UNIPLEX NANOLINE CANNULA, PLEXOLONG NANOLINE SET, STIMULONG NANOLINE SET

K053283 · Pajunk GmbH · CAZ · Jan 12, 2006 · Anesthesiology

Device Facts

Record IDK053283
Device NameUNIPLEX NANOLINE CANNULA, PLEXOLONG NANOLINE SET, STIMULONG NANOLINE SET
ApplicantPajunk GmbH
Product CodeCAZ · Anesthesiology
Decision DateJan 12, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5140
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Pajunk's electrically insulated anesthesia conduction needles -- UniPlex-NanoLine cannula -are intended for professional anesthetists to facilitate the localization of peripheral nerve bundles for transient delivery of anesthetics during regional anesthesia. Pajunk's electrically insulated anesthesia conduction needles are also provided within PlexoLong- and StimuLong and online sets intended for professional anesthetists to facilitate the transient delivery of anesthetics during regional anesthesia. The localization of the peripheral nerve bundle is facilitated via insulated needles and impulses from a neuro-stimulator. The electric current flows from the de-insulated needles and mights innervated by the stimulated nerve bundle will the de-insulated needle the suitable setting. Following nerve bundle localization, a catheter may be placed through the insulated anesthesia conduction needle close to the news for longer pain relief. After the needle has been withdrawn, the anesthesia conduction to to to to for to for the fielde nate that been withdra anesther into the negipt call relater allows for bolus injections of local anesthetics into the peripheral nervous space.

Device Story

Device consists of electrically insulated needles (UniPlex-NanoLine), conduction catheters, and adapters; used by professional anesthetists in clinical settings. Input: electrical impulses from external neuro-stimulator. Operation: needle acts as electrode to localize peripheral nerve bundles via current flow from de-insulated tip; catheter placed through needle for subsequent bolus delivery of anesthetics. Output: localized nerve stimulation and delivery of anesthetic agent. Benefit: facilitates regional anesthesia and provides longer-term pain relief via catheter placement. NanoLine coating replaces previous lacquer coating for improved material purity.

Clinical Evidence

Bench testing only. Biocompatibility testing performed with successful results. Design changes validated using protocols identical to those used for previously cleared predicate devices.

Technological Characteristics

Electrically insulated needles and catheter sets. Features NanoLine (Parylene) coating on needles. Components include luer hubs, stylets, and catheters. Materials are identical to predicate devices. Sterilization via established methods. No software or active electronic components; device is passive, used with external neuro-stimulator.

Indications for Use

Indicated for professional anesthetists to facilitate peripheral nerve bundle localization and transient delivery of local anesthetics during regional anesthesia in patients requiring regional anesthesia or pain relief.

