K121955 · Respironics, Inc. · NHJ · Dec 14, 2012 · Anesthesiology
Device Facts
Record ID
K121955
Device Name
COUGHASSIST T70 DEVICE
Applicant
Respironics, Inc.
Product Code
NHJ · Anesthesiology
Decision Date
Dec 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Respironics CoughAssist T70 assists patients in clearing retained bronchopulmonary secretions by gradually applying a positive pressure to the airway, then rapidly shifting to a negative pressure. This rapid shift in pressure, via a facemask, mouthpiece or an endotracheal or tracheostomy tube, produces a high expiratory flow rate from the lungs, simulating a cough.
Device Story
Portable electric secretion clearance device; uses blower and valve to apply alternating positive/negative pressure to airway; simulates cough. Input: patient airway pressure; Output: high expiratory flow. Used in hospital, institution, or home by patients, caregivers, or clinicians. Features 'CoughTrak' to trigger therapy based on spontaneous patient breathing effort. Includes graphical user interface, SD card data logging, and oximetry accessory interface. Benefits patients unable to clear secretions due to conditions like muscular dystrophy, myasthenia gravis, or spinal cord injury.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including black-box performance testing and white-box software verification. Testing confirmed device meets design specifications under worst-case scenarios, including extreme flow rates and sensor malfunctions.
Technological Characteristics
Portable electric blower/valve system. Pressure range: 0 to 70 cmH2O (inhale/exhale). Connectivity: SD card, wired remote, oximetry accessory, external DC power. Software: PID-controlled motor/valve system with adaptive gain. Standards: ISO 14971, ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60068, IEC 62304.
Indications for Use
Indicated for adult or pediatric patients unable to cough or clear secretions effectively. Used with facemask, mouthpiece, or adapter to endotracheal/tracheostomy tube. Intended for hospital, institutional, or home use.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
K132988 — VITAL COUGH · Lung Assist, Inc. · Mar 28, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
K121955
.
Philips Respironics CoughAssist T70
Premarket Notification -- Traditional 510(k)
# Section 4 - 510(k) Summary
# Administrative Information and Device Identification
DEC 1 4 2012
| Name and address of the<br>manufacturer and sponsor of the<br>510(k) submission: | Manufacturer: |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Respironics, Inc. |
| | 175 Chastain Meadows Court |
| | Kennesaw, GA 30144-3714 |
| | Office: (770) 423-2322 |
| | Fax: (770) 423-2300 |
| | Sponsor: |
| | |
| | Respironics |
| | 1740 Golden Mile Highway |
| | Monroeville, PA 15146 |
| | Office: 724-387-7562 |
| | Fax: 724-387-7490 |
| FDA registration number of the<br>manufacturer of the new device: | Manufacturer: 1040777 (Establishment<br>registration number) |
| | Sponsor: 2518422 (Establishment registration<br>number) |
| Official contact person for all<br>correspondence: | Colleen Witt, RAC |
| | Senior Regulatory Affairs Engineer |
| | Philips Respironics |
| | 1740 Golden Mile Highway<br>Monroeville, PA 15146 |
| | Office: 724-387-4870 |
| | Fax: 724-387-7490 |
| | Email: colleen.witt@philips.com |
| Date Prepared: | June 29, 2012 |
| Device Name: | Philips Respironics CoughAssist T70 |
| Proprietary name of new device: | Philips Respironics CoughAssist T70 |
| Common or usual name of the<br>device: | Secretion Clearance Device |
| Philips/Respironics model number: | Philips Respironics CoughAssist T70 |
| Classification of new device: | Class II |
| Classification of the predicate<br>device: | Class II |
| Classification Panel: | Anesthesiology |
| Panel Code: | NHJ - noncontinuous ventilator (IPPB) |
| CFR Regulation Number: | 21 CFR 868.5905 |
| | a) Identification. A noncontinuous ventilator<br>(intermittent positive pressure breathing – IPPB)<br>is a device intended to deliver intermittently an<br>aerosol to a patient's lungs or to assist a patient's<br>breathing. |
| | (b) Classification. Class II (performance<br>standards). |
| Predicate Device Name(s) and<br>510(k) numbers: | Emerson CoughAssist, Model CA-3000 –<br>K002598 |
| | Dima Italia Negavent DA-3 Plus Pegaso -<br>K072292 |
| Reason for submission: | Device modifications |
{1}------------------------------------------------
# Description of Device
The Philips Respironics CoughAssist T70 is a portable electric device which utilizes a blower and a valve to apply alternating positive and then a negative pressure to a patient's airway in order to assist the patient in clearing retained bronchopulmonary secretions by simulating a cough. It includes a means to adjust the pressure and suction levels applied, a pressure gauge to measure the pressures, and a means (optional) to reduce the positive pressure (inhale) flow. The air is delivered to and from the patient via a breathing circuit incorporating a flexible tube, a bacterial filter and either a facemask, a mouthpiece or an adapter to a tracheostomy or endotracheal tube.
