Disposable ECG Cable and Leadwires are intended to be used with ECG monitoring devices to measure a patient's ECG for general monitoring and/or diagnostic evaluation by heath care professional. Disposable ECG Cable and Leadwires meets the requirements of the standard ANSI/AAMI EC53, ECG TRUNK CABLES and PATIENT LEADWIRES. This leadwire set is intended for single-patient use.
Device Story
Disposable ECG cable and leadwires; interface between patient electrodes and ECG monitoring devices. Input: electrical signals from patient electrodes. Transformation: passive signal transmission via shielded bulk cables. Output: ECG signal to monitor. Used in clinical settings by healthcare professionals. Facilitates general monitoring and diagnostic evaluation of patient cardiac activity. Benefits: provides reliable signal path for patient monitoring; disposable design reduces cross-contamination risk.
Clinical Evidence
Bench testing only. Testing included electrical safety per IEC 60601-1, performance per ANSI/AAMI EC53:2013, biocompatibility per ISO 10993-1, -5, -10, and package integrity/functional performance testing following real-time aging.
Technological Characteristics
Materials: medical grade PVC and TPU cable jacket, PVC and ABS over-molded connectors. Sensing: passive electrical signal transmission. Form factor: shielded bulk cables with various connectors (trunk/lead wire, grabber/snapper). Connectivity: wired interface to patient monitor. Sterilization: non-sterile. Standards: ANSI/AAMI EC53, IEC 60601-1, ISO 10993.
Indications for Use
Indicated for patients requiring ECG monitoring or diagnostic evaluation using ECG monitoring devices. Intended for use by healthcare professionals.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
Predicate Devices
Cable/ lead-wire (ECG, EKG, SPO2 and invasive blood pressure (K120010))
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 28, 2022
Shenzhen Med-link Electronics Tech Co., Ltd. Yi Liu Regulatory Affairs Specialist 4th and 5th Floor, Building Two, Hualian Industrial Zone, Xinshi Community, Dalang Street Shenzhen, Guangdong 518109 China
Re: K220447
Trade/Device Name: Med-link Disposable ECG Cable and Leadwires Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer And Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: April 29, 2022 Received: May 9, 2022
Dear Yi Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220447
Device Name Disposable ECG Cable and Leadwires
Indications for Use (Describe)
Disposable ECG Cable and Leadwires are intended to be used with ECG monitoring devices to measure a patient's ECG for general monitoring and/or diagnostic evaluation by heath care professional.
Disposable ECG Cable and Leadwires meets the requirements of the standard ANSI/AAMI EC53, ECG TRUNK CABLES and PATIENT LEADWIRES. This leadwire set is intended for single-patient use.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: large;">☐</span> Over The Counter Use (21 CFR 801 Subpart C) </div> |
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for MedLinket. The logo is composed of the word "MedLinket" in a sans-serif font. The first letter, "M", is colored in orange, while the rest of the letters are in light blue.
Shenzhen Med-link Electronics Tech Co., Ltd.
# 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| 1. Submitter information: | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name and Address: | Shenzhen Med-link Electronics Tech Co., Ltd.<br>4th and 5th Floor, Building Two,<br>Hualian Industrial Zone, Xinshi Community,<br>Dalang Street, Longhua District, Shenzhen,<br>Guangdong 518109,<br>People's Republic of China |
| Primary Contact: | Yi Liu(Regulatory Affairs Specialist)<br>E-mail: user22@med-linket.com<br>Tel: 0086-755-61568825<br>Fax: 0086-755-61120055 |
| Data of Preparation: | 29, Apr. 2022 |
| 2. Device Information: | |
| Trade Name: | Disposable ECG Cable and Leadwires |
| Common Name: | Cable and Leadwires |
| Classification number: | 21 CFR 870.2900 |
| Classification name: | Cable, Transducer And Electrode, Patient, (Including Connector) |
| Product Code: | DSA |
| Regulatory Class: | II |
| Review Panel: | Cardiovascular |
#### 3. Identification of the Predicate Device:
| Device Name | Common Name | Manufacturer | 510(k)<br>number |
|--------------------------------------------------------------------|---------------------|-------------------------------------------------|------------------|
| Cable/ lead-wire (ECG,<br>EKG, SPO2 and invasive<br>blood pressure | Cable and Leadwires | Shenzhen Med-Link Electronics<br>Tech Co., Ltd. | K120010 |
{5}------------------------------------------------
# Medlinket
#### Intended Use and Indications for Use of the Subject Device 4.
Disposable ECG Cable and Leadwires are intended to be used with ECG monitoring devices to measure a patient's ECG for general monitoring and/or diagnostic evaluation by heath care professional.
Disposable ECG Cable and Leadwires meets the requirements of the standard ANSI/AAMI EC53, ECG TRUNK CABLES and PATIENT LEADWIRES. This leadwire set is intended for single-patient use.
#### Device Description ട്.
Disposable ECG Cable and Leadwires with specific various lengths are the replacements for similar cables manufactured by Original Equipment Manufacturers (OEM) and other third party after market manufacturers for their respective monitors.
These devices consist of connectors on each cable end and a shielded bulk cables are used to transfer the signals from the electrodes to the patient monitor.
