Blood Gas and Chemistry Analysis System
#### Indications for Use (Describe)
The i15 Blood Gas and Chemistry Analysis System (including Blood Gas and Chemistry Analyzer, Calibrant Fluid Pack, Test Cartridge) is a portable, automated system that measures pH and blood gases (*pCO2*, *pO2*), electrolytes (Na<sup>+</sup>, K<sup>+</sup>, Ca<sup>++</sup>, Cl<sup>-</sup>) and hematocrit in arterial and venous whole blood samples with lithium heparin or calcium balanced heparin. The system is intended for in-vitro diagnostic use only by trained health care professionals in a laboratory environment, near patient or point-of-care settings.
| pH, pCO <sub>2</sub> ,<br>pO <sub>2</sub> : | Whole blood measurement of certain gases in whole blood, or pH of whole blood, is used in the diagnosis and treatment of life-threatening acid-base and/or oxygenation disturbances. |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hct: | Whole blood measurements of the packed cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red blood cells) |
| Na <sup>+</sup> : | Sodium measurement is used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance. |
| K <sup>+</sup> : | Potassium measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels |
| Cl <sup>-</sup> : | Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. |
| Ca <sup>++</sup> : | Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). |
For In Vitro Diagnostic Use
Type of Use (Select one or both, as applicable)
| <span style="unicode-bidi:isolate; direction: ltr;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="unicode-bidi:isolate; direction: ltr;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
:
ल
:
.
, I
n
ﺗ
:
:
来
...
.
:
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151978
Device Name EDAN i15 Calibrant Fluid Pack
Indications for Use (Describe)
The EDAN i15 Calibrant Fluid Pack is intended for the calibration of pH, pO2, pCO2, Na * K Ca *, Cl , and Hct as part of the EDAN i15 Blood Gas and Chemistry Analysis System.
For In Vitro Diagnostic Use
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
f 1
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K151978
Device Name
EDAN i15 Blood Gas and Electrolyte Control
Indications for Use (Describe)
The EDAN i15 Blood Gas and Electrolyte Controls are external multi-analyte quality control material intended to be used for the verification of correct operation and measurement of the EDAN i15 Blood Gas and Electrolyte Analyze, together with i15 Calibrant Fluid Pack and i15 Test Cartridge for the analysis of pH, blood gases (pCO2, pO2), and electrolytes (Na , K+, Ca++, and Cl').
For In Vitro Diagnostic Use
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="font-family: Arial;"> </span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
ବ
: :
:
.
း
គ្រ
..
. . . . . . . .
: 2017-07-17 1
:
1. 1. 1. 1. 1. 1. 1.
.
100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 -
ල
:
:
.
.
"
.
ुर
申博
:
යෙ
... ...
ﻳﻮ
a katika katika mi
-
ে
.
1 100
se
.
参考
8
.
ﺎﺕ
. . . .
. . . . . . . . . . . .
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151978
Device Name EDAN i15 Hematocrit Control
Indications for Use (Describe)
The EDAN i15 Hematocrit Controls are intended to be used for the verification of correct operation and measurement of the EDAN i15 Blood Gas and Chemistry Analysis System, together with i15 Calibrant Fluid Pack and i15 Test Cartridge for the analysis of hematocrit.
For In Vitro Diagnostic Use
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for EDAN, which includes the company's name in bold, gray letters. Above the name are two leaves, one yellow and one green, symbolizing growth and nature. Below the name is the tagline "A world of potential" in a smaller, black font, suggesting the company's vision and aspirations.
# 510(k) Summary1
# (a)(1) Submitter's name, address
EDAN Instruments, Inc. 3/F-B, Nanshan Medical Equipments Park 1019# Nanhai Road Shenzhen, 518067 P.R. China
Registration Number 3003971136
Application Correspondent Tracy Yue, +86.755.76856469
# Date of Preparation of this Summary:
14 July 2016
(2) Device trade or proprietary name: EDAN i15 Blood Gas and Chemistry Analysis System, EDAN i15 Calibrant Fluid Pack, EDAN i15 Blood Gas and Electrolyte Control and EDAN i15 Hematocrit Control.
Device common or usual name: Blood Gases (pCO2, PO2) and blood pH Test System, Electrolyte Analyzer; Calibrator, multi-analyte mixture; Controls for Blood Gases, Control, Hematocrit
# Classification Name: Multiple
| | | Product | Regulation | |
|--------------------------------------------------|--------------------|---------|------------|-------|
| Classification Name | Review Panel | Code | Number | Class |
| Blood gases (PCO2, PO2) and blood pH test system | Clinical Chemistry | 75 CHL | 862.1120 | 2 |
| Sodium test system | Clinical Chemistry | 75 JGS | 862.1665 | 2 |
| Potassium test system | Clinical Chemistry | 75 CEM | 862.1600 | 2 |
| Chloride test system | Clinical Chemistry | 75 CGZ | 862.1170 | 2 |
| Calcium test system | Clinical Chemistry | 75 JFP | 862.1145 | 2 |
| Automated hematocrit instrument | Hematology | 81 GKF | 864.5600 | 2 |
| Calibrators | Clinical Chemistry | 75 JIX | 862.1150 | 2 |
| Controls for Blood Gases (Assayed and Unassayed) | Clinical Chemistry | 75 JJS | 862.1660 | 1 |
| Control, Hematocrit | Hematology | 81 GLK | 864.8625 | 2 |
# (3) Substantial Equivalence
The EDAN i15 Blood Gas and Chemistry Analysis System, including the Blood Gas and Chemistry Analyzer, Calibrant Fluid Pack, Test Cartridge, and Quality Controls) is substantially equivalent in function, safety and efficacy to currently marketed devices for the same intended use as shown in the table, below.
