ICX-Implant System is indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. ICX- Implant System is indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implants with lengths less than 7.0 mm are indicated for delayed loading only. ICX-Implant System CAD CAM abutments are intended for use with dental implants as a support for single unit or multiple unit prostheses in the maxilla of a partially or fully edentulous patient. All digitally designed abutments for use with ICX-Implant System CAD CAM abutments are intended to be manufactured at a medentis medical GmbH validated milling center.
Device Story
ICX-Implant System comprises endosseous dental implants and prosthetic components (abutments, healing caps, screws). Implants feature self-tapping apical threads, tapered body, and internal hex connection; surface is grit-blasted and acid-etched. Components are used by dentists/oral surgeons in clinical settings to restore chewing function in edentulous/partially edentulous patients. CAD/CAM abutments are digitally designed and manufactured at a validated milling center. Device provides structural support for crowns, bridges, or dentures. Benefits include restoration of oral function and aesthetics. System includes various implant lines (Premium, Diamond, Active) and abutment types (straight, angled, overdenture, adhesive base).
Clinical Evidence
No clinical data included. Substantial equivalence supported by non-clinical bench testing: sterilization validation (ISO 11137, ISO 17665), bacterial endotoxin testing (ANSI/AAMI ST72), shelf-life/packaging integrity (ASTM F1929, F88), biocompatibility (ISO 10993), surface characterization (SEM, contact angle), and mechanical dynamic loading (ISO 14801).
Indicated for partially or fully edentulous patients requiring maxillary or mandibular single-unit, multiple-unit, or overdenture dental restorations. Supports immediate loading with good primary stability and appropriate occlusal loading; implants <7.0 mm indicated for delayed loading only. CAD/CAM abutments support single or multiple-unit prostheses.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Straumann® Dental Implant System SLActive and Roxolid Product Families (K130222)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
medentis medical GmbH % Floyd Larson President PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
September 5, 2024
Re: K231566
Trade/Device Name: ICX-Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: August 7, 2024 Received: August 8, 2024
Dear Floyd Larson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name ICX-Implant System
#### Indications for Use (Describe)
ICX-Implant System is indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. ICX- Implant System is indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implants with lengths less than 7.0 mm are indicated for delayed loading only.
ICX-Implant System CAD CAM abutments are intended for use with dental implants as a support for single unit or multiple unit prostheses in the maxilla of a partially or fully edentulous patient. All digitally designed abutments for use with ICX-Implant System CAD CAM abutments are intended to be manufactured at a medentis medical GmbH validated milling center.
| Type of Use (Select one or both, as applicable) | | |
|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------|-----------------------------------------------|
| <table><tr><td></td><td>Beneficial Use (R-1, 31 CFR 336.1(b) and (d)) </td></tr></table> | | Beneficial Use (R-1, 31 CFR 336.1(b) and (d)) |
| | Beneficial Use (R-1, 31 CFR 336.1(b) and (d)) | |
| <table><tr><td></td><td>On-Site Containment Use (31 CFR 336.1(c)) </td></tr></table> | | On-Site Containment Use (31 CFR 336.1(c)) |
| | On-Site Containment Use (31 CFR 336.1(c)) | |
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary medentis medical GmbH ICX-Implant System
September 5, 2024
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | medentis medical GmbH<br>Walporzheimer Strasse 48-52<br>Bad Neuenahr-Ahrweiler, 53474, Germany<br>Telephone +49 2641 9110-171<br>Fax +49 2641 9110-120 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Alexander Scholz, CEO |
| Representative/Consultant | Floyd G. Larson, MS, MBA<br>Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: flarson@paxmed.com; kthomas@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | ICX-Implant System |
|------------------------|---------------------------------------------------------------------------------------------|
| Common Names | Endosseous dental implant |
| Regulation Number | 21 CFR 872.3640 |
| Regulation Name | Endosseous dental implant |
| Regulatory Class | Class II |
| Product Code | DZE |
| Secondary Product Code | NHA |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices) |
| Reviewing Division | Division of Health Technology 1B (Dental and ENT Devices) |
PREDICATE DEVICE INFORMATION
Primary Predicate Device K 130222, Straumann® Dental Implant System SLActive and Roxolid Product Families, Straumann USA,
Reference Devices K133510, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentarios SA K163194, Neodent Implant System - GM Line, JJGC Indústria e Comércio de Materiais Dentários SA K222288, DESS Dental Smart Solutions, Terrats Medical SL K 193096, S.I.N. Dental Implant System, S.I.N. -Sistema de Implante Nacional S.A. K213063, TLX SRAs and TLX Gold Abutments, Institut Straumann AG
{5}------------------------------------------------
K092035, Bicon Implants with a 2.5 mm Internal Connection, Bicon, L.L.C. K140440, Noris Medical Dental Implants System, Noris Medical, Ltd. K193046, Straumann® Retentive System - Novaloc TiN Abutments, Institut Straumann AG K153779, Abutment for Bridges, Altatec GmbH K203252, Multi-unit Abutments for CONELOG®, BioHorizons Implant Systems, Inc. K191123, Multi-unit Abutments, Medentika GmbH, Straumann USA, LLC K170131, TAV Medical Dental Implant System, Tav Medical Ltd. K233208, NobelProcera® Titanium ASC Abutment, Nobel Biocare AB K212108, Dynamic TiBase, Talladium España, SL
#### INDICATIONS FOR USE STATEMENT
ICX Implant Systems is indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. ICX Implant Systems is indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implants with lengths less than 7.0 mm are indicated for delayed loading only.
