K110648 · Nova Biomedical Corporation · CHL · Aug 29, 2011 · Clinical Chemistry
Device Facts
Record ID
K110648
Device Name
NOVA STAT PROFILE PHOX ULTRA ANALYZER SYSTEM
Applicant
Nova Biomedical Corporation
Product Code
CHL · Clinical Chemistry
Decision Date
Aug 29, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1120
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Nova Stat Profile pHOx Ultra Analyzer with CO-Oximeter is intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), total Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) in heparinized whole blood; Na+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma. Total Bilirubin (tBil) was not evaluated on neonatal samples. The Nova Stat Profile pHOx Ultra Analyzer without CO-Oximeter is intended for in vitro diagnostic use by health care professionals and/or point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin (Hb) in heparinized whole blood; Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma. The intended use of the Nova STP pHOx Ultra Calibrator Cartridge is for the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin (Hb) in heparinized whole blood; Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma. The intended use of the Nova Stat Profile pHOx Ultra Analyzer CO-Oximeter Calibrator Cartridge with Bilirubin and Deproteinizing Solution is for the quantitative determination of total Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) in human blood using the Nova Stat Profile pHOx Ultra Analyzer System with CO-Oximeter. Nova Stat Profile pHOx Ultra Analyzer CO-Oximeter Controls and Autocartridge QC are intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Nova Stat Profile pHOx Ultra Analyzer.
Device Story
Nova Stat Profile CCX 1+ System; point-of-care blood analyzer; measures analytes in whole blood. Modifications include touchscreen/printer resizing, sample probe length adjustment, sampler motor update, peristaltic pump material changes, and deproteinizing algorithm update. Device operates via automated sample aspiration and electrochemical/biosensor analysis. Healthcare professionals use output for clinical decision-making regarding patient metabolic status. Fundamental scientific technology remains unchanged from predicate.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by software validation and performance verification testing confirming that hardware/software modifications did not affect system safety, efficacy, or performance.
Technological Characteristics
Point-of-care blood analyzer. Hardware includes touchscreen, printer, sample probe, sampler motor, and peristaltic pump. Sensing principle: electrochemical/biosensor. Software includes deproteinizing algorithm. Design controls per 21 CFR 820.30.
Indications for Use
Indicated for use by healthcare professionals for quantitative measurement of analytes in whole blood samples in clinical settings.
Regulatory Classification
Identification
A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Predicate Devices
Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System (K061830)
Related Devices
K203790 — GEM Premier 5000 · Instrumentation Laboratory CO · Aug 20, 2021
K092686 — ABL90 FLEX SERIES · Radiometer Medical Aps · Aug 6, 2010
K223090 — GEM Premier ChemSTAT · Instrumentation Laboratory CO · Jan 27, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
K110648
AUG 2 9 2011
## 510(k) Summary
| 510(K) Owner: | Nova Biomedical Corporation |
|----------------------|---------------------------------------|
| Registration Number: | 1219029 |
| Address: | 200 Prospect St.<br>Waltham, MA 02454 |
| Phone: | 781-894-0800 |
| Fax Number: | 784-891-4806 |
| Contact Person: | Paul W. MacDonald |
| Date Prepared: | 08/24/2011 |
Proprietary Name: Nova Stat Profile pHOx Ultra Analyzer System
Common or Usual Name: Blood Gas/Electrolyte/Metabolite/CO-Oximetry Analyzer
### Classification Name: Multiple
| Classification Names: | Class No. | Reg. No. | Class |
|-----------------------------------|-----------|----------|-------|
| Blood Gases and Blood pH system | 75CHL | 862.1120 | II |
| Sodium Test System | 75JGS | 862.1665 | II |
| Potassium Test System | 75CEM | 862.1600 | II |
| Calcium Test System | 75JFP | 862.1145 | II |
| Chloride Test System | 75CGZ | 862.1170 | II |
| Creatinine Test System | 75CGL | 862.1225 | II |
| Magnesium Test System | 75CFA | 862.1495 | I |
| Glucose Test System | 75CGA | 862.1345 | II |
| Lactic Acid Test System | 75KHP | 862.1450 | I |
| Urea Nitrogen System | 75CDS | 862.1770 | II |
| Oxygen Saturation (SO2%) | 91JKS | 862.3220 | I |
| Hematocrit Measuring Device | 81JPI | 864.6400 | II |
| Hemoglobin System | 81GKR | 864.5620 | II |
| Oxyhemoglobin | 91JKS | 862.3220 | I |
| Carboxyhemoglobin | 81GHS | 864.7425 | II |
| Methemoglobin | 81JBJ | 864.7360 | II |
| Hemoglobin Standards and Controls | 81JPK | 864.8625 | II |
| Hemoglobin Calibrator | 81KRZ | 864.8165 | II |
| Calibrators | 75JIX | 862.1150 | II |
| Quality Control Materials | 81JJS | 862.1660 | I |
Product Codes: CHL, JGS, CEM, JFP, CGZ, CGL, CFA, CGA,KHP,CDS, JKS, GHS, JPI, GKR, JKS, JBJ, JPK,KRZ,JIX, JJS
Predicate Device: Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System (K061830)
{1}------------------------------------------------
#### Device Description:
As in the Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System (K061830) predicate device, the Nova Stat Profile pHOx Ultra Analyzer System combines Blood Gas/pH, Chemistry, bilirubin and CO-Oximetry testing into one Point-of-Care Analyzer. This device is analyte configurable by the end user, based on tests needed.
