DIMENSION VISTA FREE THYROXINE FLEX REAGENT CARTRIDGE, FT4, DIMENSION VISTA THYROID STIMULATING HORMONE FLEX REAGENT
K140859 · Siemens Healthcare Diagnostics, Inc. · CEC · Nov 26, 2014 · Clinical Chemistry
Device Facts
Record ID
K140859
Device Name
DIMENSION VISTA FREE THYROXINE FLEX REAGENT CARTRIDGE, FT4, DIMENSION VISTA THYROID STIMULATING HORMONE FLEX REAGENT
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
CEC · Clinical Chemistry
Decision Date
Nov 26, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1695
Device Class
Class 2
Attributes
Pediatric
Intended Use
The FT4 method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease. The TSH method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone in human serum and plasma on the Dimension Vista® System. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
Device Story
In vitro diagnostic reagent cartridges for Dimension Vista® systems; measure Free Thyroxine (FT4) and Thyroid Stimulating Hormone (TSH) in human serum/plasma; used by clinical laboratory professionals for diagnosis/monitoring of thyroid/pituitary disorders. Device utilizes existing analytical chemistry methods; no changes to reagents or hardware. Submission adds pediatric reference intervals to labeling based on analysis of 421 pediatric patient samples. Healthcare providers use quantitative results compared against established reference intervals to assess thyroid function and guide clinical decision-making.
Clinical Evidence
No new clinical performance data required. Pediatric reference intervals were established by analyzing 421 patient samples (82 infants, 191 children, 148 adolescents) using a non-parametric approach or robust symmetric method (for infants) per CLSI EP28-A3c. Analytical performance data from predicate submissions (K053531, K060090) remains applicable.
Technological Characteristics
In vitro diagnostic reagent cartridges for use on Dimension Vista® systems. No changes to materials, sensing principles, or energy sources. Assays utilize established quantitative measurement methods for FT4 and TSH. Pediatric reference intervals added to labeling; analytical measuring ranges: FT4 (0.1-8.0 ng/dL), TSH (0.005-100 µIU/mL).
Indications for Use
Indicated for quantitative measurement of TSH and Free Thyroxine in human serum and plasma for diagnosis/treatment of thyroid or pituitary disorders in pediatric populations (infants 1-23 months, children 2-12 years, adolescents 13-20 years).
Regulatory Classification
Identification
A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in serum or plasma. Levels of free thyroxine in plasma are thought to reflect the amount of thyroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
K160647 — IMMULITE 2000 Third Generation TSH, IMMULITE 2000 Free T4 · Siemens Healthcare Diagnostics, Inc. · May 31, 2016
K150403 — ADVIA Centaur TSH3-Ultra · Siemens Healthcare Diagnostics, Inc. · May 1, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a cascading manner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 26, 2014
SIEMENS HEALTHCARE DIAGNOSTICS INC. FRANCES DILLON SR. MANAGER, REGULATORY AFFAIRS PO BOX 6101, M/S 514 NEWARK DE 19714-6101
Re: K140859
Trade/Device Name: Dimension® VISTA Free Thyroxine Flex® Reagent Cartridee, FT4: Dimension® VISTA Thyroid Stimulating Hormone Flex® Reagent Cartridge, TSH Regulation Number: 21 CFR 862.1695 Regulation Name: Free thyroxine test system Regulatory Class: II Product Code: CEC. JLW Dated: November 24, 2014 Received: November 25, 2014
Dear Frances Dillon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Courtney H. Lias -S
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K140859
## Device Name
Dimension® Vista Free Thyroxine Flex® reagent cartridge, FT4
Indications for Use (Describe)
The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.
| | X Prescription Use (Part 21 CFR 801 Subpart D)<br>Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*<br>The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | |
| | | | | |
| | Department of Health and Human Services<br>Food and Drug Administration | | | |
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K140859
## Device Name
Dimension Vista® Thyroid Stimulating Hormone Flex® reagent cartridge. TSH
Indications for Use (Describe)
The TSH method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone in human serum and plasma on the Dimension Vista® System. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PRAStaff@fda.hhs.gov
Type of Use (Select one or both, as applicable)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "SIEMENS" in a sans-serif font. The text is a teal color. The letters are evenly spaced and the word is horizontally oriented.
