K152217 · Exactech, Inc. · HSN · Mar 10, 2016 · Orthopedic
Device Facts
Record ID
K152217
Device Name
Exactech Vantage Total Ankle System
Applicant
Exactech, Inc.
Product Code
HSN · Orthopedic
Decision Date
Mar 10, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3110
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vantage™ Total Ankle System is indicated for patients with ankle joints damaged by severe rheumatoid, post- traumatic, or degenerative arthritis. It is also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. The Vantage™ Total Ankle System is indicated for cemented use only.
Device Story
Fixed-bearing total ankle replacement system; comprises four sub-components: tibial plate (Ti-6Al-4V ELI), tibial insert (UHMWPE), locking component (Ti-6Al-4V ELI), and talar component (Co-28Cr-6Mo). Components feature commercially pure titanium coating on bone-contacting surfaces. Implanted by surgeons in clinical settings to replace damaged ankle joints; provides stability and range of motion. System includes instrumentation set (trials, rasps, cutting blocks) to assist surgical implantation. Designed to maximize bony support and provide anterior/posterior, medial/lateral, and rotational stability. Benefits patients by restoring joint function in cases of severe arthritis or failed prior reconstructions.
Clinical Evidence
Bench testing only. Evidence includes sizing studies, locking integrity testing, fatigue analysis, wear evaluation, contact area/stress studies, constraint evaluation, bone stability testing, range of motion studies, and finite element analysis.
Technological Characteristics
Fixed-bearing total ankle prosthesis. Materials: Ti-6Al-4V ELI (tibial plate/locking component), UHMWPE (tibial insert), Co-28Cr-6Mo (talar component). Surface: Commercially pure titanium coating. Provided sterile for single use. Mechanical design for cemented fixation.
Indications for Use
Indicated for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis, or for revision of failed previous reconstructions with sufficient bone stock and soft tissue integrity. Indicated for cemented use only.
Regulatory Classification
Identification
An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping to create a sense of depth and connection. The profiles are rendered in a simple, line-art style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2016
Exactech Incorporated Shing Jen Tai, Ph.D. Senior Regulatory Affairs Specialist 2320 N.W. 66th Court Gainesville, Florida 32653
Re: K152217
Trade/Device Name: Exactech® Vantage™ Total Ankle System Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: HSN Dated: February 16, 2016 Received: February 17, 2016
Dear Dr. Jen Tai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | K152217 Page 1 of 1 |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration | Form Approved: OMB No. 0910-0120 |
| <b>Indications for Use</b> | Expiration Date: January 31, 2017 |
| | See PRA Statement below. |
| 510(k) Number ( <i>if known</i> ) | K152217 |
| Device Name | Exactech Vantage Total Ankle System |
| Indications for Use ( <i>Describe</i> ) | The Vantage <sup>TM</sup> Total Ankle System is indicated for patients with ankle joints damaged by severe rheumatoid, post- traumatic, or degenerative arthritis. It is also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. |
| | The Vantage <sup>TM</sup> Total Ankle System is indicated for cemented use only. |
| Type of Use (Select one or both, as applicable) | <div> <span> <svg height="12" width="12"> <rect height="12" style="fill:transparent;stroke:black;stroke-width:1" width="12"></rect> <line style="stroke:black;stroke-width:1" x1="0" x2="12" y1="0" y2="12"></line> <line style="stroke:black;stroke-width:1" x1="0" x2="12" y1="12" y2="0"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> <div> <span> <svg height="12" width="12"> <rect height="12" style="fill:transparent;stroke:black;stroke-width:1" width="12"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
ງ ແດ່ຖ່ວອຊີ
I 10 I ຈອງປະຊາ
{3}------------------------------------------------
# Exactech® Vantage™ Total Ankle System Traditional 510(k)
#### 510(k) Summary
| Company: | Exactech®, Inc<br>2320 NW 66th Court<br>Gainesville, FL 32653 |
|----------------------|-------------------------------------------------------------------------------------|
| Date: | March 4, 2016 |
| Contact Person: | Shing Jen Tai, PhD<br>Sr. Regulatory Affairs Specialist |
| | Phone: (352) 327-4638<br>Fax: (352) 378-2617 |
| Proprietary Name: | Exactech® VantageTM Total Ankle System |
| Common Name: | Total Ankle Prosthesis |
| Classification Name: | 21 CFR 888.3110, Ankle Joint Metal/Polymer Semi-<br>Constrained Cemented Prosthesis |
| | Class II |
| Product Code: | HSN |
#### Legally Marketed Device to Which Substantial Equivalence Is Claimed:
- Salto Talaris™ Total Ankle Prosthesis (K090076) from Tornier
- INFINITY® Total Ankle System (K123954, and K140749 line extension to ● devices cleared in K123954) from Wright Medical Technology, Inc.