As with the predicate, this device is microprocessor-based and incorporates:
- traditional electrode technology to measure blood pH, pCO2, pO2 .
- Nova Biomedical proprietary optical reflectance technology for the measurement of oxygen . saturation
- ion selective electrode technology to measure blood sodium, chloride, ionized . calcium, ionized magnesium
- enzyme/amperometric technology for glucose, urea nitrogen, lactate and creatinine . measurements
- conductivity/Na+ correction for hematocrit .
- multi-wavelength reflectance/conductivity correction for hemoglobin. .
Calibration standards with dissolved gases are provided in sealed pouches eliminating the need for users to calibrate the blood qas electrodes using external compressed gas cylinders. Quality control materials are available as external ampules and as internal auto-cartridge quality control packs. Sampling, calibration and quality control are fully automated.
Nova will market the Nova Stat Profile pHOx Ultra Analyzer System in two configurations. The proposed Nova Stat Profile pHOx Ultra Analyzer System with CO-Ox module (Catalog #42013) will be offered with all the parameters listed above. A second configuration will be offered, called the Nova Stat Profile pHOx Ultra Analyzer System without the CO-Ox module (Catalog #42014). This configuration will not have the capability to measure Oxyhemoglobin (O₂Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), and Reduced Hemoglobin (HHb) or bilirubin.
#### Summary of the Technological Characteristics:
The Nova Stat Profile pHOx Ultra Analyzer System is identical to the previously cleared for market Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System in intended use. Modifications were made necessary due to the obsolescence and availability of certain components and hardware, which required minor modifications to the device software to support these changes. The software algorithms used in performing the analysis of samples are unchanged from the predicate device. Minor modifications have been made to simplify the User Interface. The same reagents, calibrators, and controls are used for all of the analytes. The comparison table below compares the Nova Stat Profile pHOx Ultra Analyzer System to the predicate Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System.
The results of software validation and performance verification testing confirmed that the modifications made to the hardware and software of the Nova Stat Profile pHOx Ultra Analyzer System did not affect the safety, efficacy or performance of the system is substantially equivalent to the predicate device.
{2}------------------------------------------------
## Comparison Table
| Characteristic | Predicate<br>Stat Profile Critical Care Xpress with COOX<br>Analyzer System | Proposed<br>Stat Profile pHOx Ultra with COOX Analyzer<br>System |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Stat Profile Critical Care Xpress (STP<br>CCX) Analyzer with CO-Oximeter is intended<br>for <i>in vitro</i> diagnostic use by health care<br>professionals and for point-of-care usage in the<br>quantitative determination of pH, PCO2, PO2,<br>SO2%, Hematocrit (Hct), total Hemoglobin<br>(tHb), Oxyhemoglobin (O2Hb),<br>Carboxyhemoglobin (COHb), Methemoglobin<br>(MetHb), Deoxyhemoglobin (HHb), and total<br>bilirubin (tBil) in heparinized whole blood; Na+,<br>K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate<br>(Lac), BUN (Urea), and Creatinine (Creat) in<br>heparinized whole blood, serum, or plasma.<br>Total Bilirubin (tBil) was not evaluated on<br>neonatal<br>The STP CCX Analyzer without CO-Oximeter<br>is intended for <i>in vitro</i> diagnostic use by health<br>care professionals and/or point-of-care usage<br>in the quantitative determination of pH, PCO2,<br>PO2, SO2%, Hematocrit (Hct), Hemoglobin<br>(Hb) in heparinized whole blood; Na+, K+, Cl-,<br>Ca++, Mg++, Glucose (Glu), Lactate (Lac),<br>BUN (Urea), and Creatinine (Creat) in<br>heparinized whole blood, serum, or plasma<br>samples.