K181089 · Authentic Medical · DSA · Jan 17, 2020 · Cardiovascular
Device Facts
Record ID
K181089
Device Name
SureLead Cable System
Applicant
Authentic Medical
Product Code
DSA · Cardiovascular
Decision Date
Jan 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2900
Device Class
Class 2
Indications for Use
The SureLead Disposable Cable System is intended to be used with ECG monitoring devices. The SureLead Disposable Cable System is used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for generating monitoring and/or diagnostic evaluation by a health care professional.
Device Story
SureLead Disposable Cable System consists of shielded bulk cables with connectors; serves as replacement lead-wires for ECG monitors. Inputs: electrical signals from patient-attached electrodes/sensors. Transformation: passive signal transmission from patient to monitoring device. Output: raw ECG signal for monitor display/analysis. Used in clinical settings by healthcare professionals. Benefits: enables signal acquisition for diagnostic monitoring. Device is disposable; non-sterile; features integral strain relief.
Clinical Evidence
Bench testing only. Compliance verified against ISO 10993-1, ISO 10993-5, ISO 10993-10, AAMI/ANSI EC53:2013, and IEC 60601-1:2005 (including amendments/corrections). No clinical data provided.
Technological Characteristics
Materials: medical grade PVC and PP cable jackets; medical grade PVC and PP over-molded connectors. Sensing: passive electrical signal transmission. Connectivity: wired connection between electrodes and ECG monitor. Standards: ISO 10993, AAMI/ANSI EC53, IEC 60601-1. Configuration: 3, 5, or 6 lead-wire sets; disposable.
Indications for Use
Indicated for use with ECG monitoring devices to connect electrodes, catheters, or sensors to a monitor for diagnostic evaluation by a healthcare professional. No specific age or gender restrictions stated.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2020
Authentic Medical Ricky Souza CEO 4470 Yankee Hill Rd., Ste 100 Rocklin, California 95677
Re: K181089
Trade/Device Name: SureLead Disposable Cable System Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer and Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: December 10, 2019 Received: December 11, 2019
Dear Ricky Souza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181089
Device Name SureLead Disposable Cable System
#### Indications for Use (Describe)
The SureLead Disposable Cable System is intended to be used with ECG monitoring devices. The SureLead Disposable Cable System is used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for generating monitoring and/or diagnostic evaluation by a health care professional.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
SureLead Disposable Cable System Traditional 510(k)
## SureLead Disposable Cable System Premarket Notification 510(k) Summary
| DATE PREPARED | 19 APR 2018 |
|--------------------------------------|------------------------------------------------------------------------------|
| MANUFACTURER | Authentic Medical |
| CONTACT PERSON | Ricky Souza |
| | Position: CEO/President |
| | Tel: (916) 952- 6498 |
| | Fax: (916) 644-6015 |
| | Email: rsouza@authenticmed.com |
| PANEL CODE | Cardiovascular |
| CLASSIFICATION | Patient Transducer and Electrode Cable (Including<br>Connector) |
| CLASS | Class II (Class 2) |
| COMMON NAME | Cable / Lead-wire |
| TRADE NAME | SureLead Disposable Cable System |
| PREDICATE DEVICES | K170536 Cable/Lead-Wire (ECG). |
| Identification of<br>Proposed Device | Trade Name: Surelead Disposable cable system<br>Common Name: cable/lead-wire |
### DEVICE DESCRIPTION
The SureLead Disposable Cable System with specific various lengths are the replacements for similar cables manufactured by Original Equipment Manufacturers (OEM) and other third-party after-market manufacturers for their respective monitors. These cables consist of connectors on each cable end and a shielded bulk cable. These cables are used to transfer the signals from the electrodes to the patient monitor. The SureLead Disposable Cable System uses the same type of construction and have the same technological characteristics as the predicate device. The SureLead Disposable Cable System is made of medical grade PVC and PP cable jacket with medical grade PVC and PP over molded connectors with integral relief.
## INDICATIONS FOR USE
The SureLead Disposable Cable System is intended to be used with ECG monitoring devices. The SureLead Disposable Cable System is used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for generating monitoring and/or diagnostic evaluation by a health care professional.
### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The SureLead Disposable Cable System, is substantially equivalent to the Predicate device (K170536) intended use / indications for use, materials, technological characteristics, and labelling.
{4}------------------------------------------------
## PERFORMANCE DATA
Non clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device (K170536). The test results demonstrated that the proposed device complies with the following standards:
- . ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process
- . ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests For In Vitro Cytotoxicity.
- . ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization.
- AAMI/ANSI EC53: 2013 ECG Trunk Cables and Patient Leadwires .
