MODEL 5471 & 5471L STERILE DISPOSABLE PATIENT CABLE ASSEMBLIES
Applicant
Medtronic Vascular
Product Code
DSA · Cardiovascular
Decision Date
May 22, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2900
Device Class
Class 2
Intended Use
The intended use of the patient cables is identical to previous patient cables.
Device Story
Sterile disposable patient cable assemblies; connect heart wires/leads to external cardiac pacing/monitoring equipment. Device transmits electrical signals between patient and external device. Used in clinical settings; operated by healthcare professionals. Modified connector design compared to predicate; otherwise identical function. Benefits include reliable signal transmission for cardiac monitoring/pacing.
Patient transducer and electrode cable assembly. Features modified connector for heart wires/leads. Materials biocompatible (non-cytotoxic, non-irritating). Sterilized for single use. Electrical characteristics include specified resistance and dielectric withstanding voltage.
Indications for Use
Indicated for patients requiring connection between heart wires/leads and external cardiac pacing/monitoring equipment.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
Predicate Devices
Models 5455, 5455L, 5455S, and 5455SL surgical cables
K012392 — DISPOSABLE SURGICAL EXTENSION CABLE, MODEL 4255 AND MODEL 4260 · Pace Medical · Sep 4, 2001
K220447 — Med-link Disposable ECG Cable and Leadwires · Shenzhen Med-Link Electronics Tech Co., Ltd. · Jun 28, 2022
K181089 — SureLead Cable System · Authentic Medical · Jan 17, 2020
K070926 — ELECTRODE/EXTENSION CABLES, MODELS ATAR · Oscor, Inc. · Jun 7, 2007
Submission Summary (Full Text)
{0}
MAY 22 1996
# K960446
510(k) SUMMARY
[As required by 21 CFR 807.92(a)]
## A. Submitter Information
Submitter's Name: Medtronic, Inc.
Address: 7000 Central Ave. N.E.
Minneapolis, MN 55432-3576
Telephone Number: 612/574-4654
Fax Number: 612/574-6424
Contact Person: Valerie K. Stuhr
Submission Preparation Date: January 31, 1996
## B. Device Information
Trade Name: Models 5471 and 5471L Sterile Disposable Patient Cable Assemblies
Common Name: Patient Cable
Classification Name: Patient transducer and electrode cable (including connector), Class II (21 CFR 870.2900)
Predicate Device: This 510(k) summary identifies the Models 5455, 5455L, 5455S, and 5455SL surgical cables as substantially equivalent to the proposed patient cable assemblies.
{1}
Device Description: The proposed patient cables are identical to the commercially-approved Models 5455, 5455L, 5455S, and 5455SL surgical cables, with the exception of a modified connector for heart wires and leads.
Intended Use: The intended use of the patient cables is identical to previous patient cables.
## C. Comparison of Required Technological Characteristics
The proposed patient cables are substantially equivalent to the commercially-available Models 5455, 5455L, 5455S, and 5455SL surgical cables. The proposed cables are identical to the current cables except that the proposed cables include a modified connector.
| Characteristics | Current and Proposed |
| --- | --- |
| a. Product Labeling | Substantially equivalent |
| b. Intended Use | Identical |
| c. Physical Characteristics | Substantially equivalent |
| d. Anatomical Sites | Identical |
| e. Target Population | Identical |
| f. Performance Testing | Substantially equivalent |
| g. Safety Characteristics | Substantially equivalent |
## D. Summary of Nonclinical Tests
Mechanical, electrical, and environmental/packaging tests show that the proposed patient cable assemblies are qualified and raised no new questions with regard to safety and efficacy. Mechanical test results for cable retaining force, lead, and heartwire connection withdrawal forces, and button depression all met requirements. Electrical tests for current, resistance, and dielectric withstanding voltage also met requirements. Environmental/packaging tests demonstrated that the packaging can withstand various shock, vibration, temperature, and humidity conditions. Biocompatibility tests showed that the connector does not produce cytotoxicity or dermal irritation. All tests met their respective requirements.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.