TAV Medical Dental Implants System

K170131 · Tav Medical , Ltd. · DZE · Feb 10, 2018 · Dental

Device Facts

Record IDK170131
Device NameTAV Medical Dental Implants System
ApplicantTav Medical , Ltd.
Product CodeDZE · Dental
Decision DateFeb 10, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

TAV Medical Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth and in order to restore the patient chewing function. The Implants are indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Device Story

Two-piece endosseous dental implant system; includes implants (Sirius, Silhouette, iCone), cover screws, healing caps, and abutments. Implants fabricated from medical-grade titanium alloy (ASTM F136); abutments composed of titanium alloy, PEEK, or Yttrium Stabilized Zirconia. Used in dental clinics by clinicians for one- or two-stage surgical procedures. Implants integrate with alveolar bone; abutments serve as adapters between implant and prosthetic (crowns, bridges, overdentures). System provides mechanical support for prosthetic restorations. Bench testing confirms fatigue strength, corrosion resistance, and implant-abutment compatibility. Benefits include restoration of chewing function and esthetics.

Clinical Evidence

No clinical studies were performed. Evidence consists of bench testing (fatigue, corrosion, compatibility) and biocompatibility testing per ISO 10993. Human factors study confirmed safe handling of sterile packaging.

Technological Characteristics

Materials: Titanium alloy (ASTM F136), PEEK, Yttrium Stabilized Zirconia (ISO 13356). Designs: Cylindrical/tapered bodies, internal hex or conical connections, platform switching. Surface treatment: Sandblasting and acid etching. Sterilization: Gamma irradiation. Connectivity: None (mechanical device).

Indications for Use

Indicated for surgical and restorative applications in the upper or lower jaw to support prosthetic devices (e.g., artificial teeth) and restore chewing function. Suitable for immediate loading given sufficient primary stability and appropriate occlusal loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" and the words "U.S. Food & Drug Administration". February 10, 2018 TAV Medical Ltd. Revital Shabtai General Manager - Dental Division Dora Industrial Park P.O. Box 88 Shlomi 2283202 ISRAEL Re: K170131 Trade/Device Name: TAV Medical Dental Implants System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: January 1, 2018 Received: January 4, 2018 Dear Revital Shabtai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Andrew I. Steen -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170131 Device Name TAV Medical Dental Implant System Indications for Use (Describe) TAV Medical Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth and in order to restore the patient chewing function. The Implants are indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Type of Use (Select one or both, as applicable) | <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) | |---------------|----------------------------------------------|---------------|---------------------------------------------| | <span></span> | <div style="display:inline-block;">☑</div> | <span></span> | <div style="display:inline-block;">☐</div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo and contact information for Tav Medical Ltd. The VAT registration number is 511513673. The address is Dora Industrial Park, P.O.B 88, Shlomi 2283202. The phone number is 972-4-9808615, and the fax number is 972-4-9808356. The website is www.tavmedical.com, and the email address is info@tavmedical.com. Image /page/3/Picture/1 description: The image shows the logo for TAVMedical. The logo consists of a geometric shape in the upper left corner, with the letters "TAV" in bold black font next to it. The word "Medical" is written in a light green, sans-serif font. Below the company name is the tagline "PRECISION, PASSION, PARTNERSHIP" in a smaller, lighter font. # 510(K) Summary for TAV Medical Dental Implant System #### 510(K) Number: K170131 Applicant's Name: ### TAV Medical Ltd. Dora Industrial Park P.O.B 88, Shlomi 2283202 Israel T +972 (4) 9808615/503 F +972 (4) 9808356 E-mail: info@TAVmedical.com | Contact Person: | Revital Shabtai | |-----------------|---------------------------------| | | TAV Medical Ltd. | | | Dora Industrial Park | | | P.O.B 88, Shlomi 2283202 | | | Israel | | | T +972 (4) 9808615/503 | | | F +972 (4) 9808356 | | | E-mail: revitals@TAVmedical.com | Date Prepared: 10 February 2018 Trade Name: TAV Medical Dental Implant System Classification Name: Implant, Endosseous, Root-Form Medical Specialty: Dental Product Code: DZE (primary), NHA (secondary) Device Class: Class II Regulation Number: Primary 872.3640 Review Panel: Dental {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the contact information for Tav Medical Ltd. Their VAT registration number is 511513673. Their address is Dora Industrial Park, P.O.B 88, Shlomi 2283202. Their phone number is 972-4-9808615 and their fax number is 972-4-9808356. Their website is www.tavmedical.com and their email address is info@tavmedical.com. Image /page/4/Picture/1 description: The image shows the logo for TAVMedical. The logo consists of a geometric shape in the upper left corner, with the letters "TAVMedical" to the right of it. The "Medical" portion of the name is in green. Below the name is the tagline "PRECISION, PASSION, PARTNERSHIP" in a smaller font. # Predicate Devices TAV Medical Dental Implant System including Implants and Abutments are substantially equivalent to the following Predicate Devices, divided to primary and reference devices: | Device Owner /<br>Trade name | Primary Predicate Device<br>510(k) number | Product code | |------------------------------|-------------------------------------------|--------------------------------------| | MIS Dental<br>Implant System | K040807 | DZE (primary) and NHA<br>(secondary) | The following devices are used as reference devices: | Device Owner /<br>Trade name | Reference device 510(k)<br>number | Product code | |-------------------------------------------|-----------------------------------|--------------------------------------| | MIS Dental<br>Implant System | K112162 | DZE (primary) and NHA<br>(secondary) | | Hahn Tapered<br>Implant System | K143353 | DZE | | Noris Medical<br>Dental Implant<br>System | K140440 | NHA | | SGS Dental<br>Implants System | K133362 | NHA | | AB DENTAL<br>Devices Ltd. | K132125 | NHA | | Implant Direct | K143011 | NHA | | Ditron Dental Ltd. | K140728 | DZE | ## Indications for Use TAV Medical Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth and in order to restore the patient chewing function. The Implants are indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading. {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo and contact information for Tav Medical Ltd. The VAT registration number is 511513673. The address is Dora Industrial Park, P.O.B 88, Shlomi 228320. The phone number is 972-4-9808615 and the fax number is 972-4-9808356. The website is www.tavmedical.com and the email address is info@tavmedica. Image /page/5/Picture/1 description: The image contains the logo for TAVMedical. The logo consists of a geometric shape in the upper left corner, with the letters "TAV" in bold black font to the right of the shape. The word "Medical" is written in a light green font to the right of "TAV". Below the company name is the slogan "PRECISION. PASSION. PARTNERSHIP" in a smaller font. # Implant Description TAV Medical Dental Implant System is a two piece dental implant system for one stage or two stage surgical implantation of dental implant provided with cover screw and supported by a compatible abutment systems and surgical instruments. The implant is fabricated from a medical grade titanium alloy 6Al 4V ELI, according to ASTM F136. TAV Medical implant includes a body portion and a collar. The body portion is configured to extend into and Osseo integrate with the alveolar bone. The top surface of the collar lies flush with the crest of the jawbone bone. TAV Medical abutment lies on the top surface and extends through the soft tissue, which lies above the alveolar bone. The abutment supports the final prostheses. TAV Medical implants include the following brands: # Silhouette A Cylindrical-apically tapered screw type implant. The implant has a self-tapping design and a progressive thread. The Silhouette internal profile varies depending on the implant length. The internal core of the implant has a tapered shape. Silhouette implants are designed with: - Platform switching design over the implant top geometry. ● - Implant outer surface achieved by sand blasting and acid etching processes ● - Cylindrical Implant Body in its apical region ● - Self-Tapping design ● - The implant is recommended for bone types: D1, D2 and D3. ● - The implants are suitable for both one and two stage implant procedures. . Silhouette implant is designed in two platforms: Narrow Platform (NP) - . Internal Hex Connection 2.41mm - Available in the following dimensions: ● Diameter: 3.3mm Lengths: 10mm, 11.5mm, 13mm and 16mm Standard Platform (SP): - Internal Hexagon for anti-rotation 2.44mm ● - available in the following dimensions: . Diameters: 3.75mm, 4.2mm, 5mm, and 6mm Lengths: 8mm, 10mm, 11.5mm, 13mm and 16mm # Sirius {7}------------------------------------------------ Image /page/6/Picture/0 description: The image contains two identical logos, each featuring a stylized eye symbol enclosed within a triangle. The eye symbol is detailed with concentric circles, suggesting a focus or lens. The logos are stacked vertically, creating a repetitive pattern. The overall design is simple and modern, likely representing vision, focus, or observation. VAT registration number 511513673 Dora Industrial Park, P.O.B 88, Shlomi 2283202 Tel: 972-4-9808615 Fax: 972-4-9808356 Image /page/6/Picture/3 description: The image is a logo for TAVMedical. The logo consists of a geometric shape in the upper left corner, with the text "TAVMedical" to the right of it. The word "Medical" is in a lighter green color than the rest of the text. Below the text is the tagline "PRECISION. PASSION. PARTNERSHIP" in a smaller font. TAV Medical Sirius implants has a tapered apex with three cutting notches which makes the implant's rough surface extends from bottom to the top of the implant. The internal core of the implant has a tapered shape. Sirius implants are designed with: - Platform switching design on the implant top geometry . - Additional two starts on the coronal region ● - Deep micro thread at the coronal region ● - Implant outer surface achieved by sand blasting and acid etching ● - . Self-Tapping design - The implant is recommended for bone types: D2, D3 and D4. ● - Apically tapered implant body ● - The implants are suitable for both one and two stage implant procedures. . Sirius implant is designed in two platforms: Narrow Platform (NP): - Internal Hex Connection 2.41mm - · available in the following dimensions: Diameter: 3.3mm Lengths: 10mm, 11.5mm, 13mm and 16mm Standard Platform (SP): - Internal Hexagon for anti-rotation 2.44mm ● - . available in the following dimensions: Diameters: 3.75mm, 4.2mm, 5mm, and 6mm Lengths: 8mm, 10mm, 11.5mm, 13mm and 16mm # iCone Conical Connection implants which merges cylindrical body with a conical connection. iCone implants are available in a versatile range of diameters and lengths and in two platforms, Standard and Wide platforms that are color coded for easy identification. The implants are suitable for both one and two stage implant procedures. iCone implants are designed with: - 11° cone at the interior of the implant . - Internal Hexagon features 6 indexing positions . - . Platform switching design - Micro threads in top coronal section ● - Double Thread Design ● - The implant is recommended for bone types: D1, D2 and D3. ● - Implant outer surface achieved by sand blasting and acid etching . iCone implant is designed in two platforms: {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows a logo with a stylized house shape and the text "Tav" above the text "VAT". The house shape is a simple outline with a triangular roof and a rectangular base. The text is in a simple, sans-serif font. VAT registration number: 511513673 Image /page/7/Picture/3 description: The image shows the address "Dora Industrial Park, P.O.B 88, Shlomi 2283202". The address is written in a simple, sans-serif font. The text is clear and legible, making it easy to read the address information. Tel: 972-4-9808615 Fax: 972-4-9808356 www.tavmedical.com @ info@tavmedical.com Image /page/7/Picture/5 description: The image shows the logo for TAVMedical. The logo consists of a geometric shape in shades of gray and green, followed by the text "TAVMedical" in black and green. Below the company name is the tagline "PRECISION, PASSION, PARTNERSHIP" in a smaller font size. # Standard Platform (SP): - Internal Hex Connection 2.10mm - · Available in the following dimensions: Diameters: 3.75mm and 4.2mm Lengths: 8mm, 10mm, 11.5mm, 13mm and 16mm # Wide Platform (WP): - Internal Hexagon for anti-rotation 2.50mm ● - Available in the following dimensions: . Diameters: 4.5mm and 5mm. Lengths: 9mm, 11.5mm, 13mm and 16mm. # Abutments Description TAV Medical Abutments are intended for used as an adapter between the implant and the crown. The abutments are characterized by four distinct geometrically features: Height/length, angle, diameter and internal/external hex lock or conical connection. Anodized Color coding is used for categorical consideration only. TAV Medical abutments, as other available in the market of dental prosthetics are dental components composed either of titanium (Ti 6AL 4V ELI), PEEK and Yttrium Stabilized Zirconia (ZrO2). All the abutments fixation screws are composed of Ti 6AL 4V ELI. The abutments are supplied non-sterile to be sterilized by the physician before use according to the accompanied instruction for use. Abutment device refers to the fixture that is assembled on the implant. The crown is then built on the abutment. The prosthetic parts are divided into 3 main categories: - Cemented retained restorations . - Screw retained restorations . - Overdenture retained 트 The appropriate abutment type is selected in relation to the tooth position for the proposed implant. Various abutment types are available: {9}------------------------------------------------ Image /page/8/Picture/0 description: The image contains two identical symbols, each featuring a triangle with a circle inside. Within the circle, there's a smaller, detailed design that resembles a stylized eye or a complex geometric pattern. The symbols are simple and monochromatic, suggesting they might be part of a logo, a warning sign, or an icon set. v Medical Ltd. Dora Industrial Park, P.O.B 88, Shlomi 2283202 Tel: 972-4-9808615 Fax: 972-4-9808356. Image /page/8/Picture/3 description: The image is a logo for TAVMedical. The logo consists of a geometric shape in the upper left corner, with the text "TAVMedical" to the right of it. The word "TAV" is in black, while "Medical" is in green. Below the text is the phrase "PRECISION, PASSION, PARTNERSHIP" in a smaller font. # Titanium Abutments TAV Medical Titanium Abutments are intended to be placed into TAV Medical implants of different types, diameter, lengths and platforms to provide support for prosthetic reconstructions such as crowns and bridges. # Temporary Esthetic PEEK Abutments TAV Medical temporary PEEK abutments are intended to be used with TAV Medical Implants to provide support for individual temporary crowns and are for use of a maximum duration of 180 days. # Ball attachment The ball attachment system is intended to secure a removable prosthesis in a complete or partial restoration by endosseous implants in the mandible or in the maxilla. The ball attachment includes a metal housing and retention caps offering retention for every individual situation. The ball attachments are made of full Titanium. ## Locker overdenture The Locker, commonly known in the market as "Locator""", is a Zirconia (ZrO2) overdenture attachment system designed for use with dentures-retained in whole or in parts by endosseous implants in the mandible or in the maxilla. The locker includes a titanium denture cap and retentive caps with customizable levels of retention for multi-unit restorations in full overdentures. The locker is supplied made out fully from zirconia or partially zirconia attached to a titanium base. ## Healing caps Healing caps are screwed into the implant to protect the inner configuration of the implant during the healing phase in cases of transmucosal healing protocol. {10}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo and contact information for Tav Medical Ltd. The VAT registration number is 511513673. The address is Dora Industrial Park, P.O.B 88, Shlomi 2283202. The telephone number is 972-4-9808615, and the fax number is 972-4-9808356. The website is www.tavmedical.com, and the email address is info@tavmedical.com. Image /page/9/Picture/1 description: The image shows the logo for TAVMedical. The logo consists of a geometric shape in the upper left corner, followed by the text "TAVMedical" in a sans-serif font. The "Medical" portion of the text is in a lighter green color. Below the text, there is a tagline that reads "PRECISION, PASSION, PARTNERSHIP" in a smaller font. ### Substantial Equivalence discussion The proposed TAV Medical Dental Implant System has the same indications for technological characteristics, mode of operation and performance use. specifications as its predicates devices. Both the proposed