NeuroStar Advanced Therapy for adjunctive treatment of OCD, NeuroStar Advanced Therapy System, NeuroStar TMS Therapy System, NeuroStar Advanced Therapy for Mental Health, NeuroStar
K212289 · Neuronetics, Inc. · QCI · May 6, 2022 · Neurology
Device Facts
Record ID
K212289
Device Name
NeuroStar Advanced Therapy for adjunctive treatment of OCD, NeuroStar Advanced Therapy System, NeuroStar TMS Therapy System, NeuroStar Advanced Therapy for Mental Health, NeuroStar
Applicant
Neuronetics, Inc.
Product Code
QCI · Neurology
Decision Date
May 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5802
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NeuroStar Advanced Therapy system is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).
Device Story
NeuroStar Advanced Therapy System is a computerized, electromechanical device delivering non-invasive transcranial magnetic stimulation (TMS) to the cerebral cortex. It uses an electromagnetic coil placed near the scalp to induce electric fields, causing localized axonal depolarization in targeted brain structures (specifically the dorsomedial prefrontal cortex for OCD). The system comprises a mobile console, software GUI, treatment chair, head support, coil positioning system, and a bent figure-8 coil with a ferromagnetic core. Operated by clinicians in a clinical setting, the device delivers magnetic pulses based on pre-set stimulation parameters (intensity, frequency, train duration). The output assists in modulating neuronal activity to treat OCD. Safety is managed through high-voltage and leakage current testing, and compliance with IEC60601 standards. The device provides a non-invasive therapeutic option for patients, with clinical decision-making supported by the clinician's assessment of the patient's condition and the device's standardized treatment protocols.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing, including magnetic/electric field distribution measurements, pulse timing verification, and software verification/validation. Compliance with IEC60601-1 and IEC60601-1-2 standards confirmed.
Technological Characteristics
Electromechanical TMS system; bent figure-8 coil with ferromagnetic core; air-cooled. Output: Biphasic waveform, 0.22-2.08 SMT intensity range. Connectivity: Data management system. Standards: ISO 13485:2016, IEC60601-1 (Ed 3.1), IEC60601-1-2:2014. Software: GUI-based control for treatment parameters.
Indications for Use
Indicated for adult patients with Obsessive-Compulsive Disorder (OCD) as an adjunct treatment.
Regulatory Classification
Identification
A transcranial magnetic stimulation system for neurological and psychiatric disorders and conditions is a prescription, non-implantable device that uses brief duration, rapidly alternating, or pulsed, magnetic fields to induce neural activity in the cerebral cortex. It is not intended for applying or focusing magnetic fields towards brain areas outside cerebral cortex (e.g., cerebellum). A repetitive transcranial magnetic stimulation system that is intended to treat major depressive disorder is classified in § 882.5805. A transcranial magnetic stimulation system for headache is classified in § 882.5808.
Special Controls
In combination with the general controls of the FD&C Act, the transcranial magnetic stimulation system for neurological and psychiatric disorders and conditions is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Performance testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety.
(2) Software verification, validation, and hazard analysis must be performed.
(3) The patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Magnetic pulse output testing;
(ii) Magnetic and electrical field testing;
(iii) Testing of the safety features built into the device; and
(iv) Testing of the sound levels patients are exposed to during device use.
(5) The physician and patient labeling must include the following:
(i) The risks and benefits associated with use of the device;
(ii) Detailed instructions to prevent seizures, to monitor the patient for seizure activity during treatment, and to provide seizure management care if one were to occur during treatment; and
(iii) A description of the ear protection to be worn by the patient during use of the device, including the type of protection and its noise reduction rating.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 6, 2022
Neuronetics, Inc Robin Fatzinger Sr. Director, Regulatory Affairs 3222 Phoenixville Pike Malvern, Pennsylvania 19355
Re: K212289
Trade/Device Name: NeuroStar Advanced Therapy for adjunctive treatment of OCD, NeuroStar Advanced Therapy System, NeuroStar TMS Therapy System, NeuroStar Advanced Therapy for Mental Health Regulation Number: 21 CFR 882.5802 Regulation Name: Transcranial magnetic stimulation system for neurological and psychiatric disorders and conditions Regulatory Class: Class II Product Code: QCI Dated: April 6, 2022 Received: April 7, 2022
Dear Robin Fatzinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Pamela D. Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K212289
Device Name NeuroStar Advanced Therapy System
Indications for Use (Describe)
The NeuroStar Advanced Therapy system is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| <span> <input checked="true" type="checkbox"/> </span> | <span> <input type="checkbox"/> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters "NS" in a bold, purple font. Above the letters, there is a stylized star shape formed by four figures with outstretched arms, also in purple. The figures are arranged around a central point, creating a sense of unity and movement. The overall design is simple yet eye-catching, with the purple color adding a touch of sophistication.