Like its predicates, the Respironics CoughAssist T70 device is for use on any adult or pediatric patient unable to cough or clear secretions effectively. It may be used either with a facemask or mouthpiece, or with an adapter to a patient's endotracheal or tracheostomy tube. Operators of the device are expected to be patients, lay caregivers (family members), clinicians, home nurses, homecare provider technicians, physicians and service personnel. The environment of use for the device will be for use in the hospital, institutional environment or in the home.
The CoughAssist T70 device is designed to deliver a cough cycle to a patient. A cough cycle includes a positive pressure inhale phase, a negative pressure exhale phase and a near zero pressure pause phase.
The CoughAssist T70 engineering design incorporates closed loop control algorithms through the motor controller and valve controller software algorithms. Block diagrams and a stability analysis have been provided for these closed loop algorithms.
The motor and valve controller utilize PID controllers. The PID gains are empirically tuned to achieve the necessary response and stability of the system to give best performance for all patient cases identified in the Product Requirements Document (provided in Attachment A of
{2}------------------------------------------------
the original submission). Additionally, adaptive gain changes are implemented in the Valve Controller to reject any external disturbance and to account for process and condition changes.
The CoughAssist T70 Valve (Pressure) Control system is designed to maintain stability throughout the operating flow range of the device. This includes steady-state stability with open and occluded patient circuit as well as disturbance rejection stability in cases where the device rapidly transitions from high flow to no flow and back to high flow again (open patient circuit to occluded patient circuit back to open patient circuit). Additionally, when in the inhale and exhale phases, the device pressure is continuously monitored to ensure that the pressure being delivered is the same polarity as the set point pressure. In the event of extreme flow rates (if the patient becomes disconnected from the device during therapy), the motor and valve controllers are designed to avoid any controller instability.
If the device is in Auto mode, the device will continuously trigger and cycle. If CoughTrak is enabled, the device will stay in the Pause waiting for the patient to trigger (which will not happen if the patient is not connected). In the Pause phase, the pressure is limited in order to minimize flow; a standby state is provided to preserve motor operation.
# Statement of Intended Use
The Respironics CoughAssist T70 assists patients in clearing retained bronchopulmonary secretions by gradually applying a positive pressure to the airway, then rapidly shifting to a negative pressure. This rapid shift in pressure, via a facemask, mouthpicce or an endotracheal or tracheostomy tube, produces a high expiratory flow rate from the lungs, simulating a cough.
# Indications for Use
The Respironics CoughAssist T70 is intended for use on adult or pediatric patients unable to cough or clear secretions effectively. It may be used either with a facemask or mouthpiece, or with an adapter to a patient's endotracheal or tracheostomy tube. The device is intended to be used in the hospital, institutional environment or in the home.
# Comparison of Device Technological Characteristics to Predicate Device
#### Substantial Equivalence
The Respironics CoughAssist T70 device has the following similarities to the previously cleared predicate devices:
- D Same intended use.
- Same operating principle.
- Same technology.
