Smith & Nephew, Inc. Plates and Screws Systems: EVOS, Peri-Loc, D-Rad, TC-100, VLP Mini-Mod, Compression Hip Screw (CHS), CONQUEST FN, and cannulated screws
K213126 · Smith & Nephew, Inc. · JDO · Sep 29, 2022 · Orthopedic
Device Facts
Record ID
K213126
Device Name
Smith & Nephew, Inc. Plates and Screws Systems: EVOS, Peri-Loc, D-Rad, TC-100, VLP Mini-Mod, Compression Hip Screw (CHS), CONQUEST FN, and cannulated screws
Applicant
Smith & Nephew, Inc.
Product Code
JDO · Orthopedic
Decision Date
Sep 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The EVOS MINI Plating System is indicated for adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The EVOS MINI Plating System is indicated for fracture fixation, arthrodesis, reconstruction, replantation of small bones and small bone fragments. This system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones. The EVOS Small Fragment Plating System is indicated for adult and adolescent (greater than 12-21 years of age) patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, and clavicle. The EVOS Partial Articular and Anti-Glide Plates are indicated for the treatment of partial articular fractures of the distal and proximal tibia (AO/OTA Fracture Classifications Type B), and for fracture fixation of the fibula. The EVOS Wrist Fracture Plating System is indicated for adult and pediatic patients with osteopenic bone. It is indicated for fixation of fractures, malunions, and osteotomies involving the radius and ulna. The EVOS Large Fragment Plating System is indicated for adult patients. It is indicated for fixation of long bone fractures.
Device Story
Metallic bone fixation systems (plates and screws) used for fracture fixation, arthrodesis, and reconstruction of small and long bones; pelvic, distal radius, and femoral neck fractures. Devices implanted by surgeons in clinical settings. Input: surgical need for bone stabilization. Transformation: mechanical fixation via plates and screws. Output: stabilized bone fragments. Benefit: promotes bone healing and structural integrity. Current submission updates MR safety information for existing cleared systems; no changes to design, materials, or manufacturing.
Clinical Evidence
Bench testing only. MRI compatibility testing conducted per FDA guidance and standards: ASTM F2052 (displacement force), ASTM F2213 (torque), ASTM F2182 (RF heating), and ASTM F2119 (image artifact).
Technological Characteristics
Metallic bone fixation appliances (plates and screws). Design, materials, and manufacturing are identical to previously cleared predicates. Passive implants.
Indications for Use
Indicated for fracture fixation, arthrodesis, reconstruction, and stabilization of small, long, and pelvic bones in adult, adolescent (12-21 years), and pediatric patients, including those with osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 29, 2022
Smith & Nephew Inc. Rose Beifuss Senior Regulatory Affairs Manager 1450 E Brooks Rd Memphis, Tennessee 38116
Re: K213126
Trade/Device Name: Smith & Nephew, Inc. Plates and Screws Systems: EVOS, Peri-Loc, D-Rad, TC-100, VLP Mini-Mod, Compression Hip Screw (CHS), CONQUEST FN, and cannulated screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDO, KTT, HRS, HWC, HTN, NDG, HTY Dated: August 19, 2022 Received: August 22, 2022
Dear Rose Beifuss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For: Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Device Name EVOS
#### Indications for Use (Describe)
The EVOS MINI Plating System is indicated for adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone.
The EVOS MINI Plating System is indicated for fracture fixation, arthrodesis, reconstruction, replantation of small bones and small bone fragments. This system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.
The EVOS Small Fragment Plating System is indicated for adult and adolescent (greater than 12-21 years of age) patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, and clavicle.
The EVOS Partial Articular and Anti-Glide Plates are indicated for the treatment of partial articular fractures of the distal and proximal tibia (AO/OTA Fracture Classifications Type B), and for fracture fixation of the fibula.
