K140440 · Noris Medical , Ltd. · DZE · Dec 19, 2014 · Dental
Device Facts
Record ID
K140440
Device Name
NORIS MEDICAL DENTAL IMPLANTS SYSTEM
Applicant
Noris Medical , Ltd.
Product Code
DZE · Dental
Decision Date
Dec 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Noris Medical Ltd Dental Implants System is intended to replace missing tooth/ teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Mono implants are specifically indicated for replacing maxillary lateral incisors and mandibular central and lateral incisors. They are used for immediate, non-occlusal provisionalization in single-tooth restorations. Multiple-unit restorations should be splinted together and may be used immediately when clinically appropriate.
Device Story
System comprises endosseous dental implants (Tuff, Tuff TT, Onyx, Mono) and prosthetic components (abutments, healing caps, attachments). Implants are surgically placed into jawbone by dental clinicians to support prosthetic restorations. Implants feature titanium alloy construction with SLA surface treatment. Mono implants allow one-stage surgery; others support one or two-stage procedures. System enables immediate loading if primary stability is achieved. Prosthetic components (cemented or screw-retained) connect to implants to restore chewing function. Device benefits patients by replacing missing teeth and restoring oral function. Clinicians use standard surgical and prosthetic workflows to place implants and attach superstructures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including SEM/EDS surface analysis, sterilization validation (ANSI/AAMI/ISO 11137-1), accelerated aging (ASTM F1980), and dynamic fatigue testing (ISO 14801).
Technological Characteristics
Materials: Titanium alloy (Ti 6Al 4V ELI) per ASTM F136. Surface: SLA (Sandblasted, Large grit, Acid etched). Sterilization: Radiation per ANSI/AAMI/ISO 11137-1 (SAL 10^-6). Design: Root-form endosseous implants with internal hex or one-piece configurations. Connectivity: Mechanical interface for prosthetic abutments.
Indications for Use
Indicated for edentulous or partially edentulous patients requiring replacement of missing teeth in either jaw to support prosthetic devices for restoration of chewing function. Mono implants are indicated for maxillary lateral incisors and mandibular central/lateral incisors for immediate, non-occlusal provisionalization in single-tooth restorations.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of a human figure in profile, with three overlapping faces. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the figure. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19,2014
Noris Medical, Ltd. Dr. Raanan Aloni OA/RA Director 8 Hataasia Street Nesher 3688808 ISRAEL
Re: K140440
> Trade/Device Name: Noris Medical Dental Implants System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous dental implant Regulatory Class: II Product Code: DZE, NHA Dated: November 19, 2014 Received: November 21, 2014
Dear Dr. Aloni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Dr. Aloni
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
of Surveillance and Biometrics/Division of Postmarket Surveillance.
Sincerely yours,
Susan Runno, DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a combination of red and gray. Below the text, there is a smaller line of text that reads "ENGINEERED FOR HEALTH" in gray.
# Indications for Use
510(k) Number: K140440
Device Name: Noris Medical Dental Implants System
Indications for Use:
Noris Medical Ltd Dental Implants System is intended to replace missing tooth/ teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Mono implants are specifically indicated for replacing maxillary lateral incisors and mandibular central and lateral incisors. They are used for immediate, non-occlusal provisionalization in single-tooth restorations. Multiple-unit restorations should be splinted together and may be used immediately when clinically appropriate.
Prescription Use Yes Over-The-Counter Use No AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a light gray sans-serif font. Below the text "NORIS Medical" is the text "ENGINEERED FOR HEALTH" in a smaller, darker gray sans-serif font.
