K140264 · Thornhill Research, Inc. · BSZ · Jan 5, 2015 · Anesthesiology
Device Facts
Record ID
K140264
Device Name
MADM
Applicant
Thornhill Research, Inc.
Product Code
BSZ · Anesthesiology
Decision Date
Jan 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MADM™ is intended to deliver volatile anesthetic to a patient when placed in either circle or open anesthetic circuits. It vaporizes Isoflurane and Sevoflurane and delivers the vaporized anesthetic agent into the inspiratory limb of the breathing circuit. MADM™ is also intended to monitor respiratory rate, CO2, and the anesthetic gases Isoflurane and Sevoflurane. It is intended to be connected to a patient breathing circuit for monitoring of patients to whom it is delivering volatile anesthetic gases.
Device Story
MADM™ is a portable vaporizer for delivery of volatile anesthetics (Isoflurane, Sevoflurane) into patient breathing circuits; compatible with both circle and open circuits. Device measures anesthetic concentration of incoming gas via internal sensors; automatically adjusts output to maintain set concentration, accounting for recirculated anesthetic in circle systems. Includes respiratory rate, CO2, and anesthetic gas monitoring. Features internal battery (20 min) and optional external battery base (2+ hours). Safety features include anesthetic sensor at patient mouth to trigger automatic cessation if inhaled levels exceed thresholds; safety interlock prevents delivery if sensors malfunction or are absent. Used in clinical settings by healthcare providers to manage anesthesia delivery; provides real-time monitoring and controlled vaporized agent delivery to support patient sedation.
Clinical Evidence
Bench testing only. Performance testing validated anesthetic delivery accuracy across clinical ranges of respiratory rates, fresh gas flows, and tidal volumes. Battery longevity, environmental operating range, and disturbance analysis (step changes in flow/rate) were evaluated. Anesthetic compatibility testing confirmed non-reactivity of components with Isoflurane and Sevoflurane. Biocompatibility evaluation performed per 510(k) Memorandum #G95-1. Results met all acceptance criteria and international standards (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-13, ISO 80601-2-55).
Technological Characteristics
Portable vaporizer; uses internal sensors for anesthetic gas concentration measurement. Powered by internal battery or external battery base. Compatible with Isoflurane and Sevoflurane. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-13, and ISO 80601-2-55. Materials evaluated for biocompatibility and anesthetic chemical compatibility.
Indications for Use
Indicated for adult patients weighing more than 40 kg requiring delivery of volatile anesthetic (Isoflurane or Sevoflurane) and monitoring of respiratory rate, CO2, and anesthetic gas concentrations.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized human face or profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 5,2015
Thornhill Research Inc. Cliff Ansel President 210 Dundas St. W. Suite 200 Toronto, ON, Canada M5G 2E8
Re: K140264
Trade/Device Name: MADMTM Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ, CCK Dated: December 15, 2015 Received: December 22, 2015
Dear Mr. Ansel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Ansel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runyon DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): K140264
Device Name: MADM™M
MADM™ is intended to deliver volatile anesthetic to a patient when placed in either circle or open anesthetic circuits. It vaporizes Isoflurane and Sevoflurane and delivers the vaporized anesthetic agent into the inspiratory limb of the breathing circuit.
MADM™ is also intended to monitor respiratory rate. CO2, and the anesthetic gases Isoflurane and Sevoflurane. It is intended to be connected to a patient breathing circuit for monitoring of patients to whom it is delivering volatile anesthetic gases.
#### Target Population
The intended patient population is adults who weigh more than 40 kg.
Prescription Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) SUMMARY
## SUBMITTER INFORMATION
| Company Name:<br>Company Address: | Thornhill Research Inc.<br>210 Dundas St. W, Suite 200<br>Toronto, ON, Canada M5G 2E8 |
|-----------------------------------|---------------------------------------------------------------------------------------|
| Company Phone: | (416) 597-1325 |
| Company Fax: | (416) 597-1330 |
| Contact Person: | Cliff Ansel, President |
#### DEVICE IDENTIFICATION
| Trade/Proprietary Name: | MADM™ |
|-------------------------|-------------------------|
| Classification: | II |
| Generic Device Name: | Gas-machine, Anesthesia |
#### Classification Names
| Classification Name | Product Code | Class | Regulation Number |
|-----------------------------------------|--------------|-------|-------------------|
| Gas Machine for anesthesia or analgesia | BSZ | II | 21 CFR 868.5160 |
| Carbon Dioxide Gas Analyzer | CCK | II | 21 CFR 868.1400 |
## DEVICE DESCRIPTION
The MADM™ system is a portable vaporizer capable of delivering anesthetic to the inspiratory limb of a breathing circuit. Unlike other in-line vaporizers, MADMTM can be inserted into a circle system where some exhaled anesthetic is rebreathed. The MADM™ system measures the anesthetic concentration of incoming gas and reduces the anesthetic output to ensure the anesthetic concentration of gas delivered to the patient is as set on the dial.
