Browse hierarchy Orthopedic (OR) Subpart B — Diagnostic Devices 21 CFR 888.1100 Product Code HRX K900070 — MODIFIED USES OF THE ARTHROSCOPIC SURGICAL SYSTEM
MODIFIED USES OF THE ARTHROSCOPIC SURGICAL SYSTEM
K900070 · Dyonics, Inc. · HRX · Feb 16, 1990 · Orthopedic
Device Facts
Record ID K900070
Device Name MODIFIED USES OF THE ARTHROSCOPIC SURGICAL SYSTEM
Applicant Dyonics, Inc.
Product Code HRX · Orthopedic
Decision Date Feb 16, 1990
Decision SESE
Submission Type Traditional
Regulation 21 CFR 888.1100
Device Class Class 2
Regulatory Classification
Identification An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Submission Summary (Full Text)
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
FEB 16 1990
Food and Drug Administration
1390 Piccard Drive
Rockville, MD 20850
Re: K900070
Mr. Eric Bannon
Regulatory Affairs Manager
Dyonics, Inc.
160 Dascomb Road
Andover, Massachusetts 01810
Modified Uses of the
Arthroscopic Surgical
System
Regulatory Class: II
Dated: December 28, 1989
Received: January 4, 1990
Dear Mr. Bannon:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Performance Standards) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under the Radiation Control for Health and Safety Act of 1968, or other Federal Laws or Regulations.
This letter immediately will allow you to begin marketing your device as described. An FDA finding of substantial equivalence of your device to a pre-amendments device results in a classification for your device and permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in anyway represent your device or its labeling as being approved by FDA. If you desire specific advice on the labeling for your device please contact the Division of Compliance Operations, Regulatory Guidance Branch (HFZ-323) at (301) 427-1116. Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Carl A. Larson
Carl A. Larson, Ph.D.
Director,
Division of Surgical
and Rehabilitation Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
BEST AVAILABLE COPY
{2}
[LOGO]
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
# Memorandum
From REVIEWER(S) - NAME(S)
Subject 510(k) NOTIFICATION
J. Parkhurst
K900070
To THE RECORD
It is my recommendation that the subject 510(k) Notification:
☑ (A) Is substantially equivalent to marketed devices.
☐ (B) Requires premarket approval. NOT substantially equivalent to marketed devices.
☐ (C) Requires more data.
☐ (D) Other (e.g., exempt by regulation, not a device, duplicate, etc.)
Additional Comments:
See Memo Record.
The submitter requests under 21 CFR §807.95:
☐ No Confidentiality
☐ Confidentiality for 90 days
☑ Continued Confidentiality exceeding 90 days
Predicate Product Code w/Panel and class:
87-HRX (Class II)
Additional Product Code(s) w/Panel (optional):
REVIEW: Daniel S. McLaughlin 2/7/90
(BRANCH CHIEF) (DATE)
FINAL REVIEW: Thomas J. Callahan 2/15/90
(DIVISION DIRECTOR) (DATE)
BEST AVAILABLE COPY
X
2
{3}
# 510(k) "Substantial Equivalence" Decision-Making Process (Detailed)