# 510(k) Summary
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the Noris Medical Dental Implants System
#### 1. GENERAL INFORMATION
| Date Prepared: | 11th December 2014 |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Noris Medical Dental Implants System |
| Common Name: | Endosseous Dental Implant |
| Classification Name: | Implant, Endosseous, Root-Form |
| Class: | II |
| Product Code: | DZE/NHA |
| CFR section: | 21 CFR§872.3640 |
| Device panel: | Dental |
| Legally Marketed<br>Predicate Device: | K040807-MIS, K080162-MIS, K112440-A.B. dental, K132125-<br>A.B.dental, K061477 Nobel Biocare |
| Submitter: | Noris Medical Ltd. 8 Hataasia street, Nesher 3688808, Israel |
| Contact: | Dr. Raanan Aloni, QA/RA Director<br>E-mail: aloni@norismedical.com<br>and Simha Sibony- Regulatory Affairs specialist<br>E-mail: simha.qualitech@gmail.com<br>Tel: +972-73-796-4477<br>Fax: +972-4-695-0991 |
#### 2. DEVICE DESCRIPTION
The Noris Medical Dental Implants System consists of one or two stage endosseous form dental implants, internal hexagonal and one piece implants system
Abutments and Superstructures are used in conjunction with an endosseous dental implant fixture to aid in prosthetic rehabilitation.
The implantation procedure can be accomplished in a one-stage or two-stage surgical operation for all implants type beside the Mono which is for one stage only.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a combination of red and gray. Below the text, there is a smaller line of text that reads "ENGINEERED FOR HEALTH" in gray.
# 2.A Implants sizes and dimensions:
#### Tuff Implants
With their three thread zones, Tuff implants have been uniquely designed according to the anatomy of the bone structure. The lower V-shape thread zone enables self-tapping. The middle zone square type thread is used for compressing cancellous bone. Micro threads are
on the upper zone.
Diameter 3.3, 3.75, 4.2, 5mm Length 8, 10, 11.5,13,16 mm Diameter 6 mm Length 8, 10, 11.5.13 Made of titanium alloy Ti 6Al 4V ELI
#### Tuff TT Implants
Tuff TT Implants share the same three-thread zone concept as Tuff implants. The taper top converging coronal shape allows platform switch technology .
Diameter 4.2, 5 mm ; Length 8, 10, 11.5,13,16 mm Diameter 6 mm Length 8, 10, 11.5.13 Made of titanium alloy Ti 6Al 4V ELI
#### Onyx Implants
Onyx implants are specially designed for treatments involving placement in type I and type II bone.
Diameter 3.3, 3.75, 4.2, 5, mm Length 8, 10, 11.5,13, 16 mm Diameter 6 mm Length 8, 10, 11.5,13 Made of titanium alloy Ti 6Al 4V ELI
#### Mono Implants
Mono implants are specifically indicated for replacing maxillary lateral incisors and mandibular central and lateral incisors. The implants are cleared for immediate, nonocclusal provisionalization in single-tooth restorations. Multiple-unit restorations should be splinted together. In appropriate clinical conditions Mono Implants may be loaded immediately.
Diameter 3.0, 3.3, 3.75, 4.2 ; Length 10, 11.5, 13,16 mm Diameter 5 mm Length 10, 11.5, 13 mm Made of titanium alloy Ti 6Al 4V ELI
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a combination of red and gray. Below "NORIS Medical" is the text "ENGINEERED FOR HEALTH" in gray. The logo appears to be for a medical or healthcare-related company.
The Noris Medical Dental Implants System includes prosthetics components that consist of healing caps, Cemented restorations: straight and angular abutments (narrow/narrow top/wide/shoulder/anatomic/esthetic abutment); Screw retained restorations: Multi unit, Esthetic screw abutments; Removable restorations : Vari connect abutments, Ball attachments, Flat attachments. Accessories: Cover screw, Angular Adaptor, Locator, transfer, Analog and others.
# 2.B Prosthetics size and dimensions
#### Healing Caps
Healing caps prepare the site for the superstructure insertion and "shapes" the soft tissue surrounding the implant.
Diameter 3.8, 4.6, 5.5, 6.3 mm; Length 2, 3, 4, 5, 6, 7 mm Made of titanium alloy Ti 6Al 4V ELI
#### Straight Abutments- cemented retained reconstruction
A wide variety of titanium straight abutments are available for use in different situations.
