K092484 · Cardinal Health 207, Inc. · BZR · Apr 2, 2010 · Anesthesiology
Device Facts
Record ID
K092484
Device Name
HELIOX LOW FLOW SENTRY BLENDER
Applicant
Cardinal Health 207, Inc.
Product Code
BZR · Anesthesiology
Decision Date
Apr 2, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5330
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Heliox Low Flow Sentry Blender is designed to provide a continuous heliox/oxygen gas mixture to infant, pediatric, and adult patients. It is intended as a device to mix an 79% Helium / 21% Oxygen gas source with a 100% oxygen gas source to provide low flow mixtures containing 20% to 100% oxygen with the balance of the gas being Helium for patients who are spontaneously breathing. It is a medical device intended for use by qualified, trained personnel, under the direction of a physician, in institutional environments where the delivery and monitoring of helium/oxygen mixture is required.
Device Story
Device mixes 79% Helium/21% Oxygen source with 100% Oxygen source to deliver heliox/oxygen mixtures; includes integral battery-operated oxygen concentration analyzer with user-adjustable alarms. Used in institutional environments by qualified, trained personnel under physician direction. Input: gas sources; Output: heliox/oxygen mixture and O2 concentration monitoring. Provider uses output to adjust gas delivery; benefits patient by providing controlled heliox therapy. Modified from Sentry air/oxygen microblender (K911962) with updated poppet orifice to account for helium density and modified dial knob labeling.
Clinical Evidence
Bench testing only. Performance verification included dimensional envelope, weight, interface, supply/operating pressure, O2 control, flow characteristics, blender safety features, and monitor/analyzer performance (display, controls, alarms, system accuracy). All tests passed.
Technological Characteristics
Gas mixing valve with integral O2 analyzer. Modified poppet orifice for helium density. Battery-operated monitor with digital display, user-adjustable high/low O2 alarms, calibration, and lock/unlock controls. Standalone device for institutional use.
Indications for Use
Indicated for infant, pediatric, and adult patients who are spontaneously breathing and require delivery and monitoring of heliox/oxygen gas mixtures (20-100% O2, balance Helium) in institutional settings.
Regulatory Classification
Identification
A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.
K090781 — HELIO2 BLENDER · Precision Medical, Inc. · Aug 13, 2009
K161718 — MaxBlend 2, MaxBlend Lite · Maxtec, LLC · Oct 27, 2016
K111640 — PRECISION FLOW(R) - HELIOX · Vapotherm, Inc. · Oct 7, 2011
K142291 — GMX Series Medical Air/Oxygen Blender · Gentec (Shanghai) Corporation · May 19, 2015
K990293 — ALADDIN II MONITORING OXYGEN/AIR MONITOR · Hamilton Medical, Inc. · Mar 16, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
K092484
Image /page/0/Picture/1 description: The image contains the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The text "CardinalHealth" is written in a simple, sans-serif font.
APR - 2 2010
# 510(K) Summary
# 510 (K) Summary as required by 21 CFR 807.92
| 510 (K) Submitter: | Cardinal Health 207, Inc.<br>Yorba Linda, CA 92887<br>(714) 283 - 2228 |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Monther Abushaban<br>(714) 919 - 3660<br>Monther.Abushaban@CardinalHealth.com |
| Establishment Registration Number: | 2050001 |
| Date Prepared: | April 1, 2010 |
| Name of the Device: | Heliox Low Flow Sentry Blender |
| Common/ usual name: | Oxygen Blender for Helium/Oxygen Mixtures |
| Classification: | The Heliox Low Flow Sentry Blender is classified as<br>a class II device under the following classification<br>code |
| Product code | CFR Section | Panel |
|--------------|-----------------|----------------|
| 73 BZR | 21 CFR 868.5330 | Anesthesiology |
Predicate Device: The predicate devices are:
Sentry air/oxygen microblender (K911962) Low Flow air/oxygen blender (K883038) Helio2 Blender (K090781)
Device Description: The helium/oxygen and oxygen blender/monitor combines a precision helium/oxygen (79%/21%) and oxygen mixing valve with an integral battery operated oxygen concentration analyzer and user adjustable alarms.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The graphic is made up of three curved lines that converge at a point, resembling the wings of a bird in flight. The text "CardinalHealth" is written in a simple, sans-serif font.
### Intended Use:
The Heliox Low Flow Sentry Blender is designed to provide a continuous heliox/oxygen gas mixture to infant, pediatric, and adult patients. It is intended as a device to mix an 79% Helium / 21% Oxygen gas source with a 100% oxygen gas source to provide low flow mixtures containing 20% to 100% oxygen with the balance of the gas being Helium for patients who are spontaneously breathing. It is a medical device intended for use by qualified, trained personnel, under the direction of a physician, in institutional environments where the delivery and monitoring of helium/oxygen mixture is required.
### Modifications to the predicate device:
The Heliox Low Flow Sentry Blender is the same device as the Sentry air/oxygen microblender (K911962). Modifications to Heliox Low Flow Sentry Blender are associated with this submittal are as follows:
- Intended use: Changed the gas mixture from air/oxygen to He- 02/oxygen . mixture.
- . The orifice of the poppet has been changed to address the differences in density of the helium/oxygen mixture from the density of air.
- . The dial knob label artwork has been changed from analog numbers to a amplitude range.
- In summary, the Heliox Low Flow Sentry Blender described in this submission is, in our opinion, substantially equivalent to the predicate device(s).
#### Reason for the submission:
This is a modified device with a new intended use to be marketed by Cardinal Health 207, Inc.
#### Summary of Technological Characteristics of Device Compared to the Predicate Device:
The Heliox Low Flow Sentry Blender is substantially equivalent in safety and effectiveness to the legally marketed (predicate) currently in distribution.
## Summary of Non-clinical Testing for the Device and Conclusions:
Performance testing verified that the Heliox Low Flow Sentry Blender meets it's performance requirements that this device is substantially equivalent to medical devices currently legally marketed in the United States.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized graphic above the text "CardinalHealth". The graphic is composed of three curved lines that converge at a point, resembling a bird in flight or abstract wings.
| 91727 Rev A | | ER 3348 | | Pass or Fail |
|--------------------------|--------------------------------------|------------------------------------------------|--------------------------------|--------------|
| PRS Reference<br>Numbers | Description | Verification Test Protocol<br>Reference Number | Number of<br>Units for<br>Test | |
| 4.1.1 | Dimensional Envelope | ER3348 Section 4.1.1 | 4 | Pass |
| 4.1.2 | Weight | ER3348 Section 4.1.1 | 4 | Pass |
| 4.1.3 | Interface (gas inlets and outlets) | ER3348 Section 4.1.1 | 4 | Pass |
| 4.2.1 | Nominal Supply Pressure | ER3348 Section 4.1.1 | 4 | Pass |
| 4.2.2 | Normal Operating Pressure | ER3348 Section 4.1.1 | 4 | Pass |
| 4.3 | Environmental Withstand | Refer to Sentry Blender<br>DHF | N/A | Pass |
| 4.4.1 | % O2 Control | ER3348 Sect. 5.2 | 4 | Pass |
| 4.4.2 | Flow Characteristics | ER3348 Sect. 5.2 | 4 | Pass |
| 4.4.4 | Blender Safety Features | ER3348 Sect. 5.1, 5.3 | 4 | Pass |
| 4.5.1 | Monitor Display | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 a | Oxygen Range, Resolution, & Accuracy | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 b | Low O2 Alarm Set point | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 c | High O2 Alarm Set point | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 d | "%" Flashing during calibration | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 e | Locked Indicator | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 f | Attach Sensor | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 g | Low Battery Indicator | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2 | Monitor Controls | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2 a | On/Off Switch | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2 b | Lock/Unlock Switch | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2.1 | Calibrate Button | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2.2 | High Set Button | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2.3 | Low Set Button | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2.4 | Alarm Silence Button | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.3 | Alarm/Alert Indicatons | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.3 a | Hi O2 Concentration | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.3 b | Low O2 Concentration | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.3 c | Alarm Silence | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.3 d | Alarm Loudness | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.4 | Monitor Power Source | Specified by Manufacturer<br>(not tested) | N/A | Pass |
| 4.5.5 | Oxygen Sensor | Specified by Manufacturer<br>(not tested) | N/A | Pass |
| 5.1 | Oxygen Blender Performance | ER3348 Sect. 5.2 | 4 | Pass |
| 5.2 | Monitor/Analyzer Performance | ER3348 Sect. 5.3 | 4 | Pass |
| 5.2.1 | Display, Controls, Alarms | ER3348 Sect. 5.3 | 4 | Pass |
| 5.2.2 | System Accuracy | ER3348 Sect. 5.3 | 4 | Pass |
| 6.1 | Labeling | ER3348 Section 4.1.1 | 4 | Pass |
| 6.2 | Shipping Container | ER3348 Section 4.1.1 | 4 | Pass |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Cardinal Health. The logo consists of a stylized graphic above the company name. The graphic appears to be a series of curved lines, possibly representing a bird in flight or a stylized medical symbol. The text "CardinalHealth" is written in a simple, sans-serif font.
# Conclusion:
The summary above shows that there are no new questions of safety and effectiveness of the subject device (Heliox Low Flow Sentry Blender) as compared to predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The graphic is made up of four curved lines that resemble the wings of a bird in flight. The text "CardinalHealth" is written in a simple, sans-serif font.
.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with three stripes forming its body and wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Monther Abushaban Regulatory Affairs Manager Cardinal Health 207, Incorporated 22745 Savi Ranch Parkway Yorba Linda, California 92887
APR - 2 2010
Re: K092484
Trade/Device Name: Heliox Low Flow Sentry Blender Regulation Number: 21CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: II Product Code: BZR Dated: March 3, 2010 Received: March 31, 2010
Dear Mr. Abushaban:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Mr. Abushaban
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Th for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The text "CardinalHealth" is written in a simple, sans-serif font, with the first letter capitalized.
## Indications for Use
Device Name: Heliox Low Flow Sentry Blender
Indications for Use:
The Heliox Low Flow Sentry Blender is designed to provide a continuous heliox/oxygen qas mixture to infant, pediatric, and adult patients. It is intended as a device to mix an 79% Helium / 21% Oxygen gas source with a 100% oxygen gas source to provide low flow mixtures containing 20% to 100% oxygen with the balance of the gas being Helium for patients who are spontaneously breathing. It is a medical device intended for use by qualified, trained personnel, under the direction of a physician, in institutional environments where the delivery and monitoring of helium/oxygen mixture is required.
X AND/ OR Prescription Use (Part 21CFR 801 subpart D)
Over the Counter Use: (Part 21CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Division Sign-Off) ivision of Anesthesiology, General Hospital fection Control, Dental Devices
10(k) Number: K092484
Traditional 510(k) Submission for Heliox Low Flow Sentry Blender
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.