BEST AVAILABLE COPY
* 510(k) Submissions Compare New Devices to Marketed Devices. FDA Requests Additional Information if the Relationship Between Marketed and "Predicate" (Pre-Amendments or Reclassified Post-Amendments) Devices is Unclear.
** This Decision is Normally Based on Descriptive Information Alone, But Limited Testing Information is Sometimes Required.
*** Data May Be in the 510(k), Other 510(k)s, The Center's Classification Files, or the Literature.
{4}
Vap 1 of 3
| MEMO RECORD | AVOID ERRORS PUT IT IN WRITING | DATE 2/7/90 |
| --- | --- | --- |
| FROM: Biologist | | OFFICE ODE |
| TO: File | | DIVISION BSRD-RDB |
| SUBJECT: K90 0070 - Modified Use of Anthroscopic Surgical System (Dyonics) | | |
| SUMMARY In this submission, Dyonics' intention is to modify the indications for use of these Anthroscopic Surgical System for use in the decompression of bulging discs in the lumbar region of the spine (L5-S1). The name of this modified system is the 8 precuneous Anthroscopic Micro Discectomy System (PAMDS). The PAMDS consists of a cannulated introduction set with hard instruments, video microscope set and a sterile disposable set. The cannulated intro-set/hard instruments consists of a 18 G spiral needle (ASTM 304 - stainless steel) with stylette, .031" guide wire, cannulated tapered obturator (ASTM 316 - stainless steel), 3mm and 5mm trephine (ASTM 304 - stainless steel), Cup forceps (ASTM 304 - stainless steel), Dyo-Vac suction punch, magnetic retriever (304 stainless steel), forceps detector tube and Cannula insertion stop (304 stainless steel). The video microscope set allows visualization of the disc area for procedural assessment and consists of a 2.7mm irrigation sheath and a video microscope (2.7mm x 30") ASTM 304 stainless steel. The sterile disposable set consists of a 4.5mm full radius anthrospy blade (ASTM 304L Stainless Steel). | | |
| SIGNATURE | BEST AVAILABLE COPY | DOCUMENT NO. 4 |
FORM FDA 2034 (12/80)
{5}
Page 2 of 3
| MEMO RECORD | AVOID ERRORS PUT IT IN WRITING | DATE 2/7/98 |
| --- | --- | --- |
| FROM: | | OFFICE |
| TO: | | DIVISION |
SUBJECT:
SUMMARY
a 4.5mm trimmer arthroscopy blade (ASTM 304L stainless steel),
and a tissue trap.
After I reviewed the labeling/instructions for use
of the PAMDS some questions were raised. I spoke
by phone with Eric Bannon (Reg. Affairs Mgr.) at Dyonice
to clarify my concerns and ~~it~~ obtained the
following information: there is only one "incision"
made on the patient when using the PAMDS (not two
"incisions" as initially thought); the arthroscope
(viderscope) is used to periodically ~~to view~~
the state of the procedure (in conjunction ~~with~~ with the
use of roentgenograph +MRI).
I compared the PAMDS to the Nucleotone
by Surgical Dynamics (K844131) which was found
SE on Nov 29, 1984 and to the Kambin Spinal
Instrument Set (K853678) found SE on Sept. 24, 1985.
I compared the introduction set/hand instruments
and the disposable set of the PAMDS to the surgical
SIGNATURE
BEST AVAILABLE COPY
DOCUMENT NO.
5
FORM FDA 2034 (12/80)
{6}
page 5 of 5
| MEMO RECORD | AVOID ERRORS PUT IT IN WRITING | DATE 2/7/80 |
| --- | --- | --- |
| FROM: | | OFFICE |
| TO: | | DIVISION |
SUBJECT:
SUMMARY
instruments of the predicate devices. The design + materials used in the PAMDS instruments raise no new safety + effectiveness concerns when compared to predicate devices.
In addition, I compared the video-decroscope used in the PAMDS to the Dyonics Arthroscope (K880150) found SE on March 4, 1988. Also the trimmer + full radius blades used in the PAMDS were compared to Dyonics Disposable Arthroscopy blades (K833587) found SE on Nov 14, 1983. The design + materials used to manufacture the blades are the same as predicate devices. The microscope used in the PAMDS is similar to predicate devices + raise no new safety + effectiveness concerns.
Based on the above findings, I find the Dyonics PAMDS to be SE to predicate devices.
BEST AVAILABLE COPY
SIGNATURE
J. Parkhurst
DOCUMENT NO.
K90 0070 6
FORM FDA 2034 (12/80)
{7}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Public Health Service
Food and Drug Administration
Center for Devices and
Radiological Health
Office of Device Evaluation
Document Mail Center (HFE-401)
1390 Piccard Drive
Rockville, Maryland 20850
JANUARY 8, 1990
DYONICS, INC.
ATTN: ERIC BANNON
160 DASCOMB ROAD
ANDOVER, MA 01810
D.C. Number : K900070
Received : 01-04-90
90th Day : 04-04-90
Product : MODIFIED USES OF
THE ARTHROSCOPIC
SURGICAL SYSTEM
The Premarket Notification you have submitted as required under Section 510(k) of the Federal Food, Drug, and Cosmetic Act for the above referenced device has been received and assigned an unique document control number (D.C. Number above). Please cite this D.C. Number in any future correspondence that relates to this submission.
We will notify you when the processing of this submission has been completed or if any additional information is required. You are required to wait ninety (90) days after the received date shown above or until receipt of a "substantially equivalent" letter before placing the product into commercial distribution. We intend to complete our review expeditiously and within ninety days. Occasionally, however, a submitter will not receive a final decision or a request for additional information until after ninety days has elapsed. Be aware that FDA is able to continue the review of a submission beyond the ninety day period and might conclude that the device is not substantially equivalent. A "not substantially equivalent" device may not be in commercial distribution without an approved premarket approval application or reclassification of the device. We, therefore, recommend that you not market this device before FDA has made a final decision. Thus, if you have not received a decision within ninety days, it would be prudent to check with FDA to determine the status of your submission.
All correspondence concerning your submission MUST be sent to the Document Mail Center at the above address. Correspondence sent to any address other than the one above will not be considered as part of your official premarket notification application. Telefax material will not be accepted nor considered as part of your official premarket notification application, unless specifically requested of you by an FDA official.
If you have procedural or policy questions, please contact the Division of Small Manufacturers Assistance at (301) 443-6597 or their toll-free number (800) 638-2041, or contact me at (301) 427-1190.
Sincerely yours,
Robert I. Chissler
Premarket Notification Coordinator
Office of Device Evaluation
Center for Devices and
Radiological Health
BEST AVAILABLE COPY
7
{8}
DYONICS®
K900070
Dyonics, Inc., 160 Dascomb Road, Andover, Massachusetts 01810 (508) 470-2800 TWX 710-347-0337 FAX (508) 470-2193
FDA-COM-DOC
JAN - 4 1990
December 28, 1989
FOOD AND DRUG ADMINISTRATION
Center for Devices and Radiological Health
Office of Device Evaluation
Document Mail Center (HFZ-401)
1390 Piccard Drive
Rockville, MD 20850
Re: 510(K) Premarket Notification
Dear Sir/Madam,
As required by the Federal Food Drug and Cosmetic Act, Dyonics Incorporated hereby notifies you of our intent to modify the indications for use of our Arthroscopic Surgical System for use in the decompression of bulging discs in the lumbar region of the spine. The original system was reviewed by FDA under 510(K) submission K771218A with subsequent modifications reviewed under K8202367, K833587 and K880150. The name of the modified system is the Percutaneous Arthroscopic Micro Discectomy System.
The modified system consists of a cannulated introduction set, with hand instruments, video discoscope set and a sterile disposable set. The components of each set with a description and comparison to similar products already marketed is contained in Attachment A with pertinent literature.
Drawings of each component are contained in Attachment B.
Attachment C contains the final proposed classifications for each component.
BEST AVAILABLE COPY
A MEMBER OF THE SMITH AND NEPHEW GROUP
8
{9}
Page 2
510(K) Premarket Notification
Attachment D is the proposed labeling for the set. This includes the following:
1. OUTER BOX LABEL:
To be placed on all non-sterile products. The catalog number and description will change to reflect each component.
2. INSTRUCTIONS/PROCEDURE:
Including indications, contradictions and procedural details.
3. STERILE BLADE LABEL:
For each blade.
The Percutaneous Arthroscopic Micro Discectomy System proposed by Dyonics is similar to one currently marketed by Surgical Dynamics under the name Nucleotome (R); Automated Percutaneous Lumbar Discectomy. A copy of pertinent literature is attached as Attachment E. Attachment F is a collection of pertinent articles describing the surgical procedure intended for this product. Dyonics registration number is 1216828. If there are any questions please call me at, (508) 470-2800, Ext. 348.
Sincerely,
Eric Bannon
Eric Bannon
Regulatory Affairs Manager
EB/sh
89/05
BEST AVAILABLE COPY
9
{10}
Attachment A: COMPONENT DESCRIPTIONS AND COMPARISONS TO SIMILAR PRODUCTS.
BEST AVAILABLE COPY
66
{11}
Attachment A: PERCUTANEOUS ARTHROSCOPIC MICRO DISCECTOMY
A. CANNULATED INTRODUCTION SET/HAND INSTRUMENTS: Instruments designed to provide access to the appropriate lumbar region for tissue removal. Contains the following components:
1. Spinal Needle: 18 gauge, 6" long with removeable stylette. This is similar to those contained in the Kambin spinal instrument set, marketed by Pilling and readily available from Popper and Sons, Inc. A copy of pertinent literature is attached. We intend to purchase this from Popper and Sons, Inc.
2. Guide Wire: .031" approximately 25cm in length. This is similar to those contained in the Kambin spinal instrument set marketed by Pilling. We will purchase this as an OEM item.
3. Cannulated Tapered Obturator: Approximately 5.5mm O.D. with an ID to slip fit over guide wire. Contains 1cm graduations. Similar to cannulated trocar contained in the Kambin spinal set marketed by Pilling. Current literature is attached.
4. Universal working Cannula with Removeable Fluid Seal Adapter: With 1cm graduations. Removeable fluid seal adapter for fluid control. Will also be fitted with adjustable cannula insertion stop to control cannula insertion similar to cannula contained in Kambin spinal instrument set marketed by Pilling.
5. 3mm, 5mm Trephine: 3mm, 5mm O.D. size to manually bore through annulus fibrosus. Similar to rotary cutters contained in Kambin spinal instrument set marketed by Pilling.
6. Forceps Deflector Tube: To accept 2.5mm O.D. flexible tipped cup forceps. Proximal end to accept standard suction tubing. Similar to forceps deflector contained in Kambin spinal instrument set marketed by Pilling.
7. Tissue Removal Rod/Magnetic Retriever Rod: Sized to slip fit through 5.0mm Trephine to push out trapped tissue. Similar to Kambin tissue removal rod and magnetic retriever rod. Marketed by Pilling. We plan on purchasing this item from Instrument Makar.
8. Cup Forceps: 2.5mm Flexible; Designed for use in conjunction with forcep deflector tube for manual cutting and removal of nucleus material posterior to site of fenestration. Similar to those marketed by Pilling in the Kambin spinal instrument set.
BEST AVAILABLE COPY
11
{12}
Attachment A: Page 2
9. 2mm Angled up, 3mm Straight Cup Forceps: Manual cutting instruments used in conjunction with universal cannula for removal of nucleus and annulus material. Both are similar to those in the Kambin spinal instrument set marketed by Pilling.
10. Dyo-Vac Suction Punch: 3.5mm: Hand cutting instrument properly tooled to allow for suction to be applied proximal. This is similar to those marketed by Dyonics since 1974. Pertinent literature is attached.
These components will be accompanied by a storage/sterilization.
B. Video Discoscope Set: To allow visualization of the disc area for procedural assessment. Contains the following components.
1. 2.7mm Irrigation Sheath: Designed to fit inside the working cannula. Fitted with single stopcock to allow tubing connection for suction. Similar to cannulas used as accessories to arthroscopes and powered surgical systems for arthroscopic surgery. These were reviewed by FDA under 510(K) K771218A, K820367, K880150. Pertinent literature is attached.
1. Video Discoscope: 2.7mm x 30 degree scope for visualization of disc area, to be used in conjunction with video. This is similar to arthroscopes marketed by Dyonics since 1972. These were also reviewed by FDA under K771218, K820367, K880150. Pertinent literature is attached.
This set will also be accompanied by a storage/sterilization tray.
C. Sterile Disposable Set: These represent the only components which will be sterile, one time use disposable products. They will be packaged individually.
1. 4.5mm Trimmer Blade 4.5 Full Radius Blade: For use with Dyonics powered surgical instruments. Similar to other disposable blades reviewed by FDA under 510(K) K833587. Will be packaged in PETG Thermoformed Blister and sealed with 1073 Tyvek(R) lid. Dyonics literature on disposable blades and powered instruments is attached.
2. Tissue Trap: Offered as an accessory to provide a means of collecting tissue for pathological purposes. Placed within suction tubing line. Similar to those marketed by Filtertek. Pertinent literature is attached.
BEST AVAILABLE COPY
62
{13}
# PILLING
BEST AVAILABLE COPY
13
{14}
# KAMBIN SPINAL INSTRUMENT SET
The Kambin Spinal Instrument Set, when used for percutaneous lateral discectomy*, facilitates surgical decompression of the L3 through L5 herniated lumbar discs, using a percutaneous posterior lateral approach.

This technique has the advantages of:
- Minimal morbidity
- Avoids epidural bleeding and scarring
- Directs any possibility of future rehemiation outside of the spinal canal
- Shortened hospital stay
*Covered by U.S. Patent No. 4,573,448
53-1000 Complete Kambin spinal set.
Includes 42-9602 sterilizing tray
53-1001 Spinal needle, 18 gauge,
6" long with stylet
53-1005 K-wire, .031" dia. x 9" long
53-1010 Blunt cannulated trocar, 18
cm. long with removable
Luer Lok fitting
53-1015 Cannula, 6.5 mm.
O.D., 4.9 mm. I.D. x 15
cm. with removable
Luer Lok fitting
53-1020 Rotary cutter, 3 mm.
diam. with permanently attached Luer
Lok fitting
53-1022 Rotary cutter, 5 mm.
diam. with permanently attached Luer
Lok fitting
53-1030 Forceps deflector tube for
use with flexible forceps
53-1036
BEST AVAILABLE COPY
14
{15}
# Unique characteristics of the technique include:
- Large fenestration of the annulus will not be blocked by tissue fragments and provides continuing decompression following completion of procedure.
- Diversity of instrumentation allows reaching posteriorly into the site of the herniation. The flexible forceps, deflector sheath, and angled forceps allow tissue removal in an area much larger in diameter than the 4.9 mm. inside diameter of the working cannula. This is not possible with straight probe systems which limit tissue removal to working channel diameter near the center of the disc.
- Patented suction technique moves fragmented material to the center of the disc for simple evacuation through the cannula.
- Local anesthetic allows communication with the patient during surgery and adds to safety of procedure.
- After effects on patient are minimal.
- Guide wire insertion technique makes sequential use of other instruments in the series simple and safe.
# Caution concerning flexible forceps:
The flexible forceps is 4 cm. longer than the other forceps in the set. It is designed to be used inside the deflector. Utilization of the flexible forceps without the deflector may be hazardous.

