K232521 · Blue Ortho · OLO · Nov 16, 2023 · Neurology
Device Facts
Record ID
K232521
Device Name
ExactechGPS® Total Ankle Application
Applicant
Blue Ortho
Product Code
OLO · Neurology
Decision Date
Nov 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The ExactechGPS is intended for use during preoperative planning and during orthopedic surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures provided that the requred anatomical landmarks can be identified on the patient's preoperative CT scan. The ExactechGPS Total Ankle Case Manager Application is specifically indicated for creating Total Ankle Navigation cases, uploading CT scans for reconstruction, download the reconstruction and export the reconstructed case to the Total Ankle Navigation Application. The ExactechGPS Total Ankle Navigation is specifically indicated for Total Ankle Arthroplasty using the Vantage ankle system to aid the surgeon in locating anatomical structures and aligning the tibial and talar components with the anatomical structures.
Device Story
ExactechGPS Total Ankle Application is an image-guided surgical navigation system for total ankle arthroplasty. It utilizes patient preoperative CT scan data, which undergoes segmentation to reconstruct bone models. The system comprises two software applications: Case Manager for case creation and reconstruction export, and Navigation for intraoperative guidance. The system operates on the previously cleared Blue Ortho ExactechGPS hardware platform (K213877), including trackers rigidly fixed to patient bone, a station, and a camera. Surgeons use the system to locate anatomical structures and align tibial and talar components of the Vantage ankle system. The device provides real-time navigation during surgery, assisting the surgeon in achieving accurate implant alignment, which may improve surgical outcomes and patient benefit. It is intended for use in clinical settings by orthopedic surgeons.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by in-house software verification testing to ensure design outputs meet requirements and software validation to ensure specifications conform to user needs.
Technological Characteristics
Image-guided surgical navigation system. Hardware platform includes trackers, station, and camera (cleared under K213877). Software-based navigation using preoperative CT scan data for bone model reconstruction. Connectivity via system components. No changes to hardware platform, localization system, or computer language compared to predicate.
Indications for Use
Indicated for patients undergoing Total Ankle Arthroplasty using the Vantage ankle system. Used by orthopedic surgeons to aid in locating anatomical structures and aligning tibial and talar components based on preoperative CT scan landmarks.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
November 16, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Blue Ortho Matthieu Coic QA RA Director 22 Chemin du Vieux Chene Meylan, 38240 France
Re: K232521
Trade/Device Name: ExactechGPS® Total Ankle Application Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: August 18, 2023 Received: August 18, 2023
Dear Matthieu Coic:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali-S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232521
Device Name ExactechGPS® Total Ankle Application
#### Indications for Use (Describe)
The ExactechGPS is intended for use during preoperative planning and during orthopedic surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures provided that the requred anatomical landmarks can be identified on the patient's preoperative CT scan.
The ExactechGPS Total Ankle Case Manager Application is specifically indicated for creating Total Ankle Navigation cases, uploading CT scans for reconstruction, download the reconstruction and export the reconstructed case to the Total Ankle Navigation Application.
The ExactechGPS Total Ankle Navigation is specifically indicated for Total Ankle Arthroplasty using the Vantage ankle system to aid the surgeon in locating anatomical structures and aligning the tibial and talar components with the anatomical structures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BlueOrtho, a company specializing in augmented orthopedics. The logo features a stylized blue "C" shape on the left, followed by the text "BlueOrtho" in a clean, sans-serif font. Below "BlueOrtho" is the tagline "AUGMENTED ORTHOPAEDICS" in smaller, lighter font, indicating the company's focus on advanced orthopedic solutions.
### ExactechGPS® Total Ankle Application 510(k) Summary of Safety and Effectiveness
## I. SUBMISSION DATE
August 18, 2023
### II. SUBMITTER
BLUE ORTHO 22 Chemin du Vieux Chêne 38240 Meylan France Phone: +33 (0)4 58 00 35 25 Contact person: Matthieu COIC - Mail: matthieu.coic@blue-ortho.com
### III. US LOCAL AGENT
Exactech, Inc. 2320 NW 66th Ct. Gainesville, FL. 32653 Phone: 352-377-1140
## IV. INFORMATION ON DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED
#### Primary Predicate
510(k) Number Trade or Proprietary Model Name ExactechGPS Total Shoulder Application #K213546 This predicate has not been subject to a design-related recall.
