K162433 · Oxus, Inc. · CAW · May 24, 2017 · Anesthesiology
Device Facts
Record ID
K162433
Device Name
GCE Zen-O Portable Oxygen Concentrator
Applicant
Oxus, Inc.
Product Code
CAW · Anesthesiology
Decision Date
May 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Portable Oxygen Concentrator is intended to provide supplemental oxygen in a home, institutional, or travel environment.
Device Story
Portable oxygen concentrator (POC) provides supplemental oxygen via nasal cannula; utilizes Pressure Swing Adsorption (PSA) with molecular sieve to extract oxygen from ambient air. Operates in pulse mode (1-6 LPM) or continuous flow mode (0.5-2 LPM). Input: ambient air; output: 87%-96% concentrated oxygen. Used in home, institutional, or travel settings by patients; prescription-only. Device includes rechargeable battery, AC/DC power supply, and automotive adapter. Healthcare providers use device to manage oxygen therapy for chronic respiratory conditions. Benefits include patient mobility and consistent oxygen delivery. System features LCD display, buttons for flow control, and comprehensive alarm suite (e.g., low pressure, no breath detected, battery status).
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993-1), electrical safety/EMC (AAMI ANSI ES60601-1, IEC 60601-1 series), and functional performance testing against ISO 80601-2-67 and ISO 80601-2-69. Usability testing conducted with 4 users (lay and healthcare providers).
Technological Characteristics
Pressure Swing Adsorption (PSA) with molecular sieve. Flow rates: 1-6 LPM (pulse), 0.5-2 LPM (continuous). Power: 100-240VAC or DC automotive adapter. Battery-powered. Standards: AAMI ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, ISO 80601-2-67, ISO 80601-2-69. Embedded software. Type BF applied part.
Indications for Use
Indicated for adult patients with chronic pulmonary diseases (e.g., chronic bronchitis, emphysema, asthma, lung cancer), terminal cancer, or any patient requiring supplemental oxygen. Not for life-supporting or life-sustaining use.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2017
Oxus, Incorporated Ryan Lenarcic Quality & Regulatory Manager 2676 Paldan Auburn Hills, Michigan 48326
Re: K162433
Trade/Device Name: GCE Zen-O Portable Oxygen Concentrator Model RS-00500 Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: April 19, 2017 Received: April 24, 2017
Dear Ryan Lenarcic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Lori A. Wiggins -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162433
#### Device Name
GCE Zen-O™ Portable Oxygen Concentrator Model RS-00500
Indications for Use (Describe)
The Portable Oxygen Concentrator is intended to provide supplemental oxygen in a home, institutional, or travel environment.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Oxus" in green and blue letters. The letters are arranged in a slightly curved line, with the "O" in green and the "xus" in blue. Behind the word is a light blue circle. Below the word is the text "Premarket Notification 510(k)".
Section 5 - 510(k) Summary
#### 510(k) Summary (As required by section 807.92(c)) Section V
## l. Submitter
Oxus Inc. 2676 Paldan Dr. Auburn Hills, MI 48326
| Phone: | 248-475-0925 |
|-----------------------|-----------------------------------------------------------|
| Fax: | 248-475-0938 |
| Contact Person: | Ryan Lenarcic, Quality & Regulatory Manager |
| Date Prepared: | February 22, 2017 |
| II. Device | |
| Name of Device: | GCE Zen-O™ Portable Oxygen Concentrator<br>Model RS-00500 |
| Common or Usual Name: | Portable Oxygen Concentrator |
| Classification Name: | Portable oxygen generator (21 CFR 868.5440) |
| Regulatory Class: | II |
| Product Code: | CAW |
## III. Predicate Device:
Oxus, RS-00400 Portable Oxygen Concentrator, K073242 This Predicate has not been subject to a design related recall
No reference devices were used in this submission
# IV.Device Description
Oxus, Inc. as indicated above proposes to offer the following device which we will refer to as the GCE Zen-OTM Portable Oxygen Concentrator, Model RS-00500.
