CABLE/ LEAD-WIRE (ECG, EKG, SPO2 AND INVASIVE BLOOD PRESSURE)
Applicant
Shenzhen Med-Link Electronics Tech Co., Ltd.
Product Code
DSA · Cardiovascular
Decision Date
Jan 19, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2900
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Shenzhen Med-link Cable / lead-wire are intended to be used with ECG, EKG, SpO2 and Invasive Blood Pressure monitoring devices. The Cable / lead-wire are used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for general monitoring and/or diagnostic evaluation by heath care professional.
Device Story
Passive cable/lead-wire assemblies; transfer electrical signals from patient-attached electrodes, catheters, or sensors to monitoring devices. Components include shielded/unshielded copper conductors, PVC/TPU jackets, and molded PVC/ABS connectors. Used in clinical settings by healthcare professionals. No active processing or software; serves as physical interface between patient and monitor. Benefits include enabling signal acquisition for diagnostic evaluation and patient monitoring.
Clinical Evidence
No clinical data performed. Safety and effectiveness supported by bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10) and electrical safety/performance testing (IEC 60601-1, ANSI/AAMI EC53).
Technological Characteristics
Passive cable assemblies; materials: medical grade PVC/TPU jackets, PVC/ABS over-molded connectors. Standards: ISO 10993-5/10 (biocompatibility), IEC 60601-1 (electrical safety), ANSI/AAMI EC53 (ECG cables/leadwires), DIN 42-802. Non-sterile, reusable.
Indications for Use
Indicated for patients requiring ECG, EKG, SpO2, or Invasive Blood Pressure monitoring. Used to connect sensors/electrodes to monitoring devices for general monitoring or diagnostic evaluation by healthcare professionals.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{1}------------------------------------------------
## K120010 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR§807.92.
| Submission Date: | 20 March 2011 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Shenzhen Med-link Electronics Tech Co., Ltd.<br>Building 2, HuaFu Ind. Park, HuaWang Road, LongHua, BaoAn ,<br>District 518109 Shenzhen City, P.R.China |
| Submitter Contact: | Maolin Ye<br>General Manager<br>Phone Number: 011 86 (755) 61568827<br>Fax Number: 011 86 (755) 61120755<br>E-mail Address: User18@med-linket.com |
| Manufacturing Site: | Shenzhen Med-link Electronics Tech Co., Ltd.<br>Build 2, HuaFu Ind. Park, HuaWang Road, LongHua, BaoAn ,<br>District 518109 Shenzhen City, P.R.China |
| Official Contact: | Eric Zhan<br>Shenzhen Med-link Electronics Tech Co., Ltd.<br>Build 2, HuaFu Ind. Park, HuaWang Road, LongHua, BaoAn ,<br>District 518109 Shenzhen City, P.R.China<br>Phone Number: 011 86 (755) 61120299 ext 815<br>E-mail Address: User22@med-linket.com |
| Trade Name: | Cable / lead-wire (ECG, EKG, SpO2 and Invasive Blood<br>Pressure), models as below table: |
| | Device<br>Name / Type |
| | Device Model |
| | ECG cable and<br>lead-wires |
| | EA003S5A, EA010C5A, EA007C5A, EA050C5A,<br>EA023C5A, EA035C3A, EA066S3A, EA009S5A,<br>EC021-5AI, EC040S5A, EE007-5AI, EE051C5A,<br>ED023-3AI, ED040C3A, ED040C5A, ED023-5AI,<br>EP055-6AI, EP035S3A, EL009-5AI, EL040S5A,<br>EC072M5I, EC072M5A, EC072S5I, EC072S5A |
| | EKG cable and<br>lead-wires |
| | VA008BBA, VA018BCA, VA005BBA, VA001BNA,<br>VA021BBA, VA025BAA, VE008-BAI, VE008BNA, |
| | Invasive Blood<br>Pressure cables |
| | X0014A, X0014B, X0015A, X0015B, X0018D,<br>X0018F, |
| | SpO2<br>adapter cables |
| | S0052BC-L, S0036BC-L, S0014OH-L, S0099BC-L,<br>S0002NE-L, S0020CO-L, S0125CO-L, S0010NE-L,<br>S0026OX-L, S0117OX-L, S0005NI-L, S0015OX-L,<br>S0003OX-L, S0072BC-L, S0004NE-L |
Common Name:
Cable / lead-wire
Classification Name:
Patient Transducer and electrode cable (including connector)
{2}------------------------------------------------
| Classification<br>Regulation: | 21 CFR 870.2900 | | |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|----------------------------------|
| Product Code: | DSA | | |
