K083163 · Sequal Technologies, Inc. · CAW · Oct 20, 2009 · Anesthesiology
Device Facts
Record ID
K083163
Device Name
OMNI 2 OXYGEN SYSTEM
Applicant
Sequal Technologies, Inc.
Product Code
CAW · Anesthesiology
Decision Date
Oct 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OMNI 2 Oxygen System is intended for the administration of supplemental oxygen. The device is not intended for life support nor does it provide any patient monitoring capabilities. The OMNI 2 Oxygen System is prescription legend required.
Device Story
OMNI 2 Oxygen System is a portable oxygen concentrator using pressure swing adsorption (PSA) with synthetic zeolite molecular sieve beds. Inputs: ambient air. Outputs: supplemental oxygen via continuous flow (up to 3 LPM) or pulse flow (bolus up to 96 mL). Operation: AC, DC, or lithium-ion battery power; includes air compressors, inlet/outlet filtration, and rotary valves. Pulse mode triggers oxygen delivery upon inhalation detection. Used in clinical or home settings by patients; operated by patient or caregiver. Provides supplemental oxygen to improve patient oxygenation; does not provide life support or patient monitoring. Includes audible/visual alarms for system status.
Clinical Evidence
Bench testing only. Performance data includes system performance, environmental conditions, electromagnetic compatibility (EMC), and software validation. Compliance with ISO 8359 (Oxygen Concentrator devices) and IEC 60601-1 (Medical Electrical Equipment) confirmed via independent analysis.
Technological Characteristics
Portable oxygen concentrator; PSA technology using synthetic zeolite molecular sieve beds. Components: air compressors, rotary valves, inlet/outlet filtration, electronic flow control, AC/DC power supplies, lithium-ion batteries. Connectivity: none. Sterilization: N/A. Software: electronic flow control and alarm management.
Indications for Use
Indicated for patients requiring supplemental oxygen administration. Not for life support. Prescription use only.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
Predicate Devices
OMNI Oxygen System, Model 1000 (aka: Eclipse Oxygen System) (K013931)
Related Devices
K120785 — OMNI 3 OXYGEN SYSTEM · Caire, Inc. · Oct 25, 2012
K162433 — GCE Zen-O Portable Oxygen Concentrator · Oxus, Inc. · May 24, 2017
K042262 — INTEGRA OXYGEN CONCENTRATOR, MODEL 6323A-OM-10 · Sequal Technologies, Inc. · Sep 10, 2004
K013931 — OMNI OXYGEN SYSTEM, MODEL 1000 · Sequal Technologies, Inc. · Mar 27, 2002
K223379 — Portable Oxygen Concentrator (Model: P2-E7, P2-E) · Qingdao Kingon Medical Science and Technology Co., Ltd. · Jun 22, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
K083/63
# 510(k) Summary
OCT 20 2009
## Submitter Information:
| Submitter: | SeQual Technologies, Inc.<br>11436 Sorrento Valley Road<br>San Diego, CA 92121 |
|-------------------|--------------------------------------------------------------------------------|
| Contact: | Brian Jarrell, Director of Quality and Regulatory |
| Phone: | (858) 202-3157 |
| FAX: | (858) 558-1915 |
| Date of Summary: | June 11, 2009 |
| Device Name: | |
| Proprietary Name: | OMNI 2 Oxygen System |
| Proprietary Name: | OMNI 2 Oxygen System |
|---------------------------|---------------------------------------------------|
| Common Name: | Oxygen Concentrator |
| Classification of Device: | Generator, Oxygen, Portable as per 21 CFR 868.544 |
| Product Code: | CAW |
#### Predicate Device Equivalence:
SeQual Technologies is claiming substantial equivalence to the following Iegally marketed predicate device:
K013931 - OMNI Oxygen System, Model 1000 (aka: Eclipse Oxygen System)
### Description of Device:
The OMNI 2 Oxygen System is an oxygen concentrator that provides up to 3 LPM continuous flow or in pulse mode an oxygen bolus of up to 96 mL (milli-Liter). The OMNI 2 Oxygen System is based on pressure swing adsorption (PSA) principles. The OMNI 2 Oxygen System operates from AC power, DC power, or rechargeable batteries. This device delivers supplemental oxygen for patients through the molecular sieve beds and is designed to conserve the use of oxygen while operating in pulse flow mode. During pulse flow mode, oxygen is delivered to the patient through a pulse flow valve when the start if inhalation is detected.
{1}------------------------------------------------
The OMNI 2 Oxygen System consists of pneumatic and electrical components, AC power supplies, DC cables and lithium ion batteries. The system has inlet filtration, air compressors, and Synthetic Zeolite molecular sieve beds with a rotary valves, outlet filtration, electronic flow control and audible / visual alarms.
## Intended Use:
The OMNI 2 Oxygen System is intended for the administration of supplemental oxygen. The device is not intended for life support nor does it provide any patient monitoring capabilities. The OMNI 2 Oxygen System is prescription legend required.
### Technological Characteristics:
The OMNI 2 Eclipse Oxygen System operates comparably to the listed predicate device. The PSA technology employed to generate the oxygen is well established, and therefore, raise no new questions of safety and effectiveness.
## Performance Data:
The OMNI 2 Oxygen System has a comprehensive test platform that covers system performance, environmental conditions, electromagnetic compatibility (EMC), and software validation. The test platform confirms compliance with ISO 8359 standard for Oxygen Concentrator devices.
Outside agencies provide independent analysis related to compliance with EMC and IEC 60601-1 (Medical Electrical Equipment-Part 1: General Requirements).
The test platform ensures compliance to recognized consensus standards and therefore raises no new questions of safety and effectiveness.
## Conclusion:
Based on intended use, technology characteristics, design and performance data the OMNI 2 Oxygen System is substantially equivalent to the currently marketed devices.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of a stylized eagle or bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Mr. Brian Jarrell Director of Quality and Regulatory Affairs Sequal Technologies, Incorporated 11436 Sorrento Valley Road San Diego, California 92121-1306
OCT 20 2009
Re: K083163
Trade/Device Name: OMNI 2 Oxygen System Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: June 11, 2009 Received: September 18, 2009
Dear Mr. Jarrell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2- Mr. Jarrell
Enclosure
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
for
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo or wordmark that reads "SEQUEL" in a stylized, bold font. The letters are thick and have a slightly distressed or textured appearance, giving them a rough or vintage feel. To the left of the word "SEQUEL", there is a graphic element that appears to be a stack of pancakes or a similar layered food item, also rendered in a bold, textured style.
SeQual Technologies Inc. 11436 Sorrento Valley Road, San Diego CA 92121 USA
## Indications for Use Statement
Verl 3 - 4/24/96
Applicant: SeQual Technologies Inc.
510(k) Number (if known):
Device Name: OMNI 2 Oxygen System
Indications For Use:
The OMNI 2 Oxygen System is intended for the administration of supplemental oxygen. The device is not intended for life support nor does it provide any patient monitoring capabilities.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Per 21 CFR 801.109) (Optional Format 1-2-96)
Int510k/induse
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
ੱ
510(k) Number: KD83163
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.