ADVIA Centaur Intact Parathyroid Hormone (PTH) Assay

K163658 · Siemens Healthcare Diagnostics, Inc. · CEW · May 23, 2017 · Clinical Chemistry

Device Facts

Record IDK163658
Device NameADVIA Centaur Intact Parathyroid Hormone (PTH) Assay
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeCEW · Clinical Chemistry
Decision DateMay 23, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1545
Device ClassClass 2

Intended Use

The ADVIA Centaur® Intact Parathyroid Hormone (PTH) assay is for in vitro diagnostic use in the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma using the ADVIA Centaur XP system. This assay is intended to be used to aid in the differential diagnosis of hyperparathyroidism, or disorders of calcium metabolism. This assay can be used intra-operatively.

Device Story

The ADVIA Centaur PTH assay is an in vitro diagnostic immunoassay for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma. It is designed for use on the ADVIA Centaur XP system in clinical laboratory settings. The device utilizes a sandwich chemiluminescence immunoassay protocol; it employs acridinium ester-labeled mouse monoclonal anti-human PTH antibodies and biotinylated mouse monoclonal anti-human PTH antibodies bound to streptavidin-coated paramagnetic particles. The system processes patient samples to measure PTH levels, which clinicians use to aid in the diagnosis of hyperparathyroidism, hypoparathyroidism, or calcium metabolism disorders. The assay is also validated for intra-operative use to monitor PTH levels during parathyroid surgery, assisting surgeons in confirming successful gland excision based on the 'Miami criterion' (a ≥50% drop in PTH levels post-excision).

Clinical Evidence

Clinical study evaluated 30 subjects undergoing parathyroid surgery. Performance was assessed using the 'Miami criterion' (≥50% drop in PTH at 10 minutes post-excision). The ADVIA Centaur PTH assay showed 100% positive agreement and 96.7% overall agreement with comparator hospital assays. Bench testing included precision (CLSI EP5-A2), linearity (CLSI EP06-A), interference (CLSI EP07-A2), and method comparison (r=0.99).

Technological Characteristics

Two-site sandwich chemiluminescent immunoassay. Reagents: acridinium ester-labeled mouse monoclonal anti-PTH antibody; biotinylated mouse monoclonal anti-PTH antibody on streptavidin-coated paramagnetic particles. System: ADVIA Centaur XP. Measuring range: 6.3–2000 pg/mL. Traceable to purified human PTH (1-84).

Indications for Use

Indicated for use in adults (≥21 years) to aid in the differential diagnosis of hyperparathyroidism, hypoparathyroidism, or disorders of calcium metabolism, including intra-operative monitoring.

