K063569 · Exactech, Inc. · PHX · Feb 23, 2007 · Orthopedic
Device Facts
Record ID
K063569
Device Name
EQUINOXE REVERSE SHOULDER SYSTEM
Applicant
Exactech, Inc.
Product Code
PHX · Orthopedic
Decision Date
Feb 23, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Exactech Equinoxe™ Reverse Shoulder System is indicated to relieve pain and restore function in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe™ Reverse Shoulder System is also indicated for failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
Device Story
The Equinoxe Reverse Shoulder System is a semi-constrained total shoulder prosthesis designed for joint replacement in patients with irreparable rotator cuff deficiency. The system comprises humeral stems (primary/revision), humeral adapter plates, glenospheres, humeral liners, and screws. It is implanted by orthopedic surgeons in a clinical/hospital setting. The device functions by replacing the glenohumeral joint, shifting the center of rotation to improve stability and range of motion while minimizing inferior impingement and scapular notching. Clinical benefit includes pain relief and restoration of shoulder function. The system is compatible with primary press-fit, primary cemented, and long/revision humeral stems.
Clinical Evidence
No clinical data. Evidence consists of bench testing including dynamic loading studies in polyurethane bone substitutes and cadaveric models, comparative geometric analyses of center of rotation and design parameters, and finite element analysis for fracture resistance.
Technological Characteristics
Semi-constrained reverse shoulder prosthesis. Materials: equivalent to industry standards for permanent implants. Components: humeral stems, adapter plates, glenospheres, humeral liners, screws. Sterilization: equivalent to industry standards. Design features: optimized center of rotation and geometry to maximize range of motion and minimize inferior impingement/scapular notching.
Indications for Use
Indicated for skeletally mature individuals with degenerative glenohumeral joint disease and grossly deficient, irreparable rotator cuff, or failed prior glenohumeral joint replacement with rotator cuff loss and superior humeral head migration.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
K063569
# 510(k) Summary
2320 NW 66TH COURT GAINESVILLE, F1 32653
352-377-1140 FAX 352-378-2617
FEB 2 3 2007 Company: Exactech, Inc Date: November 28, 2006 Contact Person: Amnon Talmor, Regulatory Affairs Representative Proprietary Name: Exactech Equinoxe™ Reverse Shoulder System Common Name: Reverse Shoulder Prosthesis Classification Name: Shoulder joint metal/polymer non-constrained cemented prosthesis (21 CFR 888.3650, Class II, Product Code KWT) Prosthesis, Shoulder, Semi-constrained, metal/polymer cemented (21 CFR 888.3660, Class II, Product Code KWS)
# Legally Marketed Devices to Which Substantial Equivalence Is Claimed
- Depuy Delta III Reverse Shoulder (#K021478, #K050315, #K062116) .
- Tornier Reverse Shoulder (#K030941, #K041873, #K050316) .
- Zimmer BF Reverse Shoulder (#K052906, #K060704)
- . Zimmer Anatomica Reverse Shoulder (#K053274)
- . Encore Reverse Shoulder (#K041066, #K051075, #K052086)
## Device Description
The Equinoxe™ Reverse Shoulder System includes a reverse semi-constrained prosthesis for use in total-shoulder joint replacement procedures in cases with an irreparable or nonfunctional totator cuff. The system includes primary and revision humeral stems, various sizes and types of humeral adapter plates, glenospheres, humeral liners, and screws. The reverse shoulder System is designed to function with the Equinoxe primary press-fit, primary cemented, and long/revision humeral stems. All components are supplied sterile.
#### Indications for Use
The Exactech Equinoxe™ Reverse Shoulder System is indicated to relieve pain and restore function in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe™ Reverse Shoulder System is also indicated for failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
# Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use. Equinoxe™ Reverse Shoulder System and predicate devices are intended . for use in total reverse shoulder joint replacement and have similar indications for use statements.
