CENTROX, RELIANT, ULTROX, AS-D+, AS-E, AS-G, AS-J, AS-K AND AS-L
Applicant
Airsep Corp.
Product Code
CAW · Anesthesiology
Decision Date
Mar 28, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the device is to generate and deliver USP 93% oxygen for supplemental oxygen use only or filling cylinders. The device must not be used for or with any lifesupporting applications. The device is intended to be used only by trained personnel in hospitals, health care facilities or remote locations where bottled oxygen is not readily available. The instruction manual of the device recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition, or mechanical failure,
Device Story
Centrox, Ultrox, and Reliant oxygen concentrators utilize Pressure Swing Adsorption (PSA) to generate USP 93% oxygen from ambient air. System components include an air compressor, particulate/foam filters, coalescing filters, and adsorber vessels containing inert ceramic molecular sieve. Operation involves compressing ambient air, filtering contaminants, and passing air through molecular sieve to adsorb nitrogen under pressure, allowing oxygen to exit as product gas. While one adsorber produces oxygen, the other depressurizes to exhaust waste gases. Integrated oxygen analyzers monitor purity and trigger alarms if levels fall outside safe ranges. Devices are operated by trained personnel in clinical or remote settings. Output is used for supplemental oxygen or cylinder filling. The system is not for life support; an alternate oxygen source is required for power outages or mechanical failures.
Clinical Evidence
Bench testing only. Verification and validation included internal and external testing. Performance testing confirmed oxygen flow rate and purity (USP 93%) at specified operating pressures. Testing evaluated the impact of overdrawing, component failure, and valve cycling on device parameters. Independent laboratory testing verified oxygen purity and confirmed total volatile organic compounds (VOCs) and particulates were below accepted standards. Electrical safety and EMC compliance were confirmed.
Technological Characteristics
Pressure Swing Adsorption (PSA) oxygen generator. Materials include inert ceramic molecular sieve. Features integrated oxygen analyzer with alarm. Flow rates: 17-32 scfh; delivery pressure: 20-50 psig. Powered by air compressor. Standalone operation. Validated for USP 93% purity. No software algorithm class specified.
Indications for Use
Indicated for supplemental oxygen therapy or cylinder filling in patients requiring oxygen. Intended for use by trained personnel in hospitals, healthcare facilities, or remote locations. Contraindicated for life-supporting applications.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
Portable Oxygen Generator with Medical Air (POGS 33C) (K063454)
Submission Summary (Full Text)
{0}------------------------------------------------
K101154
MAR 2 8 2011
## 510(k) Summary
#### Submitter
Airsep Corporation 401 Creekside Drive Buffalo, NY 14228 Phone: (716) 691-0202
(716) 691-4141 Fax:
Registration Number: 1319044
## Contact person
Peter Weisenborn
Email: pweisenborn@airsep.com
#### Preparation Date
March 25, 2011
#### Device
#### Trade Name:
Centrox, Ultrox and Reliant Oxygen Concentrators Classification Name: Generator, Oxygen. Regulation Number: 868.5440 Product Code: CAW Device Class: Class II Classification Panel: Anesthesiology and Respiratory Therapy Devices
#### Predicate Devices
NewLife Intensity from AirSep Corporation Product code: CA W 510(k) number: K960309
Portable Oxygen Generator with Medical Air (POGS 33C) from On Site Gas Systems, Inc. Product code: CAW 510(k) number: K063454
{1}------------------------------------------------
## Device Description
Air contains 21 % oxygen, 78% nitrogen, 0.9% argon, and 0.1% other gases. AirSep PSA Oxygen units separate this small percentage of oxygen from the compressed air through a unique Pressure Swing Adsorption (PSA) process. The compressed air flows through a filter assembly before the air enters the adsorber vessels. A particulate/foam filter removes condensed water, oil, dirt, scale, etc. from the feed air, and then, a separate coalescing filter (if present) removes additional oil and water vapor.
The oxygen generator uses in its adsorber vessels an inert ceramic material called molecular sieve to separate compressed air into oxygen and other gases. The unique properties of molecular sieve allow it to attract, or adsorb, nitrogen physically from air under pressure. This allows oxygen to exit the adsorbers as a product gas.
