The GE Datex-Ohmeda Anesthesia Delivery Unit is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control, pressure support and synchronized intermittent mandatory (SIMV) ventilation modes. The ADU is not suitable for use in a MRI environment.
Device Story
Anesthesia delivery unit provides general inhalation anesthesia and ventilatory support; used by trained medical professionals in clinical settings. Device mixes medical gases (O2, air, N2O) via mechanical mixing; user selects flows via rotary controller. Anesthetic agent delivery (Halothane, Enflurane, Isoflurane, Desflurane, Sevoflurane) managed by anesthesia computer using Aladin cassettes; valves control agent delivery percentage. Microprocessor-based ventilator provides volume/pressure control, pressure support, and SIMV modes; sensors in breathing circuit monitor ventilation, compensating for compression losses, fresh gas contribution, and leaks. PEEP regulated electronically. System displays flow, pressure, and ventilation measurements. RS-232 port enables communication with external monitors (e.g., Datex-Ohmeda S/5). Device integrates with patient monitors to assist clinicians in managing patient anesthesia and ventilation during surgical procedures.
Clinical Evidence
Bench testing only. Device compliance with voluntary standards (UL 2601, EN 740, IEC 60601-1, IEC 60601-1-2, EN 475, ASTM F1463-93, ASTM F1208-94, ASTM F1101-90, ISO 5358) provided to demonstrate safety and effectiveness.
Technological Characteristics
Microprocessor-controlled anesthesia workstation. Features mechanical gas mixing, electronic flow indicators, and Aladin cassette-based agent delivery. Ventilator uses breathing circuit sensors for pressure/volume control and PEEP regulation. Connectivity via RS-232 serial port. Complies with UL 2601, EN 740, IEC 60601-1, IEC 60601-1-2, EN 475, ASTM F1463-93, ASTM F1208-94, ASTM F1101-90, and ISO 5358.
Indications for Use
Indicated for general inhalation anesthesia and ventilatory support for a wide range of patients. Contraindicated for use in MRI environments.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
K090892 — GE DATEX-OHMEDA S/5 ADU CARESTATION · GE Healthcare Finland Oy · Apr 30, 2009
K061609 — GE DATEX-OHMEDA AISYS · Datex-Ohmeda, Inc. · Jun 27, 2006
K050626 — GE DATEX-OHMEDA AESPIRE 7900 ANESTHESIA SYSTEM · Datex-Ohmeda, Inc. · Mar 25, 2005
K132530 — GE DATEX-OHMEDA AISYS CS2 (AISYS CS^2) · Datex-Ohmeda, Inc. · Dec 20, 2013
K090233 — GE DATEX-OHMEDA AISYS · Datex-Ohmeda, Inc. · Mar 5, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
APR 1 2 2005
K050676
March 14, 2005 Date:
510(k) Summary of Safety and Effectiveness Information Subject: for the GE Datcx-Ohmeda S/5 ADU
GE Datex-Ohmeda S/5 Anesthesia Delivery Unit Proprietary:
Gas Machine, Anesthesia Common:
Anesthesiology, 73 BSZ, 21 CFR 868.5160 Classification:
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992.
The GE Datex-Ohmeda ADU is substantially equivalent to the following currently marketed device:
GE Datex-Ohmeda ADU Carestation - Class II - 21CFR868.5160, which has been the subject of a cleared 510(k) with FDA log number K042154
Datex-Ohmeda AS/3 Anesthesia Delivery Unit - Class II - 21CFR868.5160, which has been the subject of a cleared 510(k) with FDA log number K973985
The GE Datex-Ohmeda ADU is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. It is to be used only by trained and qualified medical professionals.
The GE Datex-Ohmeda ADU supplies set flows of medical gases to the breathing system using mechanical gas mixing. Gas flows are selected by the user using the rotary controller on the frame and then displayed as electronic flow indicators on the system display unit. The ADU is equipped with a traditional flow tube, as well. The ADU is also available in a pendant model. It is available with two or three gases, and up to three cylinder connections. All models have 02. The ADU comes with up to two optional gases (air, N2O). Safety features and devices within the ADU are designed to decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures.