Regulatory Classification

Identification

A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol on the right side. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 23, 2017 SIEMENS HEALTHCARE DIAGNOSTICS INC. DARIUS DARUWALA SENIOR SPECIALIST, REGULATORY AFFAIRS 511 BENEDICT AVENUE TARRYTOWN NY 10591 Re: K163658 Trade/Device Name: ADVIA Centaur Intact Parathyroid Hormone (PTH) Assay Regulation Number: 21 CFR 862.1545 Regulation Name: Parathyroid hormone test system Regulatory Class: II Product Code: CEW Dated: April 24, 2017 Received: April 25, 2017 Dear Darius Daruwala: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kellie B. Kelm -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) k163658 Device Name ADVIA Centaur® Intact Parathyroid Hormone (PTH) Assay Indications for Use (Describe) The ADVIA Centaur® Intact Parathyroid Hormone (PTH) assay is for in vitro diagnostic use in the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma using the ADVIA Centaur XP system. This assay is intended to be used to aid in the differential diagnosis of hyperparathyroidism, or disorders of calcium metabolism. This assay can be used intra-operatively. Type of Use (Select one or both, as applicable) | <span></span> <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | |-------------------------------------------------------------------------------------------------------------------| | <span></span> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990. The assigned 510(k) Number is: K163658 ## Date Prepared 1. May 22, 2017 ## Applicant Information 2. | Contact: | Darius Daruwala<br>Regulatory Affairs Senior Specialist | |----------|---------------------------------------------------------------------------------------| | Address: | Siemens Healthcare Diagnostics Inc<br>511 Benedict Avenue<br>Tarrytown, NY 10591-5097 | | Phone: | 914-524-2812 | | Fax: | 914-524-3579 | | Email: | darius.daruwala@siemens.com | ## Regulatory Information 3. # Table 1. Regulatory Information for ADVIA Centaur PTH Assay | Trade Name | ADVIA Centaur® Intact Parathyroid Hormone (PTH) | |---------------------|-------------------------------------------------| | Model Numbers | 10699154 (1-pack); 10699155 (5-pack) | | Common Name | Immunoassay, intact parathyroid hormone | | Classification Name | Parathyroid Hormone Test System | | FDA Classification | Class II | | Review Panel | Clinical Chemistry (75) | | Product Code | CEW | | Regulation Number | 862.1545 | ## Predicate Device Information 4. # ADVIA Centaur Intact Parathyroid Hormone (iPTH) Predicate Device Name: ADVIA Centaur Intact Parathyroid Hormone (iPTH) 510(k) Number: K133601 # Intended Use / Indications for Use 5. # ADVIA Centaur Intact Parathyroid Hormone (PTH) The ADVIA Centaur Intact Parathyroid Hormone (PTH) reagent is for in vitro diagnostic use in the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma using the ADVIA Centaur XP system. {4}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness This assay is intended to be used to aid in the differential diagnosis of hyperparathyroidism, hypoparathyroidism, or disorders of calcium metabolism. This assay can be used intraoperatively. # Device Description 6. | | | Table 2. Summary of Ingredients of the ADVIA Centaur PTH Assay Components | |--|--|---------------------------------------------------------------------------| | | | | | Component | Volume | Ingredients | |---------------------------------------------------------------------|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ADVIA Centaur PTH Primary Reagent ReadyPack (included in assay kit) | | | | ADVIA Centaur PTH Lite<br>Reagent | 10.0 mL/pack | Acridinium ester-labeled mouse monoclonal<br>anti-human PTH antibody (~0.6 mg/L) in<br>buffered saline with mouse gamma globulin,<br>bovine serum albumin, and preservatives | | ADVIA Centaur PTH<br>Solid Phase Reagent | 20.0 mL/pack | Biotinylated mouse monoclonal anti-human<br>PTH antibody bound to streptavidin-coated<br>paramagnetic particles (~0.4 g/L) in buffered<br>saline with bovine gamma globulin, bovine<br>serum albumin, and preservatives | | ADVIA Centaur PTH Calibrator (included in assay kit) | | | | ADVIA Centaur PTH Low<br>and High Calibrators | 1.0 mL/vial | After reconstitution, low or high levels of intact<br>PTH synthetic