The EVOS Wrist Fracture Plating System is indicated for adult and pediatic patients with osteopenic bone. It is indicated for fixation of fractures, malunions, and osteotomies involving the radius and ulna.
The EVOS Large Fragment Plating System is indicated for adult patients. It is indicated for fixation of long bone fractures.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> </div> |
|----------------------------------------------|--------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known)
K213126
Device Name PERI-LOC
#### Indications for Use (Describe)
The PERI-LOC Plate and Screw System can be used for adult and pediativ patients with osteopenic bone. PERI-LOC bone plates and bone screws are indicated for fixation of pelvic, small and long bone fractures, including those of the tibia, fibula, femur, pelvis, acetabulum, metacarpals, humerus, ulna, radius, calcaneus, and clavicle.
PERI-LOC Periarticular Locked Plating System VLP Plates and Screws can be used for adult and pediatic patients, as well as patients with osteopenic bone. PERI-LOC contoured VLP Plates and Screws are indicated for partial articular fractures (AO/OTA Fracture Classification Type B) of the distal and for fracture fixation of the firstion of the firstilula. PERI-LOC VLP One-Third Tubular Locking Plates are indicated for, but not limited to, fixation of fractures, non-unions and osteotomies of the medial malleolus, fibula, distal ulna, olecranon, calcaneus and metatarsals.
PERI-LOC Periarticular Locked Plating System Proximal Femur Bone Screws and Cable Accessories can be used for adult patients as well as patients with osteopenic bone. PERI-LOC Proximal Femur Bone Plates, Bone Screws and Cable Accessories are indicated for fractures of the trochanteric region including simple intertrochanteric, reverse oblique trochanteric, transverse trochanteric, complex multi-fragmentary, and fractures with medial cortex instability; proximal femur fractures combined with ipsilateral shaft fractures of the proximal femur including metastatic fractures: proximal fenur osteotonies: fixation of fractures in osteopenic bone; fixation of nonunions and malunions; basi/transcervical femoral neck fractures; subcapital femoral neck fractures; and subtrochanteric femur fractures.
The Smith & Nephew VLP FOOT Plating System can be used in adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The VLP FOOT Plating System is indicated for fixation, reconstruction or arthrodeses of small bones, including those in the forefoot, midfoot and hindfoot.
The Smith & Nephew PERI-LOC Ankle Fusion Plating System can be used in adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The PERI-LOC Ankle Fusion Plating System is indicated for ankle arthrodesis and fractures, including the distal tibia, talus and calcaneus.
The VLP MINI-MOD Small Bone Plating System and VLP MINI-MOD Talus Plates are indicated for adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone.
The VLP MINI-MOD Small Bone Plating System is indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bone fragments. This system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.
The VLP MINI-MOD Talus Plates are indicated for fracture fixation, reconstruction or arthrodesis of small bones, including those in the forefoot, midfoot and hindfoot.
The VLP Wrist Fracture System Radial Plates are indicated for fixation of fractures, malunions, and osteotomies involving the radius.
{4}------------------------------------------------
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K213126
Device Name D-Rad
Indications for Use (Describe) The D-RAD SMART PACK System is intended for the fixation of fractures involving the distal radius.
Type of Use (Select one or both, as applicable)
| <div> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="text-decoration: underline;"></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K213126
Device Name TC-100
Indications for Use (Describe)
Bone plates and screws from the Smith & Nephew Bone Plate System are used for adult and pediatic as indicated for pelvic, small, and long bone fracture fixations for use include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarsals, humerus, ulna, middle hand and middle foot bones; treatment of the calcaneus; hip arthrodesis, and provisional hole fixation.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### K213126
Device Name
Compression Hip Screw (CHS)
#### Indications for Use (Describe)
1. Intracapsular fractures of the femoral neck. (For high subcapsular fractures it may be more prosthesis
in lieu of internal fixation to reduce the risk of a nonunion or avascular necrosis of the femoral head.)