While one adsorber produces oxygen, the other depressurizes to exhaust the waste gases it adsorbed (collected) during the oxygen production cycle. The entire oxygen generating process is completely regenerative, which makes it both reliable and virtually maintenancefree. The molecular sieve does not normally require replacement.
The medical devices in this submission refer to the following oxygen concentrators manufactured by AirSep Corporation:
| Model | Flow (scfh) | Oxygen Delivery<br>Pressure (psig) |
|---------|-------------|------------------------------------|
| Centrox | 32 | 50 |
| Ultrox | 17 | 20 |
| Reliant | 17 | 50 |
All of these devices are designed to deliver the oxygen with USP 93 % purity. Centrox and Reliant units are also available for cylinder filling option. Ultrox is used only to fill cylinders.
Centrox, Ultrox and Reliant units are comprised of an air compressor that uses ambient air to produce higher pressure feed air.
These models have the same indications for use. Together these products are referred to as Device.
The device is intended to be USP 93% oxygen generator. As such, to assure the purity, an Oxygen Analyzer is integrated in each device and an alarm is activated if the purity is outside of the safe range.
{2}------------------------------------------------
#### Intended Use
The intended use of the device is to generate and deliver USP 93% oxygen for supplemental oxygen use only or filling cylinders. The device must not be used for or with any lifesupporting applications. The device is intended to be used only by trained personnel in hospitals, health care facilities or remote locations where bottled oxygen is not readily available. The instruction manual of the device recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition, or mechanical failure,
#### Technological Characteristics
The comparison table is provided as a summary of the technological characteristics relative to the predicate devices. Items like intended use, principle of operation, oxygen purity, inlet pressure and flow rates are compared. All major components used in the system are also compared. The primary difference between the device and the respective predicate devices is in the flow rates which is a result of the size of the units. The performance of the device to produce the desired flowrates at specified purity and delivery pressure has been validated. The summary of the comparison table demonstrates that the device has no significant differences from the predicate devices that would adversely affect product safety and effectiveness.
## Testing
The device has been tested and verified in various phases, internal testing, verification and validation as well as external testing and validation. The design was verified throughout the design process. Risk analysis was done, appropriate measures were implemented and their effectiveness verified. External test house was used to confirm compliance to EMC requirements and standards for electrical safety.
#### Performance Data
The oxygen flow rate and purity at the operating pressure as specified in the instruction manual was tested and verified. The variation of oxygen purity and output pressure with the amount of overdrawing, the variation in parameters due to failure of components, the effect of improper valve cycling on the device parameters were tested and the ability of every device to deliver USP 93% oxygen was verified.
Independent laboratory testing also verified that the oxygen purity was in accordance with USP 93% and that the total Volatile organic compounds (VOC's) and particulates were well below accepted standards.
The results of this testing prove the safety and effectiveness of the device.
{3}------------------------------------------------
## Conclusion
The device performance tests demonstrate substantial equivalence to the predicate devices. The performance tests results also confirm ability to provide USP 93% oxygen for supplemental oxygen use only and for filling cylinders safely and effectively.
.
.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Peter Weisenborn Vice President-Resources and Regulatory Affairs Airsep Corporation 401 Creekside Drive Buffalo, New York 14228-2085
MAR 2 8 2011
Re: K101154
Trade/Device Name: Centrox, Ultrox and Reliant Oxygen Concentrators ("Device") Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: March 21, 2011 Received: March 24, 2011
Dear Mr. Weisenborn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Weisenborn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-firee number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Applicant: Airsep Corporation
## 510(k) Number (if known): K101154
Device Name: Centrox, Ultrox and Reliant Oxygen Concentrators ( "Device")
# Indications For Use:
The intended use of the Device is to generate and deliver USP 93% oxygen for supplemental oxygen use only or filling cylinders. The device must not be used for or with any life-supporting applications.
The Device is intended to be used only by trained personnel in hospitals, health care facilities or remote locations where bottled oxygen is not readily available.
The instruction manual of the device recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition, or mechanical failure.
Prescription Use X AND/OR 801 Subpart D) (21 CFR 801 Subpart C)
Over-The-Counter Use _ _ _ _ (Part 21 CFR
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
h
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
(Optional Format 3-10-98)
rev A
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.