| 510(k) Number: | K212289 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | 06 May 2022 |
| Applicant: | Neuronetics, Inc.<br>3222 Phoenixville Pike<br>Malvern, PA 19355 |
| Primary Contact: | Robin Fatzinger, RAC<br>Sr. Director, Regulatory Affairs<br>Phone: 610-981-4027<br>Email: robin.fatzinger@neurostar.com |
| Secondary Contact: | Cory Anderson<br>VP, R&D and Clinical<br>Phone: 610-981-4104<br>Email: cory.anderson@neurostar.com |
| Device Trade Names: | NeuroStar, NeuroStar TMS Therapy System, NeuroStar Advanced Therapy<br>System, NeuroStar Advanced Therapy for Mental Health |
| Device Common Name: | Transcranial Magnetic Stimulation System for Neurological and<br>Psychiatric Disorders and Conditions |
| Classification: | 21 CFR 882.5802 |
| Product Code: | QCI |
| Predicate Devices: | Primary Predicate: MagVenture TMS Therapy System – for adjunctive<br>treatment of OCD (K193006)<br>Predicate: NeuroStar Advanced Therapy System (K201158, K161519,<br>K160703, K133408, K130233, K083538, DEN060153/K061053) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with the letters "NS" in a bold, purple font. To the right of the letters, there is a stylized star shape formed by three abstract figures with raised arms. The figures are also purple and appear to be connected at the center, creating the star-like design. The overall design is simple and modern.
#### Device Description / Technological Characteristics:
The NeuroStar Advanced Therapy System, for adjunctive treatment of OCD, is a computerized, electromechanical medical device that produces and delivers non-invasive magnetic fields to induce electrical currents targeting specific regions of the cerebral cortex. NeuroStar Advanced Therapy System is indicated as an adjunct for the treatment of adult patients who are suffering from Obsessive-Compulsive Disorder (OCD). NeuroStar Advanced Therapy System has previously obtained FDA clearance for treatment of major depressive disorder (MDD) in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode (K201158, K161519, K160703, K133408, K130233, K083538, DEN060153/K061053).
The subject device is substantially equivalent to the primary predicate device that was cleared under K193006. Neuronetics has implemented minor labeling enhancements to the software user interface and IFU to implement the OCD workflow. While the stimulation parameters have changed for OCD treatment, no changes alter the technical specifications of the subject device.
Consistent with the primary predicate and predicate devices, the proposed NeuroStar Advanced Therapy System, for adjunctive treatment of OCD, enables direct non-invasive activation of brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold and is directed in an appropriate orientation relative the brain's neuronal pathway, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.
#### Intended Use:
The NeuroStar Advanced Therapy system is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).
The intended use is identical to that of the Primary Predicate Device.
## Technological Characteristics and Substantial Equivalence:
The subject device, NeuroStar Advanced Therapy System, has the following similarities to the primary predicate (MagVenture TMS Therapy System - K193006) and predicate (NeuroStar Advanced Therapy System - K201158, K161519, K160703, K133408, K130233, K083538, DEN060153/K061053) devices:
- Principles of operation
- Design for delivery of Transcranial Magnetic Stimulation (TMS)
- o Output stimulation parameters (pulse width, frequency range, pulse interval range, etc.)
- Materials
The proposed changes for the NeuroStar Advanced Therapy System, for adjunctive treatment of OCD, are limited to updates to the IFU and software for user interface and treatment stimulation parameters to facilitate safe and effective treatment of patients with OCD. The proposed changes are supported by information submitted in this premarket notification and with the following rationale:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo with the letters "NS" in a bold, sans-serif font, colored in purple. To the right of the letters, there is a stylized graphic that resembles a star or a stylized human figure with outstretched arms. The graphic is also purple and adds a dynamic element to the logo.
- The subject device is substantially equivalent to the FDA-cleared MagVenture TMS Therapy . System (K193006).
- . The predicate device, NeuroStar Advanced Therapy System, was previously cleared by the FDA under K201158, K161519, K160703, K133408, K130233, K083538, and DEN060153/K061053.
The subject device changes are limited to updates to the IFU and software for user interface and treatment stimulation parameters to facilitate safe and effective treatment of patients with OCD. Therefore, the NeuroStar TMS Therapy System with the proposed changes to the treatment stimulation parameters and labeling is substantially equivalent to the predicate device.
