MagVenture TMS Therapy – for treatment of OCD, MagVenture TMS Therapy system

K193006 · Tonica Elektronik A/S · QCI · Aug 9, 2020 · Neurology

Device Facts

Record IDK193006
Device NameMagVenture TMS Therapy – for treatment of OCD, MagVenture TMS Therapy system
ApplicantTonica Elektronik A/S
Product CodeQCI · Neurology
Decision DateAug 9, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5802
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MagVenture TMS Therapy system is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

Device Story

MagVenture TMS Therapy system is a computerized, electromechanical device delivering non-invasive magnetic fields to induce electrical currents in the cerebral cortex. It uses a capacitor-driven circuit to discharge pulses through a stimulating coil, generating magnetic fields that excite neurons via Faraday's law. The system includes a MagPro stimulator, Cool D-B80 double-cone coil, trolley, patient chair, and positioning apparatus. Operated by clinicians in a clinical setting, the system targets the bilateral dorsomedial prefrontal cortex (DMPFC) using 20 Hz repetitive pulses for 18 minutes. Treatment intensity is based on the patient's Leg Motor Threshold (MT). The device provides non-invasive neuromodulation to modulate brain activity in long-range networks associated with OCD. Clinicians use the system to deliver standardized protocols, with output parameters monitored via the system's GUI. The device aims to provide therapeutic benefit for OCD patients by modulating cortical networks, offering a non-pharmacological adjunctive treatment option.

Clinical Evidence

No new pivotal clinical trial data; substantial equivalence supported by literature review of >500 subjects treated with the MagVenture system targeting the DMPFC. Studies demonstrate safety, tolerability, and effective modulation of cortical networks. Side effect profile (headache, site pain) is equivalent to the primary predicate. A small pilot trial (n=not specified) using 10 Hz stimulation showed statistically significant effects for OCD. Bench testing confirmed magnetic field characteristics and compliance with IEC 60601-1 and 60601-1-2.

Technological Characteristics

System consists of a MagPro stimulator, Cool D-B80 liquid-cooled double-cone coil, trolley, and patient fixation chair. Operates via electromagnetic induction (Faraday's law) to deliver biphasic magnetic pulses. Connectivity includes a data management system. Complies with ISO 13485:2016, IEC 60601-1, and IEC 60601-1-2. Stimulation intensity range 0-1.9 SMT. Software-controlled GUI for parameter setting and monitoring.

Indications for Use

Indicated for adult patients suffering from Obsessive-Compulsive Disorder (OCD) as an adjunct treatment.

Regulatory Classification

Identification

A transcranial magnetic stimulation system for neurological and psychiatric disorders and conditions is a prescription, non-implantable device that uses brief duration, rapidly alternating, or pulsed, magnetic fields to induce neural activity in the cerebral cortex. It is not intended for applying or focusing magnetic fields towards brain areas outside cerebral cortex (e.g., cerebellum). A repetitive transcranial magnetic stimulation system that is intended to treat major depressive disorder is classified in § 882.5805. A transcranial magnetic stimulation system for headache is classified in § 882.5808.