2. Trochanteric or subtrochanteric fractures with appropriate additional precautions about weight bearing and more than sedentary activity.
3. Osteotomies for patients with diseases or deformities of the hip.
- 4. Hip arthrodesis.
5. Supracondylar fractures and distal femoral fractures using a supracondylar plate.
6. Ipsilateral femoral shaft/neck fractures.
| CHOOLEAN COLLEGIAL And Antitle AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA | | | Company of Area Annual Andrew of Children Company |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-------------------------------------------------|---------------------------------------------------|
| | | Type of Use (Select one or both, as applicable) | |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{8}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K213126
Device Name CONQUEST FN
Indications for Use (Describe) CONQUEST FN is indicated for displaced and undisplaced intracapsular femoral neck fractures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{9}------------------------------------------------
510(k) Number (if known)
#### K213126
Device Name Cannulated Screws
#### Indications for Use (Describe)
The 4.0mm Cannulated Screws and associated washers are additionally intended for arthrodesis and osteotomies of small bones and small joints, including scaphoid and other carpal bones, metatarsals, patella, ultar styloid, capitellum, radial head and radial styloid.
The 5.5mm, 6.5mm, 7.0mm, and 8.0mm Cannulated Screws and associated washers are additionally intended for reconstruction, osteotomy, and arthrodesis of various bone fragments appropriate for the size of the device including joint fusions (arthrodesis) in the foot and ankle.
The Smith & Nephew 2.5mm, 3.0mm Cannulated and 3.0mm Headless Compression Screws are intended for fixation of intraarticular and extra-articular fractures and non-unions of small bone fragments; arthrodeses of small ioints; bunionectomies and osteotomies; scaphoid and other carpals, tarsals, metatarsals, patella, ulnar styloid, capitellum, radial head and radial styloid.
The Smith & Nephew 2.0mm OFX Screw is indicated for osteotomies of the lesser metatarsals, such as Well osteotomies. Osteotomies, fusions and fractures of the phalanges, metacarpals and carpals of the hand.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span style="font-weight: bold;"></span> </span> </span> <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span style="font-weight: bold;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </span> | <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span style="font-weight: bold;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </span> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the Smith & Nephew logo. The logo consists of an orange flower-like symbol on the left, followed by the text "We are" in gray. To the right of "We are" is the text "smith&nephew" in orange. The logo is clean and modern, with a focus on the company name.
| Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary: | August 19, 2022 |
| Contact Person and Address: | Rose Beifuss<br>Senior Manager, Regulatory Affairs<br>T 385-253-2551 |
| Name of Device: | Smith & Nephew, Inc. Plates and Screws<br>Systems: EVOS, Peri-Loc, D-Rad, TC-100, VLP<br>Mini-Mod, Compression Hip Screw (CHS),<br>CONQUEST FN, and cannulated screws |
| Common Name: | Bone Plates and Screws |
| Device Classification Name<br>and Reference: | 21 CFR 888.3030 Single/multiple component<br>metallic bone fixation appliances and accessories. |
| | 21 CFR 888.3040 Smooth or threaded metallic<br>bone fixation fastener |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | JDO, KTT, HRS, HTN, NDG, HWC, HTY |
| Predicate Device: | Smith & Nephew Plates and Screws Systems:<br>EVOS, Peri-Loc, D-Rad, TC-100, VLP Mini-Mod, Compression Hip Screw (CHS). |
CONQUEST FN, and cannulated screws
### Device Description
The purpose of this Traditional 510(k) is to update the MR safety information for the Smith & Nephew Plates and Screws Systems: EVOS, Peri-Loc, D-Rad, TC-100, VLP Mini-Mod, Compression Hip Screw (CHS), CONQUEST FN, and cannulated screws. The technological characteristics, function of the devices, packaging and sterilization remain unchanged.
No modification has been made to the device design, material, sterilization and the manufacturing processes of the previously cleared devices.
