K201158 · Neuronetics, Inc. · OBP · Nov 20, 2020 · Neurology
Device Facts
Record ID
K201158
Device Name
NeuroStar Advanced Therapy
Applicant
Neuronetics, Inc.
Product Code
OBP · Neurology
Decision Date
Nov 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NeuroStar Advanced Therapy is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Device Story
NeuroStar Advanced Therapy System is a computerized, electromechanical device delivering non-invasive, repetitive transcranial magnetic stimulation (rTMS) to the cerebral cortex. It uses brief, rapidly alternating magnetic fields to induce electrical currents in targeted cortical neurons. The system includes a mobile console, ferromagnetic treatment coil, head support system with laser-guided alignment, SenStar contact sensor, and TrakStar patient data management software. Used in clinics/hospitals by physicians to treat adult patients with Major Depressive Disorder (MDD). The system now includes 'NeuroBurst' (iTBS) protocol: 3-pulse bursts at 50Hz, 160ms interval between bursts, 5 bursts/second for 2 seconds, 8-second inter-train interval; 20 trains total (3.3 minutes). Physicians use the system to deliver anatomically focused stimulation to prefrontal cortex areas. The device provides a non-pharmacological treatment option for patients resistant to antidepressant medication.
Clinical Evidence
Bench testing only. No clinical data presented. Performance was verified through electromagnetic modeling of magnetic field characteristics (spatial distribution, strength gradients, output waveform) and system verification testing of NeuroBurst stimuli intensity. Risk management followed ISO 14971:2007 and IEC 62304:2015. Electromagnetic compatibility and electrical safety verified per IEC 60601-1-2:2014 and IEC 60601-1:2005.
Technological Characteristics
Electromechanical TMS system; mobile console with GUI, ferromagnetic coil, head support with laser alignment, and patient data management system. Uses electromagnetic induction to deliver pulsed magnetic fields. Updated power module with new micro-controller and associated firmware. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
MagVita TMS Therapy System w/ Theta Burst Stimulation (K173620)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2020
Neuronetics, Inc. Gary Johnson Director, Regulatory Affairs 3222 Phoenixville Pike Malvern, PA 19355
Re: K201158
Trade/Device Name: NeuroStar Advanced Therapy Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive transcranial magnetic stimulation system Regulatory Class: Class II Product Code: OBP Dated: October 22, 2020 Received: October 23, 2020
Dear Gary Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) N/A K201158
Device Name NeuroStar Advanced Therapy
NeuroStar Advanced Therapy is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Confidential
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for NeuroStar Advanced Therapy. The logo consists of a stylized "N" symbol in two shades of purple, followed by the word "NeuroStar" in a bold, purple font. Below the word "NeuroStar" is the phrase "Advanced Therapy" in a smaller, lighter font.
# 510(k) SUMMARY
## NeuroStar Advanced Therapy System
| 510(k) Owner: | Neuronetics, Inc. |
|------------------------|-------------------|
| 3222 Phoenixville Pike | |
| Malvern, PA 19355 | |
| Phone: 610.640.4202 | |
| Fax: 610.640.4206 | |
| Company Contact: | Gary Johnson |
|------------------|-------------------------------|
| | Director, Regulatory Affairs |
| | Phone: 610.981.4138 |
| | Fax: 610.640.4206 |
| | Email: gjohnson@neurostar.com |
| Alternative Contact: | Yelena Tropsha |
|----------------------|-----------------------------------|
| | Vice President, Commercial Access |
| Phone: | 610.981.4104 |
| Fax: | 610.640.4206 |
| Email: | ytropsha@neurostar.com |
| <b>Date Prepared:</b> | April 28, 2020 |
|--------------------------|-----------------------------------|
| <b>Proprietary Name:</b> | NeuroStar Advanced Therapy System |
| <b>Common Name:</b> | Transcranial Magnetic Stimulator |
{4}------------------------------------------------
| Classification Name: | Repetitive Transcranial Magnetic Stimulation Device |
|-----------------------------------|-----------------------------------------------------|
| 21 CFR 882.5805, Product Code OBP | |
| Predicate Device: | NeuroStar Advanced Therapy System<br>(K161519) |
|-------------------|--------------------------------------------------------------------|
| Reference Device: | MagVita TMS Therapy System w/ Theta Burst Stimulation<br>(K173620) |
### Device Description
The NeuroStar Advanced Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation (TMS).