device and the predicate devices function in the same manner providing support for prosthetic devices in the upper or lower jaw. The following table summarizes the equivalence comparison between TAV Medical Dental Implants, its primary predicate device and its reference devices. #### Substantial Equivalence Table - Implants | Feature | TAV Medical<br>[K170131]<br>Subject<br>Device | MIS<br>(K040807)<br>primary | MIS<br>(K112162)<br>reference | Hahn<br>(K143353)<br>reference | |--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name: | Sirius | Seven (internal<br>hex) | | Hahn Tapered<br>Implant System | | | Silhouette | M4 (internal hex) | | | | | iCone | | Conical<br>connection | | | K# | K170131 | K040807<br>(Internal Hex –<br>Seven & M4) | K0112162<br>(Conical<br>Connection) | K143353 | | Product Code | DZE | DZE | DZE | DZE | | Manufacturer | TAV Medical Ltd. | MIS Implant<br>Technologies Ltd. | MIS Implant<br>Technologies Ltd. | Prismatik<br>dentalcraft, inc | | Device<br>Description | Sirius<br>Titanium dental<br>implant with<br>apically tapered<br>body (conical<br>shape body) | Seven<br>Titanium dental<br>implant with<br>apically tapered<br>body (conical<br>shape body) | | The Hahn<br>Tapered Implant<br>System consists<br>of dental<br>implants,<br>abutments, | | Feature | TAV Medical<br>[K170131]<br>Subject<br>Device | MIS<br>(K040807)<br>primary | MIS<br>(K112162)<br>reference | Hahn<br>(K143353)<br>reference | | | Silhouette<br>Titanium implant<br>body combines a<br>cylindrical<br>shape in its apical<br>region and a<br>conical shape in<br>the coronal<br>region | M4<br>Titanium implant<br>body combines a<br>cylindrical shape<br>in its apical<br>region and a<br>conical shape in<br>the coronal<br>region | | screws, and<br>surgical<br>instruments. They<br>are manufactured<br>from Titanium<br>alloy, grade 23.<br>The implant is<br>designed with an<br>internal hex<br>connection and a<br>tapered body to<br>replace one or<br>more missing<br>teeth. | | | iCone<br>Conical<br>connection | | Conical<br>connection | | | Indications<br>for Use | The TAV Medical<br>Dental Implant<br>System is<br>indicated for use<br>in surgical and<br>restorative<br>applications for<br>placement in the<br>bone of the<br>upper or lower<br>jaw to provide<br>support for<br>prosthetic<br>devices, such as<br>artificial teeth, in<br>order to restore<br>the patient<br>chewing function.<br>The TAV Medical<br>Dental Implant<br>System is<br>indicated also for<br>immediate<br>loading when<br>good primary<br>stability is<br>achieved and<br>with appropriate<br>occlusal loading | The MIS Dental<br>Implant System<br>are intended to<br>be surgically<br>placed in the<br>bone of the<br>upper or lower<br>jaw arches to<br>provide support<br>for prosthetic<br>devices, such as<br>artificial teeth, in<br>order to restore<br>the patient's<br>chewing function. | The MIS Dental<br>Implant System<br>are intended to<br>be surgically<br>placed in the<br>bone of the<br>upper or lower<br>jaw arches to<br>provide support<br>for prosthetic<br>devices, such as<br>artificial teeth, in<br>order to restore<br>the patient's<br>chewing function.<br>When a one<br>stage surgical<br>procedure is<br>applied, the<br>implant may be<br>immediately<br>loaded when<br>good primary<br>stability is<br>achieved and the<br>functional load is<br>appropriate | Hahn Implants<br>are indicated for<br>use in maxillary<br>and mandibular<br>partially or fully<br>edentulous cases,<br>to support single,<br>multiple-unit, and<br>overdenture<br>restorations. The<br>implants are to<br>be used for<br>immediate<br>loading only in<br>the presence of<br>primary stability<br>and appropriate<br>occlusal loading. | | Feature | TAV Medical<br>[K170131]<br>Subject<br>Device | MIS<br>(K040807)<br>primary | MIS<br>(K112162)<br>reference | Hahn<br>(K143353)<br>reference | | Material | Titanium Alloy –<br>Ti 6AL4V ELI | Titanium Alloy –<br>Ti 6AL4V ELI | Titanium Alloy –<br>Ti 6AL4V ELI | Titanium Alloy –<br>Ti 6AL4V ELI | | Diameter<br>(mm) | Sirius &<br>Silhouette<br>3.3, 3.75, 4.2, 5,<br>6<br>iCone<br>3.75, 4.2, 4.5 , 5 | Seven & M4<br>3.3, 3.75, 4.2 5 | Conical<br>connection<br>3.75, 4.2, 5 | 3.0, 3.5, 4.3, 5.0,<br>and 7.0 | | Length(mm) | Sirius &<br>Silhouette<br>8, 10, 11.5, 13,<br>16<br>ICone<br>8, 9, 10, 11.5,<br>13, 16 | Seven & M4<br>8, 10, 11.5, 13,<br>16mm | Conical<br>connection<br>implans<br>8, 10, 11.5, 13,<br>16mm | 8, 10, 11.5, 13,<br>and 