ExactechGPS Total Shoulder Application, Equinoxe Planning Software
K213546 · Blue Ortho · OLO · Jan 6, 2022 · Neurology
Device Facts
Record ID
K213546
Device Name
ExactechGPS Total Shoulder Application, Equinoxe Planning Software
Applicant
Blue Ortho
Product Code
OLO · Neurology
Decision Date
Jan 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The ExactechGPS is intended for use during preoperative planning and during orthopedic surgery to aid the surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures, provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan. The ExactechGPS Total Shoulder Planning Application is specifically indicated for preoperative planning of Total Shoulder Arthroplasty using the Equinoxe system. The ExactechGPS Total Shoulder Planning Application permits to visualize, measure and reconstruct anatomical structures in order to select and place the glenoid and humeral components. The ExactechGPS Total Shoulder Navigation is specifically indicated for Total Shoulder Arthroplasty using the Equinoxe system to aid the surgeon in locating anatomical structures and aligning the glenoid component with the anatomical structures.
Device Story
ExactechGPS is an image-guided surgical navigation and preoperative planning system for Total Shoulder Arthroplasty. Input consists of patient CT-scan data, which undergoes segmentation to reconstruct 3D bone models. Surgeons use the software in a virtual environment to visualize, measure, and plan the positioning of Equinoxe glenoid and humeral implants. During surgery, the system provides navigation to assist the surgeon in locating anatomical structures and aligning the glenoid component. The device is used in the clinical/surgical setting by orthopedic surgeons. Output includes visual guidance and measurements that inform surgical decision-making regarding implant selection and placement. The system aims to improve the accuracy of component alignment during arthroplasty procedures.
Clinical Evidence
No clinical data provided. Safety and effectiveness are supported by in-house non-clinical software verification and validation testing to ensure design outputs meet specifications and user needs.
Technological Characteristics
Image-guided surgical navigation and planning software. Requires preoperative CT-scan data. Features 3D bone model segmentation and reconstruction. Registration algorithm used for anatomical alignment. No changes to hardware platform or computer language. Software-based system.
Indications for Use
Indicated for patients undergoing Total Shoulder Arthroplasty using the Equinoxe system. Used for preoperative planning and intraoperative navigation to assist surgeons in locating anatomical structures and aligning glenoid and humeral components.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Exactech Equinoxe Small Reverse Shoulder System (K180632)
Submission Summary (Full Text)
{1}------------------------------------------------
January 6, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Blue Ortho Matthieu Coic QA RA Director 22 Chemin du Vieux Chene Meylan, 38240 France
Re: K213546
Trade/Device Name: ExactechGPS Total Shoulder Application, Equinoxe Planning Software Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: October 29, 2021 Received: November 8, 2021
Dear Matthieu Coic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213546
Device Name ExactechGPS® Total Shoulder Application
#### Indications for Use (Describe)
The ExactechGPS is intended for use during preoperative planning and during orthopedic surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.
The ExactechGPS Total Shoulder Planning Application is specifically indicated for pre-operative planning of Total Shoulder Arthroplasty using the Equinoxe system. The ExactechGPS Total Shoulder Planning Application permits to visualize, measure and reconstruct anatomical structures in order to select and place the glenoid and humeral components.
The Exacted GPS Total Shoulder Navigation is specifically indicated for Total Shoulder Arthroplasty using the Equinoxe system to aid the surgeon in locating anatomical structures and aligning the glenoid component with the anatomical structures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Blue Ortho. The logo consists of a blue square with a white "C" inside of it. To the right of the square is the text "Blue Ortho" in blue, with the words "Augmented Orthopaedics" in smaller letters underneath.
## ExactechGPS® Total Shoulder Application 510(k) Summary of Safety and Effectiveness
# I. SUBMISSION DATE
October 29, 2021
# II. SUBMITTER
BLUE ORTHO 22 Chemin du Vieux Chêne 38240 Meylan France Phone: +33 (0)4 58 00 35 25 Contact person: Matthieu COIC - Mail: matthieu.coic@blue-ortho.com
### III. US LOCAL AGENT
Exactech, Inc. 2320 NW 66th Ct. Gainesville, FL. 32653 Phone: 352-377-1140
# IV. INFORMATION ON DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED
### Primary Predicate
510(k) Number Trade or Proprietary Model Name #K173372 ExactechGPS® Total Shoulder Application This predicate has not been subject to a design-related recall.
Manufacturer Blue Ortho
### Additional Predicates
510(k) Number Trade or Proprietary Model Name BLUEPRINT™ Patient Specific #K203315 Instrumentation
Manufacturer TORNIER SAS
# V. PROPOSED DEVICE DESCRIPTION
Trade or Proprietary or Model Name(s): ExactechGPS® Total Shoulder Application Common Name: Surgical navigation system
Classification Name: Orthopedic Stereotaxic Instrument (21 CFR 882.4560, product code OLO); Picture Archiving & Communications System (21CFR 892.2050, product code LLZ)
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for BlueOrtho, a company that specializes in augmented orthopedics. The logo features a stylized letter "C" in blue, followed by the words "BlueOrtho" in a sans-serif font, also in blue. Below "BlueOrtho" are the words "AUGMENTED ORTHOPAEDICS" in a smaller, sans-serif font. The logo is clean and modern, and the use of blue gives it a sense of trustworthiness and professionalism.