### Indication for Use
Indications for use of the Smith & Nephew, Inc. Plates and Screws are the following:
### Compression Hip Screw (CHS)
- 1. Intracapsular fractures of the femoral neck. (For high subcapsular fractures it may be more prudent to select a prosthesis in lieu of internal fixation to reduce the risk of a nonunion or avascular necrosis of the femoral head.)
{11}------------------------------------------------
- 2. Trochanteric or subtrochanteric fractures with appropriate additional postoperative precautions about weight bearing and more than sedentary activity.
- 3. Osteotomies for patients with diseases or deformities of the hip.
- 4. Hip arthrodesis.
- 5. Supracondylar fractures and distal femoral fractures using a supracondylar plate.
- 6. Ipsilateral femoral shaft/neck fractures.
# CONQUEST FN
CONQUEST FN is indicated for displaced and undisplaced intracapsular femoral neck fractures.
# D-Rad
The D-RAD SMART PACK System is intended for the fixation of fractures involving the distal radius.
# EVOS
The EVOS MINI Plating System is indicated for adolescent (12–18 years) and transitional adolescent (18–21 years) subpopulations and adults, as well as patients with osteopenic bone.
The EVOS MINI Plating System is indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bones and small bone fragments. This system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.
The EVOS Small Fragment Plating System is indicated for adult and adolescent (greater than 12-21 years of age) patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle.
The EVOS Partial Articular and Anti-Glide Plates are indicated for the treatment of partial articular fractures of the distal and proximal tibia (AO/OTA Fracture Classifications Type B), and for fracture fixation of the fibula.
The EVOS Wrist Fracture Plating System is indicated for adult and pediatric patients, as well as patients with osteopenic bone. It is indicated for fixation of fractures, malunions, and osteotomies involving the radius and ulna.
The EVOS Large Fragment Plating System is indicated for adult patients. It is indicated for fixation of long bone fractures.
PERI-LOC
{12}------------------------------------------------
The PERI-LOC Plate and Screw System can be used for adult and pediatric patients, as well as patients with osteopenic bone. PERI-LOC bone plates and bone screws are indicated for fixation of pelvic, small and long bone fractures, including those of the tibia, fibula, femur, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, radius, calcaneus, and clavicle,
PERI-LOC Periarticular Locked Plating System VLP Plates and Screws can be used for adult and pediatric patients, as well as patients with osteopenic bone. PERI-LOC contoured VLP Plates and Screws are indicated for partial articular fractures (AO/OTA Fracture Classification Type B) of the distal and proximal tibia, and for fracture fixation of the fibula. PERI-LOC VLP One-Third Tubular Locking Plates are indicated for, but not limited to, fixation of fractures, non-unions and osteotomies of the medial malleolus, fibula, distal ulna, olecranon, calcaneus and metatarsals.
PERI-LOC Periarticular Locked Plating System Proximal Femur Bone Plates, Bone Screws and Cable Accessories can be used for adult patients as well as patients with osteopenic bone. PERI-LOC Proximal Femur Bone Plates, Bone Screws and Cable Accessories are indicated for fractures of the trochanteric reqion including simple intertrochanteric, reverse oblique trochanteric, transverse trochanteric, complex multifragmentary, and fractures with medial cortex instability; proximal femur fractures combined with ipsilateral shaft fractures: pathological fractures of the proximal femur including metastatic fractures; proximal femur osteotomies; fixation of fractures in osteopenic bone; fixation of nonunions and malunions; basi/transcervical femoral neck fractures; subcapital femoral neck fractures; and subtrochanteric femur fractures.
The Smith & Nephew VLP FOOT Plating System can be used in adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The VLP FOOT Plating System is indicated for fracture fixation, reconstruction or arthrodeses of small bones, including those in the forefoot. midfoot and hindfoot.
The Smith & Nephew PERI-LOC Ankle Fusion Plating System can be used in adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The PERI-LOC Ankle Fusion Plating System is indicated for ankle arthrodesis and fractures, including the distal tibia, talus and calcaneus.