- D Same manufacturing process.
{3}------------------------------------------------
| The table below summarizes the key technical characteristics between the CoughAssist T70 to | |
|---------------------------------------------------------------------------------------------|--|
| those of the predicate devices listed in the submission: | |
| Characteristic<br>CoughAssist<br>Pegaso<br>Adult or pediatric patient<br>Patient<br>Adult or pediatric<br>Adult or pediatric<br>Population<br>unable to cough or clear<br>patient unable to<br>patient with an<br>secretions effectively<br>cough or clear<br>ineffective cough<br>scretions effectively<br>due to muscular | Technological | Description | Emerson | Negavent DA-3 Plus |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|-------------|------------|--------------------|
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | dystrophy, | |
| myasthenia gravis, | | | | |
| poliomyelitis, or | | | | |
| other neurologic | | | | |
| disorder with some | | | | |
| paralysis of the | | | | |
| respiratory muscles, | | | | |
| such as spinal cord | | | | |
| injury. | | | | |
| Delivery Type<br>Non Invasive or Invasive<br>Non Invasive or<br>Non Invasive or | | | | |
| Invasive<br>Invasive | | | | |
| Modes of<br>Manual and Auto<br>Manual and Auto<br>Manual and Auto | | | | |
| Operation | | | | |
| 0 to 70 cmH2O<br>Inhalation<br>0 to 60 cmH2O<br>+5 to 70 cmH2O | | | | |
| Pressure | | | | |
| 0 to -70 cmH2O<br>Exhalation<br>0 to -60 cmH2O<br>-5 to -70 cmH2O | | | | |
| Pressure | | | | |
| Inhale Flow<br>Low, Medium, High<br>Low and High<br>Low, Medium, High | | | | |
| Pause Time<br>0 to 5 seconds<br>0.1 - 9.9 seconds<br>0 to 5 seconds | | | | |
| Phases of<br>Insufflation, Exsufflation,<br>Insufflation,<br>Insufflation, | | | | |
| Therapy Cycle<br>Pause<br>Exsufflation, Pause<br>Exsufflation, Pause | | | | |
| Therapy<br>CoughTrak<br>N/A<br>N/A | | | | |
| Features | | | | |
| CoughAssist,<br>Dynamic Stability Analysis<br>N/A<br>N/A | | | | |
| including<br>Flow and Pressure Based | | | | |
| CoughTrak<br>Oscillation Detection | | | | |
| Safety Protocols<br>Extreme Flow Rate Control | | | | |
| and Response<br>Sensor Malfunction Stability | | | | |
| Remote Data<br>A secure digital (SD) card<br>N/A<br>N/A | | | | |
| provides means for data access<br>Access | | | | |
{4}------------------------------------------------
The table below provides a description of the modifications to the Respironics CoughAssist T70 device that are the subject of this 510(k) submission:
| Device Modification | Description | Device Modification | Testing Summary |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User Interface | A new graphical user interface with hierarchal menu system | User Interface including<br>displayed therapy<br>parameters | Has been verified to meet product requirements defined for the<br>CoughAssist T70 UI. Bench testing, code reviews and software unit<br>testing was performed to ensure that all display functions, user<br>controls and informational messages performed as intended. The<br>User Interface was verified to ensure that it displayed the proper<br>data and expected therapy information. |
| Delivered Pressure | The device Inhale pressure is 0 to 70 cmH2O and Exhale pressure is<br>0 to -70 cmH2O | Delivered Pressure and<br>Flow Stability | Verification testing has verified that the CoughAssist T70 device<br>delivers accurate pressure for the intended duration of the<br>insufflation, exsufflation and pause phases of therapy. The<br>CoughAssist T70 has been verified to demonstrate pressure and<br>flow stability across all patient test cases, including extreme flow<br>rates and sensor malfunctions (i.e. sensor inaccuracies and / or<br>complete sensor drop out). |
| CoughTrak | A feature for patients who can provide a spontaneous breathing<br>effort to trigger the cough sequence, instead of using manual or auto<br>modes of therapy. This software feature monitors the device outlet<br>pressure and initiates the Insufflation phase of therapy delivery<br>when the pressure decreases below a set threshold indicative of<br>patient effort. | CoughTrak | The CoughTrak feature of the CoughAssist T70 device has been<br>verified to meet product specifications with each defined patient<br>case simulation. The availability, operation and triggering<br>performance has been verified to operate across the range of patient<br>cases. |
| Displayed Therapy<br>Parameters | The device displays outlet pressure, measured peak cough flow and<br>measured inhaled tidal volume | Data Management | Data management of the CoughAssist T70 device has been verified<br>to meet product specifications for SD card data integrity. Software<br>verification, including code reviews and software verification<br>testing was conducted to verify the expected results with the SD<br>card. |