| Substantially<br>Equivalent<br>Devices: | Shenzhen Med-link<br>Product Group | Predicate<br>510(k) Number | Predicate<br>Manufacturer |
| | Cable / Lead-wire | K082959 | Unimed Medical<br>Supplies Inc. |
| | Cable / Lead-wire | K992524 | Advantage Medical<br>Cables, Inc |
| Device Description: | Med-link Cable / Lead-wire with specific various lengths are the<br>replacements for similar cables manufactured by Original<br>Equipment Manufacturers (OEM) and other third party after<br>market manufacturers for their respective monitors.<br>These cables consist of connectors on each cable end and a<br>shielded bulk cable. The cables are used to transfer the signals<br>from the electrodes to the patient monitor.<br>The Med-link cables use the same type of constructions and have<br>the same technological characteristics as the predicate devices.<br>They use a medical grade PVC and TPU cable jacket with medical<br>grade PVC and ABS over molded connectors with integral relief. | | |
| Intended Use: | Shenzhen Med-link Cable / lead-wire are intended to be used<br>with ECG, EKG, SpO2 and Invasive Blood Pressure monitoring<br>devices. The Cable / lead-wire are used to connect electrodes,<br>catheters, and/or sensors placed at appropriate sites on the<br>patient to a monitoring device for general monitoring and/or<br>diagnostic evaluation by heath care professional. | | |
| Technology<br>Comparison: | Med-Link Cable / Lead-wire employ the same technological<br>characteristics as the predicate device to determine the signals<br>between the patient and monitoring Device.<br>This method is characteristic of all Cable / Lead-wire that is the<br>subject of this submission as well as the predicate device. | | |
| Performance Testing: | | | |
| Biocompatibility<br>Testing | Patient contact with materials used in Med-Link Cable<br>/ lead-wire were tested in accordance with<br>ISO 10993-5:1999 Biological evaluation of medical devices -<br>Part 5: Tests for <i>in vitro</i> cytotoxicity and ISO<br>10993-10:2002/Amd1:2006 Biological evaluation of medical<br>devices - Part 10: Tests for irritation and delayed-type<br>hypersensitivity - Amendment 1.<br>Test results indicated that the patient contact material were | | |
{3}------------------------------------------------
non-toxic, non-sensitizing and non-irritating. Med-Link Cable / Lead-wire tested in accordance with Electrical Safety applicable clause of IEC 60601-1:1998; Aml; A2:1995, Medical electrical equipment-Part 1: General requirements for safety. ANSI/AAMI EC53:1995/(R) 2001 ECG cables and leadwires and FDA 21CFR Part 898 Final rule. Med-Link Cable / Lead-wire were tested in accordance Lead-wire to with ANIS/AAMI EC53:1995 5/(R) 2001 ECG cables and trunk cable lead wires to ensure that the cable assemblies where the interconnection patient lead wires are separable from the trunk cable with a single unshielded conductor lead wire, the interconnection between the patient lead-wires and the trunk cable whether meet the requirements of DIN 42-802. Test results indicated that the cable / lead-wires comply with the applicable clause of the Standard. Clinical performance testing is not required and was not Performance performed to demonstrate safety and effectiveness of the Testing - Clinical Shenzhen Med-link cable / lead-wire. Not Applicable: The Shenzhen Med-link cable / lead-wire are Shelf Life Testing intended to be reusable. Performance Bench testing was conducted on the Shenzhen Med-link cable Testing - Bench / lead-wire according with established protocols and test results confirm that the final product met the requirements for the safety and performance standards and its intended use. Performance Animal performance testing is not required and was not Testing -Animal performed to demonstrate safety and effectiveness of the Shenzhen Med-link cable / lead-wire.
## Comparison to Predicate Device:
The results of Comparison to Predicate Device as following:
Table 1—Comparison of cable / lead-wires has the same or similar technological characteristics as the predicate device. Table 2—Comparison of Med-link cable / lead-wire has different technological characteristics from the predicate device.