16 | | Prosthetic<br>Connection | Sirius &<br>Silhouette<br>Internal Hex<br>iCone<br>Conical | Seven & M4<br>Internal Hex | Conical<br>connection | Internal Hex | | Device<br>features | Screw type | Screw type | Screw type | Screw type | | Surface<br>treatment | Sirius &<br>Silhouette<br>Sandblasting and<br>acid etching<br>iCone<br>Sandblasting and<br>acid etching &<br>anodized | Seven & M4<br>Sandblasting and<br>acid etching &<br>anodized | Sandblasting and<br>acid etching &<br>anodized | Blasted with<br>Hydroxyl<br>Apatit | {11}------------------------------------------------ Image /page/10/Picture/0 description: The image contains two identical symbols, each consisting of a triangle with a circle inside. The circle within each triangle has a smaller, darker circle at its center. The symbols are simple and appear to be icons or logos, possibly related to water or nature due to the teardrop shape of the triangle. #### Tav Medical Ltd. VAT registration number: 511513673 Dora Industrial Park, P.O.B 88, Shlomi 2283202. Tel: 972-4-9808615 Fax: 972-4-9808356 网 www.tavmedical.com @ info@tavmedical.com Image /page/10/Picture/5 description: The image shows the logo for TAVMedical. The logo consists of a geometric shape in shades of gray and green, followed by the text "TAVMedical" in black and green. Below the company name is the text "PRECISION. PASSION. PARTNERSHIP" in a smaller font. {12}------------------------------------------------ Image /page/11/Picture/0 description: The image shows two triangular warning signs. The top sign has a symbol inside the triangle, while the bottom sign is a plain triangle. The signs are white with a red border. #### Tav Medical Ltd. VAT registration number: 511513673 Dora Industrial Park, P.O.B 88, Shlomi 2283202. Tel: 972-4-9808615 Fax: 972-4-9808356 网 www.tavmedical.com @ info@tavmedical.com Image /page/11/Picture/6 description: The image shows the logo for TAVMedical. The logo consists of a geometric shape in the upper left corner, with the text "TAVMedical" to the right. The word "Medical" is in a lighter green color. Below the text is the phrase "PRECISION, PASSION, PARTNERSHIP" in smaller letters. {13}------------------------------------------------ Image /page/12/Picture/0 description: The image shows two logos. The top logo features a triangle with a circle inside, and a smaller circle inside that. The bottom logo is a simple triangle. The text next to the logos is cut off, so it is not possible to determine what the logos represent. VAT registration number: 51.151.3673 Dora Industrial Park, P.O.B 88, Shlomi 228320 Tel: 972-4-9808615 Fax: 972-4-9808356 .tavmedical.com @ info@tavmedical.com Image /page/12/Picture/5 description: The image contains the logo for TAVMedical. The logo consists of a stylized geometric shape in shades of gray and green, followed by the text "TAVMedical" in a combination of black and green. Below the company name, there is a tagline that reads "PRECISION. PASSION. PARTNERSHIP" in a smaller, sans-serif font. | Feature | TAV Medical<br>[K170131]<br>Subject<br>Device | MIS<br>(K040807)<br>primary | MIS<br>(K112162)<br>reference | Hahn<br>(K143353)<br>reference | |--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Components | TAV Medical<br>Dental Implants<br>System consists<br>of one and two<br>stage endosseous<br>form dental<br>implants, internal<br>hexagonal and<br>conical<br>connection; cover<br>screws and<br>healing caps;<br>abutment<br>systems &<br>superstructures;<br>surgical<br>instruments. | The MIS Dental<br>Implant System<br>consists of one<br>and two stage<br>implants, internal<br>and external<br>hexagonal,<br>conical; cover<br>screw and<br>healing caps;<br>abutment<br>systems and<br>suprastructures;<br>surgical<br>instruments. | The MIS Dental<br>Implant System<br>consists of one<br>and two stage<br>implants, internal<br>and external<br>hexagonal,<br>conical; cover<br>screw and<br>healing caps;<br>abutment<br>systems and<br>suprastructures;<br>surgical<br>instruments. | The implant is<br>designed with an<br>internal hex<br>connection and a<br>tapered body to<br>replace one or<br>more missing<br>teeth.