- Narrow abutments ●
- Wide abutments
- Anatomic abutments
- Narrow shoulder abutments
Diameter 5.2 mm; Length 9, 10, 11, 12 mm; shoulder 1, 2, 3, 4 mm Diameter 4.5 mm; Length 8.5, 9.5, 10.5, 11.5 mm; shoulder 1, 2, 3, 4 mm Diameter 4.5 mm: Length 8.5. 9.5. 10.5. 11.5. 12.5 mm Diameter 3.75 mm; Length 8.5 mm; shoulder 0.5, 1.5, 2.75 mm Diameter 3.8 mm; Length 6, 8 mm Diameter 5.5 mm; Length 9, 11 mm Made of titanium alloy Ti 6Al 4V ELI
*Implant – abutment connection i same 3.75mm platform for all abutments .
#### Angular Abutments - cemented retained reconstruction
The abutments are available with angles of 15° and 25° degrees.
- Angular, narrow abutments are used in minimal prosthetic space.
- Angular anatomic abutments have a shape that is contoured to the gingiva. This enables an individualization of the prosthetic unit. Variety of shoulder heights provides solutions for the alignment of the abutment with the soft tissue topography.
- Narrow top angular abutment allows minimal technical preparation. ●
{7}------------------------------------------------
Anatomic - 15° and 25°:
Diameter 5.4 mm; Length 9.5, 10.5, 11.5, 12.5 mm; shoulder 1, 2, 3, 4
mm
Anatomic - 15°: Diameter 5.4 mm; Length 8.5 mm; shoulder 0.8 mm
Standard - 15° and 25°: Diameter 5.4 mm; Length 9, 11 mm
Narrow - 15°: Diameter 4 mm; Length 9 mm
Narrow Top - 15° and 25°: Length 9 mm
Made of titanium alloy Ti 6A1 4V ELI
# Esthetic Abutments
The esthetic abutments are available in different shapes, sizes and angulations.
Straight - Diameter 5.4 mm; Length 8.5, 9.5, 10.5 mm; shoulder 1, 2, 3 mm Angulated - 15° and 25°: Diameter 4.5 mm; Length 9, 11 mm Made of titanium alloy Ti 6Al 4V ELI
Color, using an anodizing technique, for categorical considerations only.
# Esthetic Screw Abutments - screw retained restoration
Esthetic Screw Abutment is designed for the screw retained rehabilitation process on single or multiple units.
Base: Length 0.5, 1.5, 2.5 mm Screw: Length 10.5, 11.5, 12.5 mm Made of titanium alloy Ti 6Al 4V ELI
Plastic Castable Abutments on Titanium Bases enable the dental laboratory to cast on an accurate Titanium base. The machined Titanium base provides an accurate fit to the implant.
Hexed: Diameter 4.5 mm; Length 10.5 mm Made of titanium alloy Ti 6A14V ELI + Delrin
No angular correction may be fabricated on Titanium base plastic castable abutments
# Multi-Unit - screw retained reconstruction
The Multi-Unit system comprises a full range of sizes for both the upper and lower jaws. Straight, 17°, 30° adaptors, in a variety of heights, connects to a wide range of complementary products.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the words "NORIS Medical" in red and gray. Below the words "NORIS Medical" is the phrase "ENGINEERED FOR HEALTH" in gray.
Straight - Length 1, 2, 3 mm Angulated 17° and 30° - Length 2, 3, 4 mm Made of titanium alloy Ti 6Al 4V ELI
#### Vari-Connect - screw retained or removable reconstruction
The Vari-Connect system presents a complete solution for removable prostheses on tilted implants. Straight, 17°, 30° adaptors are available. Complementary products are fixed to the adaptors by the adaptor thread.
Angulated 17° and 30° - Length 2, 3, 4 mm Made of titanium alloy Ti 6Al 4V ELI
#### Ball attachment - removable
The ball attachment superstructure is intended to secure a removable prosthesis.
