IDS-iSYS Intact PTHN

K161158 · Immunodiagnostic Systems , Ltd. · CEW · Jan 31, 2017 · Clinical Chemistry

Device Facts

Record IDK161158
Device NameIDS-iSYS Intact PTHN
ApplicantImmunodiagnostic Systems , Ltd.
Product CodeCEW · Clinical Chemistry
Decision DateJan 31, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1545
Device ClassClass 2

Intended Use

The IDS-iSYS Intact PTHⁿ assay is an in vitro diagnostic device intended for the quantitative determination of intact PTH in human serum or plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism.

Device Story

IDS-iSYS Intact PTHⁿ assay is a chemiluminescent magnetic particle immunoassay; utilizes polyclonal goat anti-human antibodies; capture antibody recognizes C-terminal region (39-84); detection antibody (acridinium-labeled) recognizes N-terminal region (1-84); system also detects large PTH fragment (7-84). Input: human serum or plasma samples. Process: patient samples incubated with antibodies; streptavidin-coated magnetic particles added; magnet captures antibody-antigen complexes; wash steps performed; reagent added to trigger chemiluminescence. Output: light emission proportional to PTH concentration; measured by IDS-iSYS Multi-Discipline Automated System. Used in clinical laboratory settings; operated by trained laboratory personnel. Healthcare providers use quantitative results to assess calcium metabolism disorders; aids in differential diagnosis of hypercalcemia and hypocalcemia.

Clinical Evidence

No clinical trials performed. Evidence consists of analytical bench testing: precision (n=80 per sample, CV% 1.5-9.9%), linearity (R²=1.00, range 5-3500 pg/mL), LoB/LoD/LoQ (0.9/2.3/4.5 pg/mL), interference/cross-reactivity studies, and method comparison against predicate (n=312, y=1.02x-0.7, r=1.00). Reference interval established using 129 healthy subjects.

Technological Characteristics

Chemiluminescent immunoassay; magnetic particle separation. Reagents: streptavidin-coated magnetic particles, acridinium-labeled anti-PTH (13-34), biotinylated anti-PTH (39-84). Standards: CLSI EP05-A3, EP06-A, EP07-A2, EP09-A3, EP17-A2, EP28-A3, CEN 13640. Form factor: reagent cartridge for IDS-iSYS automated system. Connectivity: integrated with IDS-iSYS analyzer.

Indications for Use

Indicated for the quantitative determination of intact PTH in human serum or plasma to aid in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism.

Regulatory Classification

Identification

A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 IMMUNODIAGNOSTIC SYSTEMS LTD MICK HENDERSON REGULATORY AFFAIRS OFFICER 10 DIDCOT WAY BOLDON BUSINESS PARK BOLDON NE35 9PD, GREAT BRITAIN January 31, 2017 Re: K161158 Trade/Device Name: IDS-iSYS Intact PTH® Regulation Number: 21 CFR 862.1545 Regulation Name: Parathyroid hormone test system Regulatory Class: II Product Code: CEW Dated: January 9, 2017 Received: January 11, 2017 Dear Mick Henderson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Katherine Serrano -S For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 510(k) Number (if known) k161158 Device Name IDS-iSYS Intact PTHN Indications for Use (Describe) IDS-iSYS Intact PTHN The IDS-iSYS Intact PTH+ assay is an in vitro diagnostic device intended for the quantitative determination of intact PTH in human serum or plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) | |---|----------------------------------------------|--|---------------------------------------------| | X | | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) SUMMARY IDS-iSYS Intact PTHN Assay This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92. | 1. | 510(k) Number: | k161158 | |----|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------| | 2. | Submitter: | Immunodiagnostic Systems Ltd<br>10 Didcot Way<br>Boldon Business Park<br>Boldon<br>Tyne and Wear<br>NE35 9PD<br>United Kingdom | | 3. | Contact Person: | Mick Henderson<br>Phone: +44(0)191 5190660<br>Fax: +44(0) 191 5190760<br>Email: mick.henderson@idsplc.com | | 4. | Secondary Contact: | Heather Pham<br>Phone: +1(480)-414-7175<br>Fax: +44(0) 191 5190760<br>Email: heather.pham@idsplc.com | | 5. | Date Prepared: | 09 January 2017 | | 6. | Proprietary and Established Name:<br>IDS-iSYS Intact PTHN | | | 7. | Regulatory Information: | Classification: 21CFR 862.1545<br>Device Class Class II<br>Product Code: CEW<br>Panel: Clinical Chemistry (75) | {4}------------------------------------------------ #### 8. Device Descriptions: The IDS-iSYS Intact PTH" assay is based on chemiluminescence technology. 100 uL of patient sample is incubated with a biotinylated polyclonal anti-PTH (39-84) antibody and an acridinium labelled PTH (13-34) antibody. Streptavidin labelled magnetic particles are added prior to a second incubation step. The magnetic particles are captured using a magnet and a wash step performed to remove any unbound analyte. Trigger reagents are added; the resulting light emitted by the acridinium label is directly proportional to the concentration of Intact PTH in the original sample. IDS-iSYS Intact PTH reagent kit consists of one (1) Immunoassay Cartridge, two (2) vials each of 2 concentration Calibrator levels and a mini-CD containing the Instructions For Use (IFU), CRY files and Certificate of Analysis. IDS-iSYS Intact PTHN Cartridge, sufficient for 100 tests, consists of reagents provided in individual compartment within a plastic container called the Cartridge. The IDS-iSYS Intact PTHN Cartridge contains the following ready to use reagents: - · Magnetic particles coated with streptavidin in a phosphate buffer containing sodium azide as preservative (<0.1%), 1 bottle. - Goat polyclonal antibody against 13-34 PTH labelled with an acridinium ester derivative, in buffer containing goat serum with sodium azide as preservative (<0.1%), 1 bottle. - Goat polyclonal antibody against 39-84 PTH 1abelled with biotin, in buffer containing bovine and goat proteins with sodium azide as preservative (<0.1%), 1 bottle. IDS-iSYS Intact PTH Calibrators are included in the reagent kit. - · Calibrator A: Two vials of lyophilized porcine serum matrix buffer containing low level PTH and sodium azide as preservative >1% (w/w%). - · Calibrator B: Two vials of lyophilized porcine serum matrix buffer containing high level PTH and sodium azide as preservative >1% (w/w%). The submission is due to a new source of antibody for the assay. {5}------------------------------------------------ #### 9. Predicate Devices: IDS-iSYS Intact PTH, K103325 #### 10. Special Conditions for Use: For in vitro diagnostic use. Rx Only ### 11. Special instrument Requirements: IDS-iSYS Multi-Discipline Automated System (K091849) #### 12. Intended Use: The IDS-iSYS Intact PTHN assay is an in vitro diagnostic device intended for the quantitative determination of intact PTH in human serum or plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism. {6}------------------------------------------------ # 13. Substantial Equivalence Information A comparison of the similarities and differences between the modified IDS-iSYS Intact PTH® assay and the original IDS-iSYS Intact PTH assay (K103325) are provided in the following table: | Assay Similarities and Differences | | | |------------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| | Characteristics | Predicate Device<br>IDS-iSYS Intact PTH<br>(K103325) | Candidate Device<br>IDS-iSYS Intact PTHN | | Intended use | intended for the quantitative<br>determination of PTH in human<br>serum or plasma. | Same | | Measured analyte | Intact PTH | Same | | Assay type | Chemiluminescence<br>immunoassay | Same | | Antibody | Goat polyclonal<br>a) anti- PTH (13-34)<br>b) anti-PTH (39-84) | Same | | Sample matrix | Plasma (K2-EDTA,<br>lithium heparin) and<br>serum | Same | | Sample size | 100 μL | Same | | Sample<br>handling/processing | Automated<br>IDS-iSYS Multi-Discipline<br>Automated System | Same | | Calibrator | 2 levels<br>Included with kit | Same | | Calibration interval | 2 point calibration curve stable<br>for 21 days | 2 point calibration curve stable<br>for 15 days | | Reagent storage | On-board or at 2-8°C | Same | | Measurement range | 5 to 5000 pg/mL | 5 to 3500 pg/mL | | Reference interval | 11.5 to 78.4 pg/mL | 10.3 to 80.5 pg/mL | | Assay Similarities and Differences | | | | Characteristics | Predicate Device<br>IDS-iSYS Intact PTH<br>(K103325) | Candidate Device<br>IDS-iSYS Intact PTHN | | Interfering Substances | | | | - Bilirubin, conjugated | N/A | 22 mg/dL | | - Bilirubin, unconjugated | 20 mg/dL | 40 mg/dL | | - Biotin | 300 nmol/L | Same | | - Cholesterol | N/A | 395 mg/dL | | - HAMA | 1000 ng/mL | Same | | - Hemoglobin | 250 mg/dL | 500 mg/dL | | - Rheumatoid Factor | 1500 IU/mL | 1836 IU/mL | | - Total Protein | N/A | 10 g/dL | | - Triglycerides | 3000 mg/dL | Same | | - Acetaminophen | N/A | 200 µg/mL | | - Carbamazepine | N/A | 200 µg/mL | | - Ibuprofen | N/A | 500 µg/mL | | Specificity | | | | - PTH (1-84) | 100% | Same | | - PTH (7-84) | 60% | 83.9% | | - PTH (1-34) | 0.5% | <0.01% | | - PTH (39-84) | Not detectable | <0.01% | | - PTH (53-84) | 2% | <0.01% | | - PTH (39-68) | N/A | <0.01% | | - PTH (44-68) | 2% | <0.01% | | - Human Calcitonin | 4% | <0.01% | | - CTX-1 (β CrossLaps) | 2% | <0.01% | | - Osteocalcin | 2% | <0.01% | | Hook Effect | Not observed up to<br>95,000 pg/mL | Not observed up to<br>100,000 pg/mL | | Sensitivity | | | | - Limit of Blank (LoB) | 1.2 pg/mL | 0.9 pg/mL | | - Limit of Detection (LoD) | 2.5 pg/mL | 2.3 pg/mL | | - Limit of Quantitation (LoQ) | 4.5 pg/mL | 4.5 pg/mL | | Precision | in the range of 13.3 to<br>3807 pg/mL, n =80 | in the range of 15.4 to<br>2229 pg/mL, n =80 | | - Within-run | 1.1% to 6.3% | 1.5% to 9.9% | | - Total | 4.1% to 8.2% | 1.8% to 9.9% | | Linearity | $y = 1.002 x – 4.748;$<br>$R^2 = 1.00$ | K2 EDTA:<br>$y = 0.96 x − 0.1; R^2 = 1.00$<br>Serum:<br>$y = 1.02x – 0.2; R^2 = 1.00$ | {7}------------------------------------------------ {8}------------------------------------------------ ### 14. Performance Characteristics # Method Comparison: The IDS-iSYS Intact PTH" Assay was compared against a commercially available quantitative chemiluminescence Intact PTH assay, following CLSI EP-9A2, "Method Comparison and Bias Estimation Using Patient Samples". A total of 312 serum samples (291 native, 21 spiked), selected to represent a wide range of Intact PTH concentrations [6.3 - 3378.4 pg/mL], was assayed by each method. Deming regression analysis was performed on the comparative data: | n | Slope | 95% CI | Intercept<br>(pg/mL) | 95% Cl | Correlation coefficient<br>(r) | |-----|-------|--------------|----------------------|-------------|--------------------------------| | 312 | 1.02 | 0.99 to 1.04 | -0.7 | -5.4 to 4.1 | 1.00 | #### Sample Matrix: The matrix comparison study was performed to evaluate the difference across tube types [serum without additive (red top), serum in separator tubes (SST), and lithium heparin plasma] versus the control samples (K2 EDTA plasma) following the CLSI EP9-A3 guideline. A total of 52 samples (45 native, 7 spiked) matched samples with Intact PTH concentrations ranging from of 10.6 to 2759.0 pg/mL were tested with the IDS-iSYS Intact PTH assay. Passing-Bablok regression analysis was performed on the comparative data: | Sample Type | n | Slope | 95% Cl. | Intercept (pg/mL) | 95% Cl. | Corr. Coeff. (r) | |-----------------|----|-------|--------------|-------------------|---------------|------------------| | Serum (Red Top) | 52 | 0.94 | 0.92 to 0.97 | 2.55 | 0.86 to 3.16 | 1.00 | | Serum (SST) | 52 | 0.93 | 0.91 to 0.96 | 2.38 | 1.25 to 3.15 | 1.00 | | Lithium Heparin | 52 | 0.98 | 0.95 to 0.99 | 0.42 | -0.43 to 1.63 | 1.00 | The type of specimen used (serum, EDTA plasma, or Lithium Heparin plasma) may influence PTH measurement. During routine monitoring of PTH levels, use the same specimen type throughout the monitoring period to avoid bias in the results. {9}------------------------------------------------ #### Reference Interval: The 95% reference interval for the following group was calculated by a nonparametric method following the CLSI C28-A3, "How to Define and Determine Reference Intervals in the Clinical Laboratory". The K2 EDTA plasma samples were collected from 129 individuals (67 males, 62 females; 21 to 89 years of age) with normal calcium, phosphate, and TSH values from the north, central and south regions of the United States. | No. of subjects | 129 | |----------------------------|--------------| | Mean (pg/mL) | 35.4 | | SD (pg/mL) | 16.6 | | Median (pg/mL) | 32.7 | | Reference Interval (pg/mL) | 10.3 to 80.5 | The above ranges should be considered as guidelines only; it is recommended that each laboratory establish its own expected range based upon its own patient population. # Sensitivity The limit of blank (LoB), limit of detection (LoD) and limit of quantitation (LoQ) were determined with guidance from CLSI EP17-A, "Protocols for Determination of Limits of Detection and Limits of Quantitation" using 60 blanks and 10 low level serum samples. The following limits were determined with the IDS-iSYS Intact PTH™ Assay: | Sensitivity | Concentration (pg/mL) | |-----------------------------|-----------------------| | Limit of Blank (LoB) | 0.9 | | Limit of Detection (LoD) | 2.3 | | Limit of Quantitation (LoQ) | 4.5 | # Precision: Precision testing was evaluated in accordance with a modified protocol based on CLSI EP-5A2, "Evaluation of Precision Performance of Quantitative Measurement Methods''. Seven (7) samples were assayed using one lot of reagent in duplicate twice per day for 20 days on one System. | | | Mean | Within run | | Total | | |----------------------|----|---------|------------|-----|-------|-----| | Sample | n | (pg/mL) | SD | CV% | SD | CV% | | Serum Sample 1 | 80 | 15.4 | 0.5 | 6.1 | 1.0 | 6.7 | | EDTA Plasma Sample 1 | 80 | 17.9 | 1.8 | 9.9 | 1.8 | 9.9 | | EDTA Plasma Sample 2 | 80 | 73.7 | 1.3 | 1.7 | 2.3 | 3.2 | | EDTA Plasma Sample 3 | 80 | 337.9 | 5.1 | 1.5 | 6.2 | 1.8 | | EDTA Plasma Sample 4 | 80 | 765.1 | 13.7 | 1.8 | 16.4 | 2.1 | | EDTA Plasma Sample 5 | 80 | 1376.6 | 20.4 | 1.5 | 40.7 | 3.0 | | EDTA Plasma Sample 6 | 80 | 2229.1 | 35.2 | 1.6 | 58.6 | 2.6 | {10}------------------------------------------------ # Linearity: Linearity was evaluated based on CLSI EP-6A, "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach". Samples were prepared by diluting a high patient sample with a low patient sample prior to assay. High intact PTH human K2 EDTA plasma and serum samples were diluted with low samples prior to assay. Each sample was measured in 4 replicates. The K2 EDTA plasma samples with Intact PTH concentration levels above 80 pg/mL show a maximum -8.8 % deviation from linearity; samples below 80 pg/mL show a maximum deviation of -6.1 pg/mL. The weighted linear regression of the Observed versus the Expected concentrations is: Observed = 0.96 x (Expected) = 0.1 pg/mL Regression coefficient R2: 1.00 The serum samples with intact PTH concentration levels above 80 pg/mL show a maximum deviation from linearity of 5.1%. The samples below 80 