ICX Implant Systems CAD-CAM abutments are intended for use with dental implants as a support for single-unit or multiple-unit prostheses in the maxilla or mandible of a partially or fully edentulous patient. All digitally designed abutments for use with ICX Implant Systems CAD-CAM abutments are intended to be manufactured at a medentis medical GmbH validated milling center.
#### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to obtain marketing clearance for an endosseous dental implant and abutment system. ICX-Implant System, from medentis medical GmbH. The ICX-Implant System includes a range of endosseous dental implants and prosthetic components. All implants have a selftapping apical thread with a tapered body and root-form designs with an internal hex implant/abutment connection. The implant body surface is blasted and acid-etched.
Abutments are available in multiple designs, including straight and angled abutments intended for single tooth and multi-unit restorations.
Table 1. ICX-Implant System Overview contains an overview of the system images and features.
{6}------------------------------------------------
| Implants | | | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|--------------------------|--|
| Implant Line | ICX-Premium | ICX-Diamond Premium | ICX-Active Master | ICX-Diamond Active Master | ICX-Active Liquid | |
| Image<br>Not to Scale | Image: ICX-Premium | Image: ICX-Diamond Premium | Image: ICX-Active Master | Image: ICX-Diamond Active Master | Image: ICX-Active Liquid | |
| Placement | Bone Level<br>Tissue Level | Bone Level<br>Tissue Level | Bone Level<br>Tissue Level | Bone Level<br>Tissue Level | Bone Level | |
| Collar Design | Micro-threaded collar | Micro-threaded collar | Threaded to top | Threaded to top | Threaded to top | |
| Packaging | Air | Saline | Air | Saline | Saline | |
| Material | Unalloyed titanium | Unalloyed titanium | Unalloyed titanium | Unalloyed titanium | Unalloyed titanium | |
| Body Ø (mm) | 3.43 - 4.8 | 3.43 - 4.8 | 3.43 - 4.8 | 3.43 - 4.8 | 3.43 - 4.8 | |
| Platform Ø (mm) | 3.3 - 3.7 | 3.3 - 3.7 | 3.3 - 3.7 | 3.3 - 3.7 | 3.3 - 3.7 | |
| Endosseous<br>Lengths* (mm) | 8 - 15<br>5 - 12.5 | 8 - 15<br>5 - 12.5 | 8 - 15 | 8 - 15 | 8 - 15 | |
| *Tissue Level implants have a 1.7 mm gingival collar (2.7 mm for 5 mm length). Overall length of Tissue Level implants is 1.7 mm (2.7 mm) greater than shown. | | | | | | |
| Healing Components, Abutments, Screws | | | | | | |
| ICX-Cover Screw ICX-Healing Cap ICX-Titanium Abutment ICX-Titanium Aesthetic Abutment ICX-Massive Abutment ICX-Universal Abutment ICX-Bar System Abutment ICX-Maximus Abutment ICX-UCLA Abutment ICX-Multi Abutment ICX-Adhesive Base Abutment (CAD/CAM) Healing Components, Abutments | | | ICX-3.3 Connection Screw ICX-3.3 Multi Connection Screw ICX Connection Screw Type A ICX Connection Screw Type B ICX Multi Connection Screw ICX Multi Prosthetics Screw ICX Maximus Connection Screw Screws | | | |
#### Table 1. ICX-Implant System Overview
#### MATERIAL COMPOSITION
All subject device dental implants are manufactured from cold worked unalloyed titanium Grade 4 conforming to ASTM F67 Standard Specification for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700) and Grade 4B conforming to ISO 5832-2 Implants for surgery – Metallic materials – Part 2: Unalloyed titanium. The threaded endosseous surface of the implant is blasted and acid-etched from the implant collar to the apex.