Special Controls

In combination with the general controls of the FD&C Act, the transcranial magnetic stimulation system for neurological and psychiatric disorders and conditions is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Performance testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety. (2) Software verification, validation, and hazard analysis must be performed. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Magnetic pulse output testing; (ii) Magnetic and electrical field testing; (iii) Testing of the safety features built into the device; and (iv) Testing of the sound levels patients are exposed to during device use. (5) The physician and patient labeling must include the following: (i) The risks and benefits associated with use of the device; (ii) Detailed instructions to prevent seizures, to monitor the patient for seizure activity during treatment, and to provide seizure management care if one were to occur during treatment; and (iii) A description of the ear protection to be worn by the patient during use of the device, including the type of protection and its noise reduction rating.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 9, 2020 Tonica Elektronik A/S Sanne Jessen Medical Advisor, MSc, PhD Lucernemarken 15 DK-3520 Farum, Denmark Re: K193006 Trade/Device Name: MagVenture TMS Therapy - for adjunctive treatment of OCD, MagVenture TMS Therapy system Regulation Number: 21 CFR 882.5802 Regulation Name: Transcranial Magnetic Stimulation System For Neurological And Psychiatric Disorders And Conditions Regulatory Class: Class II Product Code: QCI Dated: April 29, 2020 Received: May 6, 2020 Dear Sanne Jessen: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrb/ofdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K193006 Device Name MagVenture TMS Therapy System Indications for Use (Describe) The MagVenture TMS Therapy System is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD). | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary ### Submitter's Information | Name of 510(k) owner: | Tonica Elektronik A/S<br>Lucernemarken 15<br>DK-3520 Farum, Denmark | |-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Phone/ Fax: | +45 4499 8444 / +45 4499 1544 | | Contact person: | Sanne Barsballe Jessen<br>Medical Advisor<br>Email: sj@magventure.com<br>Office: +453840 8449<br>Cell: +45 3119 9516<br>Other:<br>Jan Kjoeller<br>Email: jk@tonica.dk | | Preparation date: | August 8, 2020 | | Trade names: | MagVenture TMS Therapy system<br>MagVenture TMS Therapy system – for adjunctive treatment of OCD | | Common name: | Transcranial Magnetic Stimulator | | Classification name: | Transcranial Magnetic Stimulation System for Neurological<br>and Psychiatric Disorders and Conditions [21 CFR 882.5802<br>[Product Code QCI - Transcranial Magnetic Stimulation<br>System for Obsessive-Compulsive Disorder] | | Classification: | Class II Medical Device | | Predicate Devices: | <b>Primary Predicate</b><br>Brainsway Deep Transcranial Magnetic Stimulation (DTMS)<br>system, HAC – H7 coil (DEN170078, K183303)<br>21 CFR 882.5802, Transcranial Magnetic Stimulation System<br>for Neurological and Psychiatric Disorders and Conditions<br>Product Code: QCI - Transcranial Magnetic Stimulation | {4}------------------------------------------------ Class II #### Predicates MagVenture TMS Therapy system (K150641, K171481, K171967, K172667, K173620) 21 CFR 882.5805, Repetitive Transcranial Magnetic Stimulation Product code: OBP Device Class: II #### Device description The MagVenture TMS Therapy system - for adjunctive treatment of OCD is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic fields to induce electrical currents targeting specific regions of the cerebral cortex. The MagVenture TMS Therapy system - for treatment of OCD is indicated as an adjunct for the treatment of adult patients who are suffering from Obsessive-Compulsive Disorder (OCD). MagVenture TMS Therapy system has previously obtained FDA clearance for treatment of major depressive disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode (K150641, K171481, K171967, K172667, K173620). Transcranial magnetic stimulation (TMS) is a non-invasive technique for stimulating brain and neural tissue. The principle of magnetic stimulation is implicit in Faraday's law. The pulses of current are generated with a circuit containing a capacitor connected to the stimulating coil. With the capacitor charged to a certain level, the conducting state will cause the discharging of the capacitor through the coil. A magnetic field is