{4}------------------------------------------------
Table 1--- Comparison of cable / lead-wires has the same or similar technological characteristics as the predicate device.
| Comparison Item | Shenzhen Med-link | Advantage medical | Unimed Medical |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | Shenzhen Med-link Cable / lead-wire are intended to be used with ECG, EKG, SpO2 and Invasive Blood Pressure monitoring devices. The Cable / lead-wire are used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for general monitoring and/or diagnostic evaluation by heath care professional. | These patient cables and leadwires are intended to be used with ECG, EKG, SpO2 and Blood Pressure monitoring devices. | The Unimed patient cables and lead wires are intended to be used with ECG, EKG, SpO2 and BP monitoring devices. The patient cables and lead wire are used to connect electrodes, catheters, and / or sensors placed at appropriate sites on the patient to a monitoring device for general monitoring and/or diagnostic evaluation by heath care professional. |
| Patent Usage | Reusable | Reusable | Reusable |
| Anatomical Sites | The ECG and EKG cable attached to sensors places at standard specified locations on the chest wall. | The ECG and EKG cable attached to sensors places at standard specified locations on the chest wall. | The ECG and EKG cable attached to sensors places at standard specified locations on the chest wall. |
| Design / Appearance | ECG and EKG Cables with various connectors (monitor, trunk / lead wire, electrode grabber & snapper)<br>Invasive Blood Pressure cables with various connectors (Yoke type: Utah 4pin, Instrument connector: Spacelabs 6 Pin etc.)<br>SpO2 adapter cables with various connectors for connect for the OEM Instrument connector of SpO2 monitor and SpO2 probe. (DB9F, Din 8Pin and Philips etc.) | ECG and EKG Cables with various connectors (monitor, trunk / lead wire, electrode grabber & snapper)<br>Invasive Blood Pressure cables with various connectors (Yoke type: Utah 4pin, Instrument connector: Spacelabs 6 Pin etc.)<br>SpO2 adapter cables with various connectors for connect for the OEM Instrument connector of SpO2 monitor and SpO2 probe. (DB9F, Din 8Pin and Philips round 8Pin etc.) | ECG and EKG Cables with various connectors (monitor, trunk / lead wire, electrode grabber & snapper)<br>Invasive Blood Pressure cables with various connectors (Yoke type: Utah 4pin, Instrument connector: Spacelabs 6 Pin etc.)<br>SpO2 adapter cables with various connectors for connect for the OEM Instrument connector of SpO2 monitor and SpO2 probe. (DB9F, Din 8Pin and Philips round 8Pin etc.) |
| | round 8Pin etc.) | | |
| Cable length | Various specified<br>standard lengths | Various specified<br>standard lengths | Various specified<br>standard lengths |
| Wire<br>material | Shielded &<br>Unshielded Copper<br>with PVC or TPU<br>Jacket | Shielded & Unshielded<br>Copper with PVC Jacket | Shielded & Unshielded<br>Copper with PVC Jacket |
| Sterility | Non sterile | Non sterile | Non sterile |
| Connector<br>Retention<br>Force | ANSI / AAMI EC53:<br>1995 /(R) 2001 | ANIS/AAMI EC53-1995 | ANIS / AAMI<br>EC53A-1998(R)2001<br>EC53A-1998<br>(Amendment) |
| Electrical<br>Performance<br>and Safety | ANSI / AAMI EC53:<br>1995 /(R) 2001 and<br>IEC 60601-1:1998;<br>Am1; A2:1995 | ANSI / AAMI<br>EC53-1995 and IEC<br>60601-1:1998;Am1;<br>A2:1995 | ANSI / AAMI<br>EC53A-1998(R) 2001<br>and IEC 60601-1:<br>1998; Am1; A2:1995 |
{5}------------------------------------------------
Table 2-Comparison of Med-link cable / lead-wire has different technological characteristics from the predicate device.