| Data Management | Therapy data will be stored on an SD card, which will be<br>compatible with both DirectView and Encore | Oximetry Accessory | The Oximetry Accessory has been tested to verify the behavior of<br>the pulse oximetry data and the proper display of data received by<br>the oximetry accessory. |
| Oximetry Accessory | Device shall interface with the Oximetry accessory to display<br>current SpO2 and heart rate data received from the oximeter | Wired Remote Control | Verification testing has been conducted to verify the availability and<br>operation of the CoughAssist T70 wired remote control device. |
| Wired Remote Control | A remote control accessory (foot pedal) is provided to initiate<br>manual therapy by means of a wired remote control interface | External Suctioning<br>Accessory Interface | The CoughAssist T70 device has been verified to demonstrate that<br>the device provides power to a suctioning accessory as specified. |
| External Suctioning<br>Accessory Interface | Device shall provide a DC output intended to power an external<br>suctioning accessory | Power Management | The CoughAssist T70 has been verified to properly interface with<br>the detachable battery, properly display the battery information,<br>properly charge the battery and indicate a low battery charge. The<br>CoughAssist T70 has also been verified to be compatible with an<br>external DC power supply, to properly display the power supply<br>information and to properly indicate a low external battery. |
| Power Management | Options include a Detachable battery, External DC battery and<br>Universal Battery pack | | |
The software design and algorithm descriptions for each of these modifications to the CoughAssist T70 device have been provided as part of this 510(k) submission.
# Performance Data
#### Non-Clinical Testing
Black-box performance testing was conducted for the CoughAssist T70 device using side by side bench testing methodologies to demonstrate that the CoughAssist T70 device performs to design input specifications and is equivalent to the predicate device. This bench testing included both open and closed loop conditions from the defined patient test cases to verify that the CoughAssist T70 software algorithms performed to specification. Bench testing conducted for the CoughAssist T70 device characterized the types of worst case scenario inputs that would be experienced in the intended use environment such as extreme flow rates. Additional white-box testing verified proper operation under conditions of sensor malfunctions and inaccurate or complete sensor dropout. This combination of testing demonstrated that the CoughAssist T70 closed loop control algorithms function safely and effectively under worst case scenarios.
{5}------------------------------------------------
# Device Modification Testing Summary
Verification activities have been performed to verify that the device modifications stated above did not affect the safety and effectiveness of the subject device. This included bench testing, software unit testing and code reviews.
The table below provides a summary of the testing that was conducted on the modifications to the Respironics CoughAssist T70 device that are the subject of this 510(k) submission:
{6}------------------------------------------------
# Standards Evaluation
This device has been tested to appropriate ISO, ASTM, and IEC standards and other applicable requirements passing all test protocols. The Respironics CoughAssist T70 device was designed and tested according to:
I. ISO 14971 Medical devices - Application of risk management to medical devices
2. ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
3. IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements of Safety
4. IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: Electromagnetic Compatibility
5. IEC 60601-1-6 Medical Electrical Equipment -- Part 1-6: Usability
6. IEC 60068 Environmental Testing (Part 2-6, 2-27, 2-64): Mechanical Vibration and Shock Resistance testing
7. IEC 62304 Medical device software - Software life cycle processes
The CoughAssist T70 device complies with the applicable requirements referenced in the FDA guidance documents:
- · FDA's Draft Reviewer Guidance for Premarket Notification Submissions -Anesthesiology and Respiratory Devices Branch; Division of Cardiovascular, Respiratory, and Neurological Devices (November 1993); and
- FDA "Guidance for the Content of Premarket Submissions for Software . Contained in Medical Devices" (May 11, 2005).