generated proportional to this current. The rapid change in the magnetic field induces a current in conducting materials e.g. the body tissue. If the current induced in the human body is of sufficient amplitude and duration, it will excite neurons. The standard-of-care FDA cleared TMS protocol for treatment of OCD uses repetitive transcranial magnetic pulses applied at a frequency of 20 Hz. The safety and effectiveness for treatment of OCD have been established in a clinical trial that lead to an FDA De Novo clearance of the primary predicate device, the Brainsway DTMS system. The present 510(k) does not include new pivotal data, but includes clinical trial data on more than 500 subjects treated with the Mag Venture TMS Therapy System, in order to demonstrate performance and safety. Treatment of OCD is applied to the human brain of the bilateral dorsomedial prefrontal cortex (DMPFC) using 20 Hz TMS for 18 min. The treatment parameters are identical to those recommended by the Primary Predicate Device. {5}------------------------------------------------ The MagVenture TMS Therapy system – for adjunctive treatment of OCD is an integrated system consisting of the following components: - MagPro Stimulator and Trolley . - MagPro Family (R30, R30 w. MagOption, X100, X100 w. MagOption) o - Trolley with holding arrangements o - Coil for MT determination and OCD treatment . - Coil Cool D-B80 with Coil Cooler Unit o - Marking apparatus for locating treatment area . - Pen for marking, Cap, Ruler o - Patient head fixation . - Treatment Chair o - Vacuum Pump and Vacuum pillow o - Super Flexible Arm mounted on the trolley o - Isolation Transformer . #### Intended Use The MagVenture TMS Therapy system is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD). The intended use is identical to that of the Primary Predicate Device. #### Performance Standards: The MagVenture TMS Therapy system - for adjunctive treatment of OCD has been tested and conforms with the following standards - . ISO 13485:2016 - IEC60601-1 • - IEC60601-1-2 . #### Non-Clinical performance data: The contents of this 510(k) complies with the FDA Guidance Document: "Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems - Guidance for Industry and Food and Drug Administration Staff". The non-clinical performance testing of the components of the MagVenture TMS Therapy system - for treatment of OCD has been tested as required according to the standards listed above. All components, except the Cool D-B80 coil, have previously been cleared by the FDA, see K150641, K171481, K172667 and K173620. The MagVenture TMS Therapy system – for adjunctive treatment of OCD consists of components that are identical to those of the predicate device, with the exception of the Cool D-B80 coil. The new coil is identical to the predicate device in terms of biocompatibility, design elements, such as cable lengths, coil materials, cooling media, {6}------------------------------------------------ isolation design and functionalities. All coils are subject to high-voltage tests and leakage current tests to ensure safety. To establish substantial equivalency for the new coil, Cool D-B80, and the primary predicate device, especially the HAC – H7 coil, we have performed substantial equivalence comparisons for testing and performance as described in Section 4 of the FDA's Class II Special Controls Guidance document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems. We have performed a comparative testing of the magnetic field distribution for the Cool D-B80 coil and compared to that of the primary predicate device. This comparison determines that the magnetic spatial distribution is substantially equivalent. Both coils are so-called double cone coils, containing two coils that do not overlap, and allows for a deeper and broader stimulation of the cortex. We have provided information about magnetic field characteristics, including linearity of output level, magnetic field strength gradients, output waveform and magnetic field spatial distribution according to the Special Controls Guidance document, Section 4 as mentioned above. In addition, we have provided magnetic field characteristics for other, different relevant clinical depths of the human brain than those specified in the Special Controls Guidance and in the De Novo Summary of the Primary Predicate Device, which limits information to 2 cm depth. In addition, we have modelled the magnetic properties of the New Device and the Primary Predicate Device, the HAC-7 coil. Our coil model calculations are based