peptide in buffered saline with<br>human EDTA plasma (10%), surfactants, and<br>preservatives | ## Purpose of the Submission 7. The purpose of this submission is a premarket notification for a new device: ADVIA Centaur Intact Parathyroid Hormone (PTH) assay. # 8. Comparison of Candidate Device and Predicate Device | Item | ADVIA Centaur Intact Parathyroid<br>Hormone (PTH) Assay<br>(Candidate Device) | ADVIA Centaur Intact Parathyroid<br>Hormone (iPTH) Assay<br>(Predicate Device) | |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | For <i>in vitro</i> diagnostic use in the<br>quantitative determination of intact<br>parathyroid hormone (PTH) in human<br>serum and plasma using the ADVIA<br>Centaur XP system. | For <i>in vitro</i> diagnostic use in the<br>quantitative determination of intact<br>parathyroid hormone (iPTH) in<br>EDTA plasma or serum using the<br>ADVIA Centaur systems. | | Indications for Use | This assay is intended to be used to<br>aid in the differential diagnosis of<br>hyperparathyroidism,<br>hypoparathyroidism, or disorders of<br>calcium metabolism. This assay can be<br>used intra-operatively. | This assay is intended to be used to<br>aid in the differential diagnosis of<br>hyperparathyroidism and<br>hypoparathyroidism. | | Measurement | Quantitative | Same | # Table 3. Comparison of ADVIA Centaur PTH Assay to Predicate {5}------------------------------------------------ | ltem | ADVIA Centaur Intact Parathyroid<br>Hormone (PTH) Assay<br>(Candidate Device) | ADVIA Centaur Intact Parathyroid<br>Hormone (iPTH) Assay<br>(Predicate Device) | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | Methodology | Chemiluminescence | Same | | Assay Protocol | Sandwich immunoassay | Same | | Traceability/<br>Standardization | Internal standards. Values have been<br>assigned to correlate to the Predicate<br>device. | Same | | Specimen Type | Human serum and plasma | Same | | Sample Volume | 50 uL | 200 µL | | Measuring Range | 6.3 to 2000 pg/mL | 6.3 — 1900 pg/mL | | Calibration | 2-point calibration | Same | | Detection Antibody | Mouse monoclonal antibody<br>conjugated to Acridium Ester in the<br>Lite Reagent | Goat polyclonal antibody<br>conjugated to Acridium Ester in the<br>Lite Reagent | | Capture Antibody | Biotinylated mouse monoclonal anti-<br>human PTH antibody with<br>streptavidin-coated paramagnetic<br>particles pre-formed in Solid Phase<br>Reagent | Biotinylated goat polyclonal anti-<br>human PTH antibody with<br>streptavidin-coated paramagnetic<br>particles pre-formed in Solid Phase<br>Reagent | | Expected Values | 18.4 - 80.1 pg/mL (plasma)<br>18.5 – 88.0 pg/mL (serum) | 13.8 - 85.0 pg/mL (plasma)<br>12.4 - 76.8 pg/mL (serum) | # Standard/Guidance Document References 9. The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission: - Evaluation of Precision of Quantitative Measurement Procedures; Approved S Guideline - Third Edition (CLSI EP05-A3, 2014; Recognition Number 7-251) - S Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach: Approved Guideline (CLSI EP06-A, 2003; Recognition Number 7-193) - Interference Testing in Clinical Chemistry: Approved Guideline Second Edition S (CLSI EP07-A2, 2005; Recognition Number 7-127) - S Measurement Procedure Comparison And Bias Estimation Using Patient Samples --Third Edition (CLSI EP9-A3, 2013; Recognition Number 7-245) - Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: S Approved Guideline -- Second Edition (CLSI EP17-A2, 2013; Recognition Number 7-233) - S Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition (CLSI EP28-A3c - formerly C28-A3c, 2010; Recognition Number 7-224) - S Medical devices – Application of risk management to medical devices (ANSI/AAMI/ISO 14971:2007/(R)2010; Recognition Number 5-70) {6}------------------------------------------------ # 10. Performance Characteristics: ADVIA Centaur PTH #### 10.1 Precision A 20-day precision study was performed according to CLSI EP5-A2. Six EDTA plasma samples and three levels of controls were tested. | Sample | Mean | | Repeatability<br>(Within-Run) | | | Within-Lab<br>(Total Precision) | | | |------------------------------------|---------|----------|-------------------------------|----------------|------|---------------------------------|----------------|------| | | (pg/mL) | (pmol/L) | SD<br>(pg/mL) | SD<br>(pmol/L) | % CV | SD<br>(pg/mL) | SD<br>(pmol/L) | % CV | | Sample 1 | 16.9 | 1.79 | 0.9 | 0.09 | 5.16 | 1.1 | 0.12 | 6.78 | | Sample 2 | 50.3 | 5.34 | 1.4 | 0.15 | 2.78 | 2.3 | 0.24 | 4.57 | | Sample 3 | 146.4 | 15.52 | 1.9 | 0.20 | 1.30 | 2.9 | 0.30 | 1.95 | | Sample 4 | 409.2 | 43.37 | 6.8 | 0.72 | 1.67 | 16.9 | 1.79 | 4.13 | | Sample 5 | 596.3 | 63.20 | 7.3 | 0.78 | 1.23 | 9.7 | 1.02 | 1.62 | | Sample 6 | 1080.4 | 114.52 | 11.8 | 1.25 | 1.09 | 25.1 | 2.66 | 2.32 | | ADVIA Centaur PTH Quality Controls | | | | | | | | | | Control 1 | 41.6 | 4.41 | 0.4 | 0.04 | 0.91 | 1.4 | 0.15 | 3.36 | | Control 2 | 235.9 | 25.00 | 1.9 | 0.20 | 0.81 | 4.4 | 0.46 | 1.85 | | Control 3 | 874.0 | 92.64 | 8.0 | 0.85 | 0.92 | 18.4 | 1.96 | 2.11 | #### 10.2 Linearity A linearity study was performed according to CLSI EP06-A using 11 serially diluted samples spanning the assay range. The mean was taken from each sample tested in triplicate. As presented below, the bias from the linear fit estimate was <10%. The assay is linear from 6.3 to 2000 pg/mL. | Sample | Expected Dose (pg/mL) | Observed Dose (pg/mL) | Recovery (Observed vs Expected) | Weighted Linear Fit Estimate | Deviation from Linear Fit | |--------|-----------------------|-----------------------|---------------------------------|------------------------------|---------------------------| | A | 2.7 | 2.7 | 100% | 2.6 | 3% | | B | 3.8 | 3.5 | 92% | 3.7 | -4% | | C | 7.1 | 6.6 | 93% | 6.8 | -4% | | D | 20.3 | 19.2 | 95% | 19.5 | -1% | | E | 73.3 | 68.9 | 94% | 70.2 | -2% | | F | 143.9 | 137.0 | 95% | 137.8 | -1% | | G | 285.1 | 264.3 | 93% | 273.0 | -3% | | H | 567.5 | 529.9 | 93% | 543.4 | -3% | | I | 850.0 | 819.8 | 96% | 813.9 | 1% | | J | 1132.4 | 1089.9 | 96% | 1084.3 | 1% | | K | 2262.1 | 2262.1 | 100% | 2166.0 | 4% | {7}------------------------------------------------ ## 10.3 Dilution Recovery A low sample pool (~ 22 pq/mL PTH) was spiked with a commercially available synthetic PTH peptide to reach levels of approximately 3,000 pg/mL (318 pmol/L), 6,000 pg/mL (636 pmol/L) and 9,000 pg/mL (954 pmol/L). Each sample was diluted 1:5 with ADVIA Centaur Multi-Diluent 13 on-board the ADVIA Centaur XP system. All samples were run in triplicate. | Sample | Observed Mean<br>(pg/mL) | Expected<br>(pg/mL) | Observed Mean<br>(pmol/L) | Expected<br>(pmol/L) | Recovery<br>(%) | |--------|--------------------------|---------------------|---------------------------|----------------------|-----------------| | 1 | 3068.8 | 3020.9 | 325.3 | 320.2 | 101.6 | | 2 | 6270.8 | 6019.8 | 664.7 | 638.1 | 104.2 | | 3 | 9253.2 | 9018.7 | 980.8 | 956.0 | 102.6 | | Mean | | | | | 102.8 | ## 10.4 Method Comparison Method Comparison studies were done with 349 EDTA plasma patient samples distributed over the assay range to demonstrate equivalence to the Predicate (ADVIA Centaur iPTH). The ADVIA Centaur PTH assay shows good correlation in sample results compared to the Predicate. The regression equation from the analysis is presented below. ADVIA Centaur PTH = 1.02(iPTH) – 2.18 pg/mL (r = 0.99) #### 10.5 Matrix Comparison The ADVIA Centaur PTH assay was evaluated using different specimen matrices and tube collection types. A specimen collection study was performed using 56 matched specimens drawn in different tube types including serum red top, serum separator tube (SST), EDTA, lithium heparin, and sodium heparin, on two reagent lots. No significant difference between tube types was observed. The following results were obtained: | Comparisona | Regression Equation | r | |--------------------------------------------------|-----------------------------------------|-------| | Serum vs. Dipotassium EDTA plasma | Serum = 0.99 EDTA – 1.85 | 0.996 | | Serum separator tube vs. Dipotassium EDTA plasma | Serum separator tube = 1.03 EDTA + 0.20 | 0.996 | | Lithium Heparin vs. Dipotassium EDTA plasma | Lithium Heparin = 0.99 EDTA + 1.95 | 0.998 | | Sodium Heparin vs. Dipotassium EDTA plasma | Sodium Heparin = 1.00 EDTA + 1.03 | 0.997 | I his study was performed using Becon Dickinson tubes. Siements that laboratories evaluate performance when using other manufacturers' tubes. #### 10.6 Reference Intervals Reference intervals for the ADVIA Centaur PTH assay were established according to CLSI EP28-A3c. A total of one hundred forty-two (142) paired serum