21 CFR 882.5802 — Transcranial Magnetic Stimulation System For Obsessive-Compulsive Disorder

Neurology (NE) · Part 882 Subpart F—Neurological Therapeutic Devices · § 882.5802

Identification

A transcranial magnetic stimulation system for neurological and psychiatric disorders and conditions is a prescription, non-implantable device that uses brief duration, rapidly alternating, or pulsed, magnetic fields to induce neural activity in the cerebral cortex. It is not intended for applying or focusing magnetic fields towards brain areas outside cerebral cortex (e.g., cerebellum). A repetitive transcranial magnetic stimulation system that is intended to treat major depressive disorder is classified in § 882.5805. A transcranial magnetic stimulation system for headache is classified in § 882.5808.

Classification Rationale

Class II (special controls). The special controls for this device are:

Product Codes

Product CodeDevice NameClassDevicesAttributes
QCITranscranial Magnetic Stimulation System For Obsessive-Compulsive Disorder29SaMD
QMDTranscranial Magnetic Stimulation System For Smoking Cessation22

Special Controls

QCI — Transcranial Magnetic Stimulation System For Obsessive-Compulsive Disorder

In combination with the general controls of the FD&C Act, the transcranial magnetic stimulation system for neurological and psychiatric disorders and conditions is subject to the following special controls:

De Novo Order DEN170078

QCI — Transcranial Magnetic Stimulation System For Obsessive-Compulsive Disorder

*Classification.* Class II (special controls). The special controls for this device are:(1) Performance testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety. (2) Software verification, validation, and hazard analysis must be performed. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Magnetic pulse output testing; (ii) Magnetic and electrical field testing; (iii) Testing of the safety features built into the device; and (iv) Testing of the sound levels patients are exposed to during device use. (5) The physician and patient labeling must include the following: (i) The risks and benefits associated with use of the device; (ii) Detailed instructions to prevent seizures, to monitor the patient for seizure activity during treatment, and to provide seizure management care if one were to occur during treatment; and (iii) A description of the ear protection to be worn by the patient during use of the device, including the type of protection and its noise reduction rating.

eCFR

QCI — Transcranial Magnetic Stimulation System For Obsessive-Compulsive Disorder

(1) Performance testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety. (2) Software verification, validation, and hazard analysis must be performed. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Magnetic pulse output testing; (ii) Magnetic and electrical field testing; (iii) Testing of the safety features built into the device; and (iv) Testing of the sound levels patients are exposed to during device use. (5) The physician and patient labeling must include the following: (i) The risks and benefits associated with use of the device; (ii) Detailed instructions to prevent seizures, to monitor the patient for seizure activity during treatment, and to provide seizure management care if one were to occur during treatment; and (iii) A description of the ear protection to be worn by the patient during use of the device, including the type of protection and its noise reduction rating.

Ecfr Llm

QMD — Transcranial Magnetic Stimulation System For Smoking Cessation

*Classification.* Class II (special controls). The special controls for this device are:(1) Performance testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety. (2) Software verification, validation, and hazard analysis must be performed. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Magnetic pulse output testing; (ii) Magnetic and electrical field testing; (iii) Testing of the safety features built into the device; and (iv) Testing of the sound levels patients are exposed to during device use. (5) The physician and patient labeling must include the following: (i) The risks and benefits associated with use of the device; (ii) Detailed instructions to prevent seizures, to monitor the patient for seizure activity during treatment, and to provide seizure management care if one were to occur during treatment; and (iii) A description of the ear protection to be worn by the patient during use of the device, including the type of protection and its noise reduction rating.

eCFR

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...