The Fingertip Pulse Oximeter is intended for spot-checking oxygen saturation and pulse rate, and the device is a reusable device and intended to be used with the finger of adults or children over three years old in healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
Device Story
Fingertip pulse oximeter; reusable; intended for healthcare environments. Input: red and infrared light transmitted through finger tissue. Principle: 2-wavelength relative optical absorption; photodetector measures light volume transmitted through tissue, which varies with blood pulse and hemoglobin oxygen saturation. Microprocessor calculates SpO2 and pulse rate. Output: SpO2%, pulse rate, pulse rate bar graph, pulse waveform, and battery status on OLED or TFT display. Operated by healthcare professionals. Benefits: provides non-invasive, real-time monitoring of oxygen saturation and pulse rate for clinical assessment. Device includes 20-set data memory; powered by 2x1.5V AAA batteries.
Clinical Evidence
Clinical testing conducted per Annex EE of ISO 80601-2-61:2011 to evaluate SpO2 accuracy in human subjects. Bench testing confirmed compliance with ISO 80601-2-61:2017 for pulse rate accuracy and basic safety. Biocompatibility (cytotoxicity, sensitization, irritation) per ISO 10993-1, electrical safety per ANSI/AAMI ES60601-1, and EMC per IEC 60601-1-2 were verified.
Technological Characteristics
Reusable fingertip pulse oximeter; probe with red/IR LEDs and photodetector. Power: 2x1.5V AAA batteries. Display: OLED or TFT. Biocompatibility: ISO 10993-1/5/10. Electrical safety: AAMI/ANSI ES60601-1, IEC 60601-1-11. EMC: IEC 60601-1-2. Performance: ISO 80601-2-61. Ingress protection: IP22. Type BF applied part.
Indications for Use
Indicated for spot-checking oxygen saturation and pulse rate in adults and children over three years old in healthcare environments. Contraindicated for use under motion or low perfusion scenarios.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Shenzhen Yimi Life Technology Co., Ltd., YM101/YM201/YM301 Pulse Oximeter (K191430)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. The full name of the agency is "U.S. Food & Drug Administration".
Shenzhen BSX Technology Electronics Co., Ltd. Fang Kai General Manager Rm301.3F 8th Building, LiHao Industrial Area, No.78 AiNan Road, Longgang Shenzhen, Guangdong 518116 China
Re: K202324
Trade/Device Name: Fingertip Pulse Oximeter, Models BSX221, BSX223, BSX261, BSX263, BSX281, BSX283, BSX231, BSX233, BSX251, BSX253
Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: November 20, 2020 Received: November 27, 2020
Dear Fang Kai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202324
Device Name
Fingertip Pulse Oximeter, Models BSX221, BSX261, BSX261, BSX283, BSX281, BSX231, BSX231, BSX253
Indications for Use (Describe)
The Fingertip Pulse Oximeter is intended for spot-checking oxygen saturation and pulse rate, and the device is a reusable device and intended to be used with the finger of adults or children over three years old in healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Prepared in accordance with the requirements of 21 CFR Part 807.92 K202324
- 1. Submitter: Shenzhen BSX Technology Electronics Co., Ltd. Rm301.3F 8th Building, LiHao Industrial Area, No.78 AiNan Road, Longgang District, Shenzhen, 518116, P.R. China TEL: +86 755 28719103 Contact Person: Fang Kai Prepare date: 2020-09-12 2. Device name Device Name: Fingertip Pulse Oximeter and classification Models: BSX221, BSX223, BSX261, BSX263, BSX281, BSX283, BSX231, BSX233, BSX251, BSX253 Classification Name: 21 CFR 870.2700 Oximeter Product code: DQA Regulatory Class: Class II Class III Summary and Certification: Not applicable IVD Classification: Not applicable 3. Reason for New Application. No prior submission associated with the current Submission submission. 4. Predicate Shenzhen Yimi Life Technology Co., Ltd., YM101/YM201/YM301 Pulse Device(s) Oximeter / K191430 5. Device The oximeter consists of probe, electronic circuits, and display and plastic Description enclosures. And one side of probe is designed to locate light emitting diodes and a light detector (called a photo-detector). Red and Infrared lights are shone through the tissues from one side of the probe to the other. Then parts of the light emitted absorbed by blood and tissues. The light absorbed by the blood varies with the oxygen saturation of haemoglobin. After that, the photo-detector detects the light volume transmitted through the tissues which depends on blood pulse, Hereafter, the microprocessor calculates a value for the oxygen saturation (SpO2). The subject device is a reusable device, and need to reprocess as suggested in the user manual after each use. And the device is intended to be used on the finger, and powered by 2*1.5V AAA battery. The differences between different models only includes the color of the non-patient contacting shell part and the size. And the color display screen of BSX221/BSX223/BSX261/BSX263/BSX281/BSX283 is OLED, while that of BSX231/BSX233/BSX251/BSX253 is TFT. 6. Indications for The Fingertip Pulse Oximeter is intended for spot-checking oxygen Use saturation and pulse rate, and the device is a reusable device and intended to be used with the finger of adults or children over three years old in
{4}------------------------------------------------
healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
# 7. Predicate Device Comparison
Please refer to following table to find differences between the subject device and predicate device.