and EDTA plasma samples from apparently healthy donors were analyzed. Based on a 95% confidence interval, the following reference intervals were established. | Sample<br>Type | Lower Limit<br>(95% Cl) | Upper Limit<br>(95% CI) | |----------------|-------------------------|-------------------------| | EDTA Plasma | 18.4 pg/mL | 80.1 pg/mL | | Serum | 18.5 pg/mL | 88.0 pg/mL | {8}------------------------------------------------ ## 10.7 Detection Limit The limit of blank (LoB). limit of detection (LoD). and the limit of quantitation (LoQ) were determined as described in CLSI protocol EP17-A2. The LoB is defined as the highest measurement result that is likely to be observed for a blank sample. The LoD is defined as the lowest concentration of PTH that can be detected with 95% probability. The LoQ is defined as the lowest concentration of PTH that can be detected at a total CV of 20%. The ADVIA Centaur PTH assay has a LoB of 1.5 pg/mL, a LoD of 3.2 pg/mL, and a LoQ of 4.6 pg/mL. #### 10.8 Interference Interference studies were performed according to CLSI EP07-A2. Two EDTA sample pools were tested. One sample pool had approximately 75 pg/mL PTH. The second sample pool had approximately 600 pg/mL PTH. These sample pools were spiked with potential interferents. There was no indication of interference (< 10% effect) up to the interferent levels claimed. Results are presented below. | Endogenous Substance | Endogenous Substance<br>Concentration | Interference<br>(%) | |--------------------------|---------------------------------------|---------------------| | Hemoglobin | 500 mg/dL | -4.5 | | Triglycerides | 3275 mg/dL | -0.4 | | Bilirubin (unconjugated) | 60 mg/dL | -2.2 | | Bilirubin (conjugated) | 60 mg/dL | 5.9 | | Biotin | 1000 ng/mL | 6.9 | | Cholesterol | 500 mg/dL | -6.5 | | Total Protein (high) | 12 g/dL | -4.2 | | Total Protein (low) | 6 g/dL | -7.5 | | IgG | 6 g/dL | 1.3 | | Calcitrol | 360 pg/mL | 2.5 | | Furosemide | 181 µmol/L | -2.6 | | Caffeine | 308 µmol/L | 2.3 | | Aliskiren | 200 µg/mL | 5.3 | | Enalaprilat | 0.86 µmol/L | -2.6 | | Epoetin alfa | 15 mU/L | -1.2 | | Fosrenol | 20 ng/mL | 3.4 | Six human plasma samples containing human anti-mouse antibodies (HAMA) were tested at 2 PTH concentrations. Samples at a PTH concentration of 60 pg/mL (6.36 pmol/L) demonstrated a difference of -11.5% to 3.2% from the expected values. Samples at a PTH concentration of 340 pg/mL (36.04 pmol/L) demonstrated a difference of -3.8% to 2.4% from the expected values. ## 10.9 Cross-Reactivity Cross reactivity was evaluated in the ADVIA Centaur iPTH immunoassay using a normal human EDTA plasma sample and blank assay diluent for each test compound. An aliquot of {9}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness each sample was spiked with test compound so that the final test sample contained the compound at the required concentration. A second aliquot of the base pool is spiked with just the diluent to serve as a control sample. Multiple replicates of the test and control samples are processed. Cross-reactivity was calculated as the % difference between the mean test and control sample results, with respect to the test compound concentration. Results are presented below. | Cross-Reactant<br>(Conc.) | Normal Patient (pg/mL) | Blank (MD13) (pg/mL) | | | |----------------------------------|------------------------|----------------------|---------|----------| | Calcitonin<br>(100,000 pg/mL) | Test | 30.39 | Test | 0.34 | | | Control | 30.33 | Control | 0.00 | | | %XR | 0.0001% | %XR | 0.0003% | | Beta-CrossLaps<br>(10,000 pg/mL) | Test | 30.40 | Test | 0.00 | | | Control | 31.15 | Control | 0.00 | | | %XR | -0.0075% | %XR | 0.0000% | | Osteocalcin<br>(50,000 pg/mL) | Test | 30.56 | Test | 0.00 | | | Control | 30.71 | Control | 0.00 | | | %XR | -0.0003% | %XR | 0.0000% | | PTH (1-34)<br>(12,000 pg/mL) | Test | 30.87 | Test | 0.00 | | | Control | 30.79 | Control | 0.00 | | | %XR | 0.0007% | %XR | 0.0000% | | PTH (39-68)<br>(100,000 pg/mL) | Test | 30.66 | Test | 0.00 | | | Control | 30.69 | Control | 0.00 | | | %XR | 0.0000% | %XR | 0.0000% | | PTH (53-84)<br>(100,000 pg/mL) | Test | 29.81 | Test | 0.00 | | | Control | 30.74 | Control | 0.00 | | | %XR | -0.0009% | %XR | 0.0000% | | PTH (44-68)<br>(100,000 pg/mL) | Test | 30.57 | Test | 0.00 | | | Control | 30.24 | Control | 0.00 | | | %XR | 0.0003% | %XR | 0.0000% | | PTH (39-84)<br>(100,000 pg/mL) | Test | 30.24 | Test | 0.00 | | | Control | 29.82 | Control | 0.00 | | | %XR | 0.0004% | %XR | 0.0000% | | PTH-RP (1-34)<br>(100,000 pg/mL) | Test | 30.13 | Test | 0.00 | | | Control | 29.95 | Control | 0.00 | | | %XR | 0.0002% | %XR | 0.0000% | | PTH (7-84)<br>(300 pg/mL) | Test | 141.94 | Test | 52.51 | | | Control | 29.85 | Control | 0.00 | | | %XR | 37.36% | %XR | 17.5033% | # 10.10 High-Dose Hook Effect Patient samples with high intact PTH levels can cause a paradoxical decrease in the RLUs (high-dose hook effect). In this assay, patient samples with intact PTH levels as high as 100,000 pg/mL (10,600 pmol/L) are reported as > 2000 pg/mL (> 212 pmol/L). #### 10.11 Stability The onboard stability of the ADVIA Centaur PTH reagents is 28 days with a calibration interval of 21 days. Unopened reagents and calibrators are stable until the date printed on the box label when stored at 2-8°C. {10}------------------------------------------------ ## 10.12 Clinical Study A clinical study was performed to confirm the effectiveness of the ADVIA Centaur® Intact Parathyroid Hormone (PTH) assay for Intra-operative (IO) use. Sets of specimens from 30 subjects that qualified for the primary endpoints (described below), that met the inclusion criteria, and which were not excluded based on exclusion criteria, were analyzed using the ADVIA Centaur PTH assay on the Centaur XP system. The following Inclusion/exclusion criteria were used: Inclusion Criteria - Patients undergoing parathyroid surgery with intra-operative parathyroid (PTH) . testing. - Patients who have available preoperative calcium and PTH measurements . - Patients who have baseline PTH and PTH monitored on the day of surgery on a . clinical PTH test device cleared by the FDA for use intra-operatively. # Exclusion Criteria - Patients who do not meet inclusion criteria. - Specimens for which handling and storage quidelines aren't followed. - Specimens which do not have sufficient quantity available for testing. - Specimens from pediatric subjects (<21 years old.) Specimens underwent testing in single replicates on both the ADVIA Centaur XP PTH assay and the comparator (hospital) assays. For this clinical study, the 'Miami' criterion was utilized to determine 'success'. Miami Criterion: A successful surgery is defined to be a 50% or greater drop in PTH level from the greater of the pre-incision or pre-excision baseline values to the 10 minute post-excision test result after the last parathyroid gland excision. The analyses were based on agreement between the ADVIA Centaur PTH Assay (investigational device) and the assays used by participating surgeons/sites (comparator devices), utilizing the same criteria of a successful surgery for each. Concordance between the surgeon's assays and PTH is presented in the table below. | | | PTH Assay Used During<br>Surgery | | |----------------------|--------------|----------------------------------|--------------| | | | Successful | Unsuccessful | | ADVIA Centaur<br>PTH | Successful | 29 | 1* | | | Unsuccessful | 0 | 0 | The above table displays success based on the Miami criteria for the first 30 set of patient samples that have a 10 minute post-excision draw. Primary Endpoint Positive Agreement = 29/29 = 100% Primary Endpoint Overall Agreement = 29/30 = 96.7% * Note that for one subject, local PTH device did not reach 50% drop from baseline until a blood draw after the 10 minute draw, therefore by the Miami criterion for the surgery is defined as a failure'. That subject's specimens yielded only a 48% drop in PTH values by 10 minutes post-excision with the assay used during surgery, but a 55% drop with the ADVA {11}------------------------------------------------ Centaur PTH assay run onsite. This was the only set of patient samples that were discrepant based on the Miami criterion definition of success. # Conclusions 11. Based on the results of comparative testing, the new ADVIA Centaur Intact Parathyroid Hormone (PTH) assay is substantially equivalent in principle and performance to the currently-marketed predicate device, the ADVIA Centaur Intact Parathyroid Hormone (iPTH) assay, cleared under 510(k) k133601.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...