| <b>Predicate Devices:</b> K002738 – Siemens (Bayer) Rapidpoint 400 System, including calibrators |
|--------------------------------------------------------------------------------------------------|
| K943754 – RNA Medical QC823 Blood Gas, Electrolyte, Metabolite Control |
| K955630 – RNA Medical QC900 Hematocrit Control |
<sup>4</sup> This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
{9}------------------------------------------------
# Comparison of Predicate and Candidate Devices
| Predicate Device: | K002738 Siemens (Bayer) Rapidpoint 400 System | | | Candidate Device: | EDAN i15 Blood Gas and Chemistry Analysis System | |
|-------------------|------------------------------------------------------------------------------------------|---------------|---------------|-----------------------------------------|--------------------------------------------------------------------------|---------------|
| | Characteristic | | | | | |
| | Intended Use | | | | | |
| | For the measurement of pH, pCO2, pO2, Na+, K+, Ca++, Cl- and Hct in whole blood samples. | | | Same | | |
| | Test Principle | | | | | |
| | pH and pCO2: potentiometric | | | Same | | |
| | pO2: amperometric | | | Same | | |
| | Na+, K+, Ca++, Cl-: potentiometric, ion-selective | | | Same | | |
| | Hematocrit: conductivity | | | Same | | |
| | Sensor Technology | | | | | |
| | Amperometric and potentiometric thick film microelectrodes | | | Same | | |
| | Sample Types | | | | | |
| | Heparinized whole blood | | | Same | | |
| | Sample Introduction | | | | | |
| | Syringe and capillary, by aspiration | | | Same | | |
| | Sample Volumes | | | | | |
| | 100 μL (syringe and capillary) | | | 140 μL (syringe and capillary) | | |
| | Measurement Range | | | | | |
| | Parameter | Range | Default units | Parameter | Range | Default units |
| | pH | 6.500 - 7.800 | (pH units) | pH | 6.500 - 7.800 | (pH units) |
| | pCO2 | 5.0 - 200.0 | mmHg | pCO2 | 10 - 150 | mmHg |
| | pO2 | 10.0 - 700.0 | mmHg | pO2 | 10 - 700 | mmHg |
| | Na+ | 100.0 - 200.0 | mmol/L | Na+ | 100 - 180 | mmol/L |
| | K+ | 0.50 - 15.00 | mmol/L | K+ | 2.0 - 9.0 | mmol/L |
| | Ca++ | 0.20 - 5.00 | mmol/L | Ca++ | 0.25 - 2.50 | mmol/L |
| | Cl- | 65 - 140 | mmol/L | Cl- | 65 - 140 | mmol/L |
| | Hct | 12 - 75 | % | Hct | 13 - 72 | % |
| | Glu | 20 - 750 | mg/dL | N/A | | |
| | User Interface, Printer | | | | | |
| | Menu-driven color touch screen display | | | Same | | |
| | Bar-code scanner | | | Same | | |
| | Thermal printer | | | Same | | |
| | System Operating Temperature, Humidity | | | | | |
| | 15 - 32°C, 5% to 85% (non-condensing) | | | 10 – 31° C, 25% to 80% (non-condensing) | | |
| Predicate Device: | K002738 Siemens (Bayer) 400<br>Measurement Cartridge | | | Candidate Device: | EDAN i15 Blood Gas and Chemistry<br>Analysis System Calibrant Fluid Pack | |
| Device: | Measurement Cartridge | Device: | Analysis System Calibrant Fluid Pack |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Intended Use | | | |
| The measurement cartridge is intended to be used<br>for the calibration of sensors used for the<br>quantitative measurement of pH, blood gases,<br>electrolytes and hematocrit. | | Same | |
| Configuration | | | |
| Three solutions for the calibration of the sensors<br>and a reference solution contained in Mylar and<br>aluminum foil bags, housed in a protective, self-<br>contained, molded plastic disposable container. | | One solution for the calibration of the sensors<br>contained in a Mylar and aluminum foil bag, housed<br>in a protective, self-contained, molded plastic<br>disposable container. | |
{10}------------------------------------------------
| Predicate Device: | K943754 RNA Medical QC823 Blood Gas, Electrolyte and Metabolite Quality Control | Candidate Device: | EDAN Blood Gas and Chemistry Quality Control |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---------------------------------------------------------------------------|
| Intended Use | To evaluate the operation of the analyzer and to monitor the measurement of pH and blood gases (pCO2, pO2, and electrolytes (Na+, K+, Ca++, Cl-) | | Same |
| Configuration | QC 823 is provided in three (3) levels for monitoring analyzer performance at different points within the clinical range. It is packaged in sealed glass ampoules, each containing 2.5 mL of solution. Ampoules are packaged thirty (30) per box. | | Same, except packaged with 5 ampoules per box. |
| Analytes | pH, pCO2, pO2, Na+, K+, Cl-, Ca++, Mg++, glucose, lactate, and BUN (urea) | | Same, except assay provided only for pH, pCO2, pO2, Na+, K+, Cl- and Ca++ |
| Predicate: | K955630 RNA Medical Hematocrit Controls | Candidate: | EDAN Blood Gas and Chemistry System Hematocrit Controls |
| Intended Use | To evaluate the operation and to monitor the measurement of hematocrit by electrical conductivity. | | Same |
| Configuration | QC 900 is provided in two (2) levels (Low and High) for monitoring analyzer performance at different points within the clinical range. It is packaged in sealed glass ampoules, each containing 1.7 mL of solution. Ampoules are packaged twenty (20) per box (10 of each level). | | Same, except each single level is packaged in a convenient 5-ampoule box. |
| Analytes | Hematocrit (%) | | Same |
| Active Ingredients | Salts and conductivity modifiers | | Same |
# (4) Description of the new device.
The EDAN i15 Blood Gas and Chemistry Analysis System, including the Blood Gas and Chemistry Analyzer, Calibrant Fluid Pack, Test Cartridge, and Quality Controls) is a system for in-vitro analysis of whole blood, delivering quantitative results for panels of tests determined by the Test Cartridge type used in the measurement. The Analyzer incorporates a large graphical user interface with a large color touch screen interfacing the analyzer electronic, Test Cartridge (containing the sensors) and Calibrant Fluid Pack. The sensors, flow path and waste are integrated into a single-use Test Cartridge. The calibration solution is contained in the Calibrant Fluid Pack which is available in versions for 50 or 100 tests. The product, consumables, installation instructions and packaging are designed for easy customer installation and operation.
The EDAN i15 Analyzer accepts lithium- or calcium-balanced heparinized whole blood samples from syringes and capillary tube. The minimum volume for samples from both syringe and capillary tube is 140 µL.
# Test Cartridge
As with the predicate device, the EDAN i15 Blood Gas and Chemistry Analysis System is microprocessor-based and incorporates traditional sensor technology for measurement.
{11}------------------------------------------------
- pH, pCO2, Na*, K*, Cl , and Ca** (potentiometric measurement) .
- pO2 (amperometric measurement)
- Hct (conductivity measurement) ●
The Test Cartridge is available in three (3) versions, with different test panels, but with identical sample paths and calibration fluid:
| Test Cartridge Type | Measured Parameters |
|---------------------|----------------------------------------|
| BG8 | pH, pCO2, pO2, Na+, K+, Cl-, Ca++, Hct |
| BG3 | pH, pCO2, pO2 |
| BC4 | Na+, K+, Cl-, Ca++, Hct |
The Test Cartridge aspirates the sample directly from a syringe, or capillary fitted with an adaptor, and requires a minimum sample volume of 140 µL.
# Measured Parameters
The EDAN i15 Blood Gas and Chemistry Analysis System measures blood pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium and hematocrit.
# Calculated Parameters
| Symbol | Description |
|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| cH+ | Hydrogen ion concentration |
| cH+(T) | Hydrogen ion concentration corrected for entered patient temperature |
| pH (T) | pH value corrected for entered patient temperature |
| pCO2 (T) | pCO2 corrected for entered patient |
| pO2 (T) | pO2 corrected for entered patient temperature |
| HCO3 act | Bicarbonate ion concentration |
| HCO3 std | Bicarbonate ion concentration normalized to a pCO2 of 40mmHg |
| BB (B) | Buffer base |
| BE (B) | Base excess (B) BE (ecf) Base excess (ecf) ctCO2 Total carbon dioxide |
| Ca++ (7.4) | The ionized calcium concentration of blood normalized to pH 7.4 |
| AnGap | An approximation of the difference between measured cations and measured<br>anions in the sample |
| tHb (est) | An estimation of the hemoglobin contained in the sample |
| sO2 (est) | An estimation of hemoglobin oxygen saturation: a ratio of the amount of<br>hemoglobin bound to oxygen to the total amount of hemoglobin able to bind<br>oxygen |
| pO2 (A-a) | Alveolar-arterial oxygen tension difference |
| pO2 (A-a) (T) | Alveolar-arterial oxygen tension difference corrected for entered patient<br>temperature |
| pO2 (a/A) | Arterial-alveolar oxygen tension ratio |
| pO2 (a/A) (T) | Arterial-alveolar oxygen tension ratio corrected for entered patient temperature |
| RI | Respiratory index: the ratio of the alveolar-arterial blood oxygen-pressure difference<br>to arterial pO2 |
| RI (T) | Respiratory index: the ratio of the alveolar-arterial blood oxygen-pressure difference<br>to arterial pO2 when both values are corrected for patient temperature |
| pO2/FIO2 | The ratio of arterial pO2 to the fraction of inspired oxygen |
| pO2 (T)/FIO2 | The ratio of arterial pO2 to the fraction of inspired oxygen corrected for the entered<br>patient temperature |
# Indications for Use
The i15 Blood Gas and Chemistry Analysis System (including Blood Gas and Chemistry Analyzer, Calibrant Fluid Pack, Test Cartridge) is a portable, automated system that measures pH and blood gases (pCO2, pO2),
{12}------------------------------------------------
electrolytes (Na , K , C , Cl ) and hematocrit in arterial and venous whole blood samples with lithium heparin or calcium balanced heparin. The system is intended for in-vitro diagnostic use only by trained health care professionals in a laboratory environment, near patient or point-of-care settings.
pH, pCO2, Whole blood measurement of certain gases in whole blood, or pH of whole blood, is used in pO2: the diagnosis and treatment of life-threatening acid-base and/or oxygenation disturbances.
- Hct: Whole blood measurements of the packed cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red blood cells)
- Nat: Sodium measurement is used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance.
- Kt: Potassium measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels
- Cl : Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
- Ca**: Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
The EDAN i15 Calibrant Fluid Pack is intended for the calibration of pH, pO2, pCO2, Na*, K* Ca**, Cl , and Hct as part of the EDAN i15 Blood Gas and Chemistry Analysis System.
The EDAN i15 Blood Gas and Electrolyte Controls are external multi-analyte quality control material intended to be used for the verification of correct operation and measurement of the EDAN i15 Blood Gas and Electrolyte Analyze, together with i15 Calibrant Fluid Pack and i15 Test Cartridge for the analysis of pH, blood gases (pCO2 pO2), and electrolytes (Na*, K*, Ca**, and Cl').
The EDAN i15 Hematocrit Controls are intended to be used for the verification of correct operation and measurement of the EDAN i15 Blood Gas and Chemistry Analysis System, together with i15 Calibrant Fluid Pack and i15 Test Cartridge for the analysis of hematocrit.
# Summary of the Technological Characteristics
The EDAN i15 Blood Gas and Chemistry Analysis System is substantially equivalent to the previously cleared Siemens RapidPoint 400 Test System in intended use. The test principle, sensor technology, sample types and mode of sample introduction are the same for these system, and measurement is shown to be equivalent in this submission. The results of software validation and performance verification testing confirmed the EDAN i15 Blood Gas and Chemistry Analysis System, including Blood Gas and Chemistry Analyzer, Calibrant Fluid Pack, and Test Cartridge, is safe and effective for it intended use and the EDAN i15 Blood Gas and Chemistry Analysis system is substantially equivalent to that of the Siemens RapidPoint 400 Blood Gas Electrolyte System, including the Measurement Cartridge, K002738 (predicate device).
The EDAN i15 Calibrator Fluid Pack is substantially equivalent to the Siemens 400 Measurement Cartridge previously cleared with the Siemens RapidPoint 400 Test System in configuration and in support for the calibration and measurement of equivalent panels of measurands.
The EDAN i15 Blood Gas and Electrolyte Control is substantially equivalent to the previously cleared RNA
{13}------------------------------------------------
Medical Blood Gas, Electrolyte and Metabolite Control K943754 (predicate device) in formulation, measurands and primary (glass ampoule) packaging, but packaged in convenient 5-ampoule packages.
The EDAN i15 Hematocrit Control is substantially equivalent to the previously cleared RNA Medical Hematocrit Control K955630 (predicate device) in formulation, measurands and primary (glass ampoule) packaging, but packaged in convenient 5-ampoule packages.
# Summary of Performance Testing
Bench testing was completed to demonstrate the EDAN i15 System is substantially equivalent to the Siemens RapidPoint 400 Blood Gas and Electrolyte System in performance, safety and efficacy. The bench testing included:
- Method Comparison Studies, Linearity and Analytical Measuring Range ●
- Imprecision/Reproducibility Studies
- Specificity/Interference Testing
- Detection Limit ●
The results of that testing confirm the performance of the EDAN i15 Blood Gas and Chemistry Analysis System, (including Blood Gas and Chemistry Analyzer, Calibrant Fluid Pack, and Test Cartridge), is substantially equivalent to that of the previously cleared Siemens (Bayer) RapidPoint 400 Test System, including Measurement Cartridge (predicate device).
# a. Linearity Studies
Whole, venous blood samples collected from healthy volunteers in vacuum blood collection tubes with lithium heparin were modified to obtain a range of values for each measurand approaching their analytical measurement range by the addition of isotonic electrolyte solution for electrolytes, tonometry with various gas mixtures for blood gases and pH, and by mixing with erythrocyte concentrated for depleted plasma to obtain an range of hematocrit. Criteria for acceptance is a slope between 0.95 and 1.05 and r-value ≥ 0.975. Electrolyte values were compared to those calculated from serial dilution of high and low concentration blood samples, pO2 and pCO2 against those determined by tonometry, pH was compared against parallel measurement on a Radiometer ABL 800, while hematocrit was compared the PCV% obtained by microhematocrit centrifuge. The evaluation demonstrated equivalent, linear performance over the analytical measurement range for all measurands in samples from syringes and from glass capillary tubes.
| Whole Blood from Syringe | | | | | | | | |
|--------------------------|-----------------|----------------------------|----------------|--------------------------|--------|-----------|---------|--|
| Measurand | No of<br>Levels | Claimed<br>Measuring Range | Specimen Range | % of<br>claimed<br>range | Slope | Intercept | r value | |
| Na+ | 7 | 100 - 180 | 94 - 185 | 114% | 0.9923 | -1.38 | 0.9983 | |
| K+ | 7 | 2.0 - 9.0 | 1.7 - 10.2 | 120% | 0.9886 | 0.08 | 0.9997 | |
| Cl- | 7 | 65 - 140 | 58 - 176 | 158% | 1.0028 | -1.59 | 0.9993 | |
| Ca++ | 7 | 0.25 - 2.50 | 0.23 - 2.94 | 120% | 0.9848 | -0.05 | 0.9981 | |
| pH | 7 | 6.500 - 7.800 | 6.455 - 7.944 | 115% | 0.9842 | 0.11 | 0.9996 | |
| <i>p</i> CO2 | 7 | 10 - 150 | 3 - 149 | 104% | 0.9533 | 0.62 | 0.9969 | |
| <i>p</i> O2 | 7 | 10 - 700 | 6 - 716 | 103% | 0.9969 | -6.16 | 0.9993 | |
| Hct | 7 | 13 - 72 | 3 - 77 | 125% | 0.9687 | 1.54 | 0.9981 | |
Linearity across claimed measurement range in samples introduced from Syringe.
{14}------------------------------------------------
| Whole Blood from Capillary | | | | | | | |
|----------------------------|--------|-----------------|----------------|---------|--------|-----------|---------|
| | | | | % of | | | |
| | No of | Claimed | | claimed | | | |
| Measurand | Levels | Measuring Range | Specimen Range | range | Slope | Intercept | r value |
| Na+ | 7 | 100 - 180 | 94 - 185 | 114% | 1.0059 | -3.06 | 0.9986 |
| K+ | 7 | 2.0 - 9.0 | 1.9 - 10.0 | 116% | 1.0013 | -0.01 | 0.9998 |
| Cl- | 7 | 65 - 140 | 59 - 175 | 155% | 0.9965 | -1.59 | 0.9992 |
| Ca++ | 7 | 0.25 - 2.50 | 0.24 - 2.97 | 121% | 0.9771 | -0.06 | 0.9968 |
| pH | 7 | 6.500 - 7.800 | 6.464 - 7.978 | 116% | 1.0039 | -0.03 | 0.9995 |
| pCO2 | 7 | 10 - 150 | 2 - 150 | 105% | 0.9999 | -0.06 | 0.9992 |
| pO2 | 7 | 10 - 700 | 7 - 717 | 103% | 1.0174 | -5.40 | 0.9997 |
| Hct | 7 | 13 - 72 | 4 - 76 | 122% | 0.9623 | 1.51 | 0.9984 |
Linearity across claimed measurement range in samples introduced from glass Capillary.