NeuroStar Advanced Therapy is a non-invasive tool for the stimulation of cortical neurons for the treatment of adult patients with Major Depressive Disorder (MDD) who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. NeuroStar Advanced Therapy is used for patient treatment by prescription only under the supervision of a licensed physician and can be used in both inpatient and outpatient settings including physician's offices, clinics, and hospitals.
NeuroStar Advanced Therapy uses a well-understood property of physics regarding the law of electromagnetic induction, which states that a time-varving or moving magnetic field will induce an electric current in an adjacent conductive substance with the electric current traveling in a direction perpendicular to the magnetic field. NeuroStar Advanced Therapy uses a generated magnetic field as a vector for delivering an electrical current to a target conductor of interest, which in therapeutic application, is the brain. By
{5}------------------------------------------------
using this method of delivering electrical current to the brain the electrical current produced in the brain can be delivered in an anatomically focused manner at discrete cortical areas of interest and can avoid areas of the brain that are not relevant for its therapeutic actions.
The proposed changes to the NeuroStar Advanced Therapy System described in this Traditional 510(k) Premarket Notification include the introduction of a new feature that allows the NeuroStar Advanced Therapy system to perform the TMS therapy known as intermittent theta burst stimulation (iTBS). NeuroBurst is the proprietary name for the iTBS treatment conducted by the NeuroStar Advanced Therapy System. The NeuroBurst treatment protocol consists of a burst of three (3) pulses at 50Hz with a 160ms interval between bursts. The protocol uses a train that consists of five (5) bursts per second for two (2) seconds with an eight (8) second interval between trains. A treatment session lasts for 20 trains or 3.3 minutes.
The NeuroStar Advanced Therapy System is an integrated system consisting of a combination of the following components:
- Mobile Console for housing the electronics and includes a software controlled . graphical user interface, display monitor, display arm, and gantry that supports the treatment coil.
- Ferromagnetic Coil for delivering treatment. .
- Head Support System for positioning the treatment coil and includes a laser-guided . alignment system.
- Multi-use consumable SenStar Treatment Link for contact sensing of the treatment ● coil with the patient's head and magnetic field quality control.
- TrakStar Patient Data Management System for recording patient data and includes ● a stand-alone computer and data management software.
## Intended Use / Indications for Use
NeuroStar Advanced Therapy is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
{6}------------------------------------------------
## Technological Characteristics
The key technological characteristics of the NeuroStar Advanced Therapy System are substantially equivalent to NeuroStar TMS Therapy System as described in submission K161519 (predicate device) regarding:
- Indications for Use .
- Technological characteristics, including design for delivery of TMS .
- Principles of operation and performance specifications .
- . Materials and standards
The proposed changes to the NeuroStar Advanced Therapy System include a modification to the power module subsystem, which was updated with a new micro-controller due to obsolescence. The updated micro-controller is provided by the same manufacturer and is a drop-in replacement for the current micro-controller. Firmware was also updated to support the micro-controller hardware component change. All other materials and the process for delivery of TMS are the same as the predicate device.
Both the NeuroStar Advanced Therapy System and the predicate device consist of the same components, which include: the mobile console with software, treatment coil, head support system, multi-use consumable, and patient data management system. Both devices utilize the same operational procedures for system setup, patient preparation, motor threshold determination, treatment coil positioning, and patient treatment. Both devices use the same principles of operation to apply TMS treatment at a defined intensity as repetitive pulse trains delivered as brief rapidly alternating magnetic fields to induce electrical currents to the prefrontal cortex. Both devices use the same system performance specifications to ensure the device performs as intended for the same indications for use. The only difference in technological characteristics between the NeuroStar Advanced Therapy system and predicate device is the NeuroBurst treatment specific aspects and stimulation parameters, which differ from the standard treatment stimulation parameters of the predicate device only in terms of pulse train duration, intertrain interval, and pulses per treatment.