15
{16}
# ORDERING INFORMATION
| ITEM NO. | CATALOG NO. | DESCRIPTION |
| --- | --- | --- |
| 1. | 53-1000 | Complete Kambin spinal set, consisting of items 2 through 13 below |
| 2. | 53-1001 | Spinal needle, 18 gauge, 6" long, with stylet |
| 3. | 53-1005 | K-wire, .031" diam. x 9" long |
| 4. | 53-1010 | Blunt cannulated trocar 18 cm. long with removable Luer Lok fitting |
| 5. | 53-1015 | Cannula, 6.5 mm. O.D., 4.9 mm. I.D. x 15 cm. with removable Luer Lok fitting |
| 6. | 53-1020 | Rotary cutter, 3 mm. diam. with permanently attached Luer Lok fitting |
| 7. | 53-1022 | Rotary cutter, 5 mm. diam. with permanently attached Luer Lok fitting |
| 8. | 53-1035 | Kambin 3 mm. cup forceps, rigid, straight |
| 9. | 53-1040 | Kambin 4 mm. cup forceps, rigid, straight |
| 10. | 53-1050 | Kambin 2 mm. cup forceps, rigid, angled upward |
| 11. | 53-1030 | Forceps deflector tube for use with flexible forceps 53-1036 |
| 12. | 53-1036 | Kambin flexible 2.5 mm. cup forceps. Use only with 53-1030 deflector tube |
| 13. | 42-9602 | Instrument storage and sterilizing tray, 10" x 15" |
# ACCESSORIES
| 14. | 53-1060 | Kambin tissue removal rod for 3 mm. rotary cutter |
| --- | --- | --- |
| 15. | 53-1065 | Kambin tissue removal rod for 5 mm. rotary cutter |
| 16. | 53-1070 | Kambin magnetic retriever rod |
Note: Items 2 through 7 are normally used in the sequence listed.
# BIBLIOGRAPHY
1. Kambin, Parviz and Harris Gellman. "Lateral Discectomy of the Lumbar Spine." Clinical Orthopaedics. April, 1983, Vol. 174.
2. Kambin, Parviz and Steven Sampson. "Postlateral Percutaneous Suction-Excision of Herniated Lumbar Intervertebral Discs: Report of Interim Results." Clinical
Orthopaedics. June, 1986, Vol. 207.
3. Kambin, Parviz and Steven Sampson. "Laminectomy Versus Percutaneous Lateral Discectomy: A Comparative Study." Orthopaedic Transactions. The Journal of Bone and Joint Surgery. Fall, 1984, Vol. 8.
# PILLING
420 Delaware Drive
P.O. Box 7514
Fort Washington, PA 19034 U.S.A.
To order call our Customer Service Department or contact your Pilling representative at our toll free numbers 800-523-6507 (outside Pa.) or 800-492-2387 (Pa. only).
8:00 AM-7:00 PM Eastern Time
TELEX 6711330 FAX 215-646-0340
BEST AVAILABLE COPY
LIT NO. 99-3074
5M 187 ATA PRINTED IN U.S.A.
{17}
PILLING
PRICE LIST
JANUARY 1987
# KAMBIN™ SPINAL SET
# FOR PERCUTANEOUS LATERAL DISCECTOMY*
A new method for decompressing herniated intervertebral discs in the lumbar region
CATALOG #
DESCRIPTION
53-1000 COMPLETE KAMBIN™ SPINAL SET $2,425.00
Consists of:
53-1001 Spinal needle, 18 gauge, 6" long with stylet $20.00
53-1005 K-wire, .031" diameter x 9" long $6.00
53-1010 Blunt Cannulated Trocar, 18 cm long with removable Luer Lok fitting $205.00
53-1015 Cannula, 6.5 mm O.D., 4.9 mm I.D. x 15 cm with removal Luer Lok fitting $210.00
53-1020 Rotary Cutter, 3 mm diameter with permanently attached Luer Lok fitting $225.00
53-1022 Rotary Cutter, 5 mm diameter with permanently attached Luer Lok fitting $230.00
53-1035 Kambin 3 mm Cup Forceps, rigid, straight $390.00
53-1040 Kambin 4 mm Cup Forceps, rigid, straight $392.00
53-1050 Kambin 2 mm Cup Forceps, rigid, angled upward $275.00
53-1030 Forceps Deflector Tube for use with flexible forceps 53-1036 $225.00
53-1036 Kambin Flexible Cup Forceps, 2.5 mm (Use only with 53-1030 Deflector Tube) $334.00
42-9602 Instrument Storage and Sterilizing Tray 10" x 15" $211.00
# ACCESSORIES
53-1060 Kambin Tissue Removal Rod for 3 mm Rotary Cutter $50.00
53-1065 Kambin Tissue Removal Rod for 5 mm Rotary Cutter $50.00
53-1070 Kambin Magnetic Retriever Rod $60.00
*Covered by U.S. Patent No. 4,573,448
iNk
BEST AVAILABLE COPY
420 DELAWARE DRIVE P.O. BOX 7514 FORT WASHINGTON, PENNSYLVANIA 19034 215-643-2600 800-523-2579
TELEX 6711330
CABLE: SURGICAL—FORT WASHINGTON
FAX 215-646-0340
{18}
# Surgical Equipment and Supplies for Arthroscopy
'89 Catalog

To order
call toll free!
1-800-248-4668
In Michigan 1-800-678-3434
Local (517) 332-3593
FAX (517) 332-2043
BEST AVAILABLE COPY
C8
INSIDE:
New Products
and Services
{19}

Section Two
THE COAGULATOR
# IM Coagulators
fit easily through
cannulas
## Shoulder Cannulas
Shoulder Inflow Cannula Set prevents fluid extravasation.
Eliminate the complication of fluid extravasation during shoulder arthroscopy with Instrument Makar's Shoulder Cannula Set. The 4.0mm cannula's only opening is its distal tip; side holes have been removed. A handy inflow adaptor which accepts tubing on one end is also easily screwed on and off the cannula during the procedure.
Item #10771
Shoulder Cannula Set
(includes Cannula, Sharp and Blunt Trocars, Inflow Adaptor), $230.


## Golden Retriever
Breaking a blade tip or part of a jaw in the joint is a situation easily resolved with the Golden Retriever. Combining the pull of a 1¼" magnet and suction allows you to quickly remove metal pieces from the joint. After placing suction tubing on one end, slip the Golden Retriever through your cannula, attract the metal fragment, then remove it from the joint
using the cannula walls for protection.
Be confident you've got the means to solve the broken fragment problem on the spot—get the Golden Retriever (we hope you'll never have to use it).
Item #10100
Golden Retriever, $250.
14
BEST AVAILABLE COPY
19
{20}
DyoVac®
Suction Punch
DYONICS®
Combines the convenience of a reliable and efficient manual cutting instrument with the advantage of simultaneous tissue aspiration.
- Effective for meniscal tissue and synovial plicae resection, articular cartilage debridement, and morselizing of loose bodies.
- Unique jaw design facilitates rapid contouring of the meniscal edge.
- 90° Angled and Straight styles provide complete cutting capabilities.
- Convenient blade latch keeps jaws closed for safe introduction and removal from the joint.
- 2.5mm has been designed specifically for use in the Wrist and TM Joint

Designed in conjunction with Terry L. Whipple, M.D., Richmond, Virginia. DyoVac is a registered trademark of DyoVacs, Inc., Andover, Massachusetts.
BEST AVAILABLE COPY
20 X
{21}
# STRAIGHT SUCTION PUNCH


Cutting assembly shown actual size.
- Effective for excising tears in the posterior and central thirds of the meniscus.
- Lower jaw slips under the meniscus for easy resection.
- User replaceable shear pin in the handle protects against device components breaking off in the joint.
- 5.2mm is most effective for plicae, loose bodies, and tears in the central portion of the meniscus.
- 3.4mm is ideal for use in posterior horn tears and in tight knees.
- 2.5mm is designed for use in Wrist and TMJ arthroscopy
# 90° ANGLED SUCTION PUNCH
- Facilitates resection of tears in the anterior and central thirds of the meniscus.
- Exclusive closed-loop handle provides smooth cutting action and improves tactile feedback.
- Available in 90° Left and 90° Right angles to access lesions in both medial and lateral meniscal compartments.

# ORDERING INFORMATION
| Product Description | Part Number |
| --- | --- |
| Complete DyoVac® Suction Punch System (Includes: 3.4mm and 5.2mm Straight, 90° Right and Left) | 3322 |
| 4.5mm DyoVac® Suction Punch Set (Includes: 90° Right and Left) | 3321 |
| 5.2mm DyoVac® Suction Punch (Straight) | 2685 |
| 3.4mm DyoVac® Suction Punch (Straight) | 2707 |
| 2.5mm DyoVac® Suction Punch (Straight) | 3499 |
| 4.5mm DyoVac® Suction Punch, 90° Left | 3196 |
| 4.5mm DyoVac® Suction Punch, 90° Right | 3197 |
| 5.2mm Cleaning Brush Kit | 2697 |
| 2.5mm Cleaning Brush Kit | 3577 |
| 3.4mm Cleaning Brush Kit | 2715 |
| 4.5mm Cleaning Brush Kit | 3349 |
| 5.2mm Blade Retention Screw Kit | 2699 |
| 3.4mm Blade Retention Screw Kit | 2714 |
| 3.4mm Replacement Blade Assembly | 3585 |
| 5.2mm Replacement Blade Assembly | 3586 |
# SERVICE
Dyonics maintains a complete service department.
Toll free number: 800-343-5717.
# WARRANTY
One year warranty against manufacturer's defects on all parts and labor.
BEST AVAILABLE COPY
21

