FREE TESTOSTERONE ENZYMEIMMUNOASSAY (EIA) KIT

K030730 · Diagnostics Biochem Canada, Inc. · CDZ · Aug 18, 2003 · Clinical Chemistry

Device Facts

Record IDK030730
Device NameFREE TESTOSTERONE ENZYMEIMMUNOASSAY (EIA) KIT
ApplicantDiagnostics Biochem Canada, Inc.
Product CodeCDZ · Clinical Chemistry
Decision DateAug 18, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1680
Device ClassClass 1

Intended Use

The dbc CAN-fTE-260 EiAsy™ Free Testosterone enzymeimmunoassay (EiA) kit provides the reagents necessary for the direct determination of Free Testosterone in human serum. This assay is intended for in vitro diagnostic use only. Measurement of Free Testosterone are used in the diagnosis in male sex hormones (androgens) and in females hirsutism (excessive hair) and virilisation (masculinization).

Device Story

The EiAsy Free Testosterone EiA is an in-vitro diagnostic enzyme-linked immunosorbent assay (ELISA) kit. It utilizes a microtiter plate coated with polyclonal antibodies and a testosterone-horseradish peroxidase conjugate. The device processes human serum samples by adding patient serum, control serum, or standards to the wells along with the conjugate solution. Following a 60-minute incubation at 37°C, the plates are washed and dried. A TMB substrate is added, followed by a stopping solution, resulting in a color change. The plate is read in a microtiter plate reader at 450 nm. The device is intended for use in clinical laboratory settings by trained personnel. The output is a quantitative measurement of free testosterone, which assists healthcare providers in diagnosing androgen-related conditions in males and hirsutism or virilization in females.

Clinical Evidence

Bench testing only. Performance characteristics evaluated include sensitivity, specificity, and intra/inter-assay precision. Correlation with the predicate device was established using 61 human serum samples, yielding a correlation coefficient of r=0.8908.

Technological Characteristics

Enzyme-linked immunosorbent assay (ELISA) using polyclonal antibodies coated on a 96-well microtiter plate. Reagents include testosterone-horseradish peroxidase conjugate, assay buffer, wash buffer, TMB substrate, and stopping solution. Detection via microtiter plate reader at 450 nm.

Indications for Use

Indicated for the measurement of free testosterone in human serum to aid in the diagnosis of male sex hormone (androgen) disorders and female conditions including hirsutism and virilization.

Regulatory Classification

Identification

A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 2 ## AUG 1 8 2003 K030730 # dbc - DIAGNOSTICS BIOCHEM CANADA INC. 1020 Hargrieve Road, London, Ontario, Canada NGE 1P5 e-mail dbclabs@bellnet.ca Tel/Fax (519) 681-8731 G. R. Dumont M.D., M.Sc. A.J. Desrosters, D.Sc.M.Sc. J. U. Larvea, Ph. D. M.Sc. DATE: July 2, 2003 Dr. G. R. Dumont Contact Person: Tel (519) 681-8731 ## 19. 510 K Summary This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 1. The assigned 510K number is K030730 2. The classification name of the device: Enzyme immunoassay for free testosterone 3. Classification number 862.1680 4. Proprietary name: Free Testosterone by Enzymeimmunoassay (EiA) Class I 5. Use of the device: Direct determination of Free Testosterone by enzymeimmunoassay (Elisa) in human serum. It is an in-vitro diagnostic kit 6. The predicate device's name and address of which we claim equivalence: Diagnostics Products Corporation, (DPC), 5700 West 96 Street, Los Angeles, CA, 900159597 > COAT-A-Count Free Testosterone kit Cat. No. TKTF1 (100 tubes) and TKFT2 (200 tubes) 510(K) K844423 7. The free Testosterone Elisa kit consists of one polyclonal antibody which is coated on microtiter plate (96 wells per kit). The antigen Testosterone 3 carboxymethyl oxime {1}------------------------------------------------ is conjugated to an enzyme namely horse radish peroxidase. The standards are prepared in human serum matrix and all other reagents within the kit namely assay buffer, wash buffer concentrate, substrate tetramethytbenzidine (TMB) and stopping solution. 50 ul of patient serum , control serum and each standard are added to each well. 100 ul of conjugate solution is again added to each well. The incubation time is 37°C for 60 minutes. After incubation the plates are washed 3 times, each time with 300 ul of diluted wash buffer. The plates are dried. 100 ul of substrate TMB is added to each well and allowed to incubate for 10 - 15 minutes at 37°C. 50 ul of stopping solution is added into each well and the colour becomes yellow. The plate is read within 20 minutes in a microtiter plate reader at 450 nm. #### 8. Substantial equivalence: 1. The DBC In-vitro diagnostic device namely free Testosterone by enzymeimmunoassay (EiA) has the same intended use as current device on the market referred to "predicate device". 2. The dbc In-vitro diagnostic device namely free Testosterone has the same characteristics as predicate device namely antigen/antibody. 3. The dbc results obtained of a number of human serum samples substantial equivalent to the predicate device. The results need not be exactly the same but a correlation of equivalence is necessary in this case r=0.8908 4. The dbc free testosterone reflect the same equivalence as the predicate devices which can be found in our results of performance characteristics namely, sensitivity, specificity, Intra and Inter assay precision and the results of 61 human serumsamples 5. dbc free testosterone intended use, methodology, reagents materials and control serum necessary to perform the test are similar to the predicate device. 6. Our 510K presentation demonstrates substantial equivalence to predicate device in regards: - · intended use - · performance - · material - · effectiveness - · standards - · control {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows a logo with a circular border containing text, and an abstract symbol in the center. The text around the border is partially visible and appears to be in all caps. The central symbol consists of four curved lines that converge on the right side, resembling a stylized bird in flight. The image is in black and white. AUG 1 8 2003 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 G.R. Dumont, M.D., M.Sc. Director of Laboratories Diagnostics Biochem Canada, Inc. 1020 Hargrieve Road London, Ontario CANADA N6E 1P5 Re: k030730 Trade/Device Name: EiAsy 110 Free Testosterone EiA Regulation Number: 21 CFR 862.1680 Regulation Name: Testosterone test system Regulatory Class: Class I Product Code: CDZ Dated: May 28, 2003 Received: May 29, 2003 Dear Dr. Dumont: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## 18. Statements of Indication for Use K630730 510K number Device Name: EiAsyTM Free Testosterone EiA ### Indication for Use: The dbc CAN-fTE-260 EiAsy™ Free Testosterone enzymeimmunoassay (EiA) kit provides the reagents necessary for the direct determination of Free Testosterone in human serum. This assay is intended for in vitro diagnostic use only. Measurement of Free Testosterone are used in the diagnosis in male sex hormones (androgens) and in females hirsutism (excessive hair) and virilisation (masculinization). Or Prescription USE.............................................................................................................................................................. Over-The-Counter Use ......... (Per 21 CFR 801.109 Dean Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number. K030736
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...