# b . Imprecision in External Controls
Imprecision was evaluated by running 3 levels of EDAN i15 Blood Gas and Electrolyte Control and 2 levels of EDAN i15 Hematocrit Control in duplicate each day for a total of 20 runs on one EDAN i15 Blood Gas and Chemistry Analysis System. The protocol was based on methods described in CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Second edition.
| pH Precision in Aqueous Blood Gas and Electrolyte Control | | | | | | | | |
|------------------------------------------------------------|------------------------------------------------------------|----|-------------------------------------------------------------|------------|-------------------|-------------------|--|--|
| | Mean | | Within-Run | Within-Run | Total Imprecision | Total Imprecision | | |
| Sample | (pH units) | N | SD | CV | SD | CV | | |
| Level 1 | 7.153 | 80 | 0.012 | 0.17 % | 0.012 | 0.17 % | | |
| Level 2 | 7.410 | 80 | 0.013 | 0.18 % | 0.013 | 0.18 % | | |
| Level 3 | 7.598 | 80 | 0.008 | 0.11 % | 0.009 | 0.12 % | | |
| | | | pCO2 Precision in Aqueous Blood Gas and Electrolyte Control | | | | | |
| | Mean | | Within-Run | Within-Run | Total | Total Imprecision | | |
| Sample | (mmHg) | N | SD | CV | Imprecision SD | CV | | |
| Level 1 | 62.8 | 80 | 2.11 | 3.36 % | 2.40 | 3.82 % | | |
| Level 2 | 38.8 | 80 | 2.11 | 5.45 % | 2.33 | 6.02 % | | |
| Level 3 | 20.5 | 80 | 0.80 | 3.91 % | 0.85 | 4.13 % | | |
| | pO2 Precision in Aqueous Blood Gas and Electrolyte Control | | | | | | | |
| | Mean | | Within-Run | Within-Run | Total | Total Imprecision | | |
| Sample | (mmHg) | N | SD | CV | Imprecision SD | CV | | |
| Level 1 | 71.4 | 80 | 2.67 | 3.73 % | 2.68 | 3.75 % | | |
| Level 2 | 106.2 | 80 | 1.84 | 1.73 % | 2.50 | 2.35 % | | |
| Level 3 | 149.2 | 80 | 2.59 | 1.74 % | 3.33 | 2.23 % | | |
| | | | Na+ Precision in Aqueous Blood Gas and Electrolyte Control | | | | | |
| | Mean | | Within-Run | Within-Run | Total | Total Imprecision | | |
| Sample | (mmol/L) | N | SD | CV | Imprecision SD | CV | | |
| Level 1 | 111.8 | 80 | 1.29 | 1.15 % | 1.32 | 1.18 % | | |
| Level 2 | 133.0 | 80 | 0.95 | 0.71 % | 1.14 | 0.86 % | | |
| Level 3 | 158.5 | 80 | 1.51 | 0.95 % | 1.44 | 0.91 % | | |
{15}------------------------------------------------
| K+ Precision in Aqueous Blood Gas and Electrolyte Control | | | | | | |
|-------------------------------------------------------------|---------------|----|---------------|----------------|----------------------|-----------------------|
| Sample | Mean (mmol/L) | N | Within-Run SD | Within-Run CV | Total Imprecision SD | Total Imprecision CV |
| Level 1 | 1.91 | 80 | 0.040 | 2.11 % | 0.040 | 2.07 % |
| Level 2 | 4.34 | 80 | 0.047 | 1.09 % | 0.052 | 1.20 % |
| Level 3 | 6.34 | 80 | 0.073 | 1.16 % | 0.065 | 1.02 % |
| Cl- Precision in Aqueous Blood Gas and Electrolyte Control | | | | | | |
| Sample | Mean (mmol/L) | N | Within-Run SD | Within-Run CV | Total Imprecision SD | Total Imprecision CV |
| Level 1 | 75.9 | 80 | 0.91 | 1.20 % | 0.93 | 1.22 % |
| Level 2 | 92.3 | 80 | 1.07 | 1.16 % | 1.08 | 1.17 % |
| Level 3 | 121.7 | 80 | 0.66 | 0.54 % | 0.69 | 0.57 % |
| Ca++ Precision in Aqueous Blood Gas and Electrolyte Control | | | | | | |
| Sample | Mean (mmol/L) | N | Within-Run SD | Within-Run CV% | Total Imprecision SD | Total Imprecision CV% |
| Level 1 | 1.37 | 80 | 0.04 | 3.09 | 0.05 | 3.47 |
| Level 2 | 1.23 | 80 | 0.02 | 1.54 | 0.03 | 2.42 |
| Level 3 | 0.61 | 80 | 0.02 | 3.10 | 0.02 | 3.66 |
| Hct Precision in Aqueous Hematocrit Control | | | | | | |
| Sample | Mean (%) | N | Within-Run SD | Within-Run CV | Total Imprecision SD | Total Imprecision CV |