on the concept of well-established scientific methods (1). The results of the e-field modelling of the HAC-7 coil is in line with the Manufacturer information provided in the De Novo Summary for the primary predicate device, DEN170078. The results of the e-field model show the magnetic field strength in the cortex for the two coils. This shows that the Cool D-B80 coil is able to reach the deeper underlying cortical layers similarly to the HAC - H7 coil. Based on the modelling, it is therefore concluded, that the magnetic field properties of the Cool D-B80 coil is substantially equivalent to the primary predicate device, HAC - H7 in terms of magnetic field properties and realized magnetic field properties. The modelling was used to support the technological comparison provided. We have also provided information about the magnetic field spatial distribution of the new coil superimposed on T1-weighted MRI coronal, sagittal, and axial 1 cm slices. These images support the substantial equivalence comparisons determination and together with the above information, further supports the substantial equivalence of the new device compared to the primary predicate device, HAC-7. We have also tested the new device according to IEC60601 3rd edition and verified that the device complies with the specified permissible sound pressure levels. The device also {7}------------------------------------------------ complies with the permissible thresholds for exposure defined by the Occupational Safety and Health Administration (OSHA). These tests provide evidence that the Mag Venture TMS Therapy system does not pose any risk for potential hearing reduction or loss in either patients or operators. #### Clinical performance data: This 510(k) does not contain any pivotal clinical trial data related to the new device. The substantial equivalency was established based on similar technological characteristics, but we have provided some clinical trial data to support the safety of the new device. The clinical evidence submitted also supports the safety and use of the new device as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD). A number of clinically relevant scientific references including more than 500 subjects treated with MagVenture TMS Therapy using the Cool D-B80 coil have been included with full text, as well as a summary and rationale for how these references support SE determination and/or safety and effectiveness for the new device for the proposed indication for use and for the proposed treatment protocol. The primary predicate device De Novo Summary refers to a publication by Carmi et al. (2), who published the results of their pivotal trial supporting the De Novo, DEN170078. This study included a total of n=99, of which n= 50 received active TMS treatment. The proposed indication for use and treatment protocol are identical to that of the primary predicate device. The new device is already cleared for use in both Europe and Canada and has been used in clinical research for a number of years. Most of the clinical research performed using the new device has been conducted outside the USA. Most research has focused on the use of the new device for treatment of treatment-resistant depression (TRD) or major depressive disorder (MDD). There has been considerable interest in investigating the DMPFC as a target for TRD/MDD. Though MDD/TRD and OCD are distinct psychiatric disorders they share common underlying deficits in cortical networks. Thus, it is therefore relevant to demonstrate that the new device can be used to deliver TMS at the right treatment location, DMPFC, and that this can lead to modulation of deeper brain structures and long-range networks that are important in the recovery of symptoms of OCD as well as MDD. In addition, the clinical trials also demonstrate that the new device can be used efficiently and routinely to determine MT in the leg or foot, in a substantially equivalent way to the primary predicate device. Importantly, the clinical data also helps demonstrate safety for the new device. {8}------------------------------------------------ In summary, the literature referenced includes a total of 521 subjects treated with TMS or iTBS protocols effectively targeting the DMPFC using the new device with treatment intensity defined based on the use of Leg MT. It is noteworthy that most of the clinical investigations have utilized treatment intensities that are higher than those used by the primary predicate device for OCD as reported by Carmi et al (2). Thus, treatment with the new device at