DYONICS®
Dyonics, Inc., 160 Dascomb Road, Andover, MA 01810
Telephone: 800-343-8386 (sales) 800-343-5717 (service) • In Massachusetts: 508-470-2800
A member of the Smith and Nephew group
Printed in U.S.A. 10-1-88 R-66-7434 Rev. B © 1988 Dyonics®, Inc. All rights reserved.
{22}
# VIDEO ARTHROSCOPES
## 4.0mm x 30°
## 4.0mm x 70°
DYONICS®
The 4.0mm arthroscopes join directly to a video camera, eliminating the need for an additional endoscopic coupling device.
### ■ Improved high resolution optics
The Dyonics® 4.0mm VideoArthroscope's precision-fabricated optical design transmits a bright, sharp image of extreme clarity and uniform illumination. The variable focus adjustment allows the surgeon to optimize image clarity.
### ■ Eliminates fogging during an operative procedure
The VideoArthroscope is designed to form a watertight seal to the "C" motion flange of a solid state video camera. Disinfecting the Video-Arthroscope and camera together as one unit virtually eliminates lens fogging problems.
### ■ Maximum maneuverability
Attached to the DyoCam™ video camera, the Dyonics® VideoArthroscope provides the surgeon with a compact and lightweight video system that is unsurpassed for maneuverability and convenience.
### ■ Compatible with the Dyonics Cannula System
Arthroscopic surgical instruments and the Video-Arthroscopes are interchangeable within the cannula system... eliminating the need for additional puncture wounds. This is especially valuable for shoulder arthroscopy, where maintaining portals may be difficult.
Ordering Information
| Description | Part No. |
| --- | --- |
| 4.0mm x 30° VideoArthroscope Operative System | |
| 4.0mm x 30° VideoArthroscope | |
| Compatible with DyoCam 550 Camera | 3329 |
| Compatible with all other DyoCam Cameras | 3320 |
| 4.0mm x 30° VideoArthroscope | |
| Compatible with DyoCam 550 Camera | 3327 |
| Compatible with all other DyoCam Cameras | 3304 |
| 4.0mm x 70° VideoArthroscope | |
| Compatible with DyoCam 550 Camera | 3350 |
| Compatible with all other DyoCam Cameras | 3355 |
Description
| Operative Instrument Set, 4.0mm | 2701 |
| --- | --- |
| 4.0mm x 30° VideoArthroscope | |
| Irrigation Extender, 4.0mm | 2705 |
| Cannula, 4.0mm | 2704 |
| Obturator, 4.0mm | 1992 |
| Protar, 4.0mm | 1985 |
| Sealing, Storage, Day | 3178 |
| Diagnostic Cannula Set | 3156 |
Part No.
Service/Repair
| For service, please call Dyonics Customer Service Department at 800-343-9717 or in Massachusetts call 617-470-2900 for a service replacement number. |
| --- |
| Warranty |
| 30-day warranty against manufacturer's defects on parts and labor. Note: Full warranty statement of included on VideoArthroscope instructions does not |

{23}
# ARTHROSCOPES
# 4.0mm x 30°
# 4.0mm x 70°
DYONICS
Precision workmanship produces a lightweight arthroscope with exceptional image clarity and uniform illumination.
# ■ Enhanced Optics
Improved optical design provides an increased wide angle view and enhanced fiberoptic illumination. The resulting image is brighter and sharper than ever before. Outstanding color reproduction is consistently achieved with the advanced optical system.
# ■ Optimal Maneuverability
Lightweight, streamlined styling provides comfortable control of the arthroscope with increased responsiveness and reduced surgeon fatigue. All arthroscope connections are aligned with the direction of view to improve maneuverability, particularly when viewing under the patella.
# ■ Proven Reliability
Durable opto-mechanical design provides resistance to the flexing stresses associated with arthroscopic procedures. A resilient cover lens protects the arthroscope from minor scratches.
# ■ Interchangeable Cannulas
Finger-posted cannulas allow for rapid repositioning of the arthroscopes, powered surgical instrument blades, and the infusion channel. This interchangeability is especially valuable in shoulder arthroscopy where establishing and maintaining portals may be difficult.
Ordering Information
| Description | Part No. |
| --- | --- |
| 4.0mm x 30° Arthroscope Operative System | 2721 |
| Includes Arthroscope Irrigation Extender, Inc. Cannula 14.5mm, Tracer 14.5mm, Ulcerator 14.5mm and Infusion Adaptor | |
| 4.0mm x 30° Arthroscope | 2713 |
| 4.0mm x 70° Arthroscope | 5146 |
| Operative Instrument Set | 2701 |
| Includes Irrigation Extender, Inc. Cannula 14.5mm, Tracer 14.5mm, Ulcerator 14.5mm and Infusion Adaptor | |
| Diagnostic Cannula Set | 5136 |
| Irrigation Extender, 4.5mm | 2705 |
| Cannula, 4.5mm | 2704 |
| Obturator, 4.5mm | 1982 |
| Tracer, 4.5mm | 1995 |
| Sealing/Storage Tray | 3179 |
# Service/Repair
For service, please call Dyonics Customer Service Department at 800-343-5717 or in Massachusetts call 617-470-2800 for a service replacement number.
# Warranty
90-day warranty against manufacturer's defect on parts and labor. Note: Full warranty statement is included on Arthroscope instruction sheet.

Diagnostic Cannula Set
Locking Finger-posted Cannula with single-twist convenience. End beveled to match angle of Arthroscope lens.

Sharp Tracer and Blunt Obturator open a path through tests and guide cannulas into the joint cavity.
DYONICS
Dyonics, Inc., 160 Dascomb Road, Andover, MA 01810
Telephone: 800-343-8386 (sales) 800-343-5717 (service) • In Massachusetts: 617-470-2800
A member of the Smith and Nephew group
BEST AVAILABLE COPY
Printed in U.S.A. 1-1500-0-2400-6M © Copyright 1996 Dyonics Inc. All rights reserved.
{24}
BYONGS
BEST AVAILABLE COPY
24X
{25}
# DYONICS
DISPOSABLE
BLADES ARE THE
RIGHT TOOLS
- Better Cutting Performance
- Color Coded
It takes the "right tools" to make a procedure faster, safer, and more precise. Dyonics offers the Right Tools.
Dyonics Disposable Arthroscopic Blades give you so many advantages. They are color coded, providing easy distinctions in style and size. Many of them help the drive system "recognize" the blade style and "remember" the speed you've selected for that style. All of them provide superior "fluid dynamics," with better seals and stronger

suction at the tip...giving you better cutting performance. Plus, you have the confidence of beginning each procedure with blades you know are absolutely sharp
Dyonics does not merely lead the arthroscopic blade market, we virtually invented it. We've benefited from the close collaboration of surgeons who are the pioneers in arthroscopic procedures.
We introduced. The surgeons asked for more. We developed new tips and sizes
Now, you can choose just the right tips and sizes you need for each procedure...from 16 large joint blades and 7 small joint blades. As surgeons recognize additional blade needs, Dyonics will provide them.
# PS3500™ THE BLADES THAT REMEMBER
Our PS3500™ blades contain an electronic "I-D" that works with the PS3500™ Drive Unit. The Motor Drive Unit recognizes each separate

blade and its optimal speed range. You set the distinct speed you want within that range, and the Motor Drive Unit "remembers" your speed selection. The result? Faster, safer procedures, with tools that work just the way you want them to
# THE RIGHT TOOLS FOR NEW PROCEDURES
Dyonics now offers 7 Mini Blades for small joint operations. Surgeons doing wrist, ankle, elbow, and TMJ procedures can now benefit from Dyonics technology and quality.
# CHOOSE FROM SUPERIOR DRIVE UNITS
Dyonics Arthroscopic Surgery Disposable Blades fit several motor

drive units. Arthroplasty and Shaver blades are compatible with Dyonics Arthroplasty and Shaver motor drive units from the Intra-Articular Surgical System, and the Universal Motor Drive Unit of the Advanced Arthroscopic Surgical System (AASS). PS3500 blades are compatible with the AASS Motor Drive Unit and the PS3500 Motor Drive Unit. Both sets fit in Dyonics cannulas.
Small-Joint Blades are compatible with the Mini-Motor Drive Unit and the AASS Universal Motor Drive Unit.
# STERILE PACKED
FOR ADDITIONAL CONVENIENCE
All Dyonics Disposable Arthroscopic Blades are packaged sterile in convenient peel-back tray and dispenser boxes. Both components of the blade assembly are sterilized separately, ensuring complete EIO sterilization.
Dyonics Disposable Arthroscopic Blades. Better cutting performance. The convenience and speed of color coding.
The Right Tools.
DYONICS®
BEST AVAILABLE COPY
{26}
# DISPOSABLE BLADES ASSURE A SHARP EDGE EVERY TIME

Cutter...Rounded-corner design protects articular cartilage from inadvertent damage. The open-ended blade configuration is excellent for resection of meniscal tears, plica, and selected articular cartilage lesions. The 2.9mm version works well on the triangular fibrocartiledge of the wrist, and in other small joint spaces.


Trimmer...A closed, rounded end with a side cutting window near the tip make this a good blade for debridement in patellar chondromalacia and in the final smoothing of meniscal edges.


Full Radius Blade...Aggressive cutters that provide multiangle maneuverability within the joint. The bevelled-end design allows effective use as both an end and a side cutter. The smaller 2.9mm version works well in small joints, like the wrist and ankle, while the 2.0mm version handles even smaller areas like TMJ procedures.


TurboCutter™...An aggressive, open-ended cutter, used for the rapid removal of meniscal tissue. It is especially useful to resect flap, degenerative, and bucket handle tears in the central third and posterior portion of the meniscus.


TurboWhisker™...Works safely, yet aggressively, at high speeds. The 4.5mm blade works well for debridement in patellar chondromalacia or in smoothing meniscal edges and the glenoid labrum. The 2.9mm version is used in smaller joints, like the wrist and ankle, while the 2.0mm version is used in TMJ procedures.


TurboTrimmer™...A toothlike edge on the outer blade window allows maximum maneuverability and aggressive cutting. The rounded design aids efficient side cutting, while protecting articular cartilage from inadvertent damage.


ProCutter™...Designed for high speed meniscal resection. A unique distal tip configuration directs tissue into the cutting jaw to cut twice with each rotation.


Abrader...Intended for use in roughening sclerotic lesions and to remove osteophytes and attached bony "loose bodies." Large blades are designed for use in the knee, shoulder, and elbow. The 2.9mm blade is designed for use in smaller joints, like the wrist and ankle.