pg/mL show a maximum deviation of 0.6 pg/mL. The weighted linear regression of the Observed versus the Expected concentrations is: Observed = 1.02 x (Expected) = 0.2 pg/mL Regression coefficient R2: 1 .00 # High Dose Hook Effect: Testing was conducted to determine if the IDS-iSYS Intact PTH® assay is susceptible to artificially low results in the presence of very high levels of PTH (Hook Effect). Three samples (1 EDTA Plasma, 2 Serum) were spiked with 1-84 PTH to 5000 pg/mL, 25000 pg/mL, 50000 pg/mL, 75000 pg/mL and 100000 pg/mL. Each specimen was measured in four (4) replicates using 3 reagents lots. No high dose hook effect was observed for Intact PTH concentrations measured up to 100000 pg/mL. {11}------------------------------------------------ # Interferences: Interference studies were performed in accordance with the CLSI EP7-A2, "Interference testing in clinical chemistry; approved guideline". Two K2-EDTA samples at a low and high Intact PTH levels were spiked with following potential interfering substances: Triglycerides, Hemoglobin, Bilirubin (conjugated and unconjugated), Red blood cells, Biotin, Acetaminophen, Ibuprofen and Carbamazepine. Two serum samples at different Intact PTH concentrations were used to determine the potential interference of Total Protein. % Interference was calculated using the formula below: ``` % Interference = (Mean conc. of spiked sample - mean conc. of un-spiked sample) x100% Mean concentration of un-spiked sample ``` For Cholesterol (Total), Rheumatoid factor (Rf) and HAMA (human anti mouse antibodies), the interference was tested by recovery of Intact PTH from a high serum pool spiked into a serum sample with known interferent substance levels. % Recovery was calculated using the formula below: ``` Recovery value = Mean concentration of spiked - Mean concentration of un-spiked (Recovery value / Expected recovery value) x 100 % Recovery = = ``` Following potential interference compounds were tested and found not to interfere significantly with the IDS-iSYS Intact PTH assay, based on the criteria of nonsignificant interference of ≤±10% bias between the test (spiked) and control (un-spiked) samples: | Potentially Interfering Agent | Highest concentration tested that<br>demonstrated no significant interference | |----------------------------------|-------------------------------------------------------------------------------| | Bilirubin, conjugated | 22 mg/dL | | Bilirubin, unconjugated | 40 mg/dL | | Biotin | 300 nmol/L | | Cholesterol, total | 395 mg/dL | | Haemoglobin | 500 mg/dL | | Human Anti Mouse Antibody (HAMA) | 1000 ng/mL | | Rheumatoid Factor | 1836 IU/mL | | Total Protein | 10 g/dL | | Triglyceride | 3000 mg/dL | | Acetaminophen | 200 µg/mL | | Carbamazepine | 200 µg/mL | | Ibuprofen | 500 µg/mL | {12}------------------------------------------------ # Cross Reactivity: The potential cross-reacting substances studies were performed on the IDS-iS Y S Intact PTH" assay, based on CLSI EP-7A2. A serum sample and a zero calibrator matrix was spiked with the following substance human PTH fragments and structurally similar proteins at the concentration listed below. The cross reactivity was then determined using the following equation: | | % cross reactivity = (Mean conc. spiked sample - mean conc.un-spiked sample) | | | |---------------------------|------------------------------------------------------------------------------|--|--| | x100% Spike concentration | | | | | Analyte | Concentration tested<br>(pg/mL) | Cross-Reactivity | |---------------------|---------------------------------|------------------| | PTH (1-84) | 2,000 | 100% | | PTH (7-84) | 2,000 | 83.6% | | PTH (1-34) | 100,000 | <0.01% | | PTH (39-84) | 100,000 | <0.01% | | PTH (53-84) | 100,000 | <0.01% | | PTH (39-68) | 100,000 | <0.01% | | PTH (44-68) | 100,000 | <0.01% | | Human Calcitonin | 100,000 | <0.01% | | CTX-1 (β CrossLaps) | 100,000 | <0.01% | | Osteocalcin | 100,000 | <0.01% | # 15. Conclusions The IDS-iSYS Intact PTH assay, after the modification in antibody source, is substantially equivalent in principle and performance to the originally submitted IDS-iSYS Intact PTH assay (K103325).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...