The NeuroStar Advanced Therapy System, for adjunctive treatment of OCD, consists of components that are similar to those of the primary predicate device, with the exception of the treatment coil construction characteristics. The NeuroStar Advanced Therapy System coil is equivalent to the primary predicate device in terms of biocompatibility, design elements, such as cable lengths, coil materials, isolation design and functionalities, however, differs in cooling methods and coil winding configuration / surface area. All coils are subject to high-voltage tests and leakage current tests to ensure safety.
To establish substantial equivalency for the NeuroStar Advanced Therapy System coil, both modeling and performance testing was conducted. Measurements of the magnetic and electric fields of the MagVenture Cool D-B80 coil and NeuroStar coil were taken according to Section 4 of the FDA's Class II Special Controls Guidance document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems.
Measurements of the magnetic field distribution for the Cool D-B80 coil and the NeuroStar Advanced Therapy System coil were performed and determined that the magnetic spatial distribution is substantially equivalent. Both the NeuroStar and MagVenture coils are figure 8 designs, both allowing for deep and broad stimulation of the cortex. Information about magnetic field characteristics, including linearity of output level, magnetic field strength gradients, output waveform and magnetic field spatial distribution according to the Special Controls Guidance document, Section 4 as mentioned above were provided. The coil measurements are based on well-established scientific methods using standard scientific instrumentation.
Measurements of the electric field distribution for the Cool D-B80 coil and the NeuroStar Advanced Therapy System coil were performed and determined that the linearity of the electric field output and the electric field spatial distributions are substantially equivalent. The coil measurements are based on well-established scientific methods using standard scientific instrumentation.
Secondly, performance testing comparing the subject and primary predicate device was conducted to evaluate the equivalence in safety and performance based upon electric field distribution, stimulation volume, magnetic field distribution, and magnetic flux, pulse timing as required in the FDA Special Controls Guidance Document. Results of the testing demonstrate substantial equivalence and confirm that no new safety or effectiveness issues are introduced.
Based on the measurements summarized above, it is therefore concluded, that the magnetic field and electric field properties of the NeuroStar coil are substantially equivalent to the primary predicate device, Cool D-B80.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo with the letters "NS" in a stylized font, colored in purple. Behind the letters, there is a star-like shape formed by abstract figures that resemble people with outstretched arms. The figures are also purple, and they create a sense of movement and energy around the letters.
Additionally, the subject device has been tested in accordance with IEC60601 Edition 3.1 and verified to comply with the specified permissible sound pressure levels and permissible thresholds for exposure defined by the Occupational Safety and Health Administration (OSHA).
These tests provide evidence that the NeuroStar Advanced Therapy system does not pose any risk for potential hearing reduction or loss in either patients or operators.
### Performance Standards:
The NeuroStar Advanced Therapy System, for adjunctive treatment of OCD, has been tested and conforms with the following standards:
- . ISO 13485:2016
- IEC60601-1 .
- . IEC60601-1-2
The following tests were performed to validate the modifications to the device software and indications for use:
- Software verification and validation
- Performance Testing - Bench
In all instances, the subject device functions as intended and meets all the same acceptance as the primary predicate and predicate devices.
### Non-Clinical Testing:
The contents of this 510(k) complies with the FDA Guidance Document: "Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems -Guidance for Industry and Food and Drug Administration Staff". The non-clinical performance testing of the components of NeuroStar Advanced Therapy System, for treatment of OCD, has been tested as required according to the Special Controls per 21 CFRH 882.5802.
## Clinical Testing
This 510(k) does not contain any pivotal clinical trial data related to the new device. The substantial equivalency is established based on similar technological characteristics.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with the letters "NS" in a bold, sans-serif font, colored in purple. Above the letters, there is a stylized star shape, also in purple, with rounded points. The star is formed by what appears to be three abstract figures with outstretched arms, suggesting a sense of community or collaboration. The overall design is clean and modern, conveying a sense of unity and purpose.
## Comparison of Technological Characteristics
| | Subject Device | Predicate Device | Primary Predicate Device | Justification of<br>Differences |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Specification | NeuroStar Advanced Therapy System<br>(for adjunctive treatment of OCD) | NeuroStar Advanced Therapy system<br>(K201158, K161519, K160703,<br>K133408, K130233, K083538,<br>DEN060153/K061053) | MagVenture TMS Therapy<br>System (K193006)<br>Tonica Elektronik A/S, Denmark | |
| Indications<br>for use | The NeuroStar Advanced Therapy<br>System is intended to be used as an<br>adjunct for the treatment of adult<br>patients suffering from Obsessive-<br>Compulsive Disorder (OCD) | Not Applicable | The MagVenture TMS<br>Therapy System is intended<br>to be used as an adjunct for<br>the treatment of adult<br>patients suffering from<br>Obsessive- Compulsive<br>Disorder (OCD) | No Difference |
| Energy<br>Delivered<br>and<br>Performance | Treatment parameters:<br>Intensity:<br>of Leg MT (Leg Motor<br>Threshold) | Not Applicable | Treatment parameters:<br>Intensity:<br>of Leg MT<br>(Leg Motor | No Difference |
| | 100% | | 100% | |
| | Repetition Rate: | | Repetition Rate: | No Difference |
| | 20 Hz | | 20 Hz | |
| | Train Duration: | | Train Duration: | No Difference |
| | 2 sec | | 2 sec | |
| | Inter-train-Interval: | | Inter-train- | No Difference |
| | 20 secs | | 20 secs | |
| | Number of trains: | | Number of | No Difference |
| | 50 | | 50 | |
| | Number of pulses: | | Number of | No Difference |
| | 2000 | | 2000 | |
| | Total duration: | | Total duration: | No Difference |
| | 18.3 min | | 18.3 min | |
| | Treatment area:<br>Area of brain to be stimulated:<br>Dorsomedial Prefrontal Cortex | Not Applicable | Treatment area:<br>Area of brain to be<br>stimulated: | No Difference |
| | Output Stimulation Parameters:<br>Available Stimulation Intensity in terms<br>of Standard Motor Threshold<br>(SMT) units<br>Range: 0.22 - 2.08 SMT<br>Waveform: Biphasic | Output Stimulation Parameters:<br>Available Stimulation Intensity in terms<br>of Standard Motor Threshold (SMT)<br>units<br>Range: 0.22 - 2.08 SMT<br>Waveform: Biphasic | Output Stimulation<br>Parameters: Available<br>Stimulation Intensity in terms<br>of Standard Motor Threshold<br>(SMT) units<br>Range: 0 – 2.6 SMT<br>Waveform: Biphasic | Similar |
| Design | The system consists of:<br>1. Mobile console<br>2. System software with GUI<br>3. Treatment chair<br>4. Head support system<br>5. Coil positioning system<br>6. Same Coil for both MT and treatment<br>7. Coil fixture<br>8. Data management system | The system consists of:<br>1. Mobile console<br>2. System software with GUI<br>3. Treatment chair<br>4. Head support system<br>5. Coil positioning system<br>6. Same Coil for both MT and treatment<br>7. Coil fixture<br>8. Data management system | The system consists of:<br>1. Mobile console<br>2. System software with GUI<br>3. Treatment chair*<br>4. Head support system*<br>5. Coil positioning system<br>6. Same Coil for both MT and<br>treatment<br>7. Coil fixture<br>8. Data management system<br>*optional | No Difference |
| Coil | Bent figure 8 coil with ferromagnetic<br>core | Bent figure 8 coil with ferromagnetic<br>core | Figure 8 double-cone coil, Air<br>Core | Both provide effective<br>stimulation for targeted<br>region based on<br>performance testing. |
| Cooling | Air cooled.<br>Used for both MT determination<br>and treatment | Air cooled.<br>Used for both MT determination<br>and treatment | Liquid cooled<br>Used for both MT<br>determination and treatment. | Both sufficient for safety<br>with treatment duty cycle<br>based on performance<br>testing. |
| Standards | Company complies with EN ISO<br>13485:2016. | Company complies with EN ISO<br>13485:2016. | Company complies with ISO<br>13485:2016. | No Difference |
| Electrical<br>safety | Complies with IEC60601-1 Edition 3.1,<br>and IEC60601-1-2:2014. | Complies with IEC60601-1 Edition 3.1,<br>and IEC60601-1-2:2014. | Complies with IEC60601-1 v.<br>3.1, and IEC60601-1-2 | No Difference |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with the letters "NS" in a bold, purple font. Behind the letters is a stylized star shape, also in purple. The star is formed by three abstract figures with outstretched arms, suggesting movement or celebration. The overall design is simple yet dynamic, conveying a sense of energy and collaboration.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a partial view of a logo or emblem, focusing on the right side. The visible portion includes a stylized, abstract figure in purple, possibly representing a person or a group of people with arms raised or connected. The figure is positioned next to a large, bold letter 'S', also in purple. The overall design suggests a symbol for an organization or brand, potentially related to community, support, or a similar concept.
### Conclusion:
The NeuroStar Advanced Therapy System has the same intended use and indications, principles of operation, and similar technological characteristics as the previously cleared primary predicate device. The minor differences noted in the technical characteristics between the coil assemblies do not raise any new or different questions of safety or effectiveness. Performance data demonstrates that the subject device is safe and effective for its intended use. Thus, the NeuroStar Advanced Therapy System, for adjunctive treatment of OCD, is substantially equivalent to the MagVenture TMS Therapy System - for adjunctive treatment of OCD.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.