{7}------------------------------------------------
The standard treatment protocol for the NeuroBurst treatment is similar to the treatment protocol of the MagVita TMS Therapy System w/ Theta Burst Stimulation (reference device). The iTBS protocol of the reference device has been shown to produce equally effective clinical results as compared to the standard-of-care TMS therapy protocol for the treatment of adult patients with Major Depressive Disorder. Since the use of iTBS treatment was cleared by the FDA for the reference device under submission K173620 and given that the NeuroStar Advanced Therapy system and reference device share the same intended use and similar technological characteristics, the safety and effectiveness of the NeuroBurst treatment can be demonstrated by comparison of the key performance specifications between the subject device and the reference device. This comparison supports the conclusion that the iTBS treatment of the reference device and the NeuroBurst treatment of the NeuroStar Advanced Therapy System are the same and as such the NeuroBurst treatment is an appropriate treatment for Major Depressive Disorder in adult patients.
The methods for evaluating the differences in technological characteristics between the NeuroStar Advanced Therapy System and predicate device supports the conclusion that the safety and effectiveness of the NeuroStar Advanced Therapy System are substantially equivalent to that of the predicate device cleared under submission K161519 and that the changes to the NeuroStar Advanced Therapy System do not raise new questions of safety or effectiveness compared to the predicate device.
## Non-clinical Performance Data
The non-clinical performance data included in this Traditional 510(k) Premarket Notification address the performance data recommendations and relevant mitigation measures outlined in the Agency's special controls guidance entitled "Class II Special Controls Guidance Document for Repetitive Transcranial Magnetic Stimulation (rTMS) Systems" issued on July 26, 2011.
Testing was performed according to the Special Requirements Document to show Substantial Equivalence to the predicate device. NeuroStar testing showed substantially
{8}------------------------------------------------
equivalent results to the predicate device for magnetic field characteristics such as special distribution, magnetic field strength gradients, output waveform, and linear output level.
An electro-magnetic model of both system coils was created to compare the resultant magnetic fields produced by a 1 SMT pulse. The modules and test parameters were identical with the exception of the specific mechanical construction details between the two coils. Plots of the spatial extent of the magnetic field showed substantial equivalence between the NeuroStar and predicate treatment coils.
In addition, System verification testing of the NeuroStar Advanced Therapy System has been tested to verify system performance to specifications. This includes testing of the NeuroBurst stimuli intensity which was found to meet specification and is substantially equivalent to the predicate device.
Risk assessment was applied throughout the product development lifecycle process in accordance with the requirements set forth under the Agency recognized consensus standards ISO 14971:2007 and IEC 62304:2015. The results of the comprehensive risk analysis for the device indicate that there are no new hazards, harms, or safety risks introduced when compared to the predicate device. The NeuroStar Advanced Therapy System is in compliance with the requirements set forth under the Agency recognized consensus standards 60601-1-2:2014 Edition 4.0 and IEC 60601-1:2005 AMD1:2012 Edition 3.1 for electromagnetic compatibility (EMC) and electrical safety.
The device and software specifications, verification testing, validation testing, and risk management activities demonstrate that the NeuroStar Advanced Therapy System meets performance and functionality requirements and performs as intended without raising new questions of safety or effectiveness when compared to the predicate device. Performance testing was conducted to ensure that the NeuroStar Advanced Therapy System can perform NeuroBurst treatment and that system functionality had not been adversely impacted by the introduction of the new feature.
{9}------------------------------------------------
The clinical performance of iTBS treatment of the predicate device looked at stimuli of equal intensity and that the three stimuli within a burst are kept constant. Performance testing of the NeuroStar demonstrates that the intensity of the individual stimuli within a NeuroBurst treatment burst are of virtually equal intensity between bursts, which provides a consistent treatment dose that is clinically effective for NeuroBurst treatment when used to treat patients.
The non-clinical performance data supports the conclusion that the performance, functionality, safety, and effectiveness of the subject device are substantially equivalent to that of the predicate device cleared under submission K161519 and that the changes to the NeuroStar Advanced Therapy System do not raise new questions of safety or effectiveness compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.