#### Classification: Class II Device Description:
The ExactechGPS Total Shoulder Application proposed in this submission is a modification of the ExactechGPS Total Shoulder Application cleared per 510(k) #K173372.
The ExactechGPS Total Shoulder Application is an Image Guided Surgery, or Navigation, system designed to quide surgeons during the preparation of the glenoid as part of a total shoulder arthroplasty procedure. The ExactechGPS Total Shoulder Application also offers a preoperative planning feature that enables surgeons to plan a surgical intervention by evaluating implant size, type, and positioning using reconstructed patient bone models in a virtual environment. The ExactechGPS Total Shoulder Application requires patient CT-scan data to undergo segmentation prior to being imported into the software, as part of reconstructing the bone model, for both navigation and planning.
In the predicate device, the planning step was only indicated for pre-operative planning of the dlenoid part of the Equinoxe system. This submission proposes the addition of the humeral component. Therefore, the proposed ExactechGPS Total Shoulder Planning Application permits to visualize, measure and reconstruct anatomical structures in order to select and place the glenoid and humeral components.
The proposed ExactechGPS Total Shoulder Planning and Navigation Applications integrate some other modifications compared to the predicate device. These modifications are documented in Letter to Files.
The proposed modifications do not change the ExactechGPS System general intended use, general design features, or basic fundamental scientific technology. No changes to the hardware platform or system accessories are proposed by this submission.
# VI. INDICATIONS FOR USE
The ExactechGPS is intended for use during preoperative planning and during orthopedic surgery to aid the surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures, provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.
The ExactechGPS Total Shoulder Planning Application is specifically indicated for preoperative planning of Total Shoulder Arthroplasty using the Equinoxe system. The ExactechGPS Total Shoulder Planning Application permits to visualize, measure and reconstruct anatomical structures in order to select and place the glenoid and humeral components.
The ExactechGPS Total Shoulder Navigation is specifically indicated for Total Shoulder Arthroplasty using the Equinoxe system to aid the surgeon in locating anatomical structures and aligning the glenoid component with the anatomical structures.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image is the logo for BlueOrtho. The logo consists of a blue square with a white "C" inside of it, followed by the words "BlueOrtho" in blue. Below the words "BlueOrtho" are the words "AUGMENTED ORTHOPAEDICS" in a smaller font.
#### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
This submission proposes the following modifications to the ExactechGPS Total Shoulder Application:
- . Modification of the Indications for Use by proposing an extension of the preoperative feature to the humeral component, and not only the glenoid component. This requires the modification of the ExactechGPS Total Shoulder Planning Application to extend the planning feature to the humeral part and components
- . Extension of segmentation capabilities of Blue Ortho to the reconstruction of the humeral humerus
- Integration in the ExactechGPS Total Shoulder Planning Application of ● Exactech Equinoxe humeral implants cleared in the following 510(k) submissions:
| K042021 | Equinoxe Shoulder System |
|---------|----------------------------------------------------------------|
| K063569 | Equinoxe Reverse Shoulder System |
| K061454 | Exactech Equinoxe Shoulder Stems |
| K082702 | Exactech Equinoxe Reverse Shoulder System+15mm Humeral Adapter |
| | Tray |
| K093275 | Exactech Equinoxe Reverse Shoulder System 36mm Glenosphere And |
| | Humeral Liner |
| K162726 | Exactech® Equinoxe® Preserve Stem |
| K180632 | Exactech Equinoxe Small Reverse Shoulder System |
- Modification of the implementation of the registration algorithm by changing its initialization. This technical change in the registration algorithm has no impact on the global accuracy of the software nor performance.
At a high level, the subject and predicate devices are based on the following same technological elements:
- The modifications do not affect device intended use. ●
- The modifications do not affect general device features and dimensions.
- The modifications do not change the device computer language or other basic ● fundamental technologies.
- The modifications do not affect device accuracy and / or performance.
- No changes to the hardware platform or system accessories are proposed by . this submission.
# VIII. PERFORMANCE DATA
Testing information demonstrating safety and effectiveness of the ExactechGPS Total Shoulder Application is supported by testing that was conducted in-house. This submission includes or references the following non-clinical testing:
- Software verification testing to ensure all design outputs meet all specified ● requirements
- . Software validation to ensure software specifications conform to user needs and intended uses
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for BlueOrtho Augmented Orthopaedics. The logo features a blue square with a white "C" shape inside. To the right of the square is the text "BlueOrtho" in a blue sans-serif font. Below the text is the phrase "AUGMENTED ORTHOPAEDICS" in a smaller, lighter blue font.
# IX. SUBSTANTIAL EQUIVALENCE CONCLUSION
A comparison of specific features included in this submission demonstrates the proposed ExactechGPS Total Shoulder Application is substantially equivalent to the cited predicate cleared per #K173372. The devices share identical intended use, identical general design features and basic fundamental scientific technology.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.