Length; 0.5, 1, 2, 3, 4, 5, 6 mm Made of titanium alloy Ti 6Al 4V ELI
#### Flat attachment- removable
The Flat attachment superstructure is intended to secure a removable prosthesis.
Length; 0.5, 1, 2, 3, 4, 5 mm Made of titanium alloy Ti 6Al 4V ELI
#### Materials and Production:
The implants and prosthetic components are manufactured from Titanium alloy (Ti 6Al 4V ELI) complying with standard ASTM F 136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for surgical implant applications. Noris Medical employs the SLA (Sandblasted, Large grit, Acid etched) surface treatment technology.
The product is packaged using materials known in the industry to be appropriate for medical device packaging and is provided with a minimum sterility assurance level of 10° validated in compliance with ANSI/AAMI/ISO 11137-1 Sterilization of healthcare products --Radiation -- Part 1:Requirements for development, validation and routine control of a sterilization process for medical devices.
#### 3. INTENDED USE
Noris Medical Dental Implants System is intended to replace missing tooth/ teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a combination of red and gray. Below the text, there is a tagline that reads "ENGINEERED FOR HEALTH" in gray.
operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Mono implants are specifically indicated for replacing maxillary lateral incisors and mandibular central and lateral incisors. They are used for immediate, non-occlusal provisionalization in single-tooth restorations. Multiple-unit restorations should be splinted together and may be used immediately when clinically appropriate.
#### 4. SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES
Noris Medical Dental Implants System(Noris) is substantially equivalent to M.I.S K040807,and M.I.S. K080162 in terms of intended use, design, materials used, safety and performance testing.
# Identification of the Legally Marketed Predicate Devices Used to Claim Substantial Equivalence:
| Device Name | MIS Dental Implant System | Noris Medical Dental Implants System |
|--------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| 510(k)Number | K040807<br>K080162(Uno) | K140440 |
| Sponsor | M.I.S. | Noris Medical |
| Product Code | DZE/NHA | DZE/NHA |
| Patient Population | Edentulous or partially edentulous individuals | Edentulous or partially edentulous individuals |
| Material | Titanium alloy | Titanium alloy |
| Diameter(mm) | Biocom<br>3.3, 3.75, 4.2, 5, 6.<br>Seven<br>3.3, 3.75, 4.2, 5,6<br>Uno<br>3.0, 3.5 | Onyx<br>3.3, 3.75, 4.2, 5, 6<br>Tuff<br>3.3, 3.75, 4.2, 5, 6<br>Tuff TT<br>4.2, 5, 6<br>Mono<br>3.0, 3.3,3.75,4.2,5 |
| Length (mm) | Biocom<br>8, 10, 11.5, 13, 16<br>Seven<br>8, 10, 11.5, 13, 16<br>Uno<br>10,11.5, 13, 16 | Onyx<br>8,10,11.5,13,16<br>Dia 6 -Length 8,10,11.5,13<br>Tuff, Tuff TT<br>8,10,11.5,13, 16<br>Dia 6 -Length 8,10,11.5,13 |
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the words "NORIS Medical" in red and gray. Below the words "NORIS Medical" is the phrase "ENGINEERED FOR HEALTH" in gray.
| | | Mono |
|------------|-----------------------------|------------------------------------------|
| | | 10,11.5,13,16 |
| | | Dia 5 –Length 10,11.5,13 |
| Abutments° | 0,15,25 | 0,15,17,25,30 |
| Design | Root Form | Root Form |
| Prosthetic | Internal Hex | Internal Hex |
| Connection | | |
| Sterility | Gamma irradiation | Gamma irradiation |
| | Immediate loading or | Immediate loading or |
| Clinical | for loading after a | for loading after a conventional healing |
| procedure | conventional healing period | period |
The prosthetics components of Noris Medical Dental Implants System are substantially equivalent to K112440-A.B. dental, K132125-A.B.dental, K061477 Nobel Biocare in terms of intended use, design, materials used, safety and performance testing.