<br>The intended use of the Nova STP CCX CO-<br>Oximeter Calibrator Cartridge with Bilirubin and<br>Deproteinizing Solution is for the quantitative<br>determination of total Hemoglobin (tHb),<br>Oxyhemoglobin (O2Hb), Carboxyhemoglobin<br>(COHb), Methemoglobin (MetHb),<br>Deoxyhemoglobin (HHb), and total bilirubin<br>(tBil) in human blood using the STP CCX<br>Analyzer with CO-Oximeter.<br>Nova STP CCX CO-Oximeter Controls and<br>Autocartridge QC are intended for <i>in vitro</i><br>diagnostic use by healthcare professionals for<br>monitoring the performance of Nova STP CCX<br>Analyzers. | The Stat Profile pHOx Ultra Analyzer with CO-<br>Oximeter is intended for <i>in vitro</i> diagnostic use<br>by health care professionals and for point-of-<br>care usage in the quantitative determination of<br>pH, PCO2, PO2, SO2%, Hematocrit (Hct), total<br>Hemoglobin (tHb), Oxyhemoglobin (O2Hb),<br>Carboxyhemoglobin (COHb), Methemoglobin<br>(MetHb), Deoxyhemoglobin (HHb), and total<br>bilirubin (tBil) in heparinized whole blood; Na+,<br>K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate<br>(Lac), BUN (Urea), and Creatinine (Creat) in<br>heparinized whole blood, serum, or plasma.<br>Total Bilirubin (tBil) was not evaluated on<br>neonatal samples.<br>The Stat Profile pHOx Ultra Analyzer without<br>CO-Oximeter is intended for <i>in vitro</i> diagnostic<br>use by health care professionals and/or point-<br>of-care usage in the quantitative determination<br>of pH, PCO2, PO2, SO2%, Hematocrit (Hct),<br>Hemoglobin (Hb) in heparinized whole blood;<br>Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu),<br>Lactate (Lac), BUN (Urea), and Creatinine<br>(Creat) in heparinized whole blood, serum, or<br>plasma.<br>The intended use of the Nova STP pHOx Ultra<br>Calibrator Cartridge is for the quantitative<br>determination of pH, PCO2, PO2, SO2%,<br>Hematocrit (Hct), Hemoglobin (Hb) in<br>heparinized whole blood; Na+, K+, Cl-, Ca++,<br>Mg++, Glucose (Glu), Lactate (Lac), BUN<br>(Urea), and Creatinine (Creat) in heparinized<br>whole blood, serum, or plasma.<br>The intended use of the Nova STP pHOx<br>Ultra/CCX CO-Oximeter Calibrator Cartridge<br>with Bilirubin and Deproteinizing Solution is for<br>the quantitative determination of total<br>Hemoglobin (tHb), Oxyhemoglobin (O2Hb),<br>Carboxyhemoglobin (COHb), Methemoglobin<br>(MetHb), Deoxyhemoglobin (HHb), and total<br>bilirubin (tBil) in human blood using STP pHOx<br>Ultra/CCX Analyzers with CO-Oximeter<br>Nova STP pHOx Ultra/CCX ABG and CO-<br>Oximeter Controls and Autocartridge QC are<br>intended for <i>in vitro</i> diagnostic use by<br>healthcare professionals for monitoring the<br>performance of Nova STP pHOx Ultra/CCX<br>Analyzers. |
| Analyzer | Nova Stat Profile Critical Care Xpress (CCX),<br>Model 1+ Analyzer System with CO-OX -<br>37413<br>Nova Stat Profile Critical Care Xpress (CCX),<br>Model 1+ Analyzer System - 35942 | Nova Stat Profile pHOx Ultra Analyzer System<br>with CO-OX - 42013<br>Nova Stat Profile pHOx Ultra Analyzer System<br>- 42014 |
| Consumables | | |
| Calibrators | ABG, SO2, CO-Ox | Same |
| Controls | ABG Gas, ABG Chem, CO-Ox, External Lin<br>Standards | Same |
| Sensors | PH, PCO2, PO2, SO2, Na+, K+, Cl-, Ca++,<br>Mg++, Glu, Lac, Creat, BUN, REF | Same |
| Membranes | Glu, Lac, Creat, BUN, PCO2, PO2 | Same |
| Characteristic | Predicate<br>Stat Profile Critical Care Xpress with COOX<br>Analyzer System | Proposed<br>Stat Profile pHOx Ultra with COOX Analyzer<br>System |
| Pump Tubing | ABG, CO-Ox | Same |
| Acceptable Samples | Syringe, Capillary | Same |
| Sample Volumes | | |
| ABG Only | 150μL | Same |
| CO-Ox Only | 100μL | Same |
| Combined Analysis | 210µL | Same |
| Micro Sample<br>Volumes | | |
| Full | 120µL | Same |
| ABG | 60µL | Same |
| ABG Plus | 70µL | Same…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches