Pulse Oximeter

K211143 · Shenzhen Brav Electronic Technologies Co., Ltd. · DQA · Jul 17, 2023 · Cardiovascular

Device Facts

Record IDK211143
Device NamePulse Oximeter
ApplicantShenzhen Brav Electronic Technologies Co., Ltd.
Product CodeDQA · Cardiovascular
Decision DateJul 17, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2

Intended Use

The pulse oximeter is a non-invasive device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in healthcare environments. And it is not intended to be used under motion or low perfusion scenarios. The device is reusable.

Device Story

Pulse oximeter (models OP-101, OP-102) performs non-invasive spot-checking of SpO2 and pulse rate; intended for adult patients in hospitals/clinics. Device uses finger-mounted probe containing red and infrared LEDs and a photodetector. Principle of operation: light absorption by blood/tissues varies with oxygen saturation; photodetector measures transmitted light; microprocessor calculates SpO2 and pulse rate. Output displayed on LED (OP-101) or OLED (OP-102) screen; includes SpO2%, pulse rate, battery indicator, and pulse bar graph/waveform. Operated by healthcare professionals. Benefits include convenient, non-invasive monitoring of patient oxygenation status to assist in clinical decision-making.

Clinical Evidence

Clinical study conducted on 12 adult volunteers (5 male, 7 female; diverse skin tones including Fitzpatrick I-VI) per ISO 80601-2-61:2017. Compared subject device SpO2 measurements against blood gas analyzer SaO2. Results confirmed accuracy requirements; no safety issues or adverse effects reported.

Technological Characteristics

Reusable fingertip pulse oximeter. Materials: ABS shell, silicon clip pad, PMMA/PC buttons. Sensing: Red/IR LED and photodetector. Power: 2x1.5V AAA batteries. Standards: ISO 10993-1 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 80601-2-61 (performance).

