HEATED WIRE CIRCUITS FOR SCT 3000 HEATED HUMIDIFIER
K962220 · Marquest Medical Products, Inc. · BZE · Jan 14, 1997 · Anesthesiology
Device Facts
Record ID
K962220
Device Name
HEATED WIRE CIRCUITS FOR SCT 3000 HEATED HUMIDIFIER
Applicant
Marquest Medical Products, Inc.
Product Code
BZE · Anesthesiology
Decision Date
Jan 14, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5270
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Heated wire circuits to be used with the SCT 3000 are indicated as breathing circuits for patients under going anesthesia or mechanical ventilation.
Device Story
Heated wire breathing circuits and limbs designed for use with SCT 3000 Heated Humidifier; various configurations for anesthesia and volume ventilation; includes breathing tubes, patient wye connectors, and breathing bags; single or dual heated wire options; available for ventilators with or without internal exhalation valves; may be packaged with accessories like filters, face masks, and gas sampling lines. Used in hospital settings (anesthesia, ICU, respiratory therapy) by clinicians. Provides heated, humidified gas to patients to maintain airway moisture and temperature during mechanical ventilation or anesthesia. Single-use device.
Clinical Evidence
Bench testing only. Device meets ASTM breathing systems standards.
Technological Characteristics
Heated wire breathing circuits; single or dual wire configurations; various lengths and styles for anesthesia and volume ventilators. Materials identical to standard breathing circuits (exempt from PMN). Complies with ASTM breathing systems standards. Single-use.
Indications for Use
Indicated for patients, including adults, pediatrics, and neonates, undergoing anesthesia or mechanical ventilation requiring use of the SCT 3000 heated humidifier.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
K253322 — Hudson RCI® Infant Dri-TechTM Breathing Circuits (HUD99060 / Dri-Tech with 78 Infant Circuit, HUD99060KIT / Dri-Tech with 78 Infant Circuit Kit, HUD99001KIT / Single Limb Infant Circuit Dri-Tech Kit, HUD99009KIT / Dri-Tech with 60 Infant Circuit Kit, HUD99018KIT / Dri-Tech with 44 Infant Circuit Kit, HUD870NKIT / Dri-Tech Infant Accessory Bag, HUD1631 / Infant Wye Connector with Swivel) · Medline Industries, LP · Mar 25, 2026
Submission Summary (Full Text)
{0}
JAN 14 1997
The New Marq of Excellence
K962220
# Non-Confidential Summary of Safety and Effectiveness
June 3, 1996
page 1 of 3
| Marquest Medical Products, Inc.
11039 East Lansing Circle
Englewood, CO 80112 | Tel - (303) 790-4835
Fax - (303) 799-0210 |
| --- | --- |
**Official contact:** Tom Dielmann, Director, Regulatory Affairs / Quality Assurance
**Proprietary or Trade Name:** Heated wire breathing circuits and limbs
**Common/Usual Name:** Heated wire breathing circuits and limbs
**Classification Name:** 73 BZE - Breathing systems heater w/wo controller<br/>(not humidifier or nebulizer)<br/>Subsection 868.5270
**Intended device:** Breathing circuits for use with SCT 3000 Heated Humidifier
**Predicate devices:**
- Marquest SCT 3000 - K903138
- Marquest (Intec) - Heated limb - K860941
- Intertech (SIMS) - K881231
**Device description:** Various configurations of anesthesia, volume ventilator circuits and heated limbs for use with SCT 3000 Heated Humidifier.
**Indicated use:**
Heated wire circuits to be used with the SCT 3000 are indicated as breathing circuits for patients under going anesthesia or mechanical ventilation.
**Targeted population:**
Any patient utilizing the SCT 3000 heated humidifier. Adult, pediatric and neonates.
**Environment of use:** Hospital, anesthesia, ICU, respiratory therapy
page 4 of 37
{1}
Non-Confidential Summary of Safety and Effectiveness (continued)
June 3, 1996
page 2 of 3
Comparison to predicate devices:
| Attribute | Marquest Heated Wire Circuits | Intertech / SIMS Heated Wire Circuits |
| --- | --- | --- |
| Use | | |
| Indicated for anesthesia | Yes | Yes |
| Indicated for respiratory therapy | Yes | Yes |
| Population - adult, pediatric, neonate | Yes | Yes |
| Indicated for use with heated humidifiers | Yes | Yes |
| Indicated for use with only specific manufacturers heated humidifier | Yes | Yes |
| Indicated for single use | Yes | Yes |
| Design | | |
| Various patient wye connectors | Yes | Yes |
| Breathing Tubes various lengths | Yes | Yes |
| Breathing Bags - various sizes | Yes | Yes |
| Single or dual heated wires | Yes | Yes |
| Anesthesia - rebreathing and non-rebreathing styles | Yes | Yes |
| Volume ventilator - different styles for ventilator manufacturers | Yes | Yes |
| Offered for ventilators without internal exhalation valves | Yes | Yes |
| Packaged with other devices - Filters, Face Mask, Gas sampling lines, etc. | Yes | Yes |
page 5 of 37
{2}
page 6 of 37
# Non-Confidential Summary of Safety and Effectiveness (continued)
June 3, 1996
page 3 of 3
## Comparison to predicate devices: (continued)
| Attribute | Marquest Heated Wire Circuits | Intertech / SIMS Heated Wire Circuits |
| --- | --- | --- |
| **Materials**
Utilize same materials as standard breathing circuits
exempt from PMN | Yes | Yes |
| **Performance Testing**
Meets ASTM breathing systems standard | Yes | Yes |
## Differences
Any differences that do exist would not have a significant effect on the safety or effectiveness of the device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.