DIMENSION EXL WITH LM CLINICAL CHEMISTRY SYSTEM, DIMENSION FT4L FLEX REAGENT CARTRIDGE, AND LOCI THYROID CALIBRATOR, MOD
K073604 · Dade Behring, Inc. · CEC · Mar 13, 2008 · Clinical Chemistry
Device Facts
Record ID
K073604
Device Name
DIMENSION EXL WITH LM CLINICAL CHEMISTRY SYSTEM, DIMENSION FT4L FLEX REAGENT CARTRIDGE, AND LOCI THYROID CALIBRATOR, MOD
Applicant
Dade Behring, Inc.
Product Code
CEC · Clinical Chemistry
Decision Date
Mar 13, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1695
Device Class
Class 2
Intended Use
The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ with LM system. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease. The LOCI® Thyroid Calibrator is an in vitro diagnostic product for the calibration of the FT4L method on the Dimension® EXL™ with LM system. The Dimension® EXL™ with LM system is an in vitro diagnostic device that is intended to measure a variety of analytes in human body fluids. The system utilizes photometric, turbidimetric, chemiluminescence and integrated ion selective multisensor technology for chemical and immunochemical applications for clinical use.
Device Story
The Dimension® EXL™ with LM system is a fully automated, microprocessor-controlled clinical chemistry analyzer used in clinical laboratories. It processes human serum and plasma samples to measure Free Thyroxine (FT4) using a homogeneous, sequential chemiluminescent immunoassay (LOCI® technology). The system utilizes two synthetic bead reagents: Chemibeads (coated with T3 and containing chemiluminescent dye) and Sensibeads (coated with streptavidin and containing photosensitizer dye). The device measures the chemiluminescent signal generated by the interaction of these beads with biotinylated anti-T4 antibodies and sample T4; the signal is inversely proportional to the FT4 concentration. The system automates sample handling, reagent delivery, incubation at 37°C, and signal detection at 612 nm. Healthcare providers use the quantitative results to diagnose and monitor thyroid conditions. The system supports automated QC, calibration, and barcode-based specimen identification, providing clinicians with objective data to inform patient management.
Clinical Evidence
Method comparison study conducted using 180 serum samples comparing Dimension® FT4L Flex® to the predicate Dimension® Vista FT4 Flex® reagent cartridge. Analysis via least squares linear regression showed a correlation coefficient (r) of 0.999, slope of 0.97 (95% CI: 0.96-0.97), y-intercept of -0.03, and Sy,x of 0.08 ng/mL.
Technological Characteristics
Floor-model, fully automated, microprocessor-controlled clinical chemistry analyzer. Detection technologies include photometry, turbidimetry, chemiluminescence, and integrated ion-selective multisensor technology. Reagents are provided in prepackaged plastic eight-well cartridges. Calibrators are liquid, bovine serum albumin-based. System supports random access, batch, and stat processing modes.
Indications for Use
Indicated for the quantitative measurement of Free Thyroxine in human serum and plasma for the diagnosis and monitoring of thyroid disease. For prescription use only. Contraindicated for diluted samples; potential for interference from heterophile antibodies.
Regulatory Classification
Identification
A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in serum or plasma. Levels of free thyroxine in plasma are thought to reflect the amount of thyroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
Predicate Devices
Dimension® RxL Max® clinical chemistry system (K944093)
K130276 — DIMENSION EXL WITH LM SYSTEM · Siemens Healthcare Diagnostics, Inc. · Mar 22, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
## MAR 1 3 2008
## 510(k) Summary of Safety and Effectiveness Dimension® EXL™ with LM clinical chemistry system, Dimension® FT4L Flex® reagent cartridge and LOCI® Thyroid Calibrator
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: _______________________________________________________________________________________________________
### l. Manufacturer's Name, Address, Telephone, and Contact Person, Datc of Preparation
Manufacturer:
Dade Behring Inc. P.O. Box 6101 Newark, DE 19714
Contact Information:
Dade Behring Inc. P.O. Box 6101 Newark, DE 19714 Attn: Yuk-Ting Lewis Tel: 302-631-7626
Date of Preparation:
February 18, 2008
#### 2. Device Name / Classification
Dimension® EXLTM with LM clinical chemistry system / Class I Dimension® FT4L Flex® reagent cartridge / Class II LOCI® Thyroid Calibrator / Class II
#### 3. Identification of the Predicate Device
Dimension® RxL Max® clinical chemistry system (the Dimension® RxL Max is a family member of the Dimension® XL clinical chemistry system, which was cleared under K944093} Dimension Vista™ FT4 Flex® reagent cartridge, K053531 Dimension Vista™ LOCI 1 Calibrator, K053531
{1}------------------------------------------------
#### 4. Device Description
The Dimension® EXL™ with LM clinical chemistry system is a floor model, fully automated, microprocessor-controlled, integrated instrument system which uses prepackaged Dade Behring Flex® reagent cartridges to measure a variety of analytes in human body fluids. The system can process samples in random access, batch or stat modes.