As suggested by FDA's November 1993 publication entitled "Reviewer Guidance for Premarket Notification Submissions - Anesthesiology and Respiratory Devices Branch; Division of Cardiovascular. Respiratory, and Neurological Devices" the Respironics CoughAssist T70 device was tested in accordance with the applicable voluntary standards. The Respironics CoughAssist T70 met the required performance criteria and functioned as intended.
# Conclusion
The CoughAssist T70 modifications that are the subject of this 510(k) submission have been validated using non-clinical tests and have been determined to be substantially equivalent. In summary, bench testing (black-box and white-box) and software code reviews have confirmed that the CoughAssist T70 device performs equivalently to the cited predicate devices. I The indications for use, technological characteristics, and principles of operation are similar to the predicate devices. The Respironics CoughAssist T70 device is substantially equivalent to the predicate devices and the device, as changed, does not raisc any new issues of safety and effectiveness.
{7}------------------------------------------------
| | YES | NO | N/A | |
|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|--------|---|
| tronics CoughAssist T70 (K121055)<br>Required Elements for 510(k) Summary (21 CFR 807.92) | | | | |
| Clearly labeled "510(k) Summary" | page 1 | | | |
| Submitter' s name, address, phone #, a contact person | page 1 | | | |
| Date the summary was prepared | page 1 | | | |
| The name of the device/trade name/common name/classification name | page 1 | | | |
| An identification of the legally marketed Predicate | page 2 | | | |
| Description of the subject device | page 2-3 | | | |
| Statement of intended use(identical to indications for use) | page 3 | | | |
| Technological<br>characteristics<br>if same, a summary of comparison of technological characters | page 4 | | | |
| | If different, a summary of how do they compare to the Predicate | page 4 | | |
| Performance Data | Brief discussion of non-clinical data submitted, referenced, or relied on | page 5-7 | | |
| | Brief discussion of clinical data submitted, referenced, or relied on, including:Description upon whom the device was tested,Data obtained from the tests and especially:Adverse events and complicationsOther information for SE determinationConclusion that data demonstrate SE | | | X |
| Gred Elements for 510(k) Statement (21 CFR 807.93) | | | | |
| Signed verbatim statement | | | page 7 | |
:
.
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
# December 14, 2012
Ms. Colleen Witt, RAC Senior Regulatory Affairs Engineer Respironics, Incorporated 1740 Golden Mile Highway MONROEVILLE PA 15146
Re: K121955
Trade/Device Name: Philips Respironics CoughAssist T70 Device Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: NHJ Dated: October 31, 2012 Received: November 1, 2012
Dear Ms. Witt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA havinged a determination that your device complies with other requirements of the Act that I Drims Intacted and regulations administered by other Federal agencies. You must or any I odetail the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality by every of provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 You donte speefice aav/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
2012.12.14 Susan Runner DDS, MA 1 5:34:55 -05'00'
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and
Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
Premarket Notification -- Special 510(k)
Section 6.0 Indications for Use
#### Indications for Use
510(k) Number (if known): 121 955
Device Name: Philips Respironics CoughAssist T70 device
The Philips Respironics CoughAssist T70 is intended for use on adult or pediatric patients unable to cough or clear secretions effectively. It may be used either with a facemask or mouthpiece, or with an adapter to a patient's endotracheal or tracheostomy tube. The device is intended to be used in the hospital, institutional environment or in the home.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lester W. Schultheis Jr 2012.12.14,15:41:43 -05'00'
(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dental Devices
×121955 510(k) Number:_
Tab 6 Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.