Acromionizer...Specifically designed to speed up and aid arthroscopic acromionectomy. The elongated burr provides better access, to smoothly resect and contour a slope under the anterior third of the scapular acromion.

Barrel Abrader...This 2.9mm blade is very useful in small joint spaces like the wrist and ankle.

BEST AVAILABLE COPY
26
{27}
DISPOSABLE ARTHROSCOPIC SURGERY BLADES
# ORDERING INFORMATION
ARTHROSCOPIC SURGERY BLADES
| AASS Cat. No. | PS3500™ Cat. No. | Product Description | Size | Color |
| --- | --- | --- | --- | --- |
| 3399 | 3450 | Abrader Blade | 4.0mm | Aqua |
| | 3451 | Abrader Blade | 5.5mm | Black |
| 3401 | 3452 | Acromionizer Blade | 4.0mm | Mauve |
| | 3453 | Acromionizer Blade | 5.5mm | Brown |
| 3054 | 3439 | Cutter* | 3.5mm | Red |
| 2963 | 3440 | Cutter* | 4.5mm | Blue |
| | 3598 | Cutter* | 5.5mm | Lt Pink |
| 3058 | 3442 | Full Radius Blade | 3.5mm | Beige |
| 3062 | 3443 | Full Radius Blade | 4.5mm | Yellow |
| 3059 | 3444 | Full Radius Blade | 5.5mm | Orange |
| 3324 | 3447 | ProCutter™ Blade** | 4.0mm | Purple |
| 3325 | 3448 | ProCutter™ Blade** | 5.0mm | Pink |
| 2964 | 3441 | Trimmer | 4.5mm | Green |
| 3382 | 3449 | TurboCutter™ Blade | 4.0mm | Turquoise |
| 3528 | 3529 | TurboTrimmer™ Blade | 4.5mm | Powder Blue |
| 3061 | 3446 | TurboWhisker™ Blade | 4.5mm | Navy Blue |
MINI BLADES
| Cat. No. | Product Description | Size | Color |
| --- | --- | --- | --- |
| 3530 | Mini Abrader Blade | 2.9mm | Orange |
| 3553 | Mini Barrel Abrader Blade | 2.9mm | Purple |
| 3606 | Mini Cutter* | 2.9mm | Beige |
| 3410 | Mini Full Radius Blade | 2.0mm | Blue |
| 3419 | Mini Full Radius Blade | 2.9mm | Red |
| 3407 | Mini TurboWhisker™ Blade | 2.0mm | Yellow |
| 3418 | Mini TurboWhisker™ Blade | 2.9mm | Green |
*Cutter-Patent No. 4,274,414
**ProCutter™ Blade. Patent Pending.
# WARRANTY
Dyonics' products are guaranteed to be free from defects in material and workmanship.
# STERILIZATION
Product is ethylene oxide sterilized and is provided in sterile packaging. These blades are intended for single use only and should not be resterilized.

Dyonics, Inc., 160 Dascomb Road
Andover, Massachusetts USA 01810
Call TOLL FREE 1-800-343-5717
(In Massachusetts 508-470-2900)
A member of the Smith and Nephew Group
BEST AVAILABLE COPY
27
Printed in U.S.A. 6/89 10214 10M ©Copyright 1989 Dyonics®, Inc. All rights reserved.
P/N 1060112
{28}

# FILTERTEK, INC.

BEST AVAILABLE COPY
28
{29}
# choice of filter media
FILTERTEK utilizes a wide variety of thermoplastic materials to prevent the filter media best suited for the specific application. Consideration is given to the requirements, particle retention, self-cleaning properties, pressure drop, and/or use, economy.
In most cases woven monofilament screens . . . having either a small or dutch twill weave . . . perform maximally. For even more precise filtration requirements, special non-woven media is recommended.
Five popular types of media are illustrated on the following pages for comparative purposes.

ENLARGED 100X
## Reinforced Membrane Filter Media
This microporous film made of biologically safe materials is reinforced with nylon for exceptional toughness. A choice of six different pore sizes, ranging from 0.2 micron to 5 microns in size, occupies 75% of filter volume. Absolute pore sizes make this media excellent for filtering out microorganisms and providing filtrates free of submicron particulate matter. Basically hydrophilic, this media, when wet, will allow passage of aqueous solutions, but prevent the passage of gases. Selected grades are available in special sizes with water repellent (hydrophobic) characteristics which will permit passage of gases, but will retain aqueous solutions. The combination of the two grades, hydrophobic and hydrophilic, permits the design of specialized self-venting filter assemblies. Acropor*
## Woven Monofilament Screen
Rugged monofilament provides a very serviceable filter media for many applications. The example illustrated is a square weave. This media is available in nylon, polyester, polypropylene, and teflon, with mesh openings of 10 microns (.000394") and larger. The smooth thread surface provides excellent self-cleaning action and flow characteristics up to 20% better than wire cloth equivalents. The selection of this material assures complete resistance to fatiguing due to aging, vibration, or flexing.

DECT AVAILABLE COPY
29
{30}


| No. | Description |
| --- | --- |
| 557 | .45μ Reinforced Hydrophobic Membrane Media and Acrylic Frame with Elastomer Sample Port, Vented Catheter Connector and Anti-Reflux Valve Assembly. |
| 558 | .45μ Reinforced Hydrophobic Membrane Media and ABS Frame Vented Catheter Adapter. |
| 559 | .45μ Reinforced Hydrophobic Membrane Media and ABS Frame Vented Catheter Connector with Sampling Port. |
| 560 | .45μ Reinforced Hydrophobic Membrane Media with ABS Frame Vented Connector. |
| 561 | .45μ Reinforced Hydrophobic Membrane Media and ABS Frame Vented Connector. |
| No. | Description |
| --- | --- |
| 543 | 265μ Polyester Screen and ABS Frame Blood Filter. |
| 544 | 265μ Polyester Screen with Polypropylene Frame Blood Filter. |
| 545 | 170μ Nylon Screen and ABS Frame Blood Filter. |
| 546 | 265μ Polyester Screen and Polypropylene Frame Blood Filter. |
| 547 | 210μ Nylon Screen and ABS Frame Bubble Filter. |
| 548 | 150μ Polyester Screen and ABS Frame Blood Filter. |
| 549 | 100μ Nylon Screen and ABS Frame Blood Filter. |
| 550 | 265μ Polyester Screen and ABS Frame Blood Filter. |
| 551 | 5μ Reinforced Hydrophilic Membrane Media and Acrylic Frame Particulate Filter. |
| --- | --- |
| 552 | 5μ Reinforced Hydrophilic Membrane Media and Acrylic Frame Inline Particulate Filter. |
| 553 | ABS Frame with Elastomer Band Sampling Port. |
| 554 | .45μ Reinforced Hydrophobic Membrane Media and ABS Frame Line Vent Filter. |
| 555 | .45μ Reinforced Hydrophobic Membrane Media and ABS Frame Vented Filter Adapter. |
| 556 | .22μ Hydrophilic Membrane Media and Acrylic Frame Inline Filter. |
# BEST AVAILABLE COPY

10
{31}

| No. | Description |
| --- | --- |
| 521 | 5μ Nylon Screen and Nylon Frame Filter Tip, ½" dia. |
| 522 | 5μ Nylon Screen and Nylon Frame Filter Tip, 7/16" dia. |
| 523 | 5μ Nylon Screen and Nylon Frame Filter Tip, 5/16" dia. |
| 524 | 5μ Nylon Screen and Nylon Frame Filter Tip, 3/8" dia. |
| 525 | 5μ Nylon Screen and Nylon Frame Filter Tip, 3/8" dia. |
| 526 | 74μ Polyester Screen and Polypropylene Frame Cell Counter Filter. |
| 527 | 8μ Reinforced Hydrophobic Membrane Media and ABS Frame Medical Administration Filter. |
| 528 | 74μ Polyester Screen and Acrylic Frame Inline Medical Administration Filter. |
| 529 | 5μ Reinforced Hydrophobic Membrane Media and ABS Frame Drip Chamber Vent. |
| 530 | 6μ Polyester Screen and Acrylic Frame Import Filter. |
| 531 | 5μ Reinforced Hydrophobic Membrane Media and Polycarbonate Frame Needle Hub. |
| 532 | .45μ Reinforced Hydrophobic Membrane Media and Polyethylene Frame Vent Plug and Filter. |
| 533 | 10μ Nylon Screen and Nylon Frame Inline Filter Element. |
| 534 | .45μ Reinforced Hydrophobic Membrane Media and Polyethylene Frame Vent Filter Cap. |
| 535 | 8μ Reinforced Hydrophobic Membrane Media and Acrylic Frame Filter Cap. |
| 536 | 8μ Reinforced Hydrophobic Membrane Media and ABS Frame Vent Cap. |
| No. | Description |
| --- | --- |
| 537 | 5μ Reinforced Hydrophilic Membrane Media and Propionate Frame Pre-Bypass Filter. |
| 538 | 5μ Reinforced Hydrophilic Membrane Media and Propionate Frame Pre-Bypass Filter. |
| 539 | 5μ Reinforced Hydrophilic Membrane Media and Acrylic Frame Pre-Bypass Filter. |
| 540 | 48μ Nylon Screen and ABS Frame Mouthpiece with Filter. |
| 541 | 5μ Reinforced Hydrophilic Membrane Media and ABS Frame Pre-Bypass Filter Element. |
| 542 | 265μ Polyester Screen and Propionate Frame Large Particulate Medical Filter Cone. |
BEST AVAILABLE COPY

