K230717 · Exactech, Inc. · HSN · Oct 12, 2023 · Orthopedic
Device Facts
Record ID
K230717
Device Name
Exactech® Vantage® Total Ankle System
Applicant
Exactech, Inc.
Product Code
HSN · Orthopedic
Decision Date
Oct 12, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3110
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vantage® Total Ankle System is indicated for patients with ankle joints damaged by severe rheumatoid, posttraumatic, or degenerative arthritis. It is also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. The Vantage® Total Ankle System is indicated for cemented use only.
Device Story
System of orthopedic implants for total ankle replacement; includes tibial plates, tibial inserts, talar components, and locking clips. Subject device introduces additively manufactured 3D and 3D+ tibial plates and locking clips as line extensions. Implants replace damaged ankle joint surfaces; intended for cemented fixation. Used by orthopedic surgeons in clinical/OR settings. Provides structural replacement of ankle joint; restores joint function and mobility. Benefits patients by alleviating pain and improving joint function associated with severe arthritis or failed prior reconstructions.
Implants composed of Ti-6Al-4V ELI conforming to ASTM F3001-14 or ASTM F136-13. Features include additively manufactured porous surfaces. Provided sterile for single use. No software or electronic components.
Indications for Use
Indicated for patients with severe rheumatoid, post-traumatic, or degenerative ankle arthritis, or revision of failed previous reconstructions with sufficient bone stock and soft tissue integrity. Cemented use only.
Regulatory Classification
Identification
An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 12, 2023
Exactech, Inc. Pedro Ravelo Regulatory Affairs Specialist 2320 NW 66th Ct. Gainesville, Florida 32653
Re: K230717
Trade/Device Name: Exactech® Vantage® Total Ankle System Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSN Dated: September 13, 2023 Received: September 13, 2023
Dear Pedro Ravelo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Lixin Liu -S
Lixin Liu. Ph.D Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230717
Device Name Exactech® Vantage® Total Ankle System
#### Indications for Use (Describe)
The Vantage® Total Ankle System is indicated for patients with ankle joints damaged by severe rheumatoid, posttraumatic, or degenerative arthritis. It is also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.
The Vantage® Total Ankle System is indicated for cemented use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Exactech® Vantage® Total Ankle System Traditional 510(k) - 510(k) Summary
| | | Exactech, Inc.<br>2320 NW 66th Court<br>Gainesville, FL 32653 |
|----------------------|--|----------------------------------------------------------------------------------|
| Sponsor | | Phone: (352) 377-1140<br>Fax: (352) 378-2617<br>FDA Establishment Number 1038671 |
| Contact Person | | Pedro Ravelo<br>Regulatory Affairs Specialist |
| Date | | 14 March 2023 |
| Proprietary Name | | Exactech® Vantage® Total Ankle System |
| Common Name | | Ankle Arthroplasty Prosthesis |
| Regulation Number / | | 21 CFR 888.3110 – Ankle joint metal/polymer semi- |
| Classification Name | | constrained cemented prosthesis |
| Device Class | | Class II |
| Product Code | | HSN - Prosthesis, Ankle, Semi-Constrained, Cemented,<br>Metal/Polymer |
| Classification Panel | | Orthopedic |
## 510(K) SUMMARY
#### Device Description
The Vantage® Total Ankle System is a system of orthopedic implants intended for total ankle replacement. The system includes Tibial Plates, Tibial Inserts, Talar Components, and Locking Clips. This submission proposes three new components as line extensions of the Vantage Total Ankle System, which are modifications of their respective predicate components:
- 3D Tibial Plates
- 3D+ Tibial Plates ●
- 3D/3D+ Locking Clips ●
The proposed devices have the same indications for use and similar design characteristics as their predicate devices (cleared under K152217). The proposed tibial plates differ from the predicate in that they are additively manufactured, have minor geometrical modifications, and are offered in additional sizes. The proposed locking clips differ from the predicate in that they are offered in additional sizes.
{4}------------------------------------------------
### Exactech® Vantage® Total Ankle System Traditional 510(k) - 510(k) Summarv
#### Indications for Use
The Vantage® Total Ankle System is indicated for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis. It is also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.
The Vantage® Total Ankle System is indicated for cemented use only.
#### Legally Marketed Devices to Which Substantial Equivalence Is Claimed
| 510(k) Number | Trade or Proprietary Model Name | Manufacturer |
|---------------|---------------------------------------|---------------|
| K152217 | Exactech® Vantage® Total Ankle System | Exactech, Inc |
#### Legally Marketed Reference Devices
| 510(k) Number | Trade or Proprietary Model Name | Manufacturer |
|---------------|-------------------------------------------------|---------------|
| K192097 | Exactech® Equinoxe® Stemless Humeral Components | Exactech, Inc |
#### Substantial Equivalence Discussion
The proposed Vantage 3D Tibial Plates, 3D* Tibial Plates, and 3D/3D* Locking Clips are made from Ti-6A1-4V ELI that conforms to ASTM F3001-14 or ASTM F136-13. The subject devices and their respective predicates have similar technological characteristics and the minor differences do not raise any different issues of safety and effectiveness. The rationale for substantial equivalence is based on consideration of the following device use, design characteristics, and performance requirements:
#### Indications for Use
The proposed and predicate devices have the same indications for use.
#### Materials
The proposed and predicate devices are composed of the same or similar biocompatible materials.
Design Features The proposed and predicate devices have the same or similar design features.
#### Dimensions
The proposed and predicate devices are dimensionally comparable.
#### Sterilization
The proposed and predicate devices are provided sterile for single use only.
#### Performance Requirements
The proposed and predicate devices conform to the same recognized performance standards for total ankle replacement devices.
{5}------------------------------------------------
### Exactech® Vantage® Total Ankle System Traditional 510(k) - 510(k) Summary
#### Non-Clinical Testing
The following non-clinical testing and engineering analyses were performed to demonstrate that the Exactech® Vantage® 3D Tibial Plates, 3D+ Tibial Plates and 3D/3D+ Locking Clips function as intended and are substantially equivalent to the identified predicate devices:
- Tibial Plate Fatigue Testing per ASTM F2665 ●
- Tibial Plate Fixation Stability Testing ●
- Locking Mechanism Fatigue Testing ●
- Residual Particulate Count Studies
- Porous Surface Characterization Testing ●
- Stereological Evaluation per ASTM F1854 O
- Static Shear Testing per ASTM F1044 O
- Tension Testing per ASTM F1147 O
- Abrasion Resistance Testing per ASTM F1978 O
- Shear Fatigue Testing per ASTM F1160 O
- Compression Testing o
The results of this non-clinical testing show that the strength of the Vantage 3D Tibial Plates, 3D Tibial Plates and 3D/3D Locking Clips is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices. Pyrogenicity testing was also conducted and demonstrates that the endotoxin limit is less than 20 EU/device.
#### Conclusion
Based on consideration of indications for use, technological characteristics, and results of non-clinical testing, it was concluded that the Exactech® Vantage® Total Ankle System is substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.