Subject device healing components. (cover screws and healing caps) are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) or PEEK-Classix™ by Invibio.
Subject device abutments are manufactured from titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for
{7}------------------------------------------------
Surgical Implant Applications (UNS R56401), cobalt-chromium alloy conforming to ASTM F1537 Standard Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539), or gold alloy (Ceramicor). The ICX-Multi Interim Abutment is manufactured from PEEK-Classix™ by Invibio. The burn-out sleeve packaged with the ICX-UCLA Abutments and ICX-CoCr Abutment is manufactured from polyoxymethylene (POM). The retention rings that are used with the denture housing are manufactured from nylon (Grilamid).
The material for manufacture of zirconia superstructures for Adhesive Base Abutments is yttria-stabilized zirconia (Y-TZP) from Dental Direkt GmbH (K150196, K142987, K170885, K183569). The cement recommended in labeling for bonding of superstructures is Panavia F 2.0 cement by Kuraray Medical, Inc. cleared under K032455.
The subject device screws are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401).
#### SURFACE TREATMENTS
The threaded endosseous surface of the subject device implants is grit-blasted and acid-etched from the implant collar to the apex. The subject device Maximus abutments have a titanium nitride (TiN) coating applied to the coronal end of the abutment. Some subject device abutments are anodized for color coding purposes.
#### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence included:
- validation of gamma irradiation sterilization for subject devices provided sterile to the end user to a . sterility assurance level of 10 by selecting and substantiating a 25 kGy dose using method VDmax25, according to ISO 11137-1 Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose;
- . bacterial endotoxin testing including Limulus amebocyte lysate (LAL) test according to ANSI/AAMI ST72 Bacterial endotoxins – Test methods, routine monitoring, and alternatives to batch testing on samples from sterilized product to demonstrate all sterile product meets a limit of < 20 EU/device;
- . validation of the recommended moist heat sterilization cycle by the overkill method to a sterility assurance level (SAL) of 10° according to ISO 17665-1 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices, and ISO/TR 17665-2 Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ANSI/AAMI/ISO 17665-1;
- shelf life testing, including testing of samples after 5 years of real time aging according to ASTM . F1929 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dve Penetration and F88/F88M Standard Test Method for Seal Strength of Flexible Barrier Materials, and sterility testing of product;
- . biocompatibility testing of samples representing all subject device material (including zirconia from Dental Direkt GmbH) and manufacturing processes according to ISO 10993-5 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity, and ISO 10993-12
{8}------------------------------------------------
Biological evaluation of medical devices – Part 12: Sample preparation and reference materials, including ;
- characterization of the implant acid-etched surface included scanning electron microscopy (SEM) . and surface chemistry analysis;
- Characterization of TiN coated and anodized surfaces; ●
- Characterization of super hydrophilicity by contact angle measurements before and after aging; ●
- mechanical testing according to ISO 14801 Dentistry Implants Dynamic loading test for ● endosseous dental implants;
- . analysis of the surface area and initial bone to implant contact area of short implants and comparison with a reference device;
- MR compatibility testing to support labeling as MR Conditional
No clinical data were included in this submission.
#### EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the reference devices.
#### Substantial Equivalence of Indication for Use Statement (IFUS)
The Indications for Use Statement (IFUS) of the subject device implants compared to those of the primary predicate and reference devices are detailed below in Table of Substantial Equivalence – Indications for Use Statement. The IFUS for the subject device is substantially equivalent to that of the primary predicate K130222 and reference device K133510, K163194, K222288, K193096, K213063, and K092035. There is a difference in terminology used, however, the intended use is the same. All implants are intended to be used with dental implants for support for prosthetic restoration for edentulous or partially edentulous patients. Delayed loading indications have been added for the short implants (endosseous length less than 7 mm). The digitally designed abutments are to be sent to a validated milling center for manufacture similar to reference devices K163194, K22288 and K193096.