Indications for Use

Indicated for spot-checking functional arterial oxygen saturation (SpO2) and pulse rate in adult patients in healthcare environments. Not for use under motion or low perfusion conditions.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 17, 2023 Shenzhen brav electronic technologies co., Ltd Yolanda Lan Consultant R4-5/F, Block 11, Tongfuyu Industrial District, Lezhujiao, Huangmabu. Baoan Shenzhen, GuangDong 518060 China Re: K211143 Trade/Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: June 13, 2023 Received: June 13, 2023 Dear Yolanda Lan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, James J. Lee -S James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K211143 Device Name Pulse Oximeter(OP-101,OP-102) #### Indications for Use (Describe) The pulse oximeter is a non-invasive device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in healthcare environments. And it is not intended to be used under motion or low perfusion scenarios. The device is reusable. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions search existing data sources gather and maintain the data needed and complete time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ # 510(k) Summary # Prepared in accordance with the requirements of 21 CFR Part 807.92 | Submitter: | Shenzhen Brav electronic technologies co., Ltd | |--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 4-5/F, Block 11, Tongfuyu Industrial District, Lezhujiao, Huangmabu,<br>Baoan, Shenzhen,518060, China<br>+086-18682273850<br>1817914464@qq.com | | Contact Person: | Yolanda Lan<br>Yolanda.bleu@foxmail.com | | Prepare date: | 2021-09-04 | | Device name | Device Name: Pulse Oximeter | | and classification | Models: OP-101, OP -102 | | | Classification Name: 21 CFR 870.2700 Oximeter | | | Product code: DQA | | | Regulatory Class: Class II | | Reason for<br>Submission | New Application. No prior submission associated with the current<br>submission. | | Predicate<br>Device(s) | Shenzhen IMDK Medical Technology Co., Ltd., C101H1 Pulse Oximeter/<br>K173123 | | Device<br>Description | The oximeter consists of probe, electronic circuits, and display and plastic<br>enclosures. And one side of probe is designed to locate light emitting diodes<br>and a light detector (called a photo-detector). Red and Infrared lights are<br>shone through the tissues from one side of the probe to the other. Then parts<br>of the light emitted absorbed by blood and tissues. The light absorbed by the<br>blood varies with the oxygen saturation of haemoglobin. After that, the<br>photo-detector detects the light volume transmitted through the tissues which<br>depends on blood pulse, Hereafter, the microprocessor calculates a value for<br>the oxygen saturation (SpO2). | | | The subjected device is a reusable device, and need to reprocess as<br>suggested in the user manual after each use. And the device is intended to<br>be used on the finger, and powered by 2*1.5V AAA battery.OP-101 display<br>the measuring results on 1.5' LED screen, and the backlight of the three<br>models are red, white and green respectively. And the screen of OP-102<br>and OP-103 are 0.96' OLED and 1.3' OLED. Additionally, battery<br>indicator and pulse waveform can be displayed on OP-102 and OP-103.<br>The subjected device is indicated for continuous or spot check monitoring of<br>functional arterial oxygen saturation (SpO2) and pulse rate of adult patients<br>in hospitals and clinics. | | Indications for<br>Use | The pulse oximeter is a non-invasive device and intended for spot-checking<br>of oxygen saturation and pulse rate for use with the finger of adult patients in<br>healthcare environments. And it is not intended to be used under motion or<br>low perfusion scenarios. The device is reusable. | {4}------------------------------------------------ #### Predicate Device Comparison Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device. Please refer to following table to find differences between the subject device. All the differences do not affect the basic design principle, usage, effectiveness, and safety of the subject device. And no question is raised regarding to effectiveness and safety. | ITEM | Proposed Device Brav series Pulse Oximeter | Proposed Device Brav series Pulse Oximeter | Predicate Device K173123 | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Shenzhen Brav electronic technologies co., Ltd | Shenzhen Brav electronic technologies co., Ltd | Shenzhen IMDK Medical Technology Co., Ltd | | Device Name | Pulse Oximeter | Pulse Oximeter | Pulse Oximeter | | Models | OP-101 | OP-102 | C101H1 | | Appearance | Image: Pulse Oximeter OP-101 | Image: Pulse Oximeter OP-102 | Image: Pulse Oximeter C101H1 | | Indications for Use | The pulse oximeter is a non-invasive<br>device and intended for spot-checking<br>of oxygen saturation and pulse rate for<br>use with the finger of adult patients in<br>healthcare environments. And it is not<br>intended to be used under motion or<br>low perfusion scenarios. The device<br>reusable. | The pulse oximeter is a non-invasive<br>device and intended for spot-checking<br>of oxygen saturation and pulse rate for<br>use with the finger of adult patients in<br>healthcare environments. And it is not<br>intended to be used under motion or<br>low perfusion scenarios. The device<br>reusable. | Fingertip Pulse Oximeter C101H1 is a<br>non-invasive device intended for spot<br>checking of functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate (PR). This<br>portable device is indicated for use in<br>adult patients in hospitals. | | Principle | The device displays numerical values<br>for functional oxygen saturation of<br>arterial hemoglobin (SpO2) and pulse<br>rate by measuring the absorption of red<br>and infrared (IR) light passing through<br>perfused tissue. Changes in the<br>absorption caused by the pulsation of<br>blood in the vascular bed are used to<br>determine oxygen saturation and pulse<br>rate. | The device displays numerical values for<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate by<br>measuring the absorption of red and<br>infrared (IR) light passing through<br>perfused tissue. Changes in the<br>absorption caused by the pulsation of<br>blood in the vascular bed are used to<br>determine oxygen saturation and pulse<br>rate. | The device displays numerical values<br>for functional oxygen saturation of<br>arterial hemoglobin (SpO2) and pulse<br>rate by measuring the absorption of<br>red and infrared (IR) light passing<br>through perfused tissue. Changes in<br>the absorption caused by the pulsation<br>of blood in the vascular bed are used<br>to determine oxygen saturation and<br>pulse rate. | | Intended<br>patient<br>population | Adults