# Section 6. 510(k) Summary of Safety and Effectiveness
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.
The assigned 510(k) Number is: K140859
#### 1. Date Prepared
November 25, 2014
#### Applicant Information 2.
| Contact: | Frances A. Dillon.<br>Senior Manager, Regulatory Affairs |
|----------|-----------------------------------------------------------------------------------|
| Address: | Siemens Healthcare Diagnostics Inc.<br>P.O. Box 6101, M/S 514<br>Newark, DE 19714 |
| Phone: | 302-631-6951 |
| Fax: | 302-631-6299 |
| Email: | frances.a.dillon@siemens.com |
#### 3. Regulatory Information
| | FT4 | TSH |
|---------------------|-----------------------------------------------------------------|------------------------------------------------------------------------------|
| Trade Name | Dimension Vista® Free Thyroxine<br>Flex® reagent cartridge, FT4 | Dimension Vista® Thyroid Stimulating<br>Hormone Flex® reagent cartridge, TSH |
| Model Numbers | K6410 | K6412 |
| Common Name | Radioimmunoassay,<br>free thyroxine | Radioimmunoassay, thyroid stimulating<br>hormone |
| Classification Name | Free thyroxine test system | Thyroid stimulating hormone test<br>system |
| FDA Classification | Class II | Class II |
| Review Panel | Clinical Chemistry | Clinical Chemistry |
| Product Code | CEC | JLW |
| Regulation Number | 862.1695 | 862.1690 |
## Table 1. Regulatory Information for Dimension Vista® FT4 and TSH Assays
{5}------------------------------------------------
#### Predicate Device Information 4.
The inclusion of pediatric reference intervals to the labeling (Package Inserts) of the Dimension Vista® FT4 and TSH assays does not require any other device modifications (i.e. no change to design or manufacturing process). No changes were made to the reagents. Therefore, as shown in the table below, the predicate and subject devices are the same.
| | FT4 | TSH |
|------------------|-----------------------------------------------------------------|------------------------------------------------------------------------------|
| Predicate Device | Dimension Vista® Free Thyroxine<br>Flex® reagent cartridge, FT4 | Dimension Vista® Thyroid Stimulating<br>Hormone Flex® reagent cartridge, TSH |
| Subject Device | Same (As above) | Same (As above) |
| Predicate 510(k) | K053531 | K060090 |
Table 2. Summary of Predicate Devices for Dimension Vista® FT4 and TSH Assays
#### Substantial Equivalence Information 5.
The following table demonstrates substantial equivalence between the predicate Dimension Vista® thyroid assays (with unmodified labeling) and Dimension Vista® thyroid assays which have modified Instructions for Use (Package Inserts) including pediatric reference intervals.
| Item | Predicate Device<br>(Unmodified Labeling) | Subject Device<br>(With Pediatric Reference Intervals) |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Analytes | Dimension Vista® FT4<br>Free Thyroxine (FT4) | Dimension Vista® FT4<br>Same |
| | Dimension Vista® TSH<br>Thyroid Stimulating Hormone | Dimension Vista® TSH<br>Same |
| Reagents | Dimension Vista® FT4<br>Dimension Vista® Free Thyroxine<br>Flex® reagent cartridge, FT4 | Dimension Vista® FT4<br>Same |
| | Dimension Vista® TSH<br>Dimension Vista® Thyroid Stimulating<br>Hormone Flex® reagent cartridge, TSH | Dimension Vista® TSH<br>Same |
| Instruments | Dimension Vista® 500, 1000T, 1500<br>and 3000T systems | Same |
| Intended Use<br>Statements | Dimension Vista® FT4<br>The FT4 method is an in vitro<br>diagnostic test for the quantitative<br>measurement of Free Thyroxine in<br>human serum and plasma on the<br>Dimension Vista® System. | Dimension Vista® FT4<br>The FT4 method is an in vitro<br>diagnostic test for the quantitative<br>measurement of Free Thyroxine in<br>human serum and plasma on the<br>Dimension Vista® System.<br>Measurements obtained by this device<br>are used in the diagnosis and<br>treatment of thyroid diseases.* |