The anesthetic agent delivery for the ADU is controlled via an anesthesia computer through user input from that computer. An Aladin cassette is inserted into the active cassette bay. The cassette holds the agent to be delivered - Halothane, Enflurane, Isoflurane, Desflurane or Sevoflurane. Agent is delivered as a percent volume/volume. The ADU is designed to allow only one active cassette at a time. Per the user input, valves within the active cassette bay will oney one allow agent to be delivered. The agent is mixed with gas within the FGC unit. After open and the wagent to of gases and agent is delivered to the breathing system and then onto the patient.
{1}------------------------------------------------
The ADU Anesthesia Ventilator is a microprocessor based, electronically controlled, The ADO Ancomesia "Pentilator that provides patient ventilation during surgical procedures. Sensors in the breathing circuit are used to control and monitor patient ventilation. This allows Schools in the oreating of compression losses, fresh gas contribution and small leakage in the for the compensation of our spression. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered with a sitiple and becare othir pratory Pressure (PEEP) is regulated electronically. Positive ter the paintained in the breathing system so that any leakage that occurs is outward. Pressure is names for the device include Volume Mode, Pressure Control Mode, Pressure Support vith Apnea Backup Mode (Optional) and Synchronized Intermittent Mandatory Ventilation with Aplica Dackap Mode (Optional) and measurements are displayed on the system display unit.
The ADU must be used with additional monitoring that include at least inspired 02, expired volume, expired CO2 and Anesthetic Agent.
An RS-232 serial digital communications port connects to and communicates with external devices such a Datex-Ohmeda S/5 Anesthesia Monitor.
Several frame configurations are available, including one that allows for the physical integration of the Datex-Ohmeda S/5 Anesthesia Monitor (most recently cleared via K030812). Additional or the Dations allow for the mounting of various patient monitors on the top shelf of the ADU.
The GE Datex-Ohmeda ADU was designed to comply with the applicable portions of the following voluntary standards;
- 1. UL 2601 -- General requirements for Medical Electrical Equipment
- 2. EN 740 Anesthetic Work Stations
- 3. EN/IEC 60601-1: General requirements for Medical Electrical Equipment
- 4. EN/IEC 60601-1-2: 2001 Medical Electrical Equipment Electromagnetic Compatibility
- 5. EN 475 Electrically Generated Alarm Signals
- 6. ASTM F1463-93 Standard Specification for Alarm Signals
- 7. ASTM F1208-94 Anesthesia Breathing Circuit Standard
- 8. ASTM F1101-90 Standard Specification for Ventilators Intended for Use During Anesthesia
- 9. ISO 5358 Anesthetic Gas Machines
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird in flight, with three curved lines representing its wings or feathers.
## Public Health Service
APR 1 2 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Datex Ohmeda, AB c/o Mr. Dan Kosednar Manager, Regulatory Submissions Datex-Ohmeda, Incorporated Life Support Solutions P.O. Box 7550 Madison, Wisconsin 53707-7550
Re: K050676
Trade/Device Name: GE Datex-Ohmeda Anesthesia Delivery Unit (ADU) Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine, Anesthesia Regulatory Class: II Product Code: BSZ Dated: March 15, 2005 Received: March 16, 2005
Dear Mr. Kosednar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include oonerols provisions of annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 wr ), it har obe of Federal Regulations, Title 21, Parts 800 to 898. In Jour device of may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Kosednar
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the Free of any with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and roting ( - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - required.le, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), nlease contact the Office of Compliance at (240) 276-0102. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Charles
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): Kesow 7le Device Name: GE Datex-Ohmeda Anesthesia Delivery Unit (ADU)
Indications For Use:
The GE Datex-Ohmeda Anesthesia Delivery Unit is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control, pressure support and synchronized intermittent mandatory (SIMV) ventilation modes. The ADU is not suitable for use in a MRI environment.
Prescription Use _XXX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of __1__1__
Chris
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infaction Control, Dental Devices
510(k) Number K056626
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.