K173123 · Shenzhen Imdk Medical Technology Co., Ltd. · DQA · Aug 13, 2018 · Cardiovascular
Device Facts
Record ID
K173123
Device Name
Pulse Oximeter
Applicant
Shenzhen Imdk Medical Technology Co., Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Aug 13, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Intended Use
Fingertip Pulse Oximeter C101H1 is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
Device Story
The C101H1 is a portable, reusable fingertip pulse oximeter for spot-checking adult patients in clinical settings. It uses a finger sensor to emit red (660 nm) and infrared (904 nm) light through perfused tissue. A light-electronic transducer detects light absorption changes caused by arterial blood pulsation; these signals are processed by internal hardware and software to calculate and display SpO2 percentage and pulse rate (BPM). The device is not intended for continuous monitoring, life support, or use in low-perfusion conditions. Clinicians use the displayed numerical values to assess patient oxygenation status, aiding in clinical decision-making. The device is non-sterile and does not include alarms.
Clinical Evidence
Clinical trial conducted per ISO 80601-2-61:2011 on 12 healthy adult volunteers (ages 22-27) with varying skin pigmentation. Subjects underwent induced hypoxia to achieve stable oxygen saturation levels between 70% and 100%. Device performance was compared against arterial blood CO-Oximetry. Results showed an Arms of 2.12 for SpO2 accuracy in steady-state, non-motion conditions.
Technological Characteristics
Materials: Enclosure, button, and finger clip (biocompatible per ISO 10993-5/10). Sensing: Red (660 nm) and IR (904 nm) LEDs. Power: 2 AAA alkaline batteries. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 80601-2-61. Connectivity: Standalone. Software: Moderate level of concern.
Indications for Use
Indicated for spot checking SpO2 and pulse rate in adult patients in hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
AMEMO Inc. Fingertip Pulse Oximeter A310 (K153021)
Related Devices
K221979 — Pulse Oximeter (Model C101A2, C101B1, C101A3) · Shenzhen Imdk Medical Technology Co., Ltd. · Oct 16, 2022
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shenzhen IMDK Medical Technology Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. 22A, Haijing Square No. 18, Taizi Road Nanshan District, Shenzhen, 518067 Cn
Re: K173123
Trade/Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: July 12, 2018 Received: July 12, 2018
Dear Kevin Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Todd D. Courtney -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K173123
Device Name Pulse Oximeter
Indications for Use (Describe)
Fingertip Pulse Oximeter C101H1 is a non-invasive device intended for spot checking of function of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
### 510(K) Summary
### Prepared in accordance with the requirements of 21 CFR Part 807.92
### Prepared Date: 2017/09/20
#### 1. Submission sponsor
Name: Shenzhen IMDK Medical Technology Co., Ltd.
Address: C Zone, 10F, Building 16, Yuanshan Industrial B Area, Gongming Street, Guangming
District, Shenzhen, Guangdong, P.R. China
Contact person: Xia Yuan
Title: Administration Director
E-mail: vicky@estne.org
Tel: +86 755 33637905
#### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: 22A, Haijing Square, No. 18, Taizi Road, Nanshan District, Shenzhen, P.R. China 518067 Contact person: Kevin Wang E-mail: kevin@chonconn.com
Tel: +86-755 33941160
| Trade/Device Name | Pulse Oximeter |
|-------------------|---------------------------------|
| Model | C101H1 |
| Common Name | Pulse Oximeter |
| Regulatory Class | Class II |
| Classification | 21CFR 870.2700 / Oximeter / DQA |
| Submission type | Traditional 510(K) |
#### 3. Subject Device Information
#### 4. Predicate Device
AMEMO Inc. Fingertip Pulse Oximeter A310/ K153021.
The subject device has same intended use, same target patient population, and same performance effectiveness as the predicate device and there are no different questions of effectiveness and safety. So, the conclusion is that the subject device is substantial equivalent to the predicate.
#### 5. Device Description
SpO2 is based on the absorption of pulse blood oxygen to red and infrared light by means of finger sensor and SpO2 measuring unit. The light-electronic transducer in finger sensor converts the pulse red and infrared light modulated by pulse blood oxygen into electrical signal is processed by hardware and software of the unit. The PLETH curve and numeral value of SpO2 will be obtained.
The pulse oximeter, C101H1, is designed for spot checking of the pulse oxygen saturation and pulse rate
{4}------------------------------------------------
for adults in a clinic environment. This medical device can be reused. Not for continuously monitoring. The device is not for life-supporting or life-sustaining, not for implant.
The device is not provided sterile and is NOT a reprocessed single-use device.
The device is a spot-check pulse oximeter and does not include alarms.
The device does not support the measurement in the condition of low perfusion.
The device is not intended for life-supporting or life-sustaining.
The device is reusable and does not need sterilization.