Manufacturer Blue Ortho
# V. PROPOSED DEVICE DESCRIPTION
Trade or Proprietary or Model Name(s): ExactechGPS® Total Ankle Application Common Name: Surgical navigation system
Classification Name: Orthopedic Stereotaxic Instrument (21 CFR 882.4560, product code OLO); Medical Image Communications Device (21 CFR 892.2020, product code LMD)
Classification: Class II
#### Device Description:
The ExactechGPS Total Ankle Application proposed by Blue Ortho in this submission is new clinical software application.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BlueOrtho, a company specializing in augmented orthopedics. The logo features a stylized blue "C" shape on the left, followed by the text "BlueOrtho" in a clean, sans-serif font. Below "BlueOrtho" is the text "AUGMENTED ORTHOPAEDICS" in a smaller font size.
The ExactechGPS Total Ankle Application (TAA) system is intended to be used during stereotaxic surqical procedures to aid the surgeons in locating anatomical structures and aligning the endoprosthesis with the patient bony anatomy provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.
The ExactechGPS Total Ankle Application is an Image Guided Surgery, or Navigation, system designed to guide surgeons during the preparation of the tibia and talar bones as part of a total ankle arthroplasty procedure. The ExactechGPS Total Ankle Application requires patient CT-scan data to undergo segmentation prior to being imported into the software, as part of reconstructing the bone model.
It encompasses two software applications:
- The ExactechGPS Total Ankle Case Manager Application is specifically indicated for creating Total Ankle Navigation cases, uploading CT scans for reconstruction, download the reconstruction and export the reconstructed case to the Total Ankle Navigation Application.
- -The ExactechGPS Total Ankle Navigation Application is specifically indicated for Total Ankle Arthroplasty using the Vantage ankle system to aid the surgeon in locating anatomical structures and aligning the tibial and talar components with the anatomical structures.
The ExactechGPS Total Ankle Navigation is working on the Blue Ortho ExactechGPS System cleared per 510(k) #K213877. The system is composed of ExactechGPS Trackers, which are rigidly fixed to the patient bony anatomy, and other components (ExactechGPS Station and ExactechGPS camera) that are not in contact with the patient.
In the predicate device, a planning application is provided and indicated for preoperative planning of surgical procedure. This submission does not propose planning pre-operative planning application. The others main principles of operation and technological principles between the predicate and the proposed device are similar.
#### VI. INDICATIONS FOR USE
The ExactechGPS is intended for use during preoperative planning and during orthopedic surgery to aid the surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.
The ExactechGPS Total Ankle Case Manager Application is specifically indicated for creating Total Ankle Navigation cases, uploading CT scans for reconstruction, download the reconstruction and export the reconstructed case to the Total Ankle Navigation Application.
The ExactechGPS Total Ankle Navigation Application is specifically indicated for Total Ankle Arthroplasty using the Vantage ankle system to aid the surgeon in locating
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BlueOrtho Augmented Orthopaedics. The logo features a blue square with a stylized "C" shape inside it. To the right of the square is the text "BlueOrtho" in a matching blue color. Below "BlueOrtho" is the text "AUGMENTED ORTHOPAEDICS" in a smaller font size.
anatomical structures and aligning the tibial and talar components with the anatomical structures.
#### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
This proposed submission proposes the following differences compare to the ExactechGPS Total Shoulder Application predicate device:
- Different Indications for Use, as the patient joint treated during the surgical ● procedure in the proposed device is the ankle, and is the shoulder in the predicate device.
- In the predicate device, a planning application is provided and indicated for preoperative planning of surgical procedure. This submission does not propose planning pre-operative planning application, but only a software application to manage surgical cases created by the user.
- Integration in the ExactechGPS Total Ankle Navigation Application of the Exactech total ankle systems, including Vantage Fixed Bearing or Mobile Bearing, Vantage 3D and Vantage 3D+, cleared in the following 510(k) submissions:
| K152217 | Vantage Total Ankle System |
|---------|-----------------------------------------------|
| K183343 | Vantage Total Ankle Flat Cut Talar Components |
| K230717 | Vantage 3D/3D+ |
At a high level, the subject and predicate devices are based on the following same technological elements:
- The proposed device do not affect general device features and dimensions.
- The proposed device do not change the device computer language or other . basic fundamental technologies.
- No changes to the hardware platform and localization system are proposed by ● this submission.
### VIII. PERFORMANCE DATA
Testing information demonstrating safety and effectiveness of the ExactechGPS Total Ankle Application is supported by testing that was conducted in-house.
This submission includes or references the following non-clinical testing:
- Software verification testing to ensure all design outputs meet all specified ● requirements
- Software validation to ensure software specifications conform to user needs and intended uses
# IX. SUBSTANTIAL EQUIVALENCE CONCLUSION
A comparison of specific features included in this submission demonstrates the proposed ExactechGPS Total Ankle Application is substantially equivalent to the cited predicate cleared per # K213546. The devices share identical intended use, identical general design features and basic fundamental scientific technology.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.