#### Substantial Equivalence of Technological Characteristics - Implants
The subject device implant offers two (2) implant placement types: bone level and tissue level, similar to the primary predicate K130222. The subject device implant has an internal connection similar to the primary predicate K130222, however, the subject device has an internal hex and primary predicate K 130222 has an internal octagon. Any risks that may be associated with different implant placement levels and implant/abutment connection, have been mitigated by mechanical testing
The subject device implants have body diameters ranging from 3.43 mm. The smallest subject device body diameter is similar to that of the primary predicate K130222.. The largest and smallest subject device diameters are similar to those of primary predicate K130222.
The subject device implant length of 8 mm to 15 mm is similar to the primary predicate K 130222 (implants with 8 mm to 16 mm length) and the reference device K133510 (implants with 9 mm to 19 lengths).
{9}------------------------------------------------
The subject device includes short implants (defined as implants with endosseous length less than 7 mm) with endosseous lengths of 5 mm and 6.5 mm. The short implants are available only with one implant line, the ICX-Premium TL and ICX-Diamond Premium TL tissue level implant. Both implant lengths are offered with a 4.8 mm implant diameter. These dimensions are similar to those of the reference device K092035 with implants having a 5 mm length and a 4.0 mm implant diameter. A surface area evaluation and bone to implant contact computation have been conducted comparing the short subject device implants to the reference device.
The subject device implants are manufactured from the same material (unalloyed titanium Grade 4) and have the same endosseous surface treatment (grit-blasted and acid-etched) as the reference device K133510. Any risks that may be associated with the surface treatment have been mitigated by biocompatibility evaluation and surface treatment evaluation.
The subject device implants are provided sterile to the user, for single patient, single use, the same as the primary predicate and reference devices.
#### Substantial Equivalence of Technological Characteristics - Prosthetic Components
The technological characteristics of the subject device abutments are compared to the reference devices, shown below in Tables 2-10 Table of Substantial Equivalence – Technological Characteristics. Table 3. Table of Substantial Equivalence - Technological Characteristics. The primary predicate K130222 is not being used to evaluate these characteristics, as it is included only for evaluation of the implants. The reference device K092035 is not being used to evaluate these characteristics, as it is included only for evaluation of surface area of short implants.
The subject device healing component offers diameters that range from 2.6 mm to 8.0 mm with a threaded connection. The range of diameters is similar to those of the reference devices. The subject device healing components are used only for healing purposes and are placed out of occlusion, therefore mechanical strength is not a risk that requires mitigation.
The subject device abutments are provided in diameters that range from 3.7 mm with a hex connection and angulation of 0°, 15°, and 25°. The diameters, connections, and angulations are similar to those of the reference device abutments.
The subject device overdenture abutments are provided in diameters that range from 4.5 mm to 7.0 mm with a hex connection. The subject device diameters are similar to those of the reference device K133510. The subject device overdenture abutments are not intended for angulation or angle correction, and therefore mechanical strength is not a risk requiring mitigation.
The subject device UCLA abutments are provided in diameters that range from 4.4 mm to 5.0 mm with a hex and non-hex connection. The subject device diameters are similar to those of the reference device K193096. The subject device UCLA abutments are two-part cast-on abutments consisting of a cobaltchromium alloy or gold base and a polyoxymethylene (POM) burn-out sleeve. The POM burn-out sleeve can be cast in cobalt-chromium alloy or nickel-chromium alloy. This is similar to those of the reference device K193096 and K213063. The subject device UCLA abutments are not intended for angulation or angle correction.
The subject device Multi abutments are provided in diameter of 4.8 mm with a hex, non-hex, and threaded connection and angulation of 0°, 17° and 30°. The subject device diameters, connection, and
{10}------------------------------------------------
angulation are similar to those of the reference device K222288. The subject device abutment does not have a diameter smaller or larger, or an angulation greater than those of the reference device K22288.
The subject device base abutments are provided in diameters ranging from 4.4 mm to 5.1 mm with a hex and non-hex connection and angulation up to 15° using a base manufactured from titanium alloy or Co-Cr alloy and a zirconia superstructure. The subject device diameters, connection, and material are similar to those of the reference devices K163194 and K222288. The subject device abutment does not have a diameter smaller or larger, a greater angulation, or material different than those of the reference device K136194 and K222288. The two-pieces of the Ti Base abutment which compose the final abutment consist of the pre-manufactured titanium base component composed of titanium alloy and the CAD/CAM patient matched superstructure composed of zirconia.