an intensity of 100% Leg MT might be even more tolerable and safer in comparison with higher intensity. Despite this difference in intensity, the clinical data for the new device shows that the treatment overall is well-tolerated and safe. The treatment with the new device does not introduce any new adverse or serious adverse events. In fact, the side effect profile resembles that of standard TMS and iTBS treatment of the L-DLPFC and it is also equivalent to the side effect profile of the primary predicate device. We therefore conclude that the treatment with the new device is safe and provides equivalent performance to the primary predicate device in terms of safety and performance. The most common side effects reported for both devices are headache and/or pain at stimulation site. In addition, a small pilot trial using the new device also demonstrated a clinically and statistically significant effect of treatment of OCD. Despite the low number of subjects, this trial supports the performance of the new device as an adjunct for treatment of OCD, even though the protocol was utilizing lower frequency of stimulation, that is 10 Hz compared to 20 Hz. Taken together, the literature demonstrates that the new device can be used effectively and routinely for determining Leg MT, and that the treatment modulates brain activity locally in the DMPFC and strongly suggests also distal brain areas, through downstream activation of long-range networks. Though MDD and OCD are two distinct psychiatric disorders, they share some common underlying deficits in brain networks. So, though most of the research using the new device has investigated treatment of MDD, the data can support the use of the new device as an adjunct for treatment of OCD in terms of safety and given that the treatments have been performed targeting the same cortical area. DMPFC bilaterally, in a substantially equivalent way to the primary predicate device. Finally, the clinical data for the new device essentially supports the modelled e-fields presented above, and treatment with the new device is as safe as treatment with the primary predicate device. There is no reason to suspect increased risk for unwanted or severe side effects. Moreover, clinical data shows the new device's ability to deliver effective treatment of the DMPFC and potentially affecting deeper cortical brain areas. We therefore conclude that the new device is substantially equivalent to the primary predicate device in terms of performance, effectiveness and safety. {9}------------------------------------------------ #### Substantial equivalence: The MagVenture TMS Therapy system - for adjunctive treatment of OCD is substantially equivalent to the primary predicate device, the Brainsway DTMS system HAC - H7 coil. The MagVenture TMS Therapy system - for treatment of OCD and the primary predicate device have identical indication for use, and identical treatment parameters as well as treatment target. The magnetic field properties of the Cool D-B80 is substantially equivalent to the primary predicate device, the Brainsway DTMS HAC - H7 coil (DEN170078, K183303). All components of the new device, except the Cool D-B80 coil, are identical to those of the predicate device. These have all previously obtained FDA clearance for treatment of major depressive disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode (K150641, K171967 and K173620). The new coil, Cool D-B80 is also identical to that of the predicate devices, in terms of materials, design elements, liquid cooling and biocompatibility. The treatment protocol is identical to that of the primary predicate device, and applies transcranial magnetic stimulation (TMS) at an intensity of 100% of Leg Motor Threshold (MT) as repetitive pulse trains at a frequency of 20 Hz delivered as brief rapidly alternating magnetic fields to induce electrical currents over the dorsomedial prefrontal cortex (DMPFC). All labelling claims related to effectiveness and safety are based on the literature describing the pivotal clinical trial results of the primary predicate device for the proposed intended use. Both the MagVenture TMS Therapy system - for adjunctive treatment of OCD and the predicate devices consist of the same components, that is a TMS stimulator with software, an articulated arm for positioning of the treatment coil. The operational procedures including system setup, patient preparations, motor threshold determination and coil positioning are substantially equivalent. {10}------------------------------------------------ | Area | New Device | Predicate Device | Primary Predicate<br>Device | |-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | MagVenture TMS Therapy<br>system - for treatment of<br>OCD<br>Tonica Elektronik A/S,<br>Denmark | MagVenture TMS Therapy<br>system<br>(K150641, K171481, K171967,<br>K172667, K173620)<br>Tonica Elektronik A/S,<br>Denmark | Brainsway Deep Transcranial<br>magnetic stimulation (DTMS)<br>System and HAC – H7 coil<br>(DEN170078, K183303)<br>Brainsway Ltd., Israel | | Indications<br>for use | The MagVenture TMS Therapy<br>System is intended to be used as<br>an adjunct for the treatment of<br>adult patients suffering from<br>Obsessive-Compulsive Disorder<br>(OCD) | NA | The Brainsway DTMS system is<br>intended to be used as an<br>adjunct for the treatment of<br>adult patients suffering from<br>Obsessive-Compulsive Disorder<br>(OCD) | | Energy<br>Delivered<br>and<br>Performance | Treatment parameters:<br>Intensity:<br>100%<br>of Leg MT (Leg Motor<br>Threshold) | NA | Treatment parameters:<br>Intensity:<br>100%<br>of Leg MT (Leg Motor<br>Threshold) | | | Repetition rate:<br>20Hz<br>Train duration:<br>2 sec<br>Inter-train-Interval:<br>20 secs<br>Number of trains:<br>50<br>Numbers of pulses:<br>2000 | | Repetition rate:<br>20Hz<br>Train duration:<br>2 sec<br>Inter-train-Interval:<br>20 secs<br>Number of trains:<br>50<br>Numbers of pulses:<br>2000 | | | Total duration: 18.0 min.<br>Treatment area:<br>Area of brain to be stimulated:<br>Dorsomedial Prefrontal Cortex | NA | Total duration: 18.3 min.<br>Treatment area:<br>Area of brain to be stimulated:<br>Dorsomedial Prefrontal Cortex | | | Output Stimulation Parameters:<br>Available Stimulation Intensity<br>in terms of<br>Standard Motor Threshold<br>(SMT) units<br>Range: 0 - 1.9 SMT<br>Waveform: Biphasic | Output Stimulation Parameters:<br>Available Stimulation Intensity<br>in terms of<br>Standard Motor Threshold<br>(SMT) units<br>Range: 0 - 1.7 SMT<br>Waveform: Biphasic | Output Stimulation Parameters:<br>Available Stimulation Intensity<br>in terms of<br>Standard Motor Threshold<br>(SMT) units<br>Range: 0.6- 1.4 SMT<br>Waveform: Biphasic | | Design | The system consists of:<br>1. Mobile console<br>2. System software with GUI<br>3. Treatment chair*<br>4. Head support system*<br>5. Coil positioning system<br>6. Same Coil for both MT and<br>treatment<br>7. Coil Fixture<br>8. Data Management System<br>*optional | The system consists of:<br>1. Mobile console<br>2. System software with GUI<br>3. Treatment chair*<br>4. Head support system*<br>5. Coil positioning system<br>6. Coil for MT and coil for<br>treatment<br>7. Coil Fixture<br>8. Data Management System<br>*optional | The system consists of:<br>1. Mobile console<br>2. System software<br>3. Treatment chair*<br>4. Helmet with Coil for MT<br>and for treatment<br>5. Coil positioning system<br>6. | | Coil | Double-cone coil<br>Air core | Figure-of-eight coils (butterfly<br>coil)<br>Air core | Double-cone coil contained in a<br>helmet.<br>Air core | | Cooling | Liquid cooled | Liquid cooled | Air cooled. | | | Used for both MT determination | Used for both MT determination | Used for both MT determination | | | and treatment. | and treatment. | and treatment. | | Standards | Company complies with EN | Company complies with EN | Company complies with ISO | | | ISO 13485:2016. | ISO 13485:2016. | 13485:2016. | | Electrical<br>safety | Complies with<br>IEC60601-1 v. 3.1, and<br>IEC60601-1-2. | Complies with<br>IEC60601-1 v. 3.1, and<br>IEC60601-1-2. | | {11}------------------------------------------------ For a more comprehensive comparison of devices please refer to section 10, Device Description and section 12, Substantial Equivalence Comparison. ### Conclusion: The new device, MagVenture TMS Therapy system - for adjunctive treatment of OCD, is identical to the predicate device, MagVenture TMS Therapy system, except for the treatment coil, Cool D-B80, which has not previously been cleared by the FDA. All other components of the new device have previously been cleared by the FDA, most recently in 2018 (K173620). The indication for use, the target population, the TMS treatment protocol and the treatment position are all identical for the MagVenture TMS Therapy system and the primary predicate device, Brainsway DTMS System (DEN170078). The treatment parameters proposed for the new device are identical to those recommended by the primary predicate device. Clinical evidence pertaining to the new device demonstrates that the treatment is safe, well-tolerated and effective. The new device is substantially