| Item | Predicate Device<br>(Unmodified Labeling) | Subject Device<br>(With Pediatric Reference Intervals) |
| | Dimension Vista® TSH<br>The TSH method is an in vitro<br>diagnostic test for the quantitative<br>measurement of Thyroid Stimulating<br>Hormone in human serum and plasma<br>on the Dimension Vista® System. | Dimension Vista® TSH<br>The TSH method is an in vitro<br>diagnostic test for the quantitative<br>measurement of Thyroid Stimulating<br>Hormone in human serum and plasma<br>on the Dimension Vista® System.<br>Measurements of thyroid stimulating<br>hormone produced by the anterior<br>pituitary are used in the diagnosis of<br>thyroid or pituitary disorders.* |
| | Analytical<br>Measuring Range<br>(Assay Range) | Dimension Vista® FT4<br>0.1-8.0 ng/dL [1.3-103 pmol/L] |
| | Dimension Vista® TSH<br>0.005–100 µIU/mL [mIU/L] | Dimension Vista® TSH<br>Same |
| | Adult<br>Reference Intervals | Dimension Vista® FT4<br>0.76-1.46 ng/dL [9.8-18.8 pmol/L]<br>calculated non-parametrically and<br>represents the central 95% of results<br>determined from a population of 199<br>apparently healthy adults (140 males<br>and 59 females, 18 - 59 years of age) |
| | Dimension Vista® TSH<br>0.358-3.74 µIU/mL [mIU/L] calculated<br>non-parametrically and represents the<br>central 95% of results determined from<br>a population of 297 apparently healthy<br>adults (187 males and 110 females, 8-<br>65 years of age). | Dimension Vista® TSH<br>Same |
| | Pediatric<br>Reference Intervals | Dimension Vista® FT4<br>None |
| | Dimension Vista® TSH<br>None | Dimension Vista® TSH<br>Infants (01 - 23 months)<br>0.816 - 5.91 µIU/mL<br>Children (02 - 12 years)<br>0.662 - 3.90 µIU/mL<br>Adolescents (13 – 20 years<br>0.463 - 3.98 µIU/mL) |
Table 3. Summary of Substantial Equivalence for Dimension Vista® FT4 and TSH Assays
{6}------------------------------------------------
# Table 3. Summary of Substantial Equivalence for Dimension Vista® FT4 and TSH Assays
*Additional text regarding indications added as FDA requested for 510(k) K132249.
{7}------------------------------------------------
#### Standard/Guidance Document Reference 6.
Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline -- Third Edition (CLSI C28-A3c).
#### 7. Pediatric Reference Intervals
Data from a total of 421 patients (82 infants, 191 children, 148 adolescents) were analyzed to establish the Dimension Vista® FT4 and TSH assay reference intervals for the studied pediatric population. Testing was performed on a Dimension Vista® 1500 system. These pediatric reference intervals, as well as the previously-established euthyroid adult reference intervals and analytical measuring ranges, are presented below.
| | Dimension Vista® FT4 | Dimension Vista® TSH |
|------------------------|----------------------|----------------------|
| Infants (01 - 23M) | 0.88 - 1.48 ng/dL | 0.816 - 5.91 µIU/mL |
| Children (02 – 12Y) | 0.81 - 1.35 ng/dL | 0.662 - 3.90 µIU/mL |
| Adolescents (13 – 20Y) | 0.78 - 1.33 ng/dL | 0.463 - 3.98 µIU/mL |
| Euthyroid Adults* | 0.76 - 1.46 ng/dL | 0.358 - 3.74 µIU/mL |
| Assay Range* | 0.1 - 8.0 ng/dL | 0.005 - 100 µIU/mL |
# Table 4. Comparison of Pediatric and Adult 95% Reference Intervals
* Information taken from existing Instructions for Use (Package Inserts)
A non-parametric approach was used to establish the reference intervals for children and adolescents where the 2.5th and 97.5th percentiles of the distribution of values were calculated. Because fewer than 120 patient samples were included for the infant age group, the reference intervals were calculated by using a robust measure of location and spread, as developed by Horn and Pesce and recommended in CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals.