All subject device abutments (healing components, abutments, and ball abutments) are made from titanium alloy (ASTM F136), cobalt-chromium alloy (ASTM F1537), gold alloy (Ceramicor), and PEEK-CLASSIX™, which is same as the reference devices.
Overall, the subject device has the following similarities to the predicate devices:
- has the same intended use.
- . uses the same operating principle,
- . incorporates the same basic design,
- . incorporates the same or very similar materials, and
- has similar packaging and is sterilized using the same materials and processes. ●
The basis for the belief of medentis medical GmbH that the subject device is substantially equivalent to the predicate devices is summarized in the following tables, Table of Substantial Equivalence -Indications for Use Statement and Tables 3-11 Table of substantial Equivalence – Technological Characteristics.
{11}------------------------------------------------
| | Indications for Use Statement |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject device | ICX Implant Systems is indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. ICX Implant Systems is indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implants with lengths less than 7.0 mm are indicated for delayed loading only.<br><br>ICX Implant Systems CAD-CAM abutments are intended for use with dental implants as a support for single-unit or multiple-unit prostheses in the maxilla or mandible of a partially or fully edentulous patient. All digitally designed abutments for use with ICX Implant Systems CAD-CAM abutments are intended to be manufactured at a medentis medical GmbH validated milling center. |
| Primary predicate device | Straumann® Dental implants are indicated for oral endosteal implantation in the upper and lower jaw arches and for the functional and aesthetic oral rehabilitation of edentulous and partially dentate patients. Straumann® Dental implants are also indicated for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple-tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function.<br><br>The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants through the corresponding components (abutments). |
| Reference devices | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.<br><br>Indications for Use for GM implants and conventional abutments: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. |
| | Indications for Use for GM Titanium Base abutments: Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations. All digitally designed copings and/or crowns for use with the Neodent Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.<br><br>Indications for Use for GM Pro Peek Abutments: The Pro PEEK Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months. They can be used in one or two stage procedures and also immediate load when there is good primary stability |
| | |
| | Indications for Use Statement |
| K222288<br>DESS Dental Smart<br>Solutions<br>Terrats Medical SL | DESS Dental Smart Solutions abutments are intended to be used in conjunction<br>with endosseous dental implants in the maxillary or mandibular arch to provide<br>support for prosthetic restorations.<br><br>All digitally designed custom abutments for use with Ti Base abutments or Pre-<br>milled Blank abutments are to be sent to a Terrats Medical validated milling center<br>for manufacture.<br><br>For complete Indications for Use statement on OEM Compatibility see 510(k)<br>Summary for K222288 in Section 12 |
| K193096<br>S.I.N. Dental Implant<br>System<br>S.I.N. -Sistema de<br>Implante Nacional S.A. | S.I.N. Dental Implant System is intended for placement in the maxillary or<br>mandibular arch to provide support for single-unit or multi-unit restorations. When<br>a one-stage surgical approach is applied, the S.I.N. Dental Implant System is<br>intended for immediate loading when good primary stability is achieved and with<br>appropriate occlusal loading. All digitally-designed custom abutments for use with<br>Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center<br>for manufacture |