{32}

# health care filters
The biologically sterile standards of the health care industry are satisfied by a wide variety of FILTERTEK filters.
| No. | Description |
| --- | --- |
| 500 | 170μ Polyester Screen and Polycarbonate Frame Blood Filter |
| 501 | .8μ Reinforced Membrane Media and ABS Frame Medical Administration Filter |
| 502 | 10μ Nylon Screen and Nylon Frame Diagnostic Test Instrument Filter |
| 503 | 149μ Polypropylene Screen and Nylon Frame Test Instrument Filter |
| 504 | 100μ Teflon Screen and Polypropylene Body Inline Medical Filter |
| 505 | .45μ Reinforced Hydrophobic Membrane Media and Polypropylene Frame Air Venting Filter |
| 506 | 170μ Polyester Screen and Polycarbonate Frame Blood Filter |
| No. | Description |
| --- | --- |
| 507 | 263μ Nylon Screen and ABS Frame Blood Administration Filter |
| 508 | .45μ Reinforced Hydrophobic Membrane Media and ABS Frame Inline Air Venting Filter |
| 509 | 265μ Polyester Screen and Polycarbonate Frame Blood Filter |
| 510 | 120μ Nylon Screen and ABS Frame Blood Filter |
| 511 | 263μ Nylon Screen and ABS Blood Filter Frame |
| 512 | 263μ Nylon Screen and Polypropylene Frame Blood Filter |
| 513 | 170μ Nylon Screen and Propionate Frame Blood Filter |
BEST AVAILABLE COPY

32
7
{33}

| No. | Description |
| --- | --- |
| 519 | 25MM Inline Filter available with tab or threaded female luer lock inlet and male luer slip outlet. This filter is available with either hydrophobic or hydrophilic membrane filter media in retention ratings from 0.2 to 5.0 microns. The modified acrylic multipolymer housing incorporates a patented membrane sealing method which gives pressure resistance and large effective flow area. |
| No. | Description |
| --- | --- |
| 514 | 1μ Glass Fiber Filter Media and Acrylic Frame Drive Line Filter |
| 515 | 1μ Glass Fiber Filter Media and Acrylic Frame Gas Filter |
| 516 | 8μ Reinforced Hydrophilic Membrane Media and ABS Frame Medical Administration Filter |
| 517 | 48μ Nylon Screen and Nylon Frame Filter Disc |
| 518 | 8μ Reinforced Hydrophobic Membrane Media and ABS Frame Medical Administration Filter |


# BEST AVAILABLE COPY
| No. | Description |
| --- | --- |
| 520 | 37MM Inline Filter with female luer inlet and male luer slip outlet is available both with and without 0.02 micron teflon hydrophobic air vents on the inlet side of the housing. The patented membrane sealing method assures the ability to bubble point the 0.22 micron hydrophilic filter media, which is standard in this assembly. The modified acrylic multipolymer housing can be readily solvent bonded to tubing or plastic fittings as required. |
8
33
{34}
PS3500™
ARTHROSCOPIC
SURGICAL SYSTEM Manual
DYONICS®

## General Operating Instructions
Section Page
Indications 4
Specifications 5
System Components 5
Operating Instructions 6
Maintenance and Service –
Blade Assemblies 10
Maintenance and Service –
Motor Drive Unit 11
Troubleshooting 12
Cleaning, Sterilization,
Disinfection Guidelines 14
Circuit Description Inside Back Cover
Warranty Back Cover
References Back Cover
BEST AVAILABLE COPY
34
{35}

2
BEST AVAILABLE COPY
35
{36}
# PS3500™
# Arthroscopic Surgical System
Your Dyonics PS3500 Arthroscopic Surgical System* is a multi-function system. You can select the operating characteristics in your PS3500 high speed Motor Drive unit that meet your patient's surgical needs for abrasion arthroplasty, synovectomy, or intra-articular cutting and shaving. An optional Mini Motor Drive unit is available that will broaden your application for arthroscopic surgery. Preselected optimum blade speed operating ranges have been established for the system. Within each range you may select and change at any time the speed that is best for your technique. The system will automatically return to the last speed established for each blade upon reinsertion of a particular blade style during the same or subsequent operations.
Please read this manual carefully before you use the PS3500 System. You will learn what this product will do and how to take care of it. The maintenance procedures take little time and effort, and will ensure optimum performance. They will ensure many hours of reliable operation.

*U.S. Patent Number: 4,705,038
BEST AVAILABLE COPY
10
36₃
{37}
# PS3500™
# Arthroscopic Surgical System
## Indications
Under arthroscopic direction, the PS3500 System is intended to resect damaged tissue and remove extraneous matter found in articular body cavities. The system may be used to repair tears and other defects, remove loose fragments, shave away debris, and perform appropriate synovectomy procedures.
The PS3500 System has been designed with preselected speed range and torque for each blade style.
Clinical indications for abrasion arthroplasty include degenerative arthritis with a complaint of pain, especially at night or while standing or walking. Pathological confirmation includes exposed bone. Clinical indications for synovectomy include synovitis resulting from torn meniscus.
Use an abrader to debride exposed sclerotic lesions on the tibial, femoral, or patellofemoral surface.
Use a synovial resector to remove reactive synovitis and to smooth rough areas of articular cartilage.
## Contraindications
The PS3500 Arthroscopic Surgical System should not be used with patients exhibiting ankylosis without adequate joint space or distention for arthroscopic inspection.
Abrasion arthroplasty may not be effective in treating heavy patients or those with ankylosis, instability, or expectations beyond the relief of pain. Varus or valgus deformity is not of itself a contraindication in patients with good range of motion and without gross instability or extreme
malalignment (15 deg. varus, 30 deg. valgus). Intracortical abrasion arthroplasty may be contraindicated in patients not qualifying for high tibial osteotomy or total knee replacement.
Synovectomy is contraindicated where the disease has progressed beyond the phase of synovial proliferation, and in advanced rheumatoid arthritis where erosion of the articular cartilage is present.
## Precautions
Before using the PS3500 System for the first time, you should review critically all available information. The list of references on the back cover will be helpful.
A complete and comprehensive preoperative medical history and physical examination are suggested. X-ray evaluation and laboratory investigation may be included.
Before attempting abrasion arthroplasty, you should be able to perform a comprehensive arthroscopic examination. You should also be experienced in arthroscopic surgery with powered instruments such as the Advanced Arthroscopic Surgical System with the Shaver or Arthroplasty AutoSensor, the Dyonics Arthroscopic Surgery System, or the Dyonics Intra-Articular Surgical System II.
Healthy intra-articular soft tissue and cartilage can be injured by the blades and abraders. Use every available means to avoid such injury. The vascularity of subchondral bone extends into the cortical layer. Abrasion should therefore extend no deeper than 1-2mm into the cortex and not into cancellous bone.
## CAUTION:
Direct contact of the rotating cutting edge of blades or burrs with metal (e.g., cannula, arthroscope, or other instrument) can cause damage to the instrument tip. This damage can range from slight distortion and/or dulling of the cutting edge to actual fracture of the tip in vivo. If such contact should inadvertently occur, it is important to stop using the blade immediately and examine the instrument tip carefully for evidence of cracks or fractures. If there is any doubt about the condition of the blade assembly, the blade should be discarded and replaced with a new one, or in the case of reusable blades, returned to Dyonics for evaluation.
## Suggested Technique
Arthroscopic surgery of the knee requires adequate joint distention to permit a clear view. Suspend 6 liters of saline 1-2 meters above the patient. Use the Dyonics 5.5mm inflow cannula for entry into the supra-patellar pouch; gravity inflow through a Verres needle or arthroscope may not provide adequate distention.
Effective cutting requires outflow with 14-16 inches (35.6-40.6cm) of mercury in-line continuous suction pressure.
The distention and suction required when using the Mini Motor Drive unit vary with the joint.
BEST AVAILABLE COPY
4
37
{38}
# Post Operative
After abrasion arthroplasty, a two month period of non-weightbearing ambulation is mandatory. This allows proper avascular articular surface repair, and should be accompanied by active range of motion exercises three times a day. During the two months, the patient should avoid extremity loading, weightbearing, or range of motion activity with stress. The reparative articular cartilage must mature for at least six months before you can determine the success of the procedure.
# Specifications
CONTROL UNIT:
Size: 16.4"W x 11.2"D x 4.7"H
Weight: 17 pounds
FRONT PANEL:
Power Switch: On/Off (I/O)
Diagnostic Display: 16 character linear message display
Motor Speed Select Switches: Pair of momentary push switches
Speed Display: Array of LED's
Maximum and Minimum speed, selected (set) speed
Vertical bar graph indicating relative speed setting within available range Motor Select Switch: PS3500™, Mini, and Universal Drive Unit
Motor Drive Unit Connectors:
PS3500, Mini, and Universal Drive Unit Footswitch Connector
REAR PANEL:
Cooling: Exhaust fan
AC Power: Detachable cord with a three prong hospital grade connector
Access ports: Removable panel provides access to the expansion port (left). A 25 pin connector is provided for the RS232 port (right).
CAUTION:
Only Dyonics approved equipment should be connected to RS232C and/or expansion port connectors.
MOTOR DRIVE UNIT:
Length: 7.3"
Weight: 18 ounces
Equipped with 8-foot autoclavable, replaceable power cord.
FOOT PEDAL:
Size: 7"W x 5.5"D x 2"H
Weight: 3 pounds
Directional switches: Forward/reverse/oscillate
# System Components
Your PS3500 System contains four basic elements:
1. Control Unit capable of variable speed operation of the PS3500, Universal, and Mini Motor Drive units
2. Footswitch to control the power from the Control Unit to the motor drive units
3. PS3500 Motor Drive unit capable of running between 350 rpm and 3500 rpm. A spare cord is included.
4. Thermoplastic Sterilization Tray with insert Five Blade Tray
# Optional
The Mini Motor Drive unit is also available. This motor drive used in conjunction with an Accessory Kit will position you for highly efficient tissue resection and removal during operative arthroscopy procedures performed on small joints.
BEST AVAILABLE COPY
X
38
5
{39}
# Operating Instructions
## Control Unit Panels
The picture (A) shows the front panel of the Control Unit. The Power Switch (a) is at the far right of the panel. When you are not using the system keep the switch in the off (O) position.
The Diagnostic Panel (b) is a 16 character display capable of relaying various operating messages, blade type, and alarm conditions.
The Speed Selection switches (c) enable the operator to increase (Faster ▲) or decrease (Slower ▼) the most recently set speed for the motor drive unit within the factory preset range (f). The Speed Display (d) indicates the current speed selected within that range. Also, a Vertical Array Bar Graph (e) indicates the relative position of the selected speed within the min/max limits available.
The Control Unit features an automatic Blade Speed Recall (BSR) function. The system will automatically recall the last selected speed within the pre-determined ranges for each blade style. The operator will not be required to change speeds when blades are changed during a procedure. The PS3500™ system will recall speeds even after the power has been turned "off" (O) and "on" (f).
The Motor Selector Switch (g) allows you to select the motor drive unit you will use: PS3500, Mini, or Universal. The appropriate motor must be plugged into the corresponding outlet (h). The Diagnostics will detect a mismatched selection and indicate "CHECK MOTOR" when the Footswitch is depressed.
The Footpedal cord is connected to the Control Unit through the far left outlet (i). A missing or poorly connected Footswitch will be detected and indicated on the Diagnostic Display as "FOOTSWITCH".
A slide out tray labeled "Diagnostics" is located underneath the unit for your convenience in troubleshooting.