#### Device Description
Exactech Vantage Total Ankle System is a fixed-bearing total ankle replacement device that is comprised of four sub-components, a tibial plate, a tibial insert, a locking component, and a highly polished talar component that are assembled to create the subject total ankle replacement system. The tibial plate is manufactured from titanium alloy (Ti-6Al-4V ELI) with commercially pure titanium coating on the bone-contacting surface. The tibial insert is made of ultra-high molecular weight polyethylene. The highly polished talar component is manufactured from cobalt-chromium alloy (Co-28Cr-6Mo) with commercially pure titanium coating on the bone-contacting surface. The locking component is made of titanium alloy (Ti-6Al-4V ELI). Each of the four sub-components is offered in multiple sizes to accommodate the various anatomical needs of a patient's ankle joint. With the exception of the locking component, all components are anatomically designed and offered in left and right implant components. The overall design goals of the Exactech Vantage Total Ankle System are to maximize bony support while allowing impingement-free range of motion, and provide anterior/posterior,
{4}------------------------------------------------
# Exactech® Vantage™ Total Ankle System Traditional 510(k)
medial/lateral, and rotational stability when implanted in the resected/prepared distal tibia and proximal talus. Exactech Vantage Total Ankle System device is indicated for cemented use only.
The Exactech Vantage Total Ankle System is accompanied by a complete instrumentation set including trial and rasp/cutting block system to assist surgeons in implantation of the device.
### Indications for Use
The Vantage™ Total Ankle System is indicated for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis. It is also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.
The Vantage™ Total Ankle System is indicated for cemented use only.
## Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following device use and characteristics:
- . Indications for Use. The proposed Exactech Vantage Total Ankle System and the predicate devices have similar and comparable indications for use.
- Materials/Surface Coatings. The proposed Exactech Vantage Total Ankle . System and the predicate devices are composed of same or similar, biocompatible substrate materials, and the same surface coatings for permanent implants.
- . Design Features. The proposed Exactech Vantage Total Ankle System and the predicate devices share similar design features.
- Dimensions. The proposed Exactech Vantage Total Ankle System and the predicate devices are dimensionally comparable.
- Sterilization. The proposed Exactech Vantage Total Ankle System and the predicate devices are provided sterile for single use only.
- Performance Requirements. The proposed Exactech Vantage Total Ankle ● System and the predicate devices conform to recognized performance standards for total ankle joint replacement devices.
## Non-Clinical Testing
The following sizing studies, mechanical testing, and modeling analyses were performed to demonstrate that the Exactech Vantage Total Ankle System performs as intended and is substantially equivalent to the identified predicate devices:
- Sizing Studies
- Locking integrity Testing ●
- Fatigue Analysis ●
- Wear Evaluation
- Contact Area/Contact Stress Study ●
{5}------------------------------------------------
# Exactech® Vantage™ Total Ankle System Traditional 510(k)
- Constraint Evaluation ●
- . Bone Stability Testing
- . Range of Motion Study
- . Finite Element Analysis
# Substantial Equivalence Conclusion
Based on consideration of indications for use, technological characteristics, and results of combined mechanical testing, modeling analyses, and sizing studies described above, it was concluded that Exactech Vantage Total Ankle System demonstrates substantial equivalence to the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.