<br>The Implants are<br>design with two<br>(2) platform<br>connection<br>geometries | | Sterile<br>Package | Sterile barrier<br>tube with inner<br>sealed sleeve | Double container<br>sealing system | Double container<br>sealing system | Single blister,<br>unsealed inner<br>holder | | Sterilization | Gamma | Gamma | Gamma | Gamma | | method | irradiation | irradiation | irradiation | irradiation | | Intended Use | Dental Clinic | Dental Clinic | Dental Clinic | Dental Clinic | | Environment | Setting | Setting | Setting | Setting | | Abutments | Straight and | Straight and | Straight and | Straight and | | | angulated | angulated | anqulated | anqulated | Substantial equivalence discussion: As shown in the table above, TAV Medical Dental implant System is Substantial equivalent in all devices' features. The only difference is in the maximal outer diameter of the proposed device versus the maximal outer diameter of the primary predicate device. This slight difference was bridged using K143353 which was cleared with up to 7.0 mm outer diameter. TAV Medical's single sterile package slightly differs from the primary predicate device. While the MIS dental implant system contains two tubes and sleeve, TAV Dental's packaging contains one tube and an inner sealed sleeve. This difference was bridged using K140728 which utilizes the same packaging as TAV Medical and same sealing design. There are no additional differences, thus it was concluded that the subject dental implant system is substantially equivalent with the predicate devices. {14}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the logo and contact information for Tav Medical Ltd. The VAT registration number is 511513673. The address is Dora Industrial Park, POB 88, Shlomi 2283202. The phone number is 972-4-9808615, and the fax number is 972-4-9808356. The website is www.tavmedical.com, and the email address is info@tavmedical.com. Image /page/13/Picture/1 description: The image shows the logo for TAVMedical. The logo consists of a geometric shape in the upper left corner, with the text "TAVMedical" to the right of it. The word "Medical" is in a lighter green color. Below the company name is the tagline "PRECISION, PASSION, PARTNERSHIP" in a smaller font. #### Substantial Equivalence - Abutments TAV Medical abutments, similarly to its predicate devices, are intended to be placed in implants of different types, diameter, lengths and platforms to provide support for prosthetic reconstructions such as crowns, bridges and overdentures. The differences between the proposed TAV Medical Abutments and the predicate devices are minor and are reflected only in the abutments' dimensions and in the addition of the Zirconia material to some of the abutment types. These differences were tested and evaluated and no issues affecting device performances were detected. Thus it was concluded that the subject dental implant system abutments is substantially equivalent with the predicate devices | Feature | TAV Medical | MIS<br>primary | Noris Medical<br>reference | |------------------------|------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------| | Product Code | NHA | NHA | NHA | | K# | K170131 | K112162, K040807 | K140440 | | Product Name | Titanium Healing Cap | Titanium Healing Cap | Titanium Healing Cap | | Product<br>Description | Used to allow the<br>gingiva to heal around<br>the Gingiva | Used to allow the<br>gingiva to heal around<br>the Gingiva | Used to allow the<br>gingiva to heal around<br>the implant | | Material | Titanium Alloy<br>+Anodize | Titanium Alloy<br>+Anodize | Titanium Alloy | | Diameter<br>(mm) | 3.8, 4, 4.5, 4.7, 5, 5.5,<br>6.5 | 4, 4.3, 4.8, 5, 5.5, 6.5 | 3.8, 4.6 ,5.5 ,6.3 | | Height (mm) | 2, 3, 4, 5, 6 | 2, 3, 4, 5, 6, 8 | 2, 3, 4, 5, 6, 7 | | Angle (°) | 0 | 0 | 0 | | sterility | Non Sterile | Sterile | Non Sterile | #### Titanium Healing Cap Substantial equivalence discussion: As shown in the table above, TAV Medical titanium healing caps are Substantial equivalent in all features. A minor difference exists in the diameter dimensions however this difference is not significant since all subject device diameters are within the predicate devices dimensions range. One more difference exists in terms of sterility condition. This difference is covered by the reference device which is provided non-sterile similarly to TAV Medical titanium healing cap. Like in the predicate devices the {15}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the logo and contact information for Tav Medical Ltd. The VAT registration number is 511513673. The address is Dora Industrial Park, P.O.B 88, Shlomi 2283202. The telephone number is 972-4-9808615 and the fax number is 