The GCE Zen-O™ (Model RS-00500) POC is the latest design of POC from Oxus, Inc. The previous model, Oxus Model RS-00400, is FDA cleared under 510(k) K073242. During the
Portable Oxygen Concentrator
{4}------------------------------------------------
launch of the RS-00500 model in 2015, submission of a new 510(k) was not required per FDA Guidance document "Deciding When to Submit a 510(k) for a Change to an Existing Device" and a subsequent Letter to File was generated. Due to recent business partnerships that have led to the sale of the RS-00500 design from Oxus to GCE, Oxus is submitting this 510(k) to procure a stand-alone 510(k) clearance for the GCE Zen-O POC™ (Model RS-00500) separate from the Oxus RS-00400 model. Information provided within this submission references both model numbers due to the sharing of technology and previous testing; however this submission is solely for the clearance of the GCE Zen-O™ POC, Model RS-00500 The GCE Zen-O™ POC delivers 87% - 96% oxygen to a patient through a standard singlelumen nasal cannula. The GCE Zen-O™ POC detects a patient breath and operates by delivering a bolus of oxygen during the inhalation period (pulse mode) or can provide a continuous flow of oxygen (continuous mode).
The device is portable, enabling patients who need an oxygen device to be treated at home according to a clinician's prescription or direction.
The GCE Zen-O™ POC can be set to deliver flowrates between 1-6 LPM in pulse mode and 0.5-2 LPM in continuous mode; both with increments of 0.5 LPM to a patient. The device can be operated when powered by the 100-240VAC AC/DC power supply provided with the unit and designed to operate with a DC automotive adaptor. In addition, the device contains a rechargeable battery allowing it to be carried by a patient while traveling
# V. Indications for Use
The Portable Oxygen Concentrator is intended to provide supplemental oxygen in a home, institutional, or travel environment.
### Patient Population:
Adult patients with chronic pulmonary diseases such as chronic bronchitis, emphysema, asthma, or lung cancer, those in the terminal stage of cancer, or any patient requiring supplemental oxygen.
The GCE Zen-O™ POC is not intended for use in life supporting or life sustaining situations, and is provided non-sterile. It is a prescription only device, and designed for indoor and outdoor use.
### Environment of Use:
Travel, Home or Institution
## VI. Comparison of Technological Characteristics with the Predicate Device
Pressure Swing Absorption (PSA) is the technological principle for both the subject and predicate devices. It is based on molecular sieve / pressure swing absorption technology, which draws ambient air, pushes it through a sieve bed, then utilizes pressure swing absorption to convert the ambient air to pure oxygen.
{5}------------------------------------------------
Table 1 compares the key features of the proposed GCE Zen-O™ POC with the identified predicate. The comparison demonstrates that the device can be found to be substantially equivalent to the identified predicate.
# Table 1 – Comparison of Proposed Device vs. Predicate
Image /page/5/Picture/2 description: The image shows a portable oxygen concentrator. The device is light gray with a black handle on top. The control panel is also black and has a small display screen and several buttons. The concentrator is designed to provide supplemental oxygen to people who need it.
Image /page/5/Picture/3 description: The image shows a white Oxus Zen-O portable oxygen concentrator. The device has a small display screen with several buttons below it. The Oxus logo is visible on the front of the device. The concentrator is designed to provide supplemental oxygen to individuals with respiratory conditions.
| Attribute | Predicate Device:<br>K073242 RS-00400 | Proposed Device:<br>GCE Zen-O™ RS-00500 |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Oxus RS-00400 Portable<br>Oxygen Concentrator is intended<br>to provide supplemental oxygen<br>in a home, institutional, or travel<br>environment. | The GCE Zen-O™ Portable<br>Oxygen Concentrator is intended<br>to provide supplemental oxygen<br>in a home, institutional, or travel<br>environment. |
| Environments<br>of Use | Travel, Home or Institution | Travel, Home or Institution |
| Prescriptive | Yes | Yes |
| Patient<br>Population | Adult | Adult |
| Single Patient,<br>multi-use | Yes | Yes |
| Patient Interface | Cannula port | Cannula port |