| ITEM | Predicate Device<br>YM 101/YM201/YM301<br>Pulse Oximeter<br>K191430 | Proposed Device<br>BSX221, BSX223, BSX261,<br>BSX263, BSX281, BSX283,<br>BSX231, BSX233, BSX251,<br>BSX253 | Comparison<br>Result |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacture | Shenzhen Yimi<br>Life Technology Co., Ltd. | Shenzhen BSX Technology<br>Electronics Co., Ltd. | --- |
| Indications for<br>Use | The pulse oximeter is<br>intended for measure oxygen<br>saturation and pulse rate of<br>adult patients in healthcare<br>environments. | The Fingertip Pulse Oximeter<br>is intended for measure<br>oxygen saturation and pulse<br>rate of adults or children over<br>three years old. | Different |
| Intended patient<br>population | Adult | Adult or children over three<br>years old | Different |
| Contraindications | Not intended to be used under<br>motion or low perfusion<br>scenarios. | Not intended to be used under<br>motion or low perfusion<br>scenarios. | Same |
| Operational Specifications | | | |
| Intended<br>application site | Finger | Finger | Same |
| Measurement<br>Principles | 2-wavelength Relative<br>Optical Absorption | 2-wavelength Relative<br>Optical Absorption | Same |
| Signal Detection<br>Method | Photodetector | Photodetector | Same |
| Display content | SpO2%, PR, battery indicator,<br>Pulse rate<br>bar graph, pulse waveform | SpO2%, PR, battery indicator,<br>Pulse rate<br>bar graph, pulse waveform | Same |
| SpO2 Range | 0%~100% | 35%~100% | Different |
| SpO2 Accuracy | 70~100%: ±2%<br>0% to 69%: unspecified | 70% ~79%: ±3%;<br>80%~100%:±2%;<br>0% to 69%: unspecified | Different |
| SpO2 Resolution | 1% | 1% | Same |
| Pulse Rate<br>Range | 25 bpm ~ 250 bpm | 30 bpm ~ 250 bpm | Different |
| Pulse Rate<br>Accuracy | ±2 bpm | ±3 bpm | Different |
| Pulse<br>Rate Resolution | 1 bpm | 1 bpm | Same |
| Data Memory | Not available | 20 sets | Different |
| Shipped Sterile | No | No | Same |
| Power supplier | 2*1.5V AAA alkaline battery | 2*1.5V AAA alkaline battery | Same |
| Safety<br>classification | | | |
| | Class IIa, Internally powered,<br>Type BF applied part, IP22 | Class IIa, Internally powered,<br>Type BF applied part, IP22 | Same |
| Storage and<br>Transport<br>Environment | Temperature: -20°C to 60°C<br>Atmospheric Pressure: 50 kPa to 107.4 kPa<br>Relative Humidity: 10%-95% (no condensation) | Temperature: -20°Cto 55°C<br>Atmospheric Pressure: 70 kPa to 106 kPa<br>Relative Humidity: <93% (no condensation) | Different |
| Operating<br>Environment | Temperature: 15°Cto 40°C<br>Atmospheric Pressure: 70 kPa to 106 kPa<br>Relative Humidity: 15%-95% (no condensation) | Temperature: 10°Cto 40°C<br>Atmospheric Pressure: 70 kPa to 106 kPa<br>Relative Humidity: 30%-75% (no condensation) | |
| Compliance Standards | | | |
| Bio-compatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
| Electrical Safety | IEC 60601-1<br>IEC 60601-1-11 | AAMI/ANSI EC 60601-1<br>IEC 60601-1-11 | |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | |
| Performance | ISO 80601-2-61 | ISO 80601-2-61 | |
{5}------------------------------------------------
As seen in the comparison tables, the subject and predicate devices have same design principle, similar design features and performance specifications. The different technological characteristics between the subject and predicate devices will not raise different questions of safety or effectiveness.
# 8. Performance Testing
Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.
# Non-Clinical Testing:
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
The biocompatibility evaluation for the Fingertip Pulse Oximeter was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the testing included the following tests, results of which demonstrate the biocompatibility of the subject device:
- Cytotoxicity
- Skin Sensitization
- Skin Irritation
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) for safety and the IEC 60601-1-2: 2007 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests standard for EMC.
{6}------------------------------------------------
#### Bench Testing
Bench testing was conducted and the results show that the subject device complies with the ISO 80601-2-61: 2017 Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of Pulse Oximeter Equipment standard. And Pulse Rate Accuracy meets the requirements defined in ISO 80601-2-61, Clause 201.12.1.104.
#### Software Verification and Validation Testing
Software documentation including verification & validation was provided in accordance with FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern.
#### Cleaning Validation
Cleaning and disinfection validation testing was conducted in accordance with FDA quidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" issued March 17,2015. Moreover, the performance of the subject device shows no degradation after repeated cleaning and disinfection as suggested in the manual.
#### Clinical data:
Clinical testing is conducted per Annex EE Guideline for evaluating and documenting SpO2 ACCURACY in human subjects of ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
#### 9. Conclusion
Verification and validation testing was conducted on the subject device Fingertip Pulse Oximeter and all testing passed pre-specified criteria. The subject device and the predicate device has the same intended use, and the difference in technological features do not raise different questions of safety and effectiveness. This premarket notification submission demonstrates that the subject device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.