in healthcare environments | Adults in healthcare environments | Adults in a clinic environment | | Intended<br>application<br>site | Finger | Finger | Finger | | use under motion and<br>low perfusion<br>conditions | No | No | No | | Signal<br>Detection<br>Method | Photodetector | Photodetector | Photodetector | | Display screen | LED | OLED | OLED | | Display content | SpO2%,<br>PR<br>battery indication<br>Pulse bar graph | SpO2%,<br>PR<br>battery indication<br>Pulse bar graph<br>pulse waveform | SpO2%,<br>PR<br>battery indication<br>Pulse bar graph<br>Perfusion Index | | Contacting material | Shell: ABS<br>Clip Pad: Silicon<br>Button: PMMA | Shell: ABS<br>Clip Pad: Silicon<br>Button: PMMA | Shell: ABS<br>Clip pad: Silicon<br>Button: PC | | SpO2 Range | 0~100% | 0~100% | 0~100% | | SpO2 Resolution | 1% | 1% | 1% | | SpO2 Accuracy | 70~100%: ±2%<br>0% to 69%: unspecified | 70~100%: ±2%<br>0% to 69%: unspecified | 70~100%: ±3%<br>0% to 69%: unspecified | | Pulse Rate Range | 40 bpm ~ 250 bpm | 40 bpm ~ 250 bpm | 30 bpm ~ 240 bpm | | Pulse Rate Accuracy | ±3 bpm or ±1%, whichever is greater | ±3 bpm or ±1%, whichever is greater | ±1 bpm or ±1%, whichever is greater | | Pulse Rate Resolution | 1 bpm | 1 bpm | 1 bpm | | Shipped Sterile | No | No | No | | Power source | 2*1.5V AAA alkaline battery | 2*1.5V AAA alkaline battery | 2*1.5V AAA alkaline battery | | Storage and Transport<br>Environment | Temperature: -20°Cto<br>Atmospheric Pressure: 50 kPa<br>to 107.4 kPa<br>Relative Humidity: 10%-95% (no<br>condensation) | Temperature: -20°Cto<br>Atmospheric Pressure: 50 kPa<br>to 107.4 kPa<br>Relative Humidity: 10%-95% (no<br>condensation) | Temperature: -10°Cto<br>Atmospheric Pressure: 70 kPa to<br>106 kPa<br>Relative Humidity: 10%-80% (no<br>condensation) | | Operating<br>Environment | Temperature: 15°Cto40°C<br>Atmospheric Pressure: 70 kPa to 106 kPa<br>Relative Humidity: 15%-95% (no<br>condensation) | Temperature: 15°Cto40°C<br>Atmospheric Pressure: 70 kPa to 106 kPa<br>Relative Humidity: 15%-95% (no<br>condensation) | Temperature: 5°Cto40°C<br>Atmospheric Pressure: 70 kPa to 106 kPa<br>Relative Humidity: 15%-80% (no<br>condensation) | | Bio-compatibility | ISO 10993-1<br>ISO 109903-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 109903-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 109903-5<br>ISO 10993-10 | | Electrical Safety | IEC 60601-1<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-11 | | EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | | Performance | ISO 80601-2-61 | ISO 80601-2-61 | ISO 80601-2-61 | | Dimension<br>(Width*Height*Depth) | 57mm×30mm×30 mm | 57mm×30mm×30 mm | 60mm×36mm×35 mm | | Weight | <28 g (without the batteries) | <28 g (without the batteries) | 51g | Table 1 Comparison between main predicate C101H1 and the subject device {5}------------------------------------------------ {6}------------------------------------------------ {7}------------------------------------------------ Shenzhen Brav electronic technologies co., Ltd {8}------------------------------------------------ In conclusion, the differences in technological characteristics do not raise new questions of safety and effectiveness. # Performance Summary Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below. #### Non-Clinical Testing: The following performance data were provided in support of the substantial equivalence determination. #### Biocompatibility testing The biocompatibility evaluation for the Pulse Oximeter was conducted in accordance withthe International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of permanent (>30 days). And the testing included the following tests, results of which demonstrate the biological safety of the subject device: - < Cytotoxicity - � Skin Sensitization - � Skin Irritation #### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the IEC 60601-1: 2005+CORR. 1 (2006)+CORR. 2 (2007)+AM1 (2012) Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC. So the Pulse Oximeter would be safe and effective during the intended application scenarios. #### Bench Testing Bench testing was conducted and the results show that the subject device complies with the ISO 80601-2-61: 2017 Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of Pulse Oximeter Equipment standard for performance effectiveness, IEC 60601-1-11:2015 Medical electrical equipment — Part 1-11: General requirements for basic safety and essential performance — Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, which can future demonstrate the safety and effectiveness of the subject device. #### Software Verification and Validation Testing Software documentation including verification & validation was provided in accordance with FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern. And data shows thesoftware work as expected to make sure the performance stable when operated as suggested by the manufacturer. The above tests were selected to show substantial equivalence between the subject device and the predicate device. {9}------------------------------------------------ ### Clinical data: Randomized Clinical Study is conducted outside of United States per Annex EE Guideline for evaluating and documenting SpO2 ACCURACY in human subjects of ISO 80601-2-61:2017 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment and Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and Drug Administration Staff. There were 12 adult volunteers to validated the accuracy of finger pulse oximeter (Model: OP-101). 5 males and 7 females (including 4 subjects with Fitzpatrick V, VI, 8 subjects with Fitzpatrick I ~ Fitzpatrick IV). The measure result between the SpO2 measured by subject device and the SaO2 measured by the blood gas analyzer, was conformed with the requirements of this clinical trial and related standards. The pulse oximeter had not been found any safety issues, adverse effects and complications during the clinical trial. #### Summary Based on the non-clinical performance and clinical data as documented in the device development, the subject devices were found to have a safety and effectiveness profile that is equal to the predicate device. #### Conclusion The Pulse Oximeter is monitoring the patient's SpO2 and pulse rate in non-invasive and convenient. Non-clinical testing and clinical testing were conducted on the subject device and all testing passed pre-specified criteria. The risks of Pulse Oximeter also have been evaluated according to ISO 14971, the overall residual risk and side effect are acceptable, the benefit is greater than risk and hazard. Which demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed device predicate. The subject device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...