#### 6. Intended use & Indication for use
Fingertip Pulse Oximeter C101H1 is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
#### 7. Predicate Device Comparison
The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters improvement, detailed substantial equivalence discussion is included in the following tables.
| Comparison Items | Proposed device: C101H1 | Predicate device: A310<br>(K153021) |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use/ Indications<br>for use | Fingertip Pulse Oximeter C101H1<br>is a non-invasive device intended<br>for spot checking of functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate<br>(PR). This portable device is<br>indicated for use in adult patients<br>in hospitals. | Fingertip Pulse Oximeter A310 is<br>a non-invasive device intended for<br>spot checking of functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate<br>(PR). This portable device is<br>indicated for use in adult patients<br>in hospitals |
| Principle | The device displays numerical<br>values for functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate by<br>measuring the absorption of red<br>and infrared (IR) light passing<br>through perfused tissue. Changes<br>in the absorption caused by the<br>pulsation of blood in the vascular | The device displays numerical<br>values for functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate by<br>measuring the absorption of red<br>and infrared (IR) light passing<br>through perfused tissue. Changes<br>in the absorption caused by the<br>pulsation of blood in the vascular |
{5}------------------------------------------------
| | Comparison Items | Proposed device: C101H1 | Predicate device: A310<br>(K153021) |
|-----------|----------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|
| | | bed are used to determine oxygen<br>saturation and pulse rate. | bed are used to determine oxygen<br>saturation and pulse rate. |
| ecificati | LED wavelength | Red= 660 nm; Infrared=904nm | Red= 660 nm; Infrared=905nm |
| | Power source | 2 AAA alkaline batteries. | 2 AAA alkaline batteries. |
| | Display data | SpO2%; PR | SpO2%; PR |
| | SpO2 Measuring<br>Range | 0%-100% | 0%-100% |
| | SpO2 Resolution | 1% | 1% |
| | SpO2 Accuracy | 70~100%, ±3%; | 70~100%, ±3%; |
| | | | 0-69%, unspecified; |
| | PR Measuring<br>Range | 30-240BPM. | 30-235BPM. |
| | PR Resolution | 1 bpm | 1 bpm |
| | PR Accuracy | ±1 bpm or ±1%, whichever is | ±2 bpm (30-99bmp) |
| | | greater | ±1% (100-235bmp) |
| | Biocompatibility | ISO 10993-5 and ISO 10993-10 | ISO 10993-5 and ISO 10993-10 |
| 'esting | Electrical Safety | IEC 60601-1 | IEC 60601-1 |
| | | IEC 60601-1-11 | IEC 60601-1-11 |
| | Electromagnetic<br>Compatibility | IEC 60601-1-2 | IEC 60601-1-2 |
| | Performance | ISO 80601-2-61 | ISO 80601-2-61 |
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination. Biocompatibility evaluation:
The biocompatibility evaluation for the Pulse Oximeter C101H1 were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
The subject devices are made up of enclosure, button and finger clip. The finger clip will contact with patient skin. According to ISO 10993-1, the contacting classification is surface contacting and the
{6}------------------------------------------------
duration is intact skin contact less than 24 hours.
### Electrical safety and electromagnetic compatibility (EMC):
Electrical safety and EMC testing were conducted on the Pulse Oximeter C101H1, consisting of all the modules and accessories in the system. The system complies with the IEC 60601-1:2005/A1: 2012, IEC 60601-1-11: 2015 and ISO 80601-2-61: 2011 standards for electrical safety and the IEC 60601-1-2:2014 standard for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
### Performance Testing
- Pulse rate accuracy test
- Cleaning and disinfection cycle test
### Clinical data:
The clinical trial was performed according to Clause 201.12.1.101.2 and Annex EE.2 of ISO 80601-2-61:2011 Procedure for invasive laboratory testing on healthy volunteers.
The purpose of the clinical trial was to evaluate the SpO2 accuracy performance of the C101H1 Pulse Oximeter during stationary (non-motion) conditions over a wide range of arterial blood oxygen saturation levels as compared to arterial blood CO-Oximetry. 12 healthy adult volunteer subjects (ages 22-27yr, with light to dark pigmentation) were included in the study conducted to evaluate the SpO2 accuracy performance of proposed devices. Each system was evaluated during steady state/non-motion conditions with various levels of induced hypoxia resulting in stable oxygen saturation levels between 100% and 70% SaO2. Arterial blood samples were drawn during simultaneous data collection from the test devices. The blood was immediately analyzed on reference CO-Oximetry providing functional SaO2 for the SpO2 accuracy comparison. The SpO2 accuracy performance results showed the fingertip pulse oximeter to have an Arms of 2.12 during steady state conditions over the range of 70-100%.
### Summary:
Based on the non-clinical and clinical performance as documented in the system development, the subject devices were found to substantially equivalent to the predicate device.
#### 9. Conclusion
The non-clinical data support the substantial equivalence of the device and the hardware and software verification and validation demonstrate that the subject device should perform as intended in the specified use conditions. The clinical data demonstrate that the subject device performs comparably to the predicate
{7}------------------------------------------------
device that is currently marketed for the same intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.