MADM™ includes an internal battery backup capable of powering the device for 20 minutes, and an optional external battery base with hot swappable batteries capable of powering the system for over two hours. Anesthetic is stored in custom single agent anesthetic canisters.
{4}------------------------------------------------
## INTENDED USE
MADM™ is intended to deliver volatile anesthetic to a patient when placed in either circle or open anesthetic circuits. It vaporizes Isoflurane and Sevoflurane and delivers the vaporized anesthetic agent into the inspiratory limb of the breathing circuit.
MADM™ is also intended to monitor respiratory rate, CO2, and the anesthetic gases Isoflurane and Sevoflurane. It is intended to be connected to a patient breathing circuit for monitoring of patients to whom it is delivering volatile anesthetic gases.
# Intended Use Population
The intended patient population is adults who weigh more than 40 kg.
# SUBSTANTIAL EQUIVALENCE
The MADM™ is of comparable type and is substantially equivalent to the following predicate devices:
| Device | 510(k) # |
|------------------------------|----------|
| PAC Vaporizer | K850648 |
| Penlon Sigma Delta Vaporizer | K002343 |
| Tec-6 Plus Vaporizer | K000275 |
| IRMA | K123043 |
## Comparison to Predicate Device
MADM™ and the predicate devices have the same intended use. Namely, each device is intended for the controlled delivery of anesthetic vapor into the patient breathing circuit. MADM™ can operate in both flow-through and circle systems, both inside and outside the breathing circuit. A typical draw-over vaporizer cannot be inserted into a circle system as it would continually add anesthetic to that being inspired eventually leading to a hazard.
MADM™ includes an anesthetic sensor at its input which measures the content of the incoming gas, so that the delivered agent is reduced according to what is recirculated. A detailed risk analysis has been conducted to ensure that this configuration can function safely, and several safety features have been included. These include the inclusion of an anesthetic sensor at the mouth which causes automatic cessation of delivery if inhaled anesthetic exceeds a threshold above what is set. A safety interlock is in place which prevents anesthetic from being delivered if either anesthetic sensor is malfunctioning or not present.
{5}------------------------------------------------
# Compliance to Standards and Regulations
AAMI/ANSI ES60601-1 IEC 60601-1-2 IEC 60601-2-13 ISO 80601-2-55
## Summary of Performance Testing
Safety and Performance Testing was conducted in accordance with all referenced standards and regulations, and to validate all system requirements. Performance testing included:
- . Performance testing across the intended environmental operating range.
- Battery Longevity testing across the intended environmental operating range.
- Anesthetic delivery accuracy across the anticipated clinical range of respiratory ● rates, fresh gas flows, and tidal volumes.
- Disturbance analysis including step changes in Respiratory Rate, Fresh Gas Flow, and Tidal Volume.
- Anesthetic compatibility evaluation demonstrating that components in contact with ● anesthetic gas were non-reactive to Isoflurane and Sevoflurane.
- Biocompatibility Evaluation according to the 510(k) Memorandum - #G95-1
The results of performance testing demonstrate that the characteristics of MADMTM are substantially equivalent to the identified predicates and meet the acceptance criteria of international standards in terms of anesthetic delivery accuracy and gas sensing accuracy as well as all of its system requirements.
# Determination of Substantial Equivalence
Differences in technological characteristics between MADMTM and the identified predicates do not raise any new concerns of safety and efficacy.
The MADMTM has a substantially equivalent intended use to the listed predicate devices. Where differences exist in performance characteristics, these differences do not adversely affect safety and efficacy. In addition, all relevant aspects of recognized consensus performance standards have been met.
## Conclusion
The MADM™ is substantially equivalent to the identified predicate devices and does not raise any new concerns about safety and efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.