| Low | 19.2 | 80 | 0.57 | 2.97 % | 0.73 | 3.82 % |
| High | 46.8 | 80 | 0.58 | 1.24 % | 0.54 | 1.16 % |
#### Precision in Whole Blood C.
#### Hematocrit Venous Whole Blood Precision
An in-house with-in run precision study was performed on heparinized venous whole blood samples collected from eight volunteers. Six levels were prepared (one on each day of 6 days) in order to cover the measuring range of hematocrit. Each sample was run in replicates of 10 on 3 EDAN i15 analyzers, using 3 reagent lots, over a period of 6 days (one level run per day) for a total of 30 results per level. The with-in run precision results for of the six levels of whole blood tested from both syringe and capillary tube is presented in the table below:
| n = 30 | | Syringe | | Capillary | | | |
|--------|----------|---------|-----|-----------|-----|-----|--|
| Hct | | | | | | | |
| Level | Mean (%) | SD | %CV | Mean (%) | SD | %CV | |
| ਹ | 18.0 | 1.1 | 6.3 | 18.4 | 1.1 | 5.9 | |
| 2 | 30.6 | 0.6 | 1.8 | 31.1 | 0.8 | 2.4 | |
| 3 | 47.1 | 0.8 | 1.7 | 48.0 | 1.1 | 2.3 | |
| 4 | 51.4 | 0.6 | 1.2 | 52.5 | 0.7 | 1.3 | |
| ട | 54.0 | 0.6 | 1.2 | 53.9 | 0.9 | 1.6 | |
| 6 | 62.9 | 0.9 | 1.4 | 63.1 | 0.8 | 1.3 | |
Venous whole blood within-run precision for Hct
An in-house precision study was performed utilizing 3 EDAN i15 Blood Gas and Chemistry Analysis System with venous whole blood collected from 3 healthy volunteers. The whole blood was tonometered to nominal gas values. The remainder of the analytes were not altered. The samples were run in replicates of 10 on each of the three analyzers for a total of 30 results. The results of the combined precision data is summarized below:
{16}------------------------------------------------
| n = 30 | | Syringe | | | Capillary | | |
|-----------|---------------|-------------|-------|------|-----------|-------|------|
| | | Mean<br>(%) | SD | %CV | Mean (%) | SD | %CV |
| Parameter | pH | 7.325 | 0.015 | 0.20 | 7.322 | 0.016 | 0.21 |
| | pCO2 (mmHg) | 44.3 | 1.1 | 2.6 | 43.7 | 1.2 | 3.8 |
| | pO2 (mmHg) | 150.8 | 4.5 | 3.0 | 148.7 | 4.1 | 2.8 |
| | Na+ (mmol/L) | 139.1 | 0.9 | 0.7 | 139.2 | 0.9 | 0.6 |
| | K+ (mmol/L) | 3.24 | 0.06 | 1.9 | 3.35 | 0.08 | 2.5 |
| | Ca++ (mmol/L) | 1.19 | 0.03 | 2.7 | 1.17 | 0.03 | 2.3 |
| | Cl- (mmol/L) | 103.5 | 1.8 | 1.7 | 103.5 | 2.0 | 1.1 |
| | Hct (%) | 47.1 | 0.8 | 1.7 | 48.1 | 1.1 | 2.2 |
#### Venous whole blood within-run precision
#### d. Comparison of Syringe and Capillary Measurement
Whole blood venous samples from healthy volunteers were evaluated to demonstrate the measurement results from samples introduced from syringes and capillary tubes are equivalent. Six levels of blood were prepared to provide a range of measurement values for electrolytes and hematocrit, and tonometered with precision gas mixtures to attain a range of pCO2 and PO2 values and analyzed over numerous days on three (3) EDAN i15 instruments with 10 replicates on each of six (6) levels of prepared blood samples introduced from capillary tubes as compared to introduced from syringes.
| Measurand | N | sample range<br>tested | Slope | Intercept | r-value | Bias |
|---------------|-----|------------------------|--------|-----------|---------|--------|
| pH | 180 | 7.036 - 7.689 | 0.9855 | 0.1056 | 0.9936 | -0.001 |
| pCO2 (mmHg) | 180 | 6.7 - 119.1 | 0.9870 | 0.0377 | 0.9953 | -0.6 |
| pO2 (mmHg) | 180 | 15.0 - 749.0 | 1.0177 | -1.0704 | 0.9941 | 3.3 |
| Na+ (mmol/L) | 180 | 105 - 176 | 0.9716 | 4.3119 | 0.9954 | 0.3 |
| K+ (mmol/L) | 180 | 2.8 - 10.6 | 0.9848 | 0.0339 | 0.9981 | -0.04 |
| Ca++ (mmol/L) | 180 | 0.86 - 3.16 | 0.9813 | -0.0213 | 0.9928 | -0.05 |
| Cl- (mmol/L) | 180 | 76 - 164 | 0.9844 | 1.6184 | 0.9922 | -0.2 |
| Hct (%) | 180 | 16 - 64 | 0.9933 | 0.8013 | 0.9968 | 0.5 |
Correlation in whole blood samples introduced from Capillary vs Syringe
#### e. Specificity / Interference Testing
An interference study was performed according to the CLSI guideline EP7-A2. The study used spiked and diluted human whole blood samples containing potential interferents for pH, pO2, pCO2, sodium, potassium, chloride, calcium, and hematocrit. Seven replicates of both the spiked and the unspiked sample were tested on two i15 Blood Gas and Chemistry System with one lot of test cartridges. The results of each replicate of the spiked and unspiked samples were compared and the percent bias between the results was calculated for all analytes except for pH for which absolute difference was calculated. The acceptance criteria used to evaluate if the substances significantly interfere with each analyte is shown in the table below:
| Measurand | pH | pCO2 | pO2 | Na+ | K+ | Ca++ | Cl- | Hct |
|------------------------|------------------------|-----------------|-----------------|-----------------|-----------------|-----------------|-----------------|-----------------|
| Evaluation<br>Method | Absolute<br>difference | Percent<br>Bias | Percent<br>bias | Percent<br>bias | Percent<br>bias | Percent<br>bias | Percent<br>bias | Percent<br>bias |
| Acceptance<br>Criteria | < 0.02<br>pH Units | < 8% | < 9% | < 3% | < 10% | < 10% | < 5% | < 6% |
{17}------------------------------------------------
| Interfering Substance | Highest Concentration<br>Tested | Analytes Tested |
|---------------------------|---------------------------------|----------------------------------------|
| Albumin | 1.5 g/dL (added) | Na+, K+, Ca++, Cl-, Hct |
| Ascorbic acid (Vitamin C) | 6.02 mg/dL | pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Benzylkonium Chloride | 0.80 mg/dL | Na+, K+, Ca++, Cl-, Hct |
| Bilirubin | 15 mg/dL | pH, pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Calcium Chloride | 27.75 mg/dL | pH, pCO2, pO2, Na+, K+, Hct |
| Cholesterol | 500 mg/dL | pH, pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Cysteine | 12.12 mg/dL | pH, pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Hemoglobin | 500 mg/dL | pH, pCO2, pO2, Na+, Ca++, Cl-, Hct |
| Heparin lithium salt | 11.76 mg/dL | pH, pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Hydroxybutyrate | 208 mg/dL | pCO2, pO2, Na+, Ca++, Cl-, Hct |
| Hydroxyurea | 182.52 mg/dL | pH, pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Ibuprofen | 50 mg/dL | pH, pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Lactic acid | 59.40 mg/dL | pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Lithium | 2.24 mg/dL | pH, pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
| Ofloxacin | 6 µg/mL | pH, pCO2, Na+, K+, Ca++, Cl-, Hct |
| Triglycerides | 1500 mg/dL | pH, pCO2, pO2, Na+, K+, Ca++, Cl-, Hct |
The following table represents substances that were tested without significant effects on test results:
The following table represents substances that were tested that demonstrated a significant interference on test results. The table includes the range of % bias (absolute difference for pH) that was calculated between the 7 replicates spiked and unspiked samples for each analyte.
| Measurand | Interfering Substance | Concentration<br>Tested | Blood Sample<br>Value | Absolute Difference<br>in pH Units |
|-----------|-----------------------------------|-------------------------|-----------------------|------------------------------------|
| pH | Acetaminophen | 20.01 mg/dL | pH: 7.5<br>pH: 7.3 | -0.034<br>< 0.02 |
| | Bromide | 185.20 mg/dL | pH: 7.5<br>pH: 7.3 | -0.027<br>< 0.02 |
| | Calcium Chloride | 27.27 mg/dL | pH: 7.5<br>pH: 7.3 | < 0.02 <br>-0.024 |
| | Ethanol | 400 mg/dL | pH: 7.5<br>pH: 7.3 | -0.024<br>< 0.02 |
| | Hematocrit | 20% PCV | pH: 7.5<br>pH: 7.3 | -0.022<br>< 0.02 |
| | Heparin | 58.82 mg/dL | pH: 7.5<br>pH: 7.3 | < 0.02 <br>-0.034 |
| | Hydroxycarbamide<br>(Hydroxyurea) | 182.52 mg/dL | pH: 7.5<br>pH: 7.3 | < 0.02 <br>-0.031 |
| | Iodide | 37.94 mg/dL | pH: 7.5<br>pH: 7.3 | < 0.02 <br>-0.025 |
| | Potassium Chloride | 59.64 mg/dL | pH: 7.5<br>pH: 7.3 | -0.036<br>< 0.02 |
{18}------------------------------------------------
| Measurand | Interfering Substance | Concentration Tested | Blood Sample Value | Absolute Difference in pH Units |
|-----------|-----------------------|----------------------|--------------------|---------------------------------|
| pH | Sodium Chloride | 117 mg/dL | pH: 7.5 | -0.024 |
| | Sodium Chloride | 117 mg/dL | pH: 7.3 | -0.021 |
| pH | Sodium Oxalate | 168 mg/dL | pH: 7.5 | < 0.02 |
| pH | Sodium Oxalate | 168 mg/dL | pH: 7.3 | -0.032 |
| Measurand | Interfering Substance | Concentration Tested | Blood Sample Value | % Difference |
| pCO2 | Acetaminophen | 20.01 mg/dL | pCO2: 70 mmHg | 10.15% |
| | Acetaminophen | 20.01 mg/dL | pCO2: 40 mmHg | < 8% |
| pCO2 | Acetylsalicylic acid | 39.09 mg/dL | pCO2: 70 mmHg | 8.01% |