A Great Day in the O.H
Page 1 of 2
EXACTECH exists to inques the economing this activity ond independence. We do his through interest inde-station and connish one commisses of services on services on services
1'+
{1}------------------------------------------------
ે ૮ રે રેડા દિવે
- Materials. Equinoxe " Reverse Shoulder System and predicate devices are composed of . equivalent materials conforming to recognized industry standards for permanent implants.
- Dimensions. Equinoxe™ Reverse Shoulder System and predicate device components ● are available in equivalent size ranges.
- Sterilization processes. Equinoxe™ Reverse Shoulder System and predicate devices are . sterilized using equivalent sterilization processes conforming to recognized industry standards.
- Performance specifications. Equinoxe™ Reverse Shoulder System and predicate . devices conform to recognized performance standards for total shoulder joint replacement devices.
### Summary of Non-Clinical Performance Data
Mechanical tests, engineering analyses, and simulated surgical (sawbones and cadaver) implantations were conducted to demonstrate the safety and effectiveness of the devices that compose the Equinoxe™ Reverse Shoulder System and support the claim of substantial equivalence to the predicates listed above; a summary of these tests and analyses are as follows:
- A dynamic loading study in which the stability of the Equinoxe Reverse Shoulder was . assessed in a polyurethane bone substitute (TR-2006-052) and in a cadaver (TS-2006-024). The glenoid plate/glenosphere micromotion measurements obtained in this study are compared to that associated with the Depuy Delta III and Encore RSP components when subjected to a similar loading pattern.
- A comparative assessment of the glenosphere center of rotation that demonstrates the . location of the center of rotation is similar for the Equinoxe and Delta III designs. This observation in conjunction with the micromotion test results (TR-2006-052 and TS-2006-024) suggest that the low reported incidence of glenoid loosening associated with the Grammont Reverse Shoulder is applicable to the Equinoxe™ Reverse Shoulder design (TR-2006-053).
- A geomettic analysis of the Grammont Reverse Shoulder Prosthesis-an evaluation of . the relationships between prosthetic design parameters and clinical failure modes (TR-2006-028). This assessment was used to optimize the design parameters associated with the Equinoxe design in order to maximize ROM and minimize inferior impingement.
- A geometric analysis vetification study that demonstrates the Equinoxe Reverse . Shoulder achieves an increase in the amount of motion and a decrease in the amount of inferior impingement (a measure of motion and stability, indicative of scapular notching) while maintaining a similar amount of jump distance (a measure of stability, indicative of the probability of dislocation) relative to the Grammont/Delta III design (TR-2006-029).
- A finite element analysis that demonstrates the geometry of the proposed devices is not � subject to fracture when subjected to a worst-case load (TR-2006-039).
### Substantial Equivalence Conclusion
Results from mechanical tests and engineering analyses provided within this 510(k) demonstrate that the Equinoxe™ Reverse Shoulder System is substantially equivalent to the identified predicate devices.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/12 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Exactech, Inc % Mr. Amnon Talmor Regulatory Affairs Representative 2320 NW 66th Court Gainesville, Florida 32653
FEB 2 3 2007
Re: K063569
Trade/Device Name: Equinoxe Reverse Shoulder System Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: II Product Code: KWT Dated: November 27, 2006 Received: November 29, 2006
Dear Mr. Talmor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are cased in to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Amnon Talmor
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Corbace Buehup
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K663569
# Exactech®, Inc.
# Indications for Use
510(k) Number (if known):
Device Name: Equinoxe Reverse Shoulder System
## INDICATIONS FOR USE:
The Exactech Equinoxe™ Reverse Shoulder System is indicated to relieve pain and restore function in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe™ Reverse Shoulder is also indicated for failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
Prescription Use _ X (Part 21 CFR 801 Subpart D) and/or Over-The-Counter Use (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clarlare Buchnd
(Division Signer)
(Division Sign-Off) Division of General, Restorative, and Neurological General, Resto
**510(k) Number** K063569
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.