DRG SLV TESTOSTERONE ELISA TEST

K052649 · Drg Intl., Inc. · CDZ · Jan 27, 2006 · Clinical Chemistry

Device Facts

Record IDK052649
Device NameDRG SLV TESTOSTERONE ELISA TEST
ApplicantDrg Intl., Inc.
Product CodeCDZ · Clinical Chemistry
Decision DateJan 27, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1680
Device ClassClass 1

Intended Use

An Enzyme Immunoassay for the in vitro diagnostic quantitative measurement of free active testosterone in saliva. Measurement of testosterone is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

Device Story

The DRG Salivary Testosterone ELISA Kit is an in vitro diagnostic assay for professional use. It utilizes a competitive enzyme immunoassay principle to measure free active testosterone in human saliva samples. The device consists of a microtiter plate coated with mouse anti-testosterone antiserum, enzyme-conjugated testosterone (horseradish peroxidase), substrate solution (TMB), and stop solution. During the assay, testosterone in the patient sample competes with the enzyme-conjugated testosterone for binding sites on the plate. After incubation and washing, a substrate is added, producing a colorimetric signal measured at 450nm via a calibrated EIA reader. The intensity of the color is inversely proportional to the testosterone concentration in the sample. Healthcare providers use the quantitative output to assist in diagnosing and monitoring androgen-related conditions such as hypogonadism, puberty disorders, and virilization. The assay provides a standardized method for assessing hormonal status in clinical settings.

Clinical Evidence

Bench testing only. Performance validated via method comparison studies (n=99 and n=81) against a commercial LIA method, yielding correlation of 0.904 and R2=0.9866. Analytical sensitivity: 1.857 pg/mL (lowest detectable); functional sensitivity: 7.1 pg/mL. Specificity evaluated via cross-reactivity testing against 12 steroids; minimal cross-reactivity observed. Reproducibility (intra-assay, inter-assay, inter-lot) demonstrated with CVs generally <10%. Linearity established from 7.1 to 4500 pg/mL.

Technological Characteristics

Competitive enzyme immunoassay; microtiter plate coated with mouse monoclonal anti-testosterone antiserum; horseradish peroxidase conjugate; TMB substrate; colorimetric detection at 450nm. Requires calibrated EIA reader. Standards prepared in artificial saliva matrix; traceability to GC-MS methods. No electronic connectivity; standalone manual assay.

Indications for Use

Indicated for in vitro diagnostic quantitative measurement of free active testosterone in saliva for diagnosis and treatment of androgen-related disorders in males (hypogonadism, puberty disorders, impotence) and females (hirsutism, virilization due to tumors, polycystic ovaries, adrenogenital syndromes).