## Assembling the System
Connect the Footpedal cord to the labeled socket on the front of the Control Unit. Plug the PS3500 Motor into the receptacle labeled PS3500. The motor cord should be aligned with its ridge at the twelve o'clock position. The Mini and Universal Motors can also be run from the Control Unit by connecting them to the correspondingly labeled receptacle. The system has been designed so that each motor drive
unit will function only if its cord is inserted into the correct socket, and the Motor Selector Switch is in the appropriate position.
Turn the Motor Selector Switch on the Control Unit to its upper position labeled "PS3500". If another motor drive unit is to be used, turn the Motor Selector Switch to the appropriate position. Power is supplied only to the selected motor drive unit.
## Changing Blades
### Blade Insertion: PS3500 and Mini Motors

The PS3500 and Mini Motor Drives each have a spring loaded ring at the distal end of the motor, permitting blade insertion and removal (E). 1. To insert a blade, slide the release ring on the distal end of the motor drive toward the proximal end as
shown. This will reveal the key slot (a) on the motor drive. Orient the motor drive so that you can see the key slot (a) and the key (b) on the blade. Insert the blade into the motor drive so that the key goes into the slot. Let the release ring slide back to its original position. Note that the cutting aperture is aligned with the ridge on the motor drive. This feature allows for tactile orientation.
2. To remove a blade, slide the release ring toward the proximal end of the motor drive unit, and simultaneously withdraw the blade.
If suction is being applied you will perceive a slight force holding the components together.
4
| BEST AVAILABLE COPY |
39
{40}
# Variable Suction Control
The PS3500™ and Mini Motor Drive units afford variable fingertip control of fluid outflow. The surgeon can easily select minimum suction, maximum suction, or any setting that falls between these extremes.
## PS3500 Motor Drive Unit
1. Maximum flow is achieved by positioning the suction control lever to the left of the PS3500 Motor Drive Handpiece (C).
2. Minimum flow is achieved by positioning the suction control lever to the right of the PS3500 Motor Drive Handpiece (D).


## Mini Motor Drive Unit
1. Maximum flow is achieved by positioning the suction control lever perpendicular to the Mini Motor Drive Handpiece (E).
2. Minimum flow is achieved by pressing the suction control lever completely against the handpiece in either direction (F).


## Connecting the Suction
Push the tab from the suction source firmly onto the suction portal which is located at the proximal end of the Motor Drive unit (G).

## Inflow Cannula
After you have inserted the infusion cannula into the joint space, screw the infusion adaptor onto the cannula until fingertight. Push the saline inflow tube firmly onto the infusion adaptor.
# Testing the System's Operation
To check the instrument's operation, turn the Control Unit power switch to "on" (I). "MOTOR MISSING" and "FOOTSWITCH" may alternately appear on the Diagnostic Display if they are missing or incorrectly connected. Check for correct positioning of the Motor Select switch with the actual motor connected to the Control Unit. Confirm that the Footpedal cord is securely fastened to the Footswitch outlet. "Dyonics PS3500" should now appear on the Diagnostic Display.
The motor will not rotate when there is no blade in the motor and the Footpedal is depressed. The Diagnostic Display will indicate "NO BLADE".
Once a Dyonics blade has been properly secured in the PS3500 motor, the name of the blade will appear on the Diagnostic Display. Due to BSR, the blade speed will automatically return to the last selected speed within the pre-determined range for that blade style.
Depress the left and right side of the Footpedal and check that the blade goes in both the forward and reverse directions. The Diagnostic Display will indicate the correct direction of the blade. Visual confirmation of the blade direction should be consistent with the Diagnostic Display.
Depressing the center of the Footpedal will run the motor drive unit in "OSCILLATE" Mode. Depressing both sides of pedal will also run blades in oscillate mode. The Motor Drive unit will reach the selected speed during each oscillation. The oscillate interval will be shorter at slower speeds and longer at faster speeds due to the constant acceleration/deceleration of the Motor Drive unit.
NOTE: Some blades will not run in "REVERSE" or "OSCILLATE". Please refer to the blade chart for further information.
BEST AVAILABLE COPY
{41}
# Blade Speed Ranges (rpm)
All Dyonics disposable and reusable blades have been extensively tested for safety and performance. Blade speed ranges have been pre-set on your PS3500™ System based on this information.
The tissue cutting performance was determined at various speeds for each blade. This information indicates that some blades are more efficient when they are cutting in the "OSCILLATE" mode (H). Others are more efficient when running "FORWARD" (I).
Blade speed performance data has been summarized on the following table. This data includes the pre-set speed ranges for each blade style and the direction each blade is allowed to rotate.
When the Universal Motor Drive is connected to the PS3500 Control Unit the Universal AutoSensor allows all of the PS3500 blades to run from 400 to 1400 rpm. When the Universal Motor Drive is connected to the Advanced Arthroscopic Surgical System, blades run at speeds indicated on the front panel of the AASS Control Unit.
Tissue Cutting Rate

Blade Speed

Blade Speed
BEST AVAILABLE COPY
8
41
{42}
# Blade Speed Summary Table
| Blade style | Size (P/N) | Available speed range (rpm) | Blade direction Forward: F Reverse: R Oscillate: O Listed in order of efficiency |
| --- | --- | --- | --- |
| CUTTER* | 3.5mm disposable (3439) 3.5mm reusable (3457) 4.5mm disposable (3440) 4.5mm reusable (3458) 5.5mm disposable (3598) 5.5mm reusable (3459) | 350-1400 350-1400 350-1400 350-1400 350-1400 350-1400 | O, F/R |
| TRIMMER | 4.5mm disposable (3441) 4.5mm reusable (3460) | 350-2000 350-1400 | O, F/R |
| FULL RADIUS | 3.5mm disposable (3442) 3.5mm reusable (3461) 4.5mm disposable (3443) 4.5mm reusable (3462) 5.5mm disposable (3444) 5.5mm reusable (3463) | 500-2800 500-2000 500-2800 500-2000 500-2800 500-2000 | O, F/R |
| TURBOWHISKER™ | 4.5mm disposable (3446) | 500-2000 | O, F/R |
| FULL RADIUS WHISKERS | 3.5mm reusable (3468) 5.5mm reusable (3469) | 500-2000 500-2000 | |
| PROCUTTER*** | 4.0mm disposable (3447) 4.0mm reusable (3472) 5.0mm disposable (3448) 5.0mm reusable (3473) | 500-3000 500-3000 500-3000 500-3000 | F/R, O |
| TURBOCUTTER™ | 3.0mm reusable (3470) 4.0mm disposable (3449) 4.5mm reusable (3471) | 500-3500 500-3500 500-3500 | F |
| TURBOTRIMMER™ | 4.5mm disposable (3529) | 400-2800 | O, F/R |
| ABRADER | 4.0mm disposable (3450) 4.0mm reusable (3464) 5.5mm disposable (3451) 5.5mm reusable (3465) | 500-3500 500-3500 500-3500 500-3500 | F, R |
| ACROMIONIZER | 4.0mm disposable (3452) 4.0mm reusable (3466) 5.5mm disposable (3453) 5.5mm reusable (3467) | 500-3500 500-3500 500-3500 500-3500 | F, R |
| AUTOSENSORS with Universal Motor Drive | Small Joint: reusable blades (3085) Mini: disposable blades (3576) Arthroplasty: reusable/disposable blades (3379) Shaver: reusable/disposable blades (3376) Universal: reusable/disposable blades (3492) | 300-1500 300-1500 400-1400 120- 300 400-1400 | |
| MINI MOTOR DRIVE | All Blades | 350-3500 | |
*U.S. Patent Number: 4,274,414
**U.S. Patent Pending
BEST AVAILABLE COPY
9
{43}
# Maintenance and Service
## Blade Assemblies
Each blade assembly is designed and manufactured to the close tolerances required for precise surgery. In order to maintain the assembly in that condition, you need to observe a few precautions.
# CAUTION:
Don't let the rotating part of any blade or burr touch any metallic object, such as a cannula or arthroscope. Damage to both instruments is likely, and in severe cases the instrument tip can be fractured and possibly lodged in the joint. If such contact does occur, immediately examine the instrument tip for evidence of cracks, fractures, or dulling of the edge. Use an eye loupe to do this (J.K). If you are not certain the instrument tip is in good condition, return it to Dyonics Customer Service for evaluation. Provide a brief history of what happened.
## Reusable Blades
# During Use:
Each reusable burr should remain sharp through several complete procedures. Inspect the burr tip after each procedure and replace the burr at the first sign of dullness.
Each reusable blade assembly should remain sharp through twenty or more procedures. The number of procedures varies with their duration and with the type of tissue cut. A blade can ordinarily be resharpened three times. Return blades to Dyonics for resharpening. Sharpening is authorized only if it is performed by Dyonics; any unauthorized sharpening voids the warranty. Improper service could pose a safety hazard.
Unlike burrs, cutting blades are intended only for soft tissue. They will be dulled more quickly, and can be damaged by contact with bone. When you return any blade assembly to Dyonics for service, be sure to include both the inner and outer assemblies. They can be identified by their matching serial numbers.