| K213063<br>TLX SRAs and TLX Gold<br>Abutments<br>Institut Straumann AG | TLX SRAs Straumann® abutments are indicated to be placed into Straumann®<br>dental implants to provide a support structure for the functional and esthetic oral<br>rehabilitation of edentulous or partially edentulous patients with crowns, bridges,<br>or full-arch prostheses.<br>TLX Gold Abutments Straumann® abutments are indicated to be placed into<br>Straumann® dental implants to provide a support structure for the functional and<br>esthetic oral rehabilitation of edentulous or partially edentulous patients with<br>crowns, bridges or full-arch prostheses. Copings are indirectly connected to the<br>endosseous dental implant and are indicated for use as an aid in prosthetic<br>rehabilitations. |
| K092035<br>Bicon Implants with<br>2.5mm Internal Connection<br>Bicon, L.L.C. | The Bicon implant is designed for use in edentulous sites in the mandible or<br>maxilla for support of a complete denture prosthesis, a final or intermediate<br>abutment for fixed bridgework or for partial dentures, or as a single tooth<br>replacement. |
### Table 2. Table of Substantial Equivalence – Indications for Use Statement
{12}------------------------------------------------
{13}------------------------------------------------
| | Subject Device | | Primary Predicate | | Reference Device | Reference Device |
|-------------------------------------------|---------------------------------------------------|-------------------------------------------------------|--------------------------------------------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Feature | ICX-Implant System<br>medentis medical GmbH. | | K130222<br>Straumann® Dental Implant System<br>SLActive and Roxolid Product Families | | K133510<br>Neodent Implant System<br>JJGC Industria e Comercio de<br>Materiais Dentarios SA | K092035<br>Bicon Implants with 2.5mm<br>Internal Connection<br>Bicon ,L.L.C. |
| Product Code | DZE, NHA | | DZE & NHA | | DZE & NHA | DZE |
| Reason for Predicate/<br>Reference Device | n/a | | Tissue and bone level implants | | Implant material and surface<br>treatment, ball abutment | Surface area comparison |
| Implant Placement | Bone Level | Tissue Level | Bone level | Tissue Level | Bone level | Bone Level |
| Prosthetic Interface<br>Connection | Internal Hex | | Internal octagon | | External hex | Morse Taper |
| Body Diameter, mm | 3.43, 3.9, 4.25, 4.8 | | 2.9, 3.3, 4.1, 4.8 | | Titamax Smart - 3.3, 3.75, 4.0, 4.5,<br>5.0<br>Titamax Smart EX - 3.75, 4.0 | 4.0, 4.5<br>From marketing material, not<br>stated in 510(k) Summary |
| Platform Diameter, mm | 3.3, 3.7 | | Same as body | 3.5, 4.8 | 3.3, 4.5 | 3.5 |
| Endosseous Lengths,<br>mm | 8, 10, 12.5, 15<br>(no 15 mm<br>length for 3.3 Ø) | 5, 6.5, 8, 10, 12.5<br>(5 & 6.5 mm only<br>for 4.8 Ø) | 8, 10, 12, 14, 16,<br>18<br>(no 8 mm for 2.9 Ø) | 4, 6, 8, 10, 12, 14<br>(no 4, 6, 8 mm for<br>3.3 Ø) | Titamax Smart - 9, 11, 13, 15, 17<br>Titamax Smart EX - 9, 11, 13, 15,<br>17, 19 | 5, 8, 11<br>(no 5 mm for 4 Ø) |
| Implant Material | Unalloyed titanium | | Zirconium alloy (Roxolid), 15%<br>zirconium and 85% titanium | | Unalloyed titanium | Titanium alloy (Ti6Al-4V) |
| Implant Endosseous<br>Surface | Aluminum Oxide (Al2O3) Blasted, Acid<br>Etched | | Large grit sandblasted, acid-etched | | Grit blasted, acid etched | Grit blasted, acid etched, and<br>hydroxylapatite |
| Stored | Air and Saline | | Air and Saline | | Air and saline | Air |
## Table 3. Table of Substantial Equivalence – Technological Characteristics Implants
{14}------------------------------------------------
| Feature | Subject Device<br>K231566 | Reference Device<br>K133510 | Reference Device<br>K163194 | Reference Device<br>K193096 |
|---------------------|----------------------------|--------------------------------------------------------|--------------------------------------------------------|----------------------------------------------|
| | ICX-Implant System | Neodent Implant System | Neodent Implant System -<br>GM Line | S.I.N. Dental Implant System |
| | medentis medical GmbH. | JJGC Industria e Comercio<br>de Materiais Dentarios SA | JJGC Indústria e Comércio<br>de Materiais Dentários SA | S.I.N. -Sistema de Implante<br>Nacional S.A. |
| Coronal Ø, mm | 2.6-3.7 | Not stated | 4.5, 6 | 3.65 |
| Connection | Threaded / taper | Hex | Threaded / taper | Threaded / taper |
| Gingival Height, mm | 0 - 2 | Not stated | 0.8-5.5 | 2-4 |
| Material | Titanium alloy | Titanium alloy | Titanium alloy | Titanium Alloy |
| Surface treatment | Anodized, None | None | Anodized, None | Anodized, None |
| Used with implants | BL, TL | BL | BL | BL |
| How Provided | | | | |