# Identification of the Legally Marketed Predicate Devices Used to Claim Substantial Equivalence:
| | AB Dental | Noris Medical |
|---------------|------------------------------------------------------|-------------------------------------------------------|
| Feature | | |
| K510 | K112440 | Present submission |
| Product Name | P0-Healing Cap | Healing cap |
| Intended use | Used to allow the gingiva to<br>heal around implants | Used to allow the gingiva to heal around<br>implants. |
| Material | Titanium alloy | Titanium alloy |
| Diameter (mm) | 3.75 | 3.8,4.6,5.5,6.3 |
| Height (mm) | 2,3,4,5,6,7 | 2,3,4,5,6,7 |
| Angle | 0 | 0 |
| Sterility | Non sterile | Non sterile |
#### Healing cap
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in red and gray. Below the text, there is a tagline that reads "ENGINEERED FOR HEALTH" in gray.
# Straight abutment
| | AB Dental | Noris Medical |
|---------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Feature | | |
| Product Name | P3 - abutment anti rotation | Straight abutment |
| K510 | K112440 | K140440 |
| Intended use | The straight abutment is used in<br>the fabrication of cement-<br>retained restorations, single<br>crowns or bridges. | The straight abutment is used in the<br>fabrication of cement-retained<br>restorations, single crowns or bridges. |
| Material | Titanium alloy | Titanium alloy |
| Diameter (mm) | 3,3.75,5 | 3.75,3.8,4.5,5.2,5.5,9 |
| Height (mm) | 5,7,9,12,15 | 6,8,8.5,9,9.5,10,10.5,11,11.5,12,12.5,15 |
| Angle | 0 | 0 |
| Sterility | Non sterile | Non sterile |
# Vari-connect
| Feature | AB Dental | Noris Medical |
|---------------|--------------------------------------------------|-------------------------------------------------------------------------------------|
| K510 | K132125 | K140440 |
| Product Name | P5-P14 Ball for Angular Adaptor | Vari-connect |
| Intended use | To connect a removable denture to an<br>implant. | The Vari-Connect system is intended to<br>connect a removable denture to an implant |
| Material | Titanium alloy | Titanium alloy |
| Diameter (mm) | 3.5,3.9 | 4.8 |
| Height (mm) | 1,3 | 2,3,4 |
| Angle (º) | 17,30 | 0, 17, 30 |
| Sterility | Non sterile | Non sterile |
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a red sans-serif font. Below the text is the tagline "ENGINEERED FOR HEALTH" in a smaller, gray sans-serif font.
### Esthetic Screw Abutments
| Feature | AB Dental | Noris Medical |
|---------------|------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| K210 | K132125 | K140440 |
| Product Name | P7 Anti-rotation Aesthetic<br>Abutment, | Esthetic Screw Abutments |
| Intended use | Suitable for restorations of a<br>single implant or screwed<br>bridge on non-parallel implants | Esthetic Screw Abutment is intended for the<br>screw retained rehabilitation process on<br>single or multiple units. |
| Material | Titanium alloy | Titanium alloy |
| Diameter (mm) | 3.75 | 4.7 |
| Height (mm) | 1,2,3 | 0.5,1.5,2.5 |
| Angle (의) | 0 | 0 |
| Sterility | Non sterile | Non sterile |
# Ball attachment
| Feature | AB Dental | Noris Medical |
|--------------|-------------------------------------------------|-------------------------------------------------------------------------------------|
| K510 | K132125 | K140440 |
| Product Name | P5- Ball attachment<br>abutment | Ball attachment |
| Intended use | To connect a removable<br>denture to an implant | The ball attachment superstructure is<br>intended to secure a removable prosthesis. |
| Material | Titanium alloy | Titanium alloy |
| Height (mm) | 1,2,3,4,5,6 | 0.5,1,2,3,4,5,6,7 |
| Angle (º) | 0 | 0 |
| Sterility | Non sterile | Non sterile |
{13}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a sans-serif font. Below the text, there is a tagline that reads "ENGINEERED FOR HEALTH" in a smaller, lighter font.