- IEC 60601-1:2005+CORR.1: 2006+CORR. 2:2007+AM1: 2012, Medical Electrical . Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
# SUBSTANTIAL EQUIVALENCE DISCUSSION
The SureLead Disposable Cable System, (Subject Device) is substantially equivalent to K 170536 ECG Disposable Cable / lead-wire (Predicate Device).
The Subject device and Predicate Device:
- Have nearly Identical Indications for Use
- Are provided non-sterile with steam sterilization instructions
- · Use similar cable lengths and the same materials
Table 1 is provided here to show the comparison.
| | Proposed Device | Predicate Device<br>K170536 | Remarks |
|-------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|---------|
| Manufacturer | Authentic Medical | APK Technology<br>Co., Ltd | - |
| 510(k) number | K181089 | K170536<br>Cable/Lead-wire | - |
| Classification<br>Regulation | 21 CFR Part 870.2900 | 21 CFR Part 870.2900 | SE |
| Panel | Cardiovascular | Cardiovascular | SE |
| Common Name | Cable / Lead-wire | Cable / Lead-wire | SE |
| Device<br>Classification Name | Patient transducer<br>and electrode cable<br>(including connector) | Patient transducer<br>and electrode cable<br>(including connector) | SE |
| Product Code | DSA | DSA | SE |
| Class | Class II (Class 2) | Class II (Class 2) | SE |
Table 1: Substantial Equivalence Summary Table
{5}------------------------------------------------
| Indications for Use | The SureLead | The APK ECG | SE |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Disposable Cable<br>System is intended to<br>be used with ECG<br>monitoring devices.<br>The SureLead Cable<br>System is used to<br>connect electrodes,<br>catheters, and/or<br>sensors placed at<br>appropriate sites on the<br>patient to a monitoring<br>device for generating,<br>monitoring, and/or<br>diagnostic evaluation<br>by a health care<br>professional. | Disposable Lead Wires<br>are intended to be used<br>with ECG. The lead<br>wire is used to connect<br>electrodes placed at<br>appropriate sites on the<br>patient to ECG for<br>general monitoring<br>and/or diagnostic<br>evaluation by health<br>care professional. | |
| Sterility and Shelf-life | Provided non-sterile.<br>No shelf-life | Provided non-sterile.<br>No shelf-life | SE |
| Single Patient Use | Yes | Yes | SE |
| Wire Materials | Shielded&Unshielded<br>Copper with PVC or<br>PP Jacket | Polyvinyl Chloride<br>(PVC) | Analysis 3 |
| Connectors | Medical grade<br>PVC and PP over<br>molded connectors<br>with integral relief. | Medical grade PVC<br>and ABS over<br>molded connectors<br>with integral relief. | Analysis 3 |
| Cable Length<br>(Minimum) | 10 Centimeters | 0.72 Meters | Analysis 2 |
| Cable Length<br>(Maximum) | 11 Foot (3.35 Meters) | 3.6 Meters | Analysis 2 |
| Patient end<br>termination type | Clip | Clip , snap | Analysis 4 |
| Number of leadwires | 3,5,6 | Unknown | Analysis 5 |
| Disposable or<br>reusable | Disposable | Disposable | SE |
The Subject Device and the Predicate Device are substantially equivalent. The minor differences are:
Analysis (2) The minimum SureLead Cable System (Subject Device) has a shorter minimum cable length and a shorter maximum cable length. These differences are negligible.
Analysis (3) The material of proposed device is different from predicate device. However, the biocompatibility for proposed device has been evaluated and the test result conform with requirements of ISO 10993 standards. Therefore, this difference does not affect substantially equivalence.
{6}------------------------------------------------
Analysis (4) The patient end termination type of proposed device is Clip while that of predicated device is Clip or snap, yet the electric performance and safety for proposed device has been tested per IEC 60601-1 and EC 53 and the test result is acceptable. Therefore, this difference does not affect substantially equivalence.
Analysis (5) The subject device is available in 3, 5 ,6 leadwire type, while the leadwire number for predicate device is unknown and the substantially equivalence cannot be determined. However, the electric performance and safety for proposed device has been tested per IEC 60601-1 and EC 53 and the test result is acceptable. Therefore, this difference does not affect substantially equivalence.
## CONCLUSION
The Intended Use and Indications for Use of the SureLead Disposable Cable System (subject device) and the Predicate Device, Cable / Lead-wire (Premarket Notification K170536) are substantially equivalent. The technological characteristics, components and materials used for the Predicate Device and the Subject Device are the same. The Subject Device, SureLead Disposable Cable System is substantially equivalent to the Predicate Device, Cable / Lead-wire (Premarket Notification K170536), a legally marketed device.