The VLP MINI-MOD Small Bone Plating System and VLP MINI-MOD Talus Plates are indicated for adolescent (12–18 years) and transitional adolescent (18–21 years) subpopulations and adults, as well as patients with osteopenic bone.
The VLP MINI-MOD Small Bone Plating System is indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bones and small bone fragments. This system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.
The VLP MINI-MOD Talus Plates are indicated for fracture fixation, reconstruction or arthrodesis of small bones, including those in the forefoot, midfoot and hindfoot,
The VLP Wrist Fracture System Radial Plates are indicated for fixation of fractures, malunions, and osteotomies involving the radius.
{13}------------------------------------------------
# TC-100
Bone plates and screws from the Smith & Nephew Bone Plate System are used for adult and pediatric patients as indicated for pelvic, small, and long bone fracture fixation. Indications for use include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, middle hand and middle foot bones; treatment of the calcaneus; hip arthrodesis, and provisional hole fixation.
# Cannulated Screws
The 4.0mm Cannulated Screws and associated washers are additionally intended for arthrodesis and osteotomies of small bones and small joints, including scaphoid and other carpal bones, metacarpals, tarsals, metatarsals, patella, ulnar styloid, capitellum, radial head and radial styloid.
The 5.5mm. 6.5mm. 7.0mm. and 8.0mm Cannulated Screws and associated washers are additionally intended for reconstruction, osteotomy, and arthrodesis of various bones and bone fragments appropriate for the size of the device including joint fusions (arthrodesis) in the foot and ankle.
The Smith & Nephew 2.5mm, 3.0mm Cannulated and 3.0mm Headless Compression Screws are intended for fixation of intraarticular and extra-articular fractures and nonunions of small bones and small bone fragments; arthrodeses of small joints; bunionectomies and osteotomies; scaphoid and other carpal bones, metacarpals, tarsals, metatarsals, patella, ulnar styloid, capitellum, radial head and radial styloid.
The Smith & Nephew 2.0mm QFX Screw is indicated for osteotomies of the lesser metatarsals, such as Weil osteotomies, Osteotomies, fusions and fractures of the phalanges, metacarpals and carpals of the hand.
# Technological Characteristics
The device design and material of the subject device are same as the predicate Smith & Nephew Plates and Screws Systems: EVOS, Peri-Loc, D-Rad, TC-100, VLP Mini-Mod, Compression Hip Screw (CHS), CONQUEST FN, and cannulated screws cleared under the premarket notifications listed in the tables below.
# Performance Data
Below listed Magnetic resonance imaging (MRI) compatibility testing were conducted as per the FDA's guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment", December 11, 2014 and the standards listed below:
- 1. Magnetically induced displacement force (ASTM F2052)
- 2. Magnetically induced torque (ASTM F2213)
- 3. Radiofrequency (RF) induced heating (ASTM F2182)
- 4. MR image artifact (ASTM F2119)
# Substantial Equivalence Information
{14}------------------------------------------------
The Smith & Nephew Plates and Screws Systems are identical in function, design features, materials, sterilization, manufacturing methods and operational principles to the commercially available predicate devices listed in the tables below.