| Technology | Pressure Swing Adsorption with<br>molecular sieve | Pressure Swing Adsorption with<br>molecular sieve |
| Dimensions | 4.6" H x 7.5" W x 11.6" L | 6.6" H x 8.3" W x 12.3" L |
| Weight | 10 lbs | 10.25 lbs |
| Oxygen<br>Concentration | 87% - 96% | 87% - 96% |
| Equivalent Flow<br>Rates | 1-5 LPM, increments of 0.5 LPM | 1-6 LPM, increments of 0.5 LPM<br>0.5 - 2 LPM, increments of 0.5 L<br>(Continuous mode) |
| Attribute | Predicate Device:<br>K073242 RS-00400 | Proposed Device:<br>GCE Zen-O™ RS-00500 |
| Dose at<br>Specified Flow | 8.5 mL per setting | 11 mL per setting |
| Filters | Input Filter<br>Intermediate Filter<br>Patient Filter | Input Filter<br>Patient Filter |
| User Interface | Buttons<br>LCD Display | Buttons<br>LCD Display |
| Electrical | 100-240VAC, 50/60 Hz, 1.6A<br>18VDC, 7A | 100-240VAC, 50/60 Hz, 2.5A<br>24VDC, 6.25A |
| Software | Embedded | Embedded |
| Acoustic Noise | 42 dBA at 2 LPM | 42 dBA at 2 LPM |
| Alarms | • Battery empty<br>• Low Pressure<br>• No pulse<br>• High temp<br>• RAM failure<br>• Motor stall<br>• Compressor failure<br>• Fan failure<br>• Low Battery<br>• No Breath Detected<br>• MP timeout<br>• EEPROM failure<br>• EEPROM error<br>• Replace filter<br>• Clean input filter<br>• Low RTC battery | • Battery empty<br>• Low Pressure<br>• Motor communication<br>• No pulse<br>• Skipped Breaths<br>• High temp<br>• RAM failure<br>• Motor stall<br>• Compressor failure<br>• Fan failure<br>• Invalid Battery<br>• Low Flow<br>• Low Battery<br>• No Breath Detected<br>• uC timeout<br>• EEPROM failure<br>• EEPROM error |
| Status<br>Indicators | Flowrates<br>Battery Condition<br>Alarms<br>History Log<br>Diagnostics | Flowrates<br>Battery Condition<br>Alarms<br>History Log<br>Diagnostics |
| Battery<br>Duration | Approximately 4 hours at 2LPM<br>(pulse) | Approximately 4 hours at 2LPM<br>(pulse) |
| Operating<br>Environment | 5 to 40° C<br>Altitude: 0-8000 ft<br>RH: 5 - 95 % | 5 to 40° C<br>Altitude: 0-9000 ft<br>RH: 5 - 93 % |
| Shipping /<br>Storage<br>Conditions | Temperature: -20 to 60°C, Keep<br>Dry,<br>Humidity: 0 - 95 % RH | Temperature: -20 to 60°C, Keep<br>Dry,<br>Humidity: 0 - 93 % RH |
| Attribute | Predicate Device: K073242 RS-00400 | Proposed Device: GCE Zen-O™ RS-00500 |
| Safety | IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 |
| Mechanical<br>Safety | IEC 60601-1 | IEC 60601-1 |
| Chemical Safety | Not Applicable | Not Applicable |
| Thermal Safety | IEC 60601-1 | IEC 60601-1 |
| Biocompatibility | 4 VOC's less than ambient | 4 VOC's less than ambient |
| Standards Met | IEC 60601-1<br>IEC 60601-1-2<br>ISO 8359:1996<br>ASTM F-1464-93:2005 | AAMI ANSI ES60601-1<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-8<br>IEC 60601-1-11<br>ISO 80601-2-67<br>ISO 80601-2-69 |
| Type of<br>Protection<br>against electric<br>shock | Type BF - Not for cardiac | Type BF - Not for cardiac |
| Safety Markings | UL | ETL |
{6}------------------------------------------------
{7}------------------------------------------------
### Comparison Discussion
#### Indications for Use -
The indications for use are equivalent for the proposed device when compared to the predicate - K073242 - Oxus Model RS-00400 POC.
Discussion - The indications for use are equivalent to the predicate - K073242 - Oxus Model RS-00400 POC
#### Patient Population -
The patient populations are similar for the proposed device when compared to the predicate -K073242 – Oxus Model RS-00400 POC. Adult patients with chronic pulmonary diseases such as chronic bronchitis, emphysema, asthma, or lung cancer, those in the terminal stage of cancer, or any patient requiring supplemental oxygen.
Discussion - The patient populations are equivalent to the predicate - K073242 - Oxus Model RS-00400 POC
{8}------------------------------------------------
### Environment of Use -
The environment of use are equivalent for the proposed device when compared to the predicate - K073242 - Oxus Model RS-00400 POC.
Discussion - The environment of use are equivalent to the predicate - K073242 - Oxus Model RS-00400 POC
### Principle of Operation and Technology -
The design, components, and principle of operation of the POC, based on molecular sieve / pressure swing absorption technology, to draw ambient air, push it through a sieve bed, then utilize pressure swing absorption to convert the ambient air to pure oxygen have been shown to be equivalent to the predicate - K073242 - Oxus Model RS-00400 POC.