| Sterility | Sterile | Non-sterile | Sterile | Sterile |
| Usage | Single patient, single-use | Single patient, single-use | Single patient, single-use | Single patient, single-use |
# Table 4. Table of Substantial Equivalence – Technological Characteristics Cover Screws
{15}------------------------------------------------
| Healing Caps | | | |
|---------------------|--------------------------------------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------|
| | Subject Device<br>K231566 | Reference Device<br>K163194 | Reference Device<br>K222288 |
| Feature | ICX-Implant System<br>medentis medical GmbH. | Neodent Implant System - GM Line<br>JJGC Indústria e Comércio de<br>Materiais Dentários SA | DESS Dental Smart Solutions<br>Terrats Medical SL |
| Coronal Ø, mm | 3.9/4.7 - 8.0 | 3.3 - 6.5 | 3.0 - 7.7 |
| Gingival Height, mm | 1.0 - 6.0<br>(PEEK customizable<br>1.24 - 10.46) | 0.8 - 5.5 | Not stated |
| Connection | Threaded, hex | Threaded, hex | Threaded |
| Material | Titanium alloy, PEEK | Titanium alloy, PEEK | Titanium Alloy |
| Surface treatment | Anodized, None | Anodized, None | None |
| Used with implants | BL | BL | BL |
| How Provided | | | |
| Sterility | Non-sterile | Non-sterile | Non-sterile |
| Usage | Single patient, single-use | Single patient, single-use | Single patient, single-use |
## Table 5. Table of Substantial Equivalence – Technological Characteristics Healing Caps
{16}------------------------------------------------
| | Subject Device | Reference Device<br>Reference Device | | Reference Device | Reference Device |
|----------------------------------------|----------------------------------------------|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------|------------------------------------------------------|------------------------------------------------------------|
| | K231566 | K140440 | K163194 | K222288 | K233208 |
| Feature | ICX-Implant System<br>medentis medical GmbH. | Noris Medical Dental<br>Implants System<br>Noris Medical, Ltd. | Neodent Implant System -<br>GM Line<br>JJGC Indústria e Comércio<br>de Materiais Dentários SA | DESS Dental Smart<br>Solutions<br>Terrats Medical SL | NobelProcera® Titanium<br>ASC Abutment<br>Nobel Biocare AB |
| Prosthetic Platform Ø,<br>mm | 3.7 - 6.5 | Not stated | 3.3-4.8 | 4,5 - 6.5 | 3.13 - 6.0 |
| Gingival Height for BL<br>implants, mm | 0.3 – 5.6 | 0.5 - 4.0 | 0.8-5.5 | 0.4 - 6.0 | 0.3 min. |
| Gingival Height for TL<br>implants, mm | N/A (GH is in implant) | | | N/A | N/A |
| Angulation | 0°, 15°, 25° | 0°, 17°, 30° | up to 15° | 0°, 17°, 30° | 0° to 30° |
| Connection | Engaging, Non-engaging | Engaging, Non-engaging | Engaging, Non-engaging | Engaging, Non-engaging | Engaging |
| Restoration retention | Cement | Cement | Cement | Cement | Cement |
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy | Titanium Alloy | Titanium Alloy |
| Surface treatment | Anodized, None | Anodized, None | Anodized, None | Anodized, None | Anodized, None |
| Used with implants | BL, TL | BL | BL | BL, TL | BL |
| How Provided | | | | | |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Non-sterile |
| Usage | Single patient, single-use | Single patient, single-use | Single patient, single-use | Single patient, single-use | Single patient, single-use |
## Table 6. Table of Substantial Equivalence – Technological Characteristics ICX-Titanium Abutment, ICX-Titanium Aesthetic Abutment
{17}------------------------------------------------
| | Subject Device | Reference Device | Reference Device | Reference Device |
|------------------------------|----------------------------|--------------------------------------------------------|--------------------------------------------------------|--------------------------------|
| | K231566 | K133510 | K163194 | K222288 |
| Feature | ICX-Implant System | Neodent Implant System | Neodent Implant System - GM<br>Line | DESS Dental Smart<br>Solutions |
| | medentis medical GmbH. | JJGC Industria e Comercio<br>de Materiais Dentarios SA | JJGC Indústria e Comércio de<br>Materiais Dentários SA | Terrats Medical SL |
| Prosthetic Platform Ø,<br>mm | N/A | 3.3. - 5.0 | 3.3-4.8 | 3.5-6.5 |
| Gingival Height, mm | 1.14 - 2.8 | 2-5 | 0.8-5.5 | 3.5-6.5 |
| Post height, mm | 4.0 - 11. | Not stated | Not stated | Not stated |
| Angulation | None (0°) | 0°, 17°, 30° | up to 15° | 0° |
| Connection | Non-engaging | Engaging | Engaging | Engaging, Non-engaging |