The robust symmetrical method was used for the Dimension Vista FT4 analysis because the data was normally distributed. The reference interval calculated from the actual data using the robust symmetrical method was confirmed with two simulated reference intervals (nonparametric and robust symmetric). Each simulated reference interval was estimated as the means of the 95% upper and lower bounds from 1000 data sets. In each case, a normal distribution was fitted to the actual results, with either n=120 for the non-parametric analysis, or n=82 for the robust symmetric analysis.
For the Dimension Vista TSH assay, the sample distribution for infant subjects was highly positively skewed. As permitted by CLSI EP28-A3, the actual results were natural log transformed to produce a well-fitted normal distribution. The infant reference intervals were then calculated using the robust symmetric method on the log transformed data. The reference interval was confirmed with two simulated reference intervals (nonparametric and robust symmetric). Each simulated reference interval was estimated as the means of the 95% upper and lower bounds from 1000 data sets. In each case, a normal distribution was fitted to the actual results, with either n=120 for the non-parametric analysis, or n=82 for the robust symmetric analysis.
The 90% confidence intervals for the Infant 95% Reference Intervals are listed below.
{8}------------------------------------------------
| Infants (01 - 23M) | Dimension® Vista FT4 | Dimension® Vista TSH |
|-------------------------------------------|----------------------|-------------------------------------------|
| Calculation Method | Robust Symmetric | Robust Symmetric After Log<br>Transformed |
| Reference Interval | 0.88 - 1.48 ng/dL | 0.816 - 5.91 µlU/mL |
| 90% Confidence Interval<br>of Lower Limit | 0.82 - 0.95 ng/dL | 0.683 - 0.962 µIU/mL |
| 90% Confidence Interval<br>of Upper Limit | 1.42 - 1.53 ng/dL | 5.06 - 7.09 µlU/mL |
Table 5. Infant Reference Interval - 90% Confidence Intervals for Upper and Lower Limits
#### Performance Characteristics 8.
The inclusion of pediatric reference intervals in the Instructions for Use (Package Inserts) does not necessitate the collection of additional analytical performance data as no changes were made to the reagents. Therefore, all analytical performance data previously reviewed for the Dimension Vista® FT4 and TSH assays continue to apply to these assays. All performance data is cross-referenced to the predicate 510(k) submissions for the FT4 and TSH assays (K053531 and K060090).
Specifically, the following studies are not needed for the purpose of this submission:
- . Precision/Reproducibility
- l Linearity
- . Calibrator/Assay Traceability
- l Calibrator/Assay Stability
- Assay Cut-off
- . Method Comparison
- . Matrix Comparison
- Sensitivity (Detection Limits, LoB, LoD, LoQ) .
- Analytical Specificity I
#### 9. Shelf-Life
The inclusion of pediatric reference intervals in the Instructions for Use (Package Inserts) does not necessitate the collection of additional stability data as no changes were made to the reagents. Therefore, all stability methods, acceptance criteria and data previously reviewed for the Dimension Vista® FT4 and TSH assays continue to apply to these assays.
Specifically, the following stability studies are not needed for the purpose of this submission:
- I Shelf Life Stability
- . Onboard Stabilitv
- I Open Vial Stability
#### 10. Conclusions
The Dimension Vista® FT4 and TSH assays with pediatric reference intervals are substantially equivalent to the currently marketed Dimension Vista® FT4 and TSH assays.
{9}------------------------------------------------
No changes were made to the reagents. The inclusion of pediatric reference intervals in the Instructions for Use (Package Inserts) does not require a change in the device design or a change in the manufacturing process.
The addition of pediatric reference intervals for the Dimension Vista® FT4 and TSH assays is further supported by the following rationale:
- 1. Testing of pediatric patients is within the established indications for use in the diagnosis and treatment of thyroid disease), as described in 21 CFR §862.1695 and 21 CFR §862.1690.
- 2. The newly-established pediatric reference intervals are either within or are above the previously-established reference intervals for euthyroid (normal thyroid) adult populations and they are within the analytical measuring ranges of the Dimension Vista® FT4 and TSH assays. Therefore, the Dimension Vista® FT4 and TSH assays have appropriate analytical performance to test pediatric patients.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.