Discussion – The GCE Zen-O™ POC design uses molecular sieve / Pressure Swing Adsorption (PSA) technology to deliver concentrated oxygen. A series of chambers and valves allows pressurized air to enter the sieve bed assembly, effectively separating nitrogen from the air. When one chamber is receiving pressurized air, the other is purqing nitrogen back into the air. The cycle is repeated continuously. The concentrated oxygen created at each cycle is stored in a chamber to be delivered to a patient when the device detects a patient breath are equivalent to the predicate -K073242 - Oxus Model RS-00400 POC
The differences between the proposed device and predicate are:
- The GCE Zen-O™ Portable Oxygen Concentrator has additional capability compared to . the Oxus Model RS-00400 as it can deliver in either pulse mode or continuous flow modes; however the oxygen producing operations are similar.
- The two products are different from each other as indicated in the comparison table ● below in size and weight. The subject device is a slightly heavier concentrator that provides both pulse and continuous flow delivery options.
The GCE Zen-O™ POC Model RS-00500 and the predicate device both meet the requirements related to performance and safety standards applicable to the portable oxygen concentrators and the differences noted in the comparison do not raise different questions of safety or effectiveness and thus we can conclude the proposed GCE Zen-O™ POC Model RS-00500 is substantially equivalent to the predicate, K073242 - Oxus Model RS-00400 POC.
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
{9}------------------------------------------------
### Biocompatibility testing
The biocompatibility evaluation for the GCE Zen-O™ POC was conducted in accordance with the FDA Guidance Document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA.
Particulate Matter 2.5 micron (PM2.5) testing, according to United States Environmental Protection Agency (EPA), was completed and the POC emitted a mean PM2.5 level less than the EPA PM2.5 level of 15 ug/m³.
Inorganic gases (Ozone, CO2, and CO) were evaluated with no levels detected of CO2 or CO and Ozone concentration is less than the EPA allowed 0.050 ppm.
Patient contact with composite materials is limited to:
- . Device enclosure,
- . Handle,
- Battery, and ●
- Interface panel. .
The route of exposure to these materials would be through incidental contact with the patient's skin and does not contact the patient during normal device use.
The GCE Zen-O™ POC Model RS-00500 Medical Device Categorization per Annex A of ISO 10993-1:2009(E) for all composite material is:
- Category : External communicating device
- Contact : Tissue ●
- . Duration : A - Limited (≤ 24 h)
The POC is considered tissue contacting for a duration of less than 24 hours.
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the GCE Zen-O™ POC. The system complies with the AAMI ANSI ES60601-1, IEC 60601-1. IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, ISO 80601-2-67, and ISO 80601-2-69 standards for electrical safety and the IEC 60601-1-2 standard for EMC.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could result in Minor Iniurv. either to a patient or to a user of the device.
#### Mechanical and acoustic testing
Functional Performance Testing has been performed on a production unit to demonstrate that the GCE Zen-O™ Portable Oxygen Concentrator will perform as intended and additionally to
{10}------------------------------------------------
demonstrate equivalency to the Oxus RS-00400 POC. The production unit was constructed of components of the final device configuration and meet performance specifications.
Testing was performed according to the performance standards listed above and internal test protocols to identify test methods.
### Comparative Bench Testing
The GCE Zen-O™ Portable Oxygen Concentrator (POC) has been designed to be compliant with the requirements of recognized consensus standards for Oxygen Concentrators:
- . ISO 80601-2-67 Particular requirements for basic safety and essential performance of oxygen conserving equipment
- . ISO 80601-2-69 Particular requirements for basic safety and essential performance of oxygen concentrator equipment
#### Usability Studies
Usability testing was performed with 4 users on two devices, lay users and Healthcare Providers. Testing included the appropriate tasks based upon the risk analysis. Testing found that the design of the device and instructions for use were appropriate for the intended user groups.
### Animal Study
Not applicable for this device
### Clinical Studies
There was no clinical testing performed.
### VIII. Conclusions
As detailed, the indications for use, patient population, environment of use, technology or principle of operation, and performance are substantially equivalent to the predicate.
The identified differences between the proposed GCE Zen-O™ POC Model RS-00500 and the predicate - K073242 - Oxus Model RS-00400 POC, allows us to conclude that there are no different questions of safety or effectiveness that raise new risks and thus the proposed device can be determined to be substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.