| Item | Med-link<br>models | Predicate<br>device Model | | Contrast results | | |
|-----------------|--------------------|---------------------------|-------------------|------------------|---------|-------------------|
| | | AMC | Unimed<br>Medical | Med-link | AMC | Unimed<br>Medical |
| Cable<br>length | EA003S5A | CB-72500R/90 | 2540S | L:3.6m | L:3.0m | L:3.4m |
| | EC021-5AI | CB-82516R | / | L:2.7m | L:3.0m | / |
| | EC040S5A | LW-26000MX/5A | D5-90S | L:1.0m | L:0.74m | L:0.9m |
| | EC072M5A | W-26000MX/5A | D5-90S | L:0.72m | L:0.74m | L:0.9m |
| | VA008BBA | / | E10-MQ12-B | L:3.06m | / | L:3.7m |
| | VE008-BAI | CB-5110006 | / | L:2.2m | L:3.0m | / |
| | X0015B | CB-91068 | BC-6P-UT | L:3.6m | L:3.0m | L: 4.0m |
| | S0052BC-L | CB-A400-1004A | / | L:2.4m | L:3.0m | / |
| | S00150X-L | CB-A400-1006VM | U708-40 | L:2.4m | L:2.4m | L:2.2m |
## Conclusion:
- Shenzhen Med-link design is equivalent to predicate devices cleared to market under . 510(k) K082959 and K992524.
- Electrical safety and bench Testing confirms that Subject device performs as intended. .
- The proposed device does not raise new issues of safety and effectiveness. .
- The proposed device is deemed safe and effective for its intended use. .
- The bench testing can be fully proved that the different wire length of subject device . detailed as table 2 which have not produce new issue to affect the safety and effectiveness.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 1 9 2012
Shenzhen Med-Link Electronics Tech., Co., Ltd. c/o Ms. Paula Wilkerson Responsible Third Party Official Intertek Testing Services NA 2307 E. Aurora Rd., Unit B7 Twinsburg, OH 44087
K120010 Re:
Trade/Device Name: Cable/Lead-Wire (ECG, EKG, SpO2 and Invasive Blood Pressure) Regulatory Number: 21 CFR 870.2900 Regulation Name: Patient transducer and electrode cable (including connector) Regulatory Class: II (two) Product Code: DSA Dated: December 20, 2011 Received: January 3, 2012
Dear Ms. Wilkerson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use sualed in the encreated of the enactment date of the Medical Device Amendments, or to conninered prior to 1118) 2017-11-11 in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, merelore, mance of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Ms. Paula Wilkerson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
D. Zadeh, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known):
K 120010
Device Name:
Cable / lead-wire (ECG, EKG, SpO2 and Invasive Blood Pressure), models
as below table:
| Device<br>Name / Type | Device Model |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ECG cable and<br>lead-wires | EA003S5A, EA010C5A, EA007C5A, EA050C5A,<br>EA023C5A, EA035C3A, EA066S3A, EA009S5A,<br>EC021-5AI, EC040S5A, EE007-5AI, EE051C5A,<br>ED023-3AI, ED040C3A, ED040C5A, ED023-5AI,<br>EP055-6AI, EP035S3A, EL009-5AI, EL040S5A,<br>EC072M5I, EC072M5A, EC072S5I, EC072S5A |
| EKG cable and<br>lead-wires | VA008BBA, VA018BCA, VA005BBA, VA001BNA,<br>VA021BBA, VA025BAA, VE008-BAI, VE008BNA, |
| Invasive Blood<br>Pressure cables | X0014A, X0014B, X0015A, X0015B, X0018D,<br>X0018F, |
| SpO2<br>adapter cables | S0052BC-L, S0036BC-L, S0014OH-L, S0099BC-L,<br>S0002NE-L, S0020CO-L, S0125CO-L, S0010NE-L,<br>S0026OX-L, S0117OX-L, S0005NI-L, S0015OX-L,<br>S0003OX-L, S0072BC-L, S0004NE-L |
Indications of Use:
Shenzhen Med-link Cable / lead-wire are intended to be used with ECG, EKG, SpO2 and Invasive Blood Pressure monitoring devices. The Cable / lead-wire are used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for general monitoring and/or diagnostic evaluation by heath care professional.
Prescription Use __ V__
AND/OR
Over-The-Counter Use_
(Part 21 CFR 801 Subparts D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices.
| 510(k) Number | K 120010 |
|---------------|----------|
|---------------|----------|
Page 1 of 1Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.