# Flat attachment
| Feature | AB Dental | Noris Medical |
|---------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Product Name | P55 low connector | Flat attachment |
| K510 | K132125 | K140440 |
| Intended use | Over denture attachment system<br>for easy connection between the<br>denture and the implant | The Flat attachment superstructure is intended<br>to secure an overdenture |
| Material | Titanium alloy | Titanium alloy |
| Diameter (mm) | 3.75 | 3.75 |
| Height (mm) | 1,2,3,4,5,6,7,8 | 0.5,1,2,3,4,5 |
| Angle (º) | 0 | 0 |
| Sterility | Non sterile | Non sterile |
## Multi-Unit
| Feature | Nobel biocare | Noris Medical |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K510 | K061477 | K140440 |
| Product Name | Multi-unit | Multi-Unit |
| Intended use | Nobel Biocare's Multi-Unit Abutment<br>is a premanufactured prosthetic<br>component intended for use as an aid<br>in prosthetic rehabilitation. | The Multi-Unit Abutment is a prosthetic<br>device that fits only the 2.42 mm internal<br>Hex implants. The device has been<br>developed for long-term, permanent use.<br>The Multi-Unit system provides a solution<br>for screw-retained prosthetic rehabilitation |
| Material | Titanium alloy | Titanium alloy |
| Diameter (mm) | 3.5,3.9 | 4.8 |
| Height (mm) | 2,2.5,3,3.5,4,4.5,5 | 1,2,3,4 |
| Angle (º) | Up to 30 | 0, 17, 30 |
| Sterility | Non sterile | Non sterile |
{14}------------------------------------------------
Image /page/13/Picture/0 description: The image contains the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a larger, bolder font. Below the text, there is a smaller line that reads "ENGINEERED FOR HEALTH".
Evidence of equivalence has been demonstrated through:
- * The Noris Medical Dental Implants System intended use and indications for use were previously cleared by FDA for the predicate devices.
- * The technical characteristics of the Noris Medical Dental implants System are similar to those of the predicate devices.
- * Safety and performance testing of the Noris Medical Dental implants System are similar to those of the predicate devices.
Therefore, the Noris Medical Dental Implants System is substantially equivalent to the predicate devices in terms of intended use, materials used, and technological characteristics
#### 4. NON-CLINICAL TEST
The components are manufactured from medical grade Titanium alloy (Ti 6Al 4V ELI) per ASTM F136.
SEM and Surface analysis (EDS) after SLA process demonstrated the morphology and cleanliness of the final product.
Sterilization validation tests were conducted in compliance with ANSI/AAMI/ISO 11137-1:06 and EN ISO 11137-2:12 in order to ensure safety and effectiveness related to Noris Medical Dental Implants System.
Test results have demonstrated that the SAL of 10th was achieved and all testing requirements were met.
Accelerated aging per ASTM-F-1980:07 have been applied on the final packaging followed by validating durability to peel, dye and burst tests conditions in order to substantiate 5 years shelf life.
Static and dynamic compression performance test was conducted per ISO 14801: 07-Dentistry-Implants-Dynamic fatigue test for Endosseous Dental implants. The worst case scenario was chosen based on the FDA guideline "Class II Special Controls Guidance Document: Root form for Endosseous dental implants and Endosseous dental Implant Abutments".
The results of the testing indicate that the Noris Medical Dental Implants System is substantial equivalent to the predicate devices sighted in this submission.
#### 5. CLINICAL TEST
No clinical studies were performed.
## 6. CONCLUSION
The results of the testing conducted on the Noris Medical Dental Implants System demonstrated that the system is substantially equivalent to the named predicated devices in terms of functional, mechanical properties, indications for use and material.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.