Regulatory Classification

Identification

A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 2 7 2006 K052649 # SUMMARY OF SAFETY AND EFFECTIVENESS FOR DRG SALIVARY TESTOSTERONE ELISA | Manufacturer: | DRG International, Inc.<br>1167 U.S. Highway 22<br>Mountainside, NJ 07092 | |----------------------|----------------------------------------------------------------------------------------------------------------| | Contact Information: | Lehnus & Associates<br>Gary Lehnus<br>150 Cherry Lane Rd.<br>East Stroudsburg, PA 18301<br>Tel: (570) 620-0198 | #### Device Name / Classification: The device trade name is the DRG SLV Testosterone ELISA having FDA assigned name: Testosterone test system, 21 CFR, 862.1680, categorized as Class I "reserved" medical devices for the Clinical Chemistry and Clinical Toxicology Panel, as Product Code CDZ. #### Test Principle The DRG Salivary Testosterone ELISA Kit is based on the competition princible and the microplate separation. An unknown amount of free testosterone present in the sample and a fixed amount of testosterone conjugated with horseradish peroxidase compete for the binding sites of mouse monoclonal testosterone -antiserum coated onto the wells. After one hour incubation the microplate is washed to stop the competition reaction. After addition of the substrate solution the concentration of testosterone is inversely proportional to the optical density measured. #### Device Intended Use: An Enzyme Immunoassay for the in vitro diagnostic quantitative measurement of free active testosterone in saliva. Measurement of testosterone is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delaved or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes. #### Device Performance Normal Range Study In order to determine the normal range of SLV Testosterone, saliva samples from 187 adult male and 188 adult female apparently healthy subjects, ages 21 to 75 years, were collected in the morning and analyzed using the DRG SLV Testosterone ELISA kit. The following range was calculated from this study. | | Men $\male$ | | | Women $\female$ | | | |--------------------|--------------------|--------|----|--------------------|--------|----| | Age Group<br>Years | Range<br>(5 - 95%) | Median | n | Range<br>(5 - 95%) | Median | n | | 21 - 30 | 47.2 - 136.2 | 92.8 | 42 | 7.9 - 50.4 | 20.8 | 40 | | 31 - 40 | 46.8 - 106.8 | 73.6 | 37 | <7.0 - 44.8 | 17.1 | 40 | | 41 - 50 | 36.5 - 82.7 | 58.8 | 34 | <7.0 - 39.4 | 18.3 | 38 | | 51 - 60 | 19.1 - 89.0 | 44.5 | 36 | <7.0 - 29.8 | 19.2 | 38 | | 61 - 75 | 12.2 - 68.6 | 38.9 | 38 | <7.0 - 29.3 | 16.0 | 32 | {1}------------------------------------------------ ### Method comparison A study was performed that evaluated saliva samples from 99 male and female subjects ages 20 to 70 years. The saliva samples were run in duplicate on the DRG test and a commercially available LIA method to determine the concentration of free Testosterone in the samples. A correlation of 0.904 and regression formula of y = 0.9251x - 7.4369 was obtained versus this method. Another study was performed to further evaluate the substantial equivalence of the DRG SLV Testosterone to the LIA saliva test. The concentration of testosterone in 81 additional saliva samples collected from 40 - 65 year old men and women was determined using DRG SLV testosterone kit and the other method. From this study an R2 = 0.9866 was obtained with the following regression. Image /page/1/Figure/3 description: This image is a scatter plot titled "DRG Comparison". The x-axis is labeled "LIA" and ranges from 0 to 700, while the y-axis is labeled "DRG ELISA" and also ranges from 0 to 700. The scatter plot shows a positive correlation between the two variables, and a line of best fit is drawn through the data points. The equation of the line of best fit is given as y = 1.0057x - 2.4196. #### Sensitivity The lowest analytical detectable level of testosterone that can be distinquished from the Zero Standard is 1.857 pg/mL at the 95 % confidence limit. The lowest functional sensitivity of 7.1 pg/mL at the 95% confidence limit was obtained. #### Specificity The following materials have been evaluated for cross reactivity. The percentage indicates cross reactivity at 50% displacement compared to Testosterone. | Steroid | % Cross Reaction | |------------------------|------------------| | Testosterone | 100% | | 5α-Dihydrotestosterone | 0.80% | | Androstenedione | 0.90% | | 11β-hydroxysterone | 3.30% | | 17α-methyltestosterone | 0.10% | | 19-Nortestosterone | 3.30% | | Epitestosterone | 0.10% | | Estradiol | 0.10% | | Progesterone | < 0,10% | | Cortisol | < 0,10% | | Estrone | < 0,10% | | Danazol | < 0,10% | {2}------------------------------------------------ # Reproducibility ## Intra-Assay The intra-assay variation was determined by 20 replicate measurements of 5 saliva samples within one run. The within-assay variability is shown below: | Mean (pg/ml) | 144.00 | 256.15 | 81.30 | 35.35 | 12.94 | |--------------|--------|--------|-------|-------|-------| | SD | 8.974 | 17.587 | 5.545 | 2.498 | 1.787 | | CV (%) | 6.23 | 6.87 | 6.82 | 7.07 | 13.81 | | n = | 20 | 20 | 20 | 20 | 20 | #### Inter-Assay The inter-assay (between-run) variation was determined by duplicate measurements of 5 saliva samples over 10 days. | Mean (pg/mL) | 823.08 | 87.57 | 118.82 | 112.13 | 33.61 | |--------------|--------|-------|--------|--------|-------| | SD | 45.314 | 6.864 | 8.864 | 8.628 | 3.232 | | CV (%) | 5.51 | 7.84 | 7.46 | 7.69 | 9.62 | | n = | 20 | 20 | 20 | 20 | 20 | ## Inter-Lot The Inter-Lot (between-lot) variation was determined by triplicate measurements of five saliva samples in three different kit lots. The between lot variability is shown below: | | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 | |--------------|----------|----------|----------|----------|----------| | Mean (pg/ml) | 64.5 | 352.89 | 517.65 | 44.00 | 116.54 | | SD (pg/ml) | 3.77 | 13.44 | 15.01 | 1.53 | 5.00 | | CV (%) | 5.85 | 3.81 | 2.90 | 3.47 | 4.29 | | n = | 9 | 9 | 9 | 9 | 9 | ## Recovery Recovery of the DRG ELISA was determined by adding increasing amounts of the analyte to six different saliva samples containing different amounts of endogenous analyte. Each sample (native and spiked) was assayed and analyte concentrations of the samples were calculated from the standard curve. The percentage recoveries were determined by comparing expected and measured values of the samples | Sample | Measured<br>(pg/ml) | Expected<br>(pg/ml) | Recovery<br>(%) | |--------|---------------------|---------------------|-----------------| | 1 | 8.39 | - | - | | | 2396.90 | 2508.39 | 95.56 | | | 517.17 | 508.39 | 101.73 | | | 271.10 | 258.39 | 104.92 | | | 55.95 | 58.39 | 95.82 | | 2 | 46.23 | - | - | | | 2474.70 | 2546.23 | 97.19 | | | 564.37 | 546.23 | 103.32 | | | 308.97 | 296.23 | 104.30 | | | 88.87 | 96.23 | 92.35 | | 3 | 122.09 | - | - | | | 2602.41 | 2622.09 | 99.25 | | | 591.16 | 622.09 | 95.03 | {3}------------------------------------------------ | | 352.56 | 372.09 | 94.75 | |---|---------|---------|-------| | | 166.40 | 172.09 | 96.69 | | 4 | 210.00 | - | - | | | 2650.00 | 2710.00 | 97.8 | | | 700.00 | 710.00 | 98.6 | | | 240.00 | 240.00 | 100.0 | | 5 | 1250.00 | - | - | | | 3700.00 | 3750.00 | 98.7 | | | 1680.00 | 1750.00 | 96.0 | | | 1250.00 | 1280.00 | 97.7 | | 6 | 2090.00 | - | - | | | 4550.00 | 4590.00 | 99.1 | | | 2650.00 | 2590.00 | 102.3 | | | 2130.00 | 2120.00 | 100.5 | ### Linearity Six saliva samples containing different amounts of analyte were serially diluted with zero standard and assayed with the DRG ELISA. Three native samples were serially diluted, and 3 samples were spiked with testosterone and then serially diluted up to 1:128. The percentage recovery was calculated by comparing the expected and measured values for testosterone. An assay linearity of 7.1 - 4500 pg/mL has been identified as the usable range. Samples above this range must be diluted and re-run. | | | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 | Sample 6 | |-----------------------|------|----------|----------|----------|----------|----------|----------| | Concentration (pg/ml) | | 4312.00 | 1838.00 | 440.00 | 8000.0 | 4500.0 | 5500.0 | | Average % Recovery | | 97.5 | 99.6 | 98.9 | 100.3 | 100.8 | 100.7 | | Range of | from | 96.3 | 98.4 | 93.6 | 93.6 | 93.7 | 94.1 | | Recovery % | to | 98.7 | 101.0 | 106.7 | 106.7 | 107.0 | 107.8 | {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JAN 2 7 2006 DRG International, Inc. c/o Mr. Gary Lehnus Lehnus & Associates 150 Cherry Land Rd. East Stroudsburg, PA 18301 Re: k052649 Trade/Device Name: DRG SLV Testosterone ELISA Test Regulation Number: 21 CFR§ 862.1680 Regulation Name: Testosterone test system Regulatory Class: Class I Product Code: CDZ Dated: December 27, 2005 Received: January 9, 2006 Dear: Mr. Lehnus We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Alberto Garcia Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Indications for Use 510(k) Number (if known): k052649 DRG SLV Testosterone ELISA Test Device Name: Indications For Use: An enzyme immunoassay for the in vitro diagnostic quantitative measurement of free active testosterone in saliva. Measurement of testosterone is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (OlVD) Ann Chappie Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 710(k) K052649 Page of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...