## Cleaning and Lubrication
After use, remove the inner assembly from the outer tube and clean both with a brush (L). Then rinse thoroughly in water to remove saline and cleaning agents. Dry both sections inside and out. Lubricate the shaft of the inner assembly. Always lubricate before use or storage; failure to do so may result in rusting of the assembly.
## Sterilization
All PS3500® reusable blade assemblies and accessory hardware may be sterilized by ETO or steam, or soaked in disinfectant. For detailed instruction, see Guidelines for Cleaning and Sterilization in the back of this manual. NOTE: Extended soaking in solutions that contain instruments of dissimilar metals may lead to electrolytic corrosion.
## Disposable Blades
Dyonics PS3500 disposable arthroscopic surgery blades must be used with the PS3500 System. Follow the instructions provided in each box of disposable blades for assembly.
Attach the blade directly to the PS3500 Motor by following the instructions in this manual. Dyonics disposable arthroscopy blades are intended for single use only. Do not resterilize or lubricate them.
NOTE: The blue "O" rings (P/N 3590) on PS3500 reusable blades should periodically be replaced to maintain good seals with the motor drive unit.
10
BEST AVAILABLE COPY
43
{44}
# PS3500™ and
Mini Motor
Drive Units
## Cleaning and Sterilization
After each operation, disconnect the blade assembly from the PS3500 Motor Drive unit. Do not disconnect the power cord from the motor. Rinse the Motor Drive unit thoroughly. Be sure to run water through the drain tube of the Motor Drive unit. Ensure that the flow control lever is open to allow flushing through the entire length of the fluid channel. You must clear all debris from these areas. For detailed instructions see Guideline for Sterilization in the back of this manual.
### Do not disinfect or sterilize the Control Unit or the Foot Pedal.
After autoclaving your Motor Drive unit with its attached power cord, let them cool slowly to room temperature. This will take at least thirty minutes. Draping with a cold, damp sterile cloth can reduce the cooling time. Do not immerse a hot Motor Drive in cold water. A spare power cord is included with each system.
If the power cord is damaged it can be removed from the Motor Drive unit. Call our Sales Department to order a replacement power cord with installation instructions (part no. 3477).
## Replacement Cord
1. Using standard adjustable wrench, grasp opposing flat surface of stainless steel ring at proximal end of Motor Drive unit and loosen by turning it counterclockwise. Then remove old cord from unit (M).
2. Before replacing the cord, clean and dry the connector pins and baseplate of the Motor Drive unit. Improper alignment of the connector pins and Motor Drive unit will result in damage to the electronics.
3. Carefully align the sockets of the replacement cord with the pins of the unit (N).
4. Push the cord on straight, slide the ring up to the Motor Drive unit, and turn the ring clockwise by hand. Slide the wrench over the opposing flat surfaces of the stainless steel ring and securely tighten (O).
5. Test your unit according to the operating instructions in this manual. Refer to "Assembling the System".



BEST AVAILABLE COPY
11
{45}
# Troubleshooting
| When diagnostics say... | The problem may be... | And this may help... |
| --- | --- | --- |
| MOTOR SELECTOR | The Motor Selector Switch is in between positions | Turn to the desired setting "PS3500™", "Mini", or "Universal" |
| FOOTSWITCH | The Control Unit senses the Footswitch plug loose or missing | Check to see if the Footswitch cord is securely attached |
| MOTOR MISSING | The Control Unit senses the PS3500 Motor Drive unit missing | Check to see if the PS3500 Motor Drive cord is securely attached |
| CHECK MOTOR (only when Footswitch is depressed) | The Motor Selector Switch does not match the motor which has been plugged into the Control Unit If the Motor Selector and Motor Drive unit both match, the cord to the Motor Drive unit may be faulty. If the Motor Selector Switch and Motor Drive unit match and the cord has been replaced, and the message CHECK MOTOR persists | Change the Motor Selector Switch to the appropriate setting Replace the Motor Drive cord Call the Dyonics Service Department |
| HI-TEMP | The temperature sensor is indicating a high temperature condition in the Control Unit. After clearing the ventilation and cooling for 10 minutes HI-TEMP message remains on the Diagnostic Display | Check to make sure the ventilation from the back of the unit has not been covered. Clear drapes and reposition Control Unit to maximize ventilation. Call the Dyonics Service Department |
| DATA NOT READY | The PS3500 Motor Drive is not sending blade code data to the Control Unit | Call the Dyonics Service Department |
| CONTROL UNIT ER1 | Control Unit electronics failure. The Motor Drive unit will have no power | Call the Dyonics Service Department |
| DISPLAY ERROR | The fluorescent display is not ready to receive data | Reset system by turning power "off" then "on" |
| RESERVED | An inappropriate blade has been inserted into the PS3500 Motor Drive unit If a Dyonics blade is attached to the PS3500 Motor Drive unit, the blade may have a faulty sensor | Check to insure only a Dyonics blade is in the PS3500 Motor Drive unit Replace the Dyonics blade with another blade. Return faulty blade to the Dyonics Service Dept. |
| SWITCH ERROR | The Speed Control Switches or Footswitch have become corroded The Footswitch has been replaced and the message SWITCH ERROR persists | Replace Footswitch Call the Dyonics Service Department |
12
BEST AVAILABLE COPY
45
{46}
| When diagnostics say... | The problem may be... | And this may help... |
| --- | --- | --- |
| SERVICE REQ'D | The Control Unit has reached the suggested time table for the first service maintenance | Call the Dyonics Service Department |
| NO BLADE (only when Footswitch is depressed) | The Footswitch has been depressed when there is no blade in the PS3500™ Motor Drive unit or the Universal AutoSensor is missing from the Universal Motor Drive unit | Insert a Dyonics blade into the desired Motor Drive unit or insert the AutoSensor into the Universal Motor Drive unit |
| SYSTEM STATUS OK | No errors have been detected in the system. Message appears only when Speed Control Switches have been depressed simultaneously | |
| BLADE CODE ERROR | There is an electronics error in the Motor Drive which is unable to read the blade code | Call the Dyonics Service Department |
| The blade name in the Diagnostic Display does not match blade style in the PS3500 Motor Drive unit | A faulty sensor within blade The blade has been replaced. The Motor Drive may have faulty electronics | Replace blade with another of the same style Call the Dyonics Service Department |
| Miscellaneous Troubleshooting | Problem | Solution |
| --- | --- | --- |
| If the video picture becomes cloudy due to poor suction | Suction Control Lever may be on minimum If Suction Control Lever is on maximum the blade/motor may be clogged with tissue If Suction Control Lever is on maximum and the blade/motor are clear but suction is still poor | Turn Suction Control Switch to maximum Disconnect blade and clear blade and Motor Drive unit Call the Dyonics Service Department |
| If the PS3500 System will not turn on | The Circuit Breaker Switch may have been turned off If the Circuit Breaker is in the "ON" position and the system will not turn on with the Power Switch If the Circuit Breaker continues to trip | Turn Circuit Breaker located on rear panel to the "ON" position Call the Dyonics Service Department Call the Dyonics Service Department |
| If the Diagnostic Display is not functioning in its normal manner | The Diagnostics may have to be reset | Turn the system "off" and then "on" with the Power Switch |
| If the Motor Drive unit abruptly loses power after continuous heavy usage | The over-current protector has been activated | The system will recover automatically and operate normally as soon as the heavy usage is reduced |
BEST AVAILABLE COPY
A
46
13
{47}
# Cleaning, Sterilization, and Disinfection Guidelines
| Component | Cleaning | Sterilization/Disinfection |
| --- | --- | --- |
| Reusable Blades, Abraders | Clean with mild detergent and brush. Rinse thoroughly. Dry completely and lubricate before storing. During use, lubricate with lubricant compatible with chosen sterilization method | 1st Choice – Steam (gravity displacement or prevacuum) 270°F. 3 min. at temperature 2nd Choice – ETO per sterilizer instructions 3rd Choice – Soak in disinfectant per manufacturer's instructions |
| Trocars, Obturators, Cannulas | Clean with mild detergent and brush. Rinse and dry thoroughly | 1st Choice – Steam (gravity displacement or prevacuum) 270°F. 3 min. at temperature 2nd Choice – ETO per sterilizer instructions 3rd Choice – Soak in disinfectant per manufacturer's instructions |
| Disposable Blades, Abraders | Not intended for reuse | Safety may be compromised through the resterilization of a disposable blade. The plastic molding may be distorted and the lubricant between the inner and outer blade may be reduced after resterilization, compromising the proper rotation of the blade |
| Motor Drive unit PS3500™ Mini Universal | Clean unit thoroughly with soapy water. Unit may be immersed. Clean drain tube with brush. Rinse thoroughly with water. Do not use saline or solvents such as alcohol or acetone. Insure that the suction control valve is open | 1st Choice –…