| | Acetylsalicylic acid | 39.09 mg/dL | pCO2: 40 mmHg | < 8% |
| pCO2 | Ethanol | 400 mg/dL | pCO2: 70 mmHg | 9.54% |
| | Ethanol | 400 mg/dL | pCO2: 40 mmHg | < 8% |
| pCO2 | Iodide | 37.94 mg/dL | pCO2: 70 mmHg | -10.00% |
| | Iodide | 37.94 mg/dL | pCO2: 40 mmHg | -8.69% |
| pCO2 | Potassium Chloride | 59.64 mg/dL | pCO2: 70 mmHg | 11.15% |
| | Potassium Chloride | 59.64 mg/dL | pCO2: 40 mmHg | < 8% |
| pCO2 | Bicarbonate (NaHCO3) | 294 mg/dL | pCO2: 70 mmHg | -18.40% |
| | Bicarbonate (NaHCO3) | 294 mg/dL | pCO2: 40 mmHg | -17.44% |
| pO2 | Acetylsalicylic acid | 39.09 mg/dL | pO2: 100 mg/dL | < 9% |
| | Acetylsalicylic acid | 39.09 mg/dL | pO2: 70 mg/dL | -10.11% |
| pO2 | Hematocrit | 20% PCV | pO2: 100 mg/dL | < 9% |
| | Hematocrit | 20% PCV | pO2: 70 mg/dL | 12.13% |
| pO2 | Lactic Acid | 90 mg/dL | pO2: 100 mg/dL | < 9% |
| | Lactic Acid | 90 mg/dL | pO2: 70 mg/dL | 9.74% |
| pO2 | pCO2 | 60 mmHg | pO2: 100 mg/dL | < 9% |
| | pCO2 | 60 mmHg | pO2: 70 mg/dL | 9.60% |
| pO2 | Salicylic acid | 59.94 mg/dL | pO2: 100 mg/dL | < 9% |
| | Salicylic acid | 59.94 mg/dL | pO2: 70 mg/dL | 14.98% |
| pO2 | Sodium Chloride | 117 mg/dL | pO2: 100 mg/dL | < 9% |
| | Sodium Chloride | 117 mg/dL | pO2: 70 mg/dL | 9.22% |
| pO2 | Bicarbonate (NaHCO3) | 294 mg/dL | pO2: 100 mg/dL | -14.46% |
| | Bicarbonate (NaHCO3) | 294 mg/dL | pO2: 70 mg/dL | < 9% |
{19}------------------------------------------------
| Measurand | Interfering Substance | Concentration Tested | Blood Sample<br>Value | % Difference |
|-----------|--------------------------|------------------------|-----------------------|--------------|
| Na+ | Calcium Chloride | 55.50 mg/dL | Na+: 150 mmol/L | 3.26% |
| | | | Na+: 130 mmol/L | 4.99% |
| | Dobutamine hydrochloride | 22.30 mg/dL | Na+: 150 mmol/L | < 3% |
| | | | Na+: 130 mmol/L | 5.62% |
| K+ | Acetylsalicylic acid | 65.22 mg/dL | K+: 5 mmol/L | 17.21% |
| | | | K+: 3 mmol/L | < 10% |
| | Dobutamine hydrochloride | 22.30 mg/dL | K+: 5 mmol/L | < 10% |
| | | | K+: 3 mmol/L | 13.97% |
| | Hydroxybutyrate | 208 mg/dL | K+: 5 mmol/L | 12.28% |
| | | | K+: 3 mmol/L | 16.87% |
| | Iodide | 37.94 mg/dL | K+: 5 mmol/L | < 10% |
| | | | K+: 3 mmol/L | 24.20% |
| | Lactic Acid | 90 mg/dL | K+: 5 mmol/L | 13.05% |
| | | | K+: 3 mmol/L | < 10% |
| | <i>p</i> CO2 | 60 mmHg | K+: 5 mmol/L | 10.13% |
| | | | K+: 3 mmol/L | < 10% |
| | Salicylic acid | 29.97 mg/dL | K+: 5 mmol/L | 10.54% |
| | | | K+: 3 mmol/L | < 10% |
| | Bicarbonate (NaHCO3) | 294 mg/dL | K+: 5 mmol/L | 21.51% |
| | | | K+: 3 mmol/L | 10.31% |
| Ca++ | Halothane | 14.98 mg/dL | Ca++: 2 mmol/L | < 10% |
| | | | Ca++: 1 mmol/L | 11.20% |
| | Heparin | 58.82 mg/dL | Ca++: 2 mmol/L | -12.68% |
| | | | Ca++: 1 mmol/L | -14.06% |
| | Magnesium Chloride | 47.61 mg/dL | Ca++: 2 mmol/L | 13.91% |
| | | | Ca++: 1 mmol/L | 16.86% |
| | Bicarbonate (NaHCO3) | 294 mg/dL | Ca++: 2 mmol/L | -27.07% |
| | | | Ca++: 1 mmol/L | -21.17% |
| | Phosphate (NaH2PO4) | 24 mg/dL | Ca++: 2 mmol/L | -12.14% |
| | | | Ca++: 1 mmol/L | -12.34% |
| | Sodium Oxalate | 168 mg/dL | Ca++: 2 mmol/L | -94.16% |
| | | | Ca++: 1 mmol/L | -86.21% |
| Cl- | Acetylsalicylic acid | 39.09 mg/dL | Cl-: 110 mmol/L | 7.72% |
| | | | Cl-: 90 mmol/L | 5.11% |
| | Albumin | 3 g/dL (added) | Cl-: 110 mmol/L | 5.13% |
| | | Approx. 8 g/dL (total) | Cl-: 90 mmol/L | 7.19% |
| | Bromide (NaBr) | 185.20 mg/dL | Cl-: 110 mmol/L | < 5% |
| | | | Cl-: 90 mmol/L | 6.38% |
{20}------------------------------------------------
| Measurand | Interfering Substance | Concentration Tested | Blood Sample Value | % Difference |
|-----------|--------------------------|------------------------------------------|--------------------|--------------|
| Cl- | Iodide | 37.94 mg/dL | Cl-: 110 mmol/L | -7.60% |
| | Iodide | 37.94 mg/dL | Cl-: 90 mmol/L | -11.76% |
| | Potassium Thiocyanate | 20.06 mg/dL | Cl-: 110 mmol/L | 10.81% |
| | Potassium Thiocyanate | 20.06 mg/dL | Cl-: 90 mmol/L | 13.41% |
| | Salicylic acid | 29.97 mg/dL | Cl-: 110 mmol/L | 11.28% |
| | Salicylic acid | 29.97 mg/dL | Cl-: 90 mmol/L | 9.88% |
| | Bicarbonate (NaHCO3) | 294 mg/dL | Cl-: 110 mmol/L | 7.72% |
| | Bicarbonate (NaHCO3) | 294 mg/dL | Cl-: 90 mmol/L | 8.79% |
| | Sodium Oxalate | 168 mg/dL | Cl-: 110 mmol/L | 5.01% |
| | Sodium Oxalate | 168 mg/dL | Cl-: 90 mmol/L | 6.47%…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.