| Restoration retention | Cement | Cement | Cement | Cement |
| Used with implants | BL, TL | BL | BL | BL |
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy | Titanium Alloy |
| Surface Treatment | Anodized, None | | None | |
| How Provided | | | | |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Non-sterile |
| Usage | Single patient, single-use | Single patient, single-use | Single patient, single-use | Single patient, single-use |
## Table 7. Table of Substantial Equivalence – Technological Characteristics ICX-Massive Abutment, ICX-Universal Abutment
{18}------------------------------------------------
| | Subject Device<br>K231566 | Reference Device<br>K193046 | Reference Device<br>K133510 | Reference Device<br>K153779 |
|---------------------|----------------------------|--------------------------------------------------------|--------------------------------------------------------|-----------------------------|
| Feature | ICX-Implant System | Straumann® Retentive System<br>- Novaloc TiN Abutments | Neodent Implant System | Abutments for Bridges |
| | medentis medical GmbH. | Institut Straumann AG | JJGC Industria e Comercio de<br>Materiais Dentarios SA | Altatec GmbH |
| Gingival Height, mm | 1 - 5 | 1 - 6 | 2-5 | 0.4 - 2.2 |
| Connection | Engaging, non-engaging | Non-engaging | Non-engaging | Non-engaging |
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy | Titanium Alloy |
| Surface Treatment | TiN coating | TiN coating | TiN coating | None |
| Insert | Nylon Inserts | Nylon Inserts | Nylon Inserts | N/A |
| Used with implants | BL, TL | BL, TL | BL | BL |
| How Provided | | | | |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Non-sterile |
| Usage | Single patient, single-use | Single patient, single-use | Single patient, single-use | Single patient, single-use |
## Table 8. Table of Substantial Equivalence – Technological Characteristics ICX Maximus Abutment (Overdenture Attachment System), ICX-Bar System Abutment
{19}------------------------------------------------
| Feature | Subject Device | Reference Device | Reference Device |
|---------------------------|-------------------------------------|-------------------------------------------|----------------------------|
| | K231566 | K193096 | K213063 |
| | ICX-Implant System | S.I.N. Dental Implant System | Straumann TLX SRAs |
| | medentis medical GmbH | S.I.N. -Sistema de Implante Nacional S.A. | Straumann USA, LLC |
| Prosthetic Platform Ø, mm | 4.4-5.0 | 3.8, 4.5 | 4.6 |
| Gingival Height, mm | 0.5 | 0.5, 1 | 3.5 |
| Connection | Hex, Non-hex | Hex | TorcFit |
| Material | Co-Cr alloy, Gold (Ceramicor) alloy | Co-Cr alloy | Gold (Ceramicor) alloy |
| Angulation | Straight only | Straight only | Straight only |
| Used with implants | BL, TL | BL | TL |
| How Provided | | | |
| Sterility | Non-sterile | Non-sterile | Sterile |
| Usage | Single patient, single-use | Single patient, single-use | Single patient, single-use |
## Table 9. Table of Substantial Equivalence – Technological Characteristics ICX-UCLA Abutments
{20}------------------------------------------------
| | | Reference Device | | | |
|-------------------------------------|--------------------------------------|--------------------------------|--------------------------------------|-------------------------------------------|----------------------------------------|
| | Subject Device | | Reference Device | Reference Device | Reference Device |
| | K231566 | K222288 | K203252 | K191123 | K233208 |
| Feature | ICX-Implant System | DESS Dental Smart<br>Solutions | Multi-unit Abutments<br>for CONELOG® | Medentika Multi-unit<br>Abutments | NobelProcera® Titanium ASC<br>Abutment |
| | medentis medical GmbH. | Terrats Medical SL | BioHorizons Implant<br>Systems, Inc. | Medentika GmbH<br>Straumann USA, LLC | Nobel Biocare AB |
| Prosthetic Platform Ø, mm | 4.8 | 3.0 - 5.0 | 4.8 | 4.5 – 6.5 | 3.13 - 6.0 |
| Gingival Height, mm | 0.4 - 4.5 | Not stated | 2.0 - 4.0 | 0.6 - 5.5 | 0.3 min. |
| Angulation | 0°, 17°, 30° | 0°, 17°, 30° | 0°, 17°, 30° | 0°, 17°, 30° | 0° to 30° |
| Connection | Engaging, Non-engaging | Engaging, Non-engaging | Engaging | Engaging, Non-engaging | Engaging |
| Restoration retention | Screw | Screw | Screw | Screw | Screw |
| Material, MUA | Titanium Alloy | Titanium…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.