The Dimension® FT4L Flex® reagent cartridge is an in vitro diagnostic device that consists of prepackaged reagents in a plastic eight-well cartridge for use on the Dimension® EXL™ with LM system.
The LOCI® Thyroid Calibrator is a liquid, bovine serum albumin based product containing thyroxine. There are four calibrator levels with target values of 0.8, 1.6, 4.0 and 8.4 ng/dL.
#### 5. Device Intended Use
The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ with LM system. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.
The LOCI® Thyroid Calibrator is an in vitro diagnostic product for the calibration of the FT4L method on the Dimension® EXL™ with LM system.
The Dimension® EXL™ with LM system is an in vitro diagnostic device that is intended to measure a variety of analytes in human body fluids. The system utilizes photometric, turbidimetric, chemiluminescence and integrated ion selective multisensor technology for chemical and immunochemical applications for clinical use.
#### Medical device to which equivalence is claimed and comparison information 6.
The Dimension® EXL™ with LM system is substantially equivalent in intended use and technological characteristics to the Dimension® RxL Max® system. Both instruments are fully automated, microprocessor-controlled integrated instruments with similar detection technologies and modes of operation.
The Dimension® FT4L Flex® reagent cartridge is substantially equivalent in intended use and technological characteristics to the Dimension® Vista FT4 Flex® reagent cartridge. Both devices use LOCI technology and are intended for use in the quantitative measurement of Free Thyroxine in human serum and plasma. The devices have identical measuring ranges of 0.1 - 8.0 ng/dL.
The LOCI® Thyroid Calibrator is substantially equivalent in intended use and technological characteristics to the Dimension® LOCI 1 Calibrator. Both calibrators use a matrix comprised of stripped bovine albumin spiked with USP-grade Thyroxine.
{2}------------------------------------------------
### Comparison Information
Method comparison studies were conducted on the Dimension® FT4L Flex® vs. the Dimension® Vista FT4 Flex® reagent cartridge using one hundred-and-eighty (180) serum samples. The data was analyzed using least squares linear regression. The Dimension® FT4L Flex® demonstrated excellent correlation to the predicate device as evidenced by a correlation coefficient = 0.999. The resulting regressions statistics are shown below.
| Slope | 0.97 |
|--------|------------|
| 95% CI | 0.96-0.97 |
| y-int | -0.03 |
| r | 0.999 |
| Sy,x | 0.08 ng/mL |
{3}------------------------------------------------
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Dade Behring, Inc. c/o Ms. Yuk-Ting Lewis, Regulatory Affairs and Compliance Manager P.O. Box 6101, M/S 514 Newark, DE 19714-6101
MAR 13 2008
Re: k073604 Trade/Device Name: Dimension® FT4L Flex® reagént cartridge LOCI® Thyroid Calibrator Dimension® EXL™ with LOCI® Module (LM) Clinical Chemistry System Regulation Number: 21 CFR§ 862.1695 Regulation Name: Free thyroxine test system. Regulatory Class: Class II Product Code: CEC, JIT, JJE Dated: December 20, 2007 Received: December 21, 2007
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Sean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510(k) Number (if known): K073604
Device Name: Dimension® FT4L Flex® reagent cartridge LOCI® Thyroid Calibrator Dimension® EXL™ with LM clinical chemistry system
Indications For Use:
The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ with LM system. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.
The LOCI® Thyroid Calibrator is an in vitro diagnostic product for the calibration of the FT4L method on the Dimension® EXLTM with LM system.
The Dimension® EXL™ with LM system is an in vitro diagnostic device that is intended to measure a variety of analytes in human body fluids. The system utilizes photometric, turbidimetric, chemiluminescence and integrated ion selective multisensor technology for chemical and immunochemical applications for clinical use.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Carol C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k073604
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.