| | | | Table 5.1: Substantially Equivalent Predicates to the Compression Hip Screw (CHS) |
|-------------------------|--------------------------------------------------|----------------------|-----------------------------------------------------------------------------------|
| Svstem | | | |
| | | | |
| Manufacturer | Description | Submission<br>Number | Clearance Date |
| Smith & Nephew,<br>Inc. | Smith & Nephew<br>Compression Hip Screw<br>(CHS) | K933027 | 02/22/1994 |
| | | K952697 | 09/12/1995 |
| | | K954712 | 10/31/1995 |
| | | K993289 | 12/20/1999 |
| | | K210837 | 06/29/2021 |
# Table 5.2: Substantially Equivalent Predicates to the CONQUEST FN System
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|-------------------------|--------------------------------------|----------------------|----------------|
| Smith & Nephew,<br>Inc. | Smith & Nephew<br>CONQUEST FN System | K152686 | 03/17/2016 |
| Smith & Nephew, | Smith & Nephew | K172785 | 12/22/2017 |
| Inc. | CONQUEST FN System | K193029 | 02/28/2020 |
| | | K210837 | 06/29/2021 |
# Table 5.3: Substantially Equivalent Predicates to the External Fixation Systems
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|-------------------------|----------------------|-------------------------------|----------------------------------------|
| Smith & Nephew,<br>Inc. | D-Rad Plating System | K132296<br>K161665<br>K210837 | 01/07/2014<br>11/15/2016<br>06/29/2021 |
# Table 5.4: Substantially Equivalent Predicates to the EVOS Plating System
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|-------------------------|---------------------|----------------------|----------------|
| Smith & Nephew,<br>Inc. | EVOS Plating System | K140814 | 05/07/2014 |
| | | K162078 | 11/18/2016 |
| | | K170457 | 06/14/2017 |
| | | K170887 | 04/24/2017 |
| | | K173293 | 01/08/2018 |
| | | K181533 | 08/09/2018 |
| | | K190253 | 03/11/2019 |
| | | K201527 | 07/22/2020 |
| | | K201918 | 11/16/2020 |
| | | K203405 | 01/27/2021 |
| | | K210837 | 06/29/2021 |
# Table 5.5: Substantially Equivalent Predicates to the subject Systems
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|-------------------------|-------------------------------------------------------------------------------|----------------------|----------------|
| Smith & Nephew,<br>Inc. | Peri-Loc Periarticular and<br>Variable-Angled Locked Plating<br>(VLP) Systems | K033669 | 12/10/2003 |
| | | K051735 | 07/19/2005 |
| | | K061352 | 06/08/2006 |
| | | K062216 | 09/15/2006 |
| | | K071563 | 08/08/2007 |
| | | K072818 | 11/19/2007 |
{15}------------------------------------------------
| K080434 | 04/10/2008 |
|---------|------------|
| K082516 | 09/17/2008 |
| K083032 | 01/07/2009 |
| K090675 | 06/04/2009 |
| K092015 | 07/30/2009 |
| K100325 | 05/04/2010 |
| K110670 | 07/12/2011 |
| K120667 | 04/04/2012 |
| K132886 | 02/04/2014 |
| K143050 | 12/18/2014 |
| K152976 | 11/12/2015 |
| K210837 | 06/29/2021 |
| | Table 5.6: Substantially Equivalent Predicates to the TC-100 Screw and Plating System | | | | | | | |
|--|---------------------------------------------------------------------------------------|--|--|--|--|--|--|--|
|--|---------------------------------------------------------------------------------------|--|--|--|--|--|--|--|
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|-------------------------|------------------------------------|-------------------------------|----------------------------------------|
| Smith & Nephew,<br>Inc. | TC-100 Screw and Plating<br>System | K123055<br>K993106<br>K210837 | 01/25/2013<br>12/09/1999<br>06/29/2021 |
### Table 5.7: Substantially Equivalent Predicates to the Cannulated Screws
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|-------------------------|-------------------|----------------------|----------------|
| Smith & Nephew,<br>Inc. | Cannulated Screws | K060736 | 04/18/2006 |
| | | K111994 | 10/11/2011 |
| | | K133662 | 5/15/2014 |
| | | K210837 | 06/29/2021 |
# Conclusion
In summary, the only differences between the subject devices and what was previously cleared is the updated MR safety information within the package insert. The Smith & Nephew Plates and Screws Systems are identical in function, design features, materials, sterilization, manufacturing methods and operational principles to what was previously cleared. Therefore, it is concluded that the Smith & Nephew Plates and Screws Systems are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.