These devices use the same type of constructions and have the same technological characteristics as the predicate devices. They use a medical grade PVC and TPU cable jacket with medical grade PVC and ABS over molded connectors with integral relief.
| Item | Proposed Device | Predicate Device | Remark |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Comparison | Disposable<br>ECG Cable and Leadwires | Shenzhen Med-link<br>Cable/lead-wire | / |
| 510k number | K220447 | K120010 | / |
| 510(K) Submitter | Shenzhen Med-link<br>Electronics Tech Co., Ltd. | Shenzhen Med-link<br>Electronics Tech Co., Ltd. | / |
| Intended use<br>&Indications<br>for Use | Disposable ECG Cable and<br>Leadwires are intended to be used<br>with ECG monitoring devices to<br>measure a patient ' s ECG for<br>general monitoring and/or<br>diagnostic evaluation by heath<br>care professional.<br>Disposable ECG Cable and | Shenzhen Med-link<br>Cable/lead-wire is intended to be<br>used with ECG, EKG, SpO2 and<br>invasive Blood Pressure<br>monitoring devices. The<br>Cable/lead-wire are used to<br>connect electrodes, catheters,<br>and/or sensors placed at | SE |
| | Leadwires meets the requirements<br>of the standard ANSI/AAMI<br>EC53, ECG TRUNK CABLES<br>and PATIENT LEADWIRES.<br>This leadwire set is intended for<br>single-patient use. | appropriate sites on the patient to a<br>monitoring device for general<br>monitoring and/or diagnost ic<br>evaluation by health care<br>professional. | |
| Device description | Disposable ECG Cable and<br>Leadwires with specific various<br>lengths are the replacements for<br>similar cables manufactured by<br>Original Equipment<br>Manufacturers (OEM) and other<br>third party after market<br>manufacturers for their respective<br>monitors.<br>These devices consist of<br>connectors on each cable end and<br>a shielded bulk cable. The cables<br>are used to transfer the signals<br>from the electrodes to the patient<br>monitor. | Med-link Cable / Lead-wire with<br>specific various lengths are the<br>replacements for similar cables<br>manufactured by Original<br>Equipment Manufacturers (OEM)<br>and other third party after market<br>manufacturers for their respective<br>monitors.<br>These cables consist of connectors<br>on each cable end and a shielded<br>bulk cable. The cables are used to<br>transfer the signals from the<br>electrodes to the patient monitor. | SE |
| Anatomical<br>Sites | Attached to electrodes placed at<br>standard specified locations on<br>chest wall | Attached to electrodes placed at<br>standard specified locations on<br>chest wall | SE |
| Design<br>/Appearance | ECG and EKG Cables<br>with various connectors (monitor | ECG and EKG Cables<br>with various connectors (monitor | SE |
| | trunk / lead wire,electrode grabber<br>& snapper) | trunk / lead wire,electrode grabber<br>& snapper) | |
| Sterility | Non sterile | Non sterile | SE |
| Usage | Disposable | Reusable | Different1) |
| Patient end<br>termination | Clip | Clip, Snap | Different2) |
| Material | PVC | Shielded & Unshielded Copper<br>with PVC or TPU Jacket | Different3) |
| Cable Length | Various specified<br>standard lengths | Various specified<br>standard lengths | SE |
| Electrical Safety | Complied with IEC 60601-1 and<br>EC53 | Complied with IEC 60601-1 and<br>EC53 | SE |
| Biocompatibility | Complied with ISO 10993-1, ISO<br>10993-5, ISO 10993-10 | Complied with ISO 10993-1, ISO<br>10993-5, ISO 10993-10 | SE |
### 8. Comparison to the Predicate Device
{6}------------------------------------------------
## Tedli ket
{7}------------------------------------------------
## Shenzhen Med-link Electronics Tech Co., Ltd.
Justifications for differences between proposed device and the predicate device are shown as below:
Different (1): The proposed device is disposable and the predicate device is reusable. The proposed device was tested according to IEC 60601-1 and EC53. Therefore, this difference does not raise new questions of safety and effectiveness for subject device as compared to the predicate device.
Different (2): The patient end termination is different. This end is intended to connect the electrodes. The proposed device was tested according to IEC 60601-1 and EC53. Therefore, this difference does not raise new questions of safety and effectiveness for subject device as compared to the predicate device.
Different (3): The material is different. The proposed device was tested according to ISO 10993-10. Therefore, this difference does not raise new questions of safety and effectiveness for subject device as compared to the predicate device.
## 9. Non-clinical Test
Compare to predicate product specified in K120010, our device and the predicate device are same in Essential Components, raw materials, physical features, and same manufacturing processes. The biocompatibility performance equivalence evidence of proposed electrode can be demonstrated.
{8}------------------------------------------------
# MedLinket
The safety performances of subject device are demonstrated by the Third party, CTI through testing following IEC 60601-1 General requirements for basic safety and essential performance and ANSI AAMI EC53:2013 ECG Trunk Cables And Patient Leadwires.
The results of bench testing & type testing provide reasonable assurance that the proposed device has been designed and validated to assure conformance to the requirements for its indication for use.
Package integrity and functional performance testing were completed on the subject device following real time aging test to support the proposed shelf life.
### 11. Conclusion
Based on the comparison and analysis in this submission, it can be concluded that: Disposable ECG Cable and Leadwires is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.