equivalent to the primary predicate device in terms of performance, safety and effectiveness. The new coil is identical to the predicate coils, Cool-B70 and Cool-B65, in terms of design parameters, such as materials, biocompatibility, liquid cooling and pulse width. The new coil is substantially equivalent to the primary predicate device, Brainsway DTMS HAC - H7 coil in terms of magnetic properties and magnetic spatial distribution. The two coils are both so-called double cone coils, which contain two individual, nonoverlapping magnetic coils that allow for a broader and more deep stimulation of the cortex. E-field modelling of the new device compared to the primary predicate device demonstrates that the magnetic field properties and the depth penetration of the two devices in the human cortex are substantially equivalent. The clinical data submitted with this 510(k), helps support the performance and safety of the new device and shows that also the side effect profile is substantially equivalent to that of the primary predicate. {12}------------------------------------------------ The reliability of the positioning method used by the new device is based on the direct relationship of the underlying cortical brain anatomy to the patient's scalp, identical to the method used for the predicate devices. The MagVenture TMS Therapy system - for adjunctive treatment of OCD does not introduce any new safety considerations in comparison to the predicate devices. All other identified differences between the two systems are minor and without any impact on safety or efficacy. The above comparison demonstrates and supports the substantial equivalence of the MagVenture TMS Therapy system - for adjunctive treatment of OCD to the predicate devices, Brainsway DTMS System (DEN170078; primary predicate), and the MagVenture TMS Therapy system (K150641, K171967, K172667, K172667, K173620). #### Reference List 1. Deng ZD, Lisanby SH, Peterchev AV. Electric field depth-focality tradeoff in transcranial magnetic stimulation: simulation comparison of 50 coil designs. Brain Stimul. 2013;6(1):1-13. Carmi L, Tendler A, Bystritsky A, Hollander E, Blumberger DM, 2. Daskalakis J. et al. Efficacy and Safety of Deep Transcranial Magnetic Stimulation for Obsessive-Compulsive Disorder: A Prospective Multicenter Randomized Double-Blind Placebo-Controlled Trial. Am J Psychiatry. 2019:appiajp201918101180. Dunlop K. Woodside B. Olmsted M. Colton P. Giacobbe P. Downar J. 3. Reductions in Cortico-Striatal Hyperconnectivity Accompany Successful Treatment of Obsessive-Compulsive Disorder with Dorsomedial Prefrontal rTMS. Neuropsychopharmacology. 2016;41(5):1395-403. Miron JP, Feffer K, Cash RFH, Derakhshan D, Kim JMS, Fettes P, et al. 4. Safety, tolerability and effectiveness of a novel 20 Hz rTMS protocol targeting dorsomedial prefrontal cortex in major depression: An open-label case series. Brain Stimul. 2019. Schulze L, Feffer K, Lozano C, Giacobbe P, Daskalakis ZJ, Blumberger న్. DM, et al. Number of pulses or number of sessions? An open-label study of trajectories of improvement for once-vs. twice-daily dorsomedial prefrontal rTMS in major depression. Brain Stimul. 2018;11(2):327-36. Dunlop K, Sheen J, Schulze L, Fettes P, Mansouri F, Feffer K, et al. 6. Dorsomedial prefrontal cortex repetitive transcranial magnetic stimulation for treatment- {13}------------------------------------------------ refractory major depressive disorder: A three-arm, blinded, randomized controlled trial. Brain Stimul. 2019. Salomons TV, Dunlop K, Kennedy SH, Flint A, Geraci J, Giacobbe P, et al. 7. Resting-state cortico-thalamic-striatal connectivity predicts response to dorsomedial prefrontal rTMS in major depressive disorder. Neuropsychopharmacology. 2014;39(2):488-98. Kreuzer PM, Schecklmann M, Lehner A, Wetter TC, Poeppl TB, Rupprecht 8. R, et al. The ACDC pilot trial: targeting the anterior cingulate by double cone coil rTMS for the treatment of depression. Brain Stimul. 2015;8(2):240-6. 0. Schulze L. Wheeler S. McAndrews MP. Solomon CJ. Giacobbe P. Downar J. Cognitive safety of dorsomedial prefrontal repetitive transcranial magnetic stimulation in major depression. Eur Neuropsychopharmacol. 2016. 10. Bakker N, Shahab S, Giacobbe P, Blumberger DM, Daskalakis ZJ, Kennedy SH, et al. rTMS of the dorsomedial prefrontal cortex for major depression: safety, tolerability, effectiveness, and outcome predictors for 10 Hz versus intermittent theta-burst stimulation. Brain Stimul. 2015;8(2):208-15.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...