Regulatory Classification

Identification

An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 053283 JAN 1 2 2006 NanoLine - Premarket Notification Submission Image /page/0/Picture/3 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters. Below that, the word "MEDIZINTECHNOLOGIE" is in smaller, black letters. The background of the logo is black and white. # 510(k) Premarket Notification Submission: Summary of Safety and Effectiveness ## Date of Preparation: October 27, 2005 ## Submitter Information/ production site: Pajunk GmbH Karl-Hall-Straße 01 78187 Geisingen Germany Fon: +49(0)7704-9291-0 Fax: +49(0)7704-9291-610 #### Contact: Christian Quaß, Regulatory Affairs Fon: +49(0)7704-9291-533 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com ### Device Information: | Trade Names: | UniPlex NanoLine cannula,<br>PlexoLong-NanoLine set,<br>StimuLong-NanoLine set | | | | |---------------------------|-------------------------------------------------------------------------------------------------------------------|--|--|--| | Common Name: | Anesthesia conduction kit/ cannula | | | | | Classification Name: | Anesthesia conduction kit | | | | | Classification Reference: | 21 CFR 868.5140, April 1. 2003 | | | | | Proposed Classification: | Regulatory Class: II | | | | | Proposed Product Code: | CAZ | | | | | Predicate Devices: | 1. K000722: Pajunk's Unipolar needles (UniPlex facet & Sprotte tip) | | | | | | 2. K023218: Pajunk's Unipolar needles (UniPlex Tuohy tip) | | | | | | 3. K033018: Pajunk's StimuLong set | | | | | | 4. K013041, K023218 and K042979: Pajunk's PlexoLong set | | | | | | 5. K994059: HDC's Insul-Cote PTM needles and CLA kit<br>(nerve stimulation), includes HDC's Neuro-Trace insulated | | | | needle. ﯿﻨﭧ {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for PAJUNK MEDIZINTECHNOLOGIE. The word "PAJUNK" is in large, bold, white letters against a black background. Below that, the word "MEDIZINTECHNOLOGIE" is in smaller, thinner letters. ## Device Description: Pajunk's PlexoLong and StimuLong sets consist of a Pajunk unipolar needle, an open ended conduction catheter and a catheter adapter, as well as a filter. The coating of the needle will be chnaged to NanoLine. Pajunk's Unipolar needles marketed separately as single shot needles under the trade name UniPlex have a lacquer coating, which will be changed to NanoLine coating, Only the coating material of the needles is affected by this change. The anesthesia conduction catheters and the packaging materials are the same as those used for Pajunk's PlexoLong sets cleared for market by FDA under 510(k) numbers K013041, K023218 and K042979. The NanoLine coating is equivalent to HDC's Insul-Cote PTM needles and CLA kit (nerve stimulation) marketed under K994059. (In fact NanoLine is a better material because of a higher grade of purity. For a precise discussion of this topic see section 15 "Biocompatibility" of this submission of a special 510(k)). The contract sterilizer other than a company name change (was IBA Griffith Micro Science, and now is Sterigenics) and the stepally hame change (was film inford Sterice, and este sets. #### Intended Use: Pajunk's electrically insulated anesthesia conduction needles -- UniPlex-NanoLine cannula -are intended for professional anesthetists to facilitate the localization of peripheral nerve bundles for transient delivery of anesthetics during regional anesthesia. Pajunk's electrically insulated anesthesia conduction needles are also provided within PlexoLong- and StimuLong and online sets intended for professional anesthetists to facilitate the transient delivery of anesthetics during regional anesthesia. The localization of the peripheral nerve bundle is facilitated via insulated needles and impulses from a neuro-stimulator. The electric current flows from the de-insulated needles and mights innervated by the stimulated nerve bundle will the de-insulated needle the suitable setting. Following nerve bundle localization, a catheter may be placed through the insulated anesthesia conduction needle close to the news for longer pain relief. After the needle has been withdrawn, the anesthesia conduction to to to to for to for the fielde nate that been withdra anesther into the negipt call relater allows for bolus injections of local anesthetics into the peripheral nervous space. ## Technology Characteristics: Pajunk's anesthesia conduction needles & sets – UniPlex-NanoLine cannula, PlexoLong- & StimuLong -NanoLine sets – have the same technological characteristics as the predicate devices identified above. Pajunk's anesthesia conduction needles & sets – UniPlex-NanoLine cannula, PlexoLong-NanoLine set & StimuLong-NanoLine set – are equivalent (in fact they are the same) in design, physical dimensions, luer hub, stilett, metal and plastics materials of needles and catheter, filter and connect of there, there, materials of needles and catheter, filter and connector, as well as packaging to Pajunk's PlexoLong sets cleared for · ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------- {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the word "PAJUNK" in large, bold, white letters against a black background. The letters are slightly distorted, with some pixelation and jagged edges, giving the image a somewhat rough or textured appearance. The overall impression is that of a logo or brand name, possibly for a product or company. MEDIZINTECHNOLOGIE market by FDA under 510(k) numbers K013041, K023218 and K042979. The NanoLine coating material is substantial equivalent to HDC's Insul-Cote PTM needles and CLA hit (nerve stimulation) marketed under K994059. Biocompatibility testing of Pajunk's anesthesia conduction needles & sets – UniPlex-NanoLine cannula, PlexoLong- & StimuLong -NanoLine sets -- has been accomplished with best results. ## Summary of Design Control Activities Changes that could affect performance of the cannula or result in a risk to the patient or the user were validated using the same protocols used to validate cleared devices. Results can be found in the summary of Risk-Analysis and in the Validation Reports of Section 09. #### Conclusion: The electrically insulating coating on the cannula is only a change in material for Pajunk GmbH products. The comparison between the predicate devices and the proposed devices demonstrates that the proposed devices are safe and effective when used according to the instructions for use supplied with the devices, as well as substantially equivalent to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 #### JAN 1 2 2006 Mr. Christian Quaß Manager Regulatory Affairs Pajunk GmbH Karl-Hall Strasse 1 78187 Geisingen GERMANY Re: K053283 Trade/Device Name: Pajunk Anesthesia Conduction NanoLine Coated Needles Regulation Number: 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ, BSP Dated: December 21, 2005 Received: December 27, 2005 Dear Mr. Quaß: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Mr. Quaß Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Ours Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Special Premarket Notification Submission ## NanoLine Image /page/5/Picture/2 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in a smaller, sans-serif font. ## Section 04 #### Indications for use 510(k) Number: K053283 Device Name: Pajunk anesthesia conduction NanoLine coated needles Indications for Use: The change of coating material from lacqueur to Parylene does not affect the Indications for use of the predicated devices marketed under K042979 (Pajunk Plexolong Sets) K033018 (Pajunk Stimulong Plus Catheter Sets) K023218 (Pajunk Plexolong Sets) K013041 (Pajunk Plexolong Anesthesia Sets) K999722 (Unipolar Needles with Standard and Sprotte-tip) The change in coating has no effect on the Indications for use of these devices already cleared for market. Prescription Use AND/OR (Per 21 CFR 801.109) Over-The-Counter Use (21 CFR 801 Subpan C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Clues K.053283 Page 1 of 1 | PMN | | |------------|--| | CQ | | | 2006-01-05 | | | P 1 of 1 | |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...