972-4-9808356. The website is www.tavmedical.com and the email address is info@tavmedical.com. Image /page/14/Picture/1 description: The image shows the logo for TAVMedical. The logo consists of a stylized geometric shape in shades of gray and green, followed by the text "TAVMedical" in a combination of black and green. Below the company name is the tagline "PRECISION. PASSION. PARTNERSHIP" in a smaller, sans-serif font. healing cap is screwed into the implant directly in trans-mucosal healing protocol. Cover screw is used in submucosal healing protocol. | Feature | TAV Medical | MIS | Noris Medical | |------------------------|------------------------------------------------------------|------------------------------------------------------------|---------------------------------------------------------------| | Product Code | NHA | NHA | NHA | | K# | K170131 | K112162, K040807 | K140440 | | Product Name | Zirconia Healing Cap | Titanium Healing Cap | Titanium Healing Cap | | Product<br>Description | Used to allow the<br>gingiva to heal around<br>the gingiva | Used to allow the<br>gingiva to heal around<br>the gingiva | Used to allow<br>the<br>gingiva to heal around<br>the implant | | Material | Zirconia | Titanium Alloy +<br>Anodize | Titanium Alloy | | Diameter (mm) | 4, 4.7, 5.5 | 4, 4.3, 4.8, 5, 5.5, 6.5 | 3.8, 4.6 ,5.5 ,6.3 | | Height (mm) | 2, 3, 4, 5, 6 | 2, 3, 4, 5, 6, 8 | 2, 3, 4, 5, 6, 7 | | Angle (°) | 0 | 0 | 0 | | Sterility | Non Sterile | Sterile | Non Sterile | #### Zirconia Healing Cap Substantial equivalence discussion: As shown in the table above, TAV Medical zirconia healing caps are Substantial equivalent since it is the same in most features. The only features that differ from the subject device are material and sterility conditions. In terms of materials TAV Medical healing cap is made of zirconia while titanium is utilized for the predicate devices healing caps. In order to assure that this difference does not compromise the subject device performance, thorough mechanical and biological tests were performed with satisfactory results. In terms of sterility condition, this difference was covered by the reference device which is provided non-sterile similarly to TAV Medical zirconia healing cap. | Feature | TAV Medical | MIS | |--------------|-----------------------------------|-----------------------------------| | Product Code | NHA | NHA | | K# | K170131 | K112162, K040807 | | Product Name | TAV Healing Cap | Plastic healing cap | | Product | Used to allow the gingiva to heal | Used to allow the gingiva to heal | | Description | around the gingiva | around the gingiva | | Material | PEEK | PEEK | | Height (mm) | 4, 6, 8 | 4, 6, 8 | | Sterility | Non sterile | Non sterile | #### PEEK Healing cap Substantial equivalence discussion: As shown in the table above, TAV Medical PEEK healing caps are substantial equivalent in all features. {16}------------------------------------------------ Image /page/15/Picture/0 description: The image contains the text "Tav Medical Ltd.". The text appears to be a company name. The text is in a simple, sans-serif font and is left-aligned. Image /page/15/Picture/1 description: The image shows the VAT registration number and address. The VAT registration number is 511513673. The address is Dora Industrial Park, P.O.B 88, Shlomi 228320. @ info@tavmedical.com Image /page/15/Picture/3 description: The image shows the logo for TAVMedical. The logo consists of a stylized geometric shape in shades of gray and green, followed by the text "TAVMedical" in a bold, sans-serif font. The "Medical" portion of the name is in a lighter green color. Below the name is the tagline "PRECISION. PASSION. PARTNERSHIP" in a smaller, lighter font. # Titanium Abutments, Straight | Feature | TAV Medical | MIS | |---------------|-----------------------------------|-----------------------------------| | Product Code | NHA | NHA | | K# | K170131 | K112162, K040807 | | Product Name | Titanium Abutments | Titanium Abutments | | Product | An accessory to endosseous dental | An accessory to endosseous dental | | Description | implants to support a prosthetic | implants to support a prosthetic | | | device | device | | Material | Titanium Alloy + Anodized | Titanium Alloy + Anodized | | Diameter (mm) | 3.8, 4.8, 5.2, 5.5 | 4.8, 5.5, 5.6, 5.8, 